ProMinent DULCOMETER Assembly And Operating Instructions Manual

Category
Measuring, testing & control
Type
Assembly And Operating Instructions Manual
DULCOMETER
®
Compact Controller
pH/redox measured variable
Assembly and operating instructions
A0206
Part no. 986214 BA DM 190 10/11 EN
Please carefully read these operating instructions before use! · Do not discard!
The operator shall be liable for any damage caused by installation or operating errors!
Technical changes reserved.
986214, 2, en_GB
© 2011
ProMinent Dosiertechnik GmbH
Im Schuhmachergewann 5 - 11
69123 Heidelberg
Telephone: +49 6221 842-0
Fax: +49 6221 842-419
Internet: www.prominent.com
2
Table of contents
1 Introduction............................................................................................................. 5
1.1 Explanation of the safety information............................................................. 5
1.2 Users' qualifications........................................................................................ 7
2 Safety and responsibility......................................................................................... 9
2.1 General Safety Information............................................................................ 9
2.2 Correct and proper use................................................................................ 10
3 Functional description........................................................................................... 12
3.1 Flow diagram................................................................................................ 13
3.2 Overview of the first level menu................................................................... 14
4 Mounting and installation...................................................................................... 17
4.1 Scope of delivery.......................................................................................... 19
4.2 Mounting (mechanical)................................................................................. 19
4.2.1 Wall mounting............................................................................................ 19
4.2.2 Pipe mounting........................................................................................... 21
4.2.3 Control panel mounting............................................................................. 22
4.3 Installation (electrical)................................................................................... 30
4.3.1 Cable Cross-Sections and Cable End Sleeves......................................... 31
4.3.2 Installation of coaxial cable to guard terminal XE1.................................... 31
4.3.3 Installation (electrical)................................................................................ 36
4.4 Switching of inductive loads......................................................................... 36
5 Commissioning..................................................................................................... 39
5.1 Initial commissioning.................................................................................... 39
5.2 Selection of the measured variable.............................................................. 39
5.3 Setting the controller during commissioning................................................. 39
6 Operating diagram................................................................................................ 41
6.1 Overview of device /Control elements.......................................................... 41
6.2 Adjusting display contrast............................................................................. 42
6.3 Continuous display....................................................................................... 43
6.4 Info display................................................................................................... 43
6.5 Password...................................................................................................... 45
7 Operating menus for the pH and redox measured variables................................ 46
7.1 pH sensor calibration (CAL)......................................................................... 46
7.2 Redox sensor calibration (CAL)................................................................... 54
Table of contents
3
7.3 Limit value setting (LIMITS).......................................................................... 56
7.4 Control setting (CONTROL)......................................................................... 58
7.5 Input setting (INPUT).................................................................................... 61
7.6 Output setting (OUTPUT)............................................................................. 64
7.7 DEVICE setting............................................................................................ 68
8 Control parameters and functions......................................................................... 69
8.1
DULCOMETER
®
Compact Controller function states ................................. 69
8.2 STOP/START key........................................................................................ 71
8.3 Priming (PRIME).......................................................................................... 72
8.4 Hysteresis limit............................................................................................. 73
8.5 Temperature correction variable for pH........................................................ 74
8.6 Checkout time measurement variable and correction variable.................... 75
8.7 Checkout time control................................................................................... 75
8.8 Power relay "P-REL" as limit value relay...................................................... 76
8.9 Setting and functional description of "Relay Used as a Solenoid Valve" .... 77
8.10 Alarm relay................................................................................................. 79
8.11 "Error logger" operating mode.................................................................... 79
9 Maintenance......................................................................................................... 80
9.1
Changing the fuse, DULCOMETER
®
Compact Controller........................... 80
9.2 Fault reporting and troubleshooting.............................................................. 81
10
Technical data DULCOMETER
®
Compact Controller........................................... 84
10.1 Permissible ambient conditions.................................................................. 84
10.2 Sound Pressure Level................................................................................ 84
10.3 Material data............................................................................................... 85
10.4 Chemical Resistance.................................................................................. 85
10.5 Dimensions and weights............................................................................ 86
11 Electrical data....................................................................................................... 87
12
Spare parts and accessories................................................................................ 90
13 Standards complied with....................................................................................... 91
14 Disposal of used parts.......................................................................................... 92
15 Declaration of Conformity..................................................................................... 93
16 Index..................................................................................................................... 94
Table of contents
4
1 Introduction
Data and functions
These operating instructions describe the
technical data and functions of the
DULCOMETER
®
Compact Controller pH /
redox measured variable.
