Transfer System - GMAX™
[4]
2. SAFETY INSTRUCTIONS
! Any serious incident that occurs in relation to this medical device should be reported to
STERIS and the competent authority of the member state in which the user and/or patient
is established.
! warning The modification of medical device without manufacturer authorisation is prohibited.
Pinch points
! During patient positioning, moving dierent sections of the tabletop and using extensions and
accessories can cause pinch points to appear, which can injure the patient or the surgical
team. Typical pinch points are shown below:.
! During Trendelenburg and right tilt movements, take care to check that the connections
(charger, hand control or foot pedal) are plugged into the interface to prevent damage to the
cables or connectors.
Precaution during the movements
! Before any movement activation, be sure that there is no risk of contact or collision with an
object, on equipment or a person.
! Do not place any objects on the base of the column because they could become jammed when
the sliding covers on the column are lowered.
Stability
! Once the GMAXTR column is on the floor, ensure that it is stable. If not, move it if necessary to
find a flatter floor surface, or use the adjustable foot of the base.
Moving with a patient
! When moving the tabletop with a patient on it, the patient must be centred more or less over
the trolley attachment.
It is strongly recommended that you use barriers or another system to hold the patient in place.
! The trolley is not intented to cross threshold, to ascend or to descend step. Always move the
trolley very carefully.
! Before going anywhere, ensure that there are no obstacles along the way you wish to use.
When you set o, move the trolley very carefully. Adapt your speed to the environment (i.e.
corridor width, corner angles, doors, etc.).
Switching the table on
! Every time you turn the table on, check that the pre-selected patient orientation corresponds to
what you want to do. When you switch o the table, the automatic patient orientation pre-
selection is memorized, but the manual selection is not.
If the main hand control and the optional pedal are used at the same time, the hand control
overrides the pedal.
Switching the table o
! The STOP keys on the main hand control and on the backup system allow you to stop all table
movements.
They deactivate the optional pedal and other external control mechanisms.
Handling the surgical table
! Avoid handling the surgical table roughly during the operation or transfer. Do not exert excessive
lateral loads on it, do not pull or push the surgical table with the accessories or with the rails.
! Avoid rough handling of the hand control and the backup system. Do not press the keys with
too much force or using unsuitable objects. Protect the hand control from colliding with the
table.
! Do not sit on one of the two ends of the table, especially when the tabletop is not centered, as
this can cause the table to tilt.
! Avoid handling the cables roughly when connecting or disconnecting the hand control, the foot
pedal or the power cord set.
! When power cord is connected to the surgical table, and therefore is lying on the floor, pay
attention to this power cord when moving around the surgical table and remove the cable
before moving the table.
Load limits
! Do not exceed a patient+accessory weight of 250kg for the standard described positions and
respect the indications and warnings signaled.
! The use of certain extensions may decrease this patient + accessories weight. When
extensions are installed on Gmax™ tables, the user manuals of these extensions should also
be consulted to find out the possible weight restrictions related to the use of these extensions.
! It is possible to reach 360kg patient+accessory weight only for the bariatric described
positions.
! In positions where the tabletop will be significantly slide away from the column, ensure that
there is sucient stability for the type of surgical procedure requirements.
! For all other uses, please contact our network for advice. If you are working intensively on the
patient’s chest, we recommend that you bring the tabletop back to a centered position on the
column.
Positioning the patient
! The positions described in this manual are given purely as examples. The required accessories
are not always shown. Under no circumstances should these instructions replace the
instructions given by the surgical team in charge of the patient.
! Ensure the stability of the patient and the table when making combined position.
! Secure the patient properly (ex. using straps), especially when the table has a strongt tilt
inclination.
! Long surgical procedures can cause compression points on patients and have post operative
consequences. Monitor patient during surgical procedures and use gel cushions for extented
procedures.
! During table movements (rev. Trend/Trend, up/down, left/right tilt), a minimum distance should
be left between the c-arm and the patient, to avoid any contact between them during handling.
Autonomy
! A lithium-ion battery is integrated into both the column and the hand control, meaning they can
function with being plugged in. Before using the table, check that the batteries are suciently
charged.
Connections
! The sockets are equipped with locks: do not pull on the hand control or pedal cables to
disconnect them. The internal E-Serve connection is only to be used by technician sta and
with equipment approved by STERIS
! Always use the power cord provided by STERIS.
! The operator should never touch the patient when the operator is plugging or unplugging
the hand control or the the pedal, or when the operator is handling the connectors of these
devices.
Installation (for CPS02G and GMAX ST)
! The fixed components should not be installed in areas where there is a risk of explosion
(presence of flammable anaesthetic gas)
! Installation operations should only be assigned to STERIS technicians or technicians certified by
STERIS, and should be carried out according to the installation instructions prepared by STERIS.
Contact the STERIS installation department
! Disconnect all equipment before installing the table. Do not install the GMAX™ unless all
elements of the GMAX™ System have been properly locked.