  • Hello! I am an AI chatbot trained to assist you with the Roche cobas p 512 LCP1 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the 'Roche cobas p 512 LCP1'?
    What is the purpose of the 24-position centrifuge rack?
    How should the 24-position centrifuge rack be loaded?
    What are the requirements for barcode labels on sample tubes intended for the 24-position centrifuge rack?
    What are the consequences of using a 24-position centrifuge rack with displaced spring insets?
Addendum 2
For cobas p 512 pre-analytical system (64x) operator’s manual 2.1
24-position centrifuge rack
Addendum 2
24-position centrifuge rack
2
Publication information
Publication version Revision date Change description
1.0 Jan 2018 First release
Revisions
This chapter contains additional information for the cobas p 512 pre-analytical system
(64x) operator’s manual and was added due to additional information about the use of the
24-position centrifuge rack (RD ID 06623611001).
Addendum 2
24-position centrifuge rack
3
Table of contents
Affected system ................................................................................................................. 4
Additional information....................................................................................................... 4
Loading of the 24-position centrifuge rack ...................................................................... 5
Placement of barcode labels on sample tubes intended for the 24-position centrifuge
rack ................................................................................................................................. 5
Addendum 2
24-position centrifuge rack
4
Affected system
System
Operator’s manual
publication version *
Applicable
software version
Change
description
cobas p 512
(SN 63xxxxxx)
2.1 2.0.0 and higher Information about
the loading of the
rack and the
labelling of the tubes
* For US customers, this addendum is applicable to the currently approved version of the aforementioned
operator’s manual found on the US Roche Diagnostics website.
Table 1: Affected system
Roche recommends that you familiarize yourself with the new and/or revised content
provided in this addendum.
If you print these pages, Roche recommends printing them single sided. In this way,
you can easily insert the new and/or revised con
tent in its appropriate locations in
the existing document.
Additional information
This information specifically refers to the usage of 24-position centrifuge racks.
Figure 1: 24-position centrifuge rack (RD ID 06623611001)
Addendum 2
24-position centrifuge rack
5
Loading of the 24-position centrifuge rack
This information specifically refers to the usage of 24-position centrifuge racks and is an
addition to section C in the operator’s manual.
NOTICE
Damage to the system due to displaced spring insets
The gripper in the sorting area of the subsequent system may crash into the tube in cases
where the spring insets of a 24
-position centrifuge rack are displaced.
Do not use a 24-position centrifuge rack with displaced spring insets.
Avoid empty rack positions during centrifugation.
When a Hettich stand-alone centrifuge and 24-position centrifuge racks
(RD ID 06623611001) are used, make sure that:
24-position centrifuge racks are fully loaded for centrifuging.
Balance tubes are used to fill empty rack positions.
Placement of barcode labels on sample tubes intended for the 24-
position centrifuge rack
This information about the placement of barcode labels specifically refers to the usage of
24-position centrifuge racks and is an addition to the information in section A of the
operator’s manual.
When a Hettich stand-alone centrifuge and 24-position centrifuge racks
(RD ID 06623611001) are used, make sure that:
No sample tubes that are covered with more than 3 layers of label are used.
Labels run the length of the tube and are parallel to the tube edge.
The labels are applied smoothly, without dog-eared corners, wrinkles, or edges that
stick out.
/