Order number DULCOMETER
®
Compact
Controller pH / redox measured variable:
1035638
General non-discriminatory approach
In order to make it easier to read, this
document uses the male form in grammat‐
ical structures but with an implied neutral
sense. It is aimed equally at both men and
women. We kindly ask female readers for
their understanding in this simplification of
the text.
1.1
Explanation of the safety
information
Introduction
These operating instructions provide infor‐
mation on the technical data and functions
of the product. These operating instruc‐
tions provide detailed safety information
and are provided as clear step-by-step
instructions.
The safety information and notes are cate‐
gorised according to the following
scheme. A number of different symbols
are used to denote different situations.
The symbols shown here serve only as
examples.
DANGER!
Nature and source of the danger
Consequence: Fatal or very serious
injuries.
Measure to be taken to avoid this
danger
Danger!
Denotes an immediate threat‐
ening danger. If this is disre‐
garded, it will result in fatal or
very serious injuries.
WARNING!
Nature and source of the danger
Possible consequence: Fatal or very
serious injuries.
Measure to be taken to avoid this
danger
Warning!
Denotes a possibly hazardous sit‐
uation. If this is disregarded, it
could result in fatal or very
serious injuries.
Introduction
5
CAUTION!
Nature and source of the danger
Possible consequence: Slight or
minor injuries, material damage.
Measure to be taken to avoid this
danger
Caution!
Denotes a possibly hazardous sit‐
uation. If this is disregarded, it
could result in slight or minor inju‐
ries. May also be used as a
warning about material damage.
NOTICE!
Nature and source of the danger
Damage to the product or its sur‐
roundings
Measure to be taken to avoid this
danger
Note!
Denotes a possibly damaging sit‐
uation. If this is disregarded, the
product or an object in its vicinity
could be damaged.
Type of information
Hints on use and additional informa‐
tion
Source of the information, additional
measures
Information!
Denotes hints on use and other
useful information. It does not
indicate a hazardous or dam‐
aging situation.
Introduction
6
1.2 Users' qualifications
WARNING!
Danger of injury with inadequately qualified personnel!
The operator of the plant / device is responsible for ensuring that the qualifications
are fulfilled.
If inadequately qualified personnel work on the unit or loiter in the hazard zone of the
unit, this could result in dangers that could cause serious injuries and material
damage.
All work on the unit should therefore only be conducted by qualified personnel.
Unqualified personnel should be kept away from the hazard zone
Training Definition
Instructed personnel An instructed person is deemed to be a person who has been
instructed and, if required, trained in the tasks assigned to him/
her and possible dangers that could result from improper
behaviour, as well as having been instructed in the required
protective equipment and protective measures.
Trained user A trained user is a person who fulfils the requirements made of
an instructed person and who has also received additional
training specific to the system from ProMinent or another
authorised distribution partner.
Trained qualified per‐
sonnel
A qualified employee is deemed to be a person who is able to
assess the tasks assigned to him and recognize possible haz‐
ards based on his/her training, knowledge and experience, as
well as knowledge of pertinent regulations. The assessment of
a person's technical training can also be based on several
years of work in the relevant field.
Introduction
7
Training Definition
Electrician Electricians are deemed to be people, who are able to com‐
plete work on electrical systems and recognize and avoid pos‐
sible hazards independently based on his/her technical training
and experience, as well as knowledge of pertinent standards
and regulations.
Electricians should be specifically trained for the working envi‐
ronment in which the are employed and know the relevant
standards and regulations.
Electricians must comply with the provisions of the applicable
statutory directives on accident prevention.
Customer Service
department
Customer Service department refers to service technicians,
who have received proven training and have been authorised
by ProMinent to work on the system.
Note for the system operator
The pertinent accident prevention regulations, as well as all other generally acknowl‐
edged safety regulations, must be adhered to!
Introduction
8
2 Safety and responsibility
2.1 General Safety Information
WARNING!
Live parts!
Possible consequence: Fatal or very
serious injuries
Measure: Disconnect the mains
power supply prior to opening the
housing
De-energise damaged, defective
or manipulated units by discon‐
necting the mains plug
WARNING!
Unauthorised access!
Possible consequence: Fatal or very
serious injuries
Measure: Ensure that there can
be no unauthorised access to the
unit
WARNING!
Operating errors!
Possible consequence: Fatal or very
serious injuries
The unit should only be operated
by adequately qualified and tech‐
nically expert personnel
Please also observe the oper‐
ating instructions for controllers
and fittings and any other compo‐
nent groups, such as sensors,
measuring water pumps ...
The operator is responsible for
ensuring that personnel are quali‐
fied
CAUTION!
Electronic malfunctions
Possible consequence: Material
damage to destruction of the unit
The mains connection cable and
data cable should not be laid
together with cables that are
prone to interference
Measure: Take appropriate inter‐
ference suppression measures
Safety and responsibility
9
NOTICE!
Correct and proper use
Damage to the product or its sur‐
roundings
The unit is not intended to
measure or regulate gaseous or
solid media
The unit may only be used in
accordance with the technical
details and specifications pro‐
vided in these operating instruc‐
tions and in the operating instruc‐
tions for the individual
components
NOTICE!
Correct sensor operation / Run-in
time
Damage to the product or its sur‐
roundings
Correct measuring and dosing is
only possible if the sensor is
working perfectly
It is imperative that the run-in
times of the sensors are adhered
to
The run-in times should be
allowed for when planning initial
operation
It may take a whole working day
to run-in the sensor
Please read the operating instruc‐
tions for the sensor
NOTICE!
Correct sensor operation
Damage to the product or its sur‐
roundings
Correct measuring and dosing is
only possible if the sensor is
working perfectly
Check and calibrate the sensor
regularly
NOTICE!
Compensation of control deviations
Damage to the product or its sur‐
roundings
This controller cannot be used in
control circuits which require
rapid compensation (< 30 s)
2.2 Correct and proper use
NOTICE!
Compensation for control deviations
Damage to the product or its sur‐
roundings
The controller can be used in pro‐
cesses, which require compensa‐
tion of > 30 seconds
Safety and responsibility
10
NOTICE!
Correct and proper use
The unit is intended to measure and
regulate liquid media. The marking of
the measured variables is located on
the controller and is absolutely
binding.
The unit may only be used in accord‐
ance with the technical details and
specifications provided in this oper‐
ating manual and in the operating
manuals for the individual compo‐
nents (such as, for example, sensors,
fittings, calibration devices, metering
pumps etc.).
Any other uses or modifications are
prohibited.
Safety and responsibility
11
3 Functional description
Brief functional description
The DULCOMETER
®
Compact Controller
for pH and redox measured variables pro‐
vides basic functions for water treatment
applications. It has a fixed configuration
with the following features:
n Measured variables pH and redox
(can be switched over on the
DULCOMETER
®
Compact Controller)
n Language independent operation (use
of abbreviations, such as
[INPUT]
,
[OUTPUT]
,
[CONTROL]
,
[ERROR]
)
n Illuminated display
n 3 LEDs indicate the operating states
(
[f-REL]
active,
[P-REL]
active, error)
n Sensor monitoring of pH
n P or PID control characteristics
n Selectable control direction (raise or
lower measured value)
n Impulse frequency relay
[f-REL]
for
metering pump control
n Power relay
[P-REL]
, configurable as
alarm, limit value or pulse width
modulated (PWM) control output for
metering pumps
n Analogue output 0/4…20 mA, can be
configured as a measured value or
correction variable
n Suction function for all actuators
n Digital input to switch off the
DULCOMETER
®
Compact Controller
or to process a sample water limit
contact by remote control
n Temperature sensor input (Pt 1000)
for temperature compensation of the
pH value
n Protection class IP67 (wall/pipe
mounting), protection class IP54 (con‐
trol panel mounting)
Applications:
n Waste water treatment
n Treatment of drinking water
n Swimming pool water treatment
Functional description
12
3.1 Flow diagram
mA Output
Frequency relay
(f-REL)
(Pump:Metering)
Power relay
(P-REL)
Corrected
measured value
Input
temperature as
correction variable
Contact input for
PAUSE / HOLD
Input
measured value
from the sensor
Control
(P or PID)
Correction
Alarm
Limit
Actuating variable (PWM)
Actuating variable
Measured value
Correction value
Actuating variable
(metering
stroke)
Function depends on
the configuration
Limit value
monitoring
A0325
STOP
START
Manual
0 ... 25 mA
Function depends
on the configuration
Priming
16 mA
Priming
80 % of "PUMPMAX"
Priming
100 %
Fig. 1: Flow diagram
Functional description
13
3.2 Overview of the first level menu
A0276
OPERATION
pH
6.20
80.0 °C
7.20
0%
LIMIT
MENU
MENU
LIMITS
CONTROL
INPUT
LIMITS SETUP
MENU
LIMITS
CONTROL
INPUT
MENU
LIMITS
CONTROL
INPUT
MENU
OUTPUT
CONTROL
INPUT
MENU
DEVICE
OUTPUT
INPUT
ESC
ESC
ESC
ESC
ESC
CONTROL SETUP
INPUT SETUP
OUTPUT SETUP
DEVICE SETUP
STOP
pH
6.20
80.0 °C
7.20
0%
LIMIT
START
STOP
CAL pH
CAL
ESC
Fig. 2: Overview of the first level menu; shown for pH
Display view Selection
with:
Reference Function
OPERATION
pH
6.20
80.0 °C
7.20
0%
LIMIT
A0329
Ä Chapter 6 ‘Oper‐
ating diagram’
on page 41
Changes to the calibration menu.
Functional description
14
Display view Selection
with:
Reference Function
CAL pH
SLOPE 59.16 mV/pH OK
CAL=START
ZERO 0.00 mV OK
A0332
Ä Chapter 7.1 ‘pH
sensor calibration
(CAL)’ on page 46
The calibration menu enables cali‐
bration of controller and sensor.
OPERATION
pH
6.20
80.0 °C
7.20
0%
LIMIT
A0329
Stop/Start the control and
metering function.
STOP
pH
6.20
80.0 °C
7.20
0%
LIMIT
A0333
Ä Chapter 8.2
‘STOP/START key’
on page 71
By pressing the STOP key, the
control is stopped. The STOP key
can be pressed independently of
the currently displayed menu.
However the STOP state is only
shown in the continuous display.
OPERATION
pH
6.20
80.0 °C
7.20
0%
LIMIT
A0329
Ä Chapter 6.3
‘Continuous dis‐
play’ on page 43
Changes from the continuous dis‐
play to the setting menu.
MENU
LIMITS
CONTROL
INPUT
A0326
Ä Chapter 7.3
‘Limit value setting
(LIMITS)’
on page 56
Enables the setting of the limit
value for limit value monitoring.
MENU
LIMITS
CONTROL
INPUT
A0327
Ä Chapter 7.4
‘Control setting
(CONTROL)’
on page 58
Enables parameter setting for the
control.
MENU
LIMITS
CONTROL
INPUT
A0328
Ä Chapter 7.5
‘Input setting
(INPUT)’
on page 61
Enables setting of the measured
value input parameter.
Functional description
15
Display view Selection
with:
Reference Function
Ä Chapter 7.6
‘ Output setting
(OUTPUT)’
on page 64
Enables setting of the mA output
parameter.
MENU
DEVICE
OUTPUT
INPUT
A0331
Ä Chapter 7.7
‘DEVICE setting’
on page 68
Enables adjustment of the pass‐
word and the controller
[RESTART]
function.
Functional description
16
4 Mounting and installation
n User qualification, mechanical instal‐
lation: trained qualified personnel, see
Ä Chapter 1.2 ‘Users' qualifications’
on page 7
n User qualification, electrical installa‐
tion: Electrical technician, see
Ä Chapter 1.2 ‘Users' qualifications’
on page 7
CAUTION!
Possible consequence: Material
damage.
The hinge between the front and rear
part of the housing cannot absorb
high mechanical loading. When
working on the
DULCOMETER
®
Compact Controller, firmly hold the
top section of the controller housing.
CAUTION!
Check band for strain relief
Possible consequence: Material
damage.
The ribbon cable and its socket
cannot be mechanically loaded.
Hence, it is essential that during con‐
troller mounting in the control panel
mounting, the check strap (part
number 1035918) is fitted for strain
relief and mechanical securing.
Without the check strap, the ribbon
cable or its socket may be damaged if
they fall our of the controller upper
housing.
NOTICE!
Mounting position and conditions
The controller fulfils the IP 67
degree of protection requirements
(wall/pipe mounting) or IP 54
(control panel mounting). This
degree of protection is only ach‐
ieved if all seals and threaded
connectors are correctly fitted.
The installation (electrical) can
only take place after mounting
(mechanical)
Ensure that there is unimpeded
access for operation
Secure, low-vibration fixing
Avoid direct sunlight
Permissible ambient temperature
of the controller at the installation
location: -10 ... +60 °C at max.
95% relative air humidity (non-
condensing)
The permissible ambient tem‐
perature of the connected sen‐
sors and other such components
must be considered
The controller is only suitable for
operation in closed rooms. If
operated outside, the controller
must be protected against the
environment by a suitable protec‐
tive enclosure
Measured value and operating
position
Install the device in a favourable
position for reading and operating
(preferably at eye level)
Mounting and installation
17
Mounting position
Leave sufficient free space for the
cables
Packaging material
Dispose of packaging material envi‐
ronmentally. All packaging compo‐
nents are provided with their corre‐
sponding recycling
.
Mounting and installation
18
4.1 Scope of delivery
The following parts belong to the standard scope of delivery of a
DULCOMETER
®
Com‐
pact Controller.
Description Quantity
Assembled device 1
Cable connection set DMTa/DXMa (metr.) 1
Operating instructions 1
4.2 Mounting (mechanical)
The DULCOMETER
®
Compact Controller is suitable for mounting on a wall, a pipe or a
control panel.
Mounting materials (contained in the scope of delivery):
Description Quantity
Wall/pipe bracket 1
Round head screws 5x45 mm 2
Washer 5.3 2
Rawlplug Ø 8 mm, plastic 2
4.2.1 Wall mounting
Mounting (mechanical)
Mounting and installation
19
2
1
A0273
Fig. 3: Removing the wall/pipe bracket
1. Remove the wall/pipe bracket. Pull
the two snap-hooks (1) outwards
and push upwards
2. Fold out the wall/pipe bracket (2)
and pull out in a downwards direc‐
tion
3. Mark two drill holes diagonal to
each other by using the wall/pipe
bracket as a drilling template
4. Drill holes: Ø 8 mm, d = 50 mm
A0274
Fig. 4: Screwing on the wall/pipe bracket
using washers
5. Screw on the wall/pipe bracket
using the washers
6.
Suspend the DULCOMETER
®
Compact Controller at the top in the
wall/pipe bracket and push using
light pressure at the bottom against
the wall/pipe bracket. Then press
upwards until the
DULCOMETER
®
Compact Controller audibly snaps
into position.
Mounting and installation
20
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95

ProMinent DULCOMETER Assembly And Operating Instructions Manual

Category
Measuring, testing & control
Type
Assembly And Operating Instructions Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI