Roche cobas p 512 LCP1 User manual

Type
User manual
cobas p 512
pre-analytical system (64x)
Operator’s manual version 2.1
cobas p 512 pre-analytical system (64x)
Roche Diagnostics
2 Operator’s Manual · Version 2.1
Document information
Revision history
Edition notice This Operator’s Manual is for users of the cobas p 512 pre-analytical system.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche Diagnostics GmbH
reserves the right to make any changes necessary without notice as part of ongoing
product development.
Intended use The cobas p 512 pre-analytical system is a computer controlled fully automatic
system for sorting of open and closed barcoded and centrifuged and non-centrifuged
sample tubes.
It includes modules for registration and de-capping of sample tubes, liquid level
detection and sample quality assessment (optional), as well as re-capping of sample
tubes (optional).
The cobas p 512 pre-analytical system is intended for use with analyzers that perform
tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology,
urinalysis, nucleic acid testing.
In case of standalone use of the cobas p 512 pre-analytical system, the transfer of
tubes, processed on the cobas p 512 pre-analytical system, to any other workstation
within the laboratory is subject of validation performed by the laboratory itself.
Copyright © 2017, Roche Diagnostics GmbH. All rights reserved.
Off the shelf software Portions of the cobas p 512 pre-analytical system might include one or more Open
Source or Commercial Software programs. Copyright and other notices and licensing
information regarding such Software programs can be found on the cobas p 512 pre-
analytical system.
Trademarks The following trademarks are acknowledged:
COBAS is a trademark of Roche.
All other trademarks are the property of their respective owners.
Manual Version Software Version Revision date Changes
2.0 2.0.0 April 2016 First release
2.1 2.0.6 August 2017 Spin status detection
improvements and
re-labeling
Roche Diagnostics
Operator’s Manual · Version 2.1 3
cobas p 512 pre-analytical system (64x)
System approvals
The cobas p 512 pre-anaytical system meets the protection requirements specified in
IVD Directive 98/79/EC.
Compliance is demonstrated by the following marks:
Contact address
For in vitro diagnostic use.
Complies with the Directive 98/79/EC on in vitro
diagnostic medical devices
Issued by Intertek (Electrical Testing Lab) for the US and
Canada.
Manufacturer
Roche Diagnostics GmbH
Sandhofer Strasse 116
D-68305 Mannheim
Germany
cobas p 512 pre-analytical system (64x)
Roche Diagnostics
4 Operator’s Manual · Version 2.1
Roche Diagnostics
Operator’s Manual · Version 2.1 5
cobas p 512 pre-analytical system (64x)
Table of contents
General Information Part A
1 General information
Preface A–5
Using the operator’s manual A–5
Abbreviations A–8
Safety precautions A–9
Prerequisites for scanning a barcode A–23
System Description Part B
2 System description
Introduction B–7
Hardware description B–9
Consumable material B–31
Accessories B–36
Sample flow and sample processing B–38
Basic information about the software B–46
Software description B–52
Operation Part C
3Operation
Commissioning the system C–5
Daily routine C–5
Preparation C–12
Operation C–29
Terminating operation after daily routine C–48
Maintenance Part D
4Maintenance
Device care and maintenance D–5
Cleaning agents and equipment D–6
Care and maintenance schedule D–6
Weekly device care D–7
Monthly device maintenance D–16
Device care, as and when required D–20
Troubleshooting Part E
5 Troubleshooting
What do I do if ... E–5
Important questions for customer service E–15
System and error messages E–16
Appendix Part F
6 Appendix
Included in delivery F–5
Technical data F–6
Consumable materials F–10
Accessories F–11
Sample tubes F–11
Caps F–12
Racks F–13
Floorplan F–16
Glossary F–17
Roche Diagnostics
6 Operator’s Manual · Version 2.1
cobas p 512 pre-analytical system (64x)
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–3
General Information
A
Roche Diagnostics
Operator’s Manual · Version 2.1 A-3
cobas p 512 pre-analytical system (64x) 1 General information
Table of contents
General information
Preface ............................................................................................................................... A–5
Using the operator’s manual ......................................................................................... A–5
Units and conventions used in this document ..................................................... A–6
Screen graphics, signs and symbols ........................................................................ A–6
Command entries ............................................................................................... A–7
Safety symbols ........................................................................................................... A–7
Abbreviations ................................................................................................................... A–8
Safety precautions ........................................................................................................... A–9
Product IT security ................................................................................................... A–9
Operator qualification ............................................................................................ A–10
Warning labels on the system ............................................................................... A–11
Rating plate ........................................................................................................ A–13
Protective hoods ............................................................................................... A–14
Interlocks of protective hoods ........................................................................ A–14
Camera enclosure ............................................................................................. A–15
Input sorter (rear side) ..................................................................................... A–16
LASER-LLD ....................................................................................................... A–16
Decapper ............................................................................................................ A–17
Recapper ............................................................................................................. A–18
Maintenance doors ........................................................................................... A–19
Mains socket ...................................................................................................... A–19
Compressed air connection ............................................................................. A–20
Electrical safety ........................................................................................................ A–20
Samples ..................................................................................................................... A–20
Tubes ........................................................................................................................ A–21
Processing of open tubes ........................................................................................ A–21
LASER exposure ...................................................................................................... A–21
LASER-LLD ....................................................................................................... A–21
Barcode scanner ................................................................................................ A–22
In this chapter
Chapter
1
Roche Diagnostics
A-4 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Table of contents
Drawers .................................................................................................................... A–22
Sorting unit .............................................................................................................. A–22
Acrylic glass cover ................................................................................................... A–22
Prerequisites for scanning a barcode .......................................................................... A–23
Barcode quality, printing and labeling material ................................................. A–23
Barcode label ..................................................................................................... A–24
Specifications for the barcode labels .................................................................... A–25
Placement of the barcode label ............................................................................. A–27
Sample images of correct and incorrect barcode labels ..................................... A–29
Blood sample tubes with faulty or wrongly placed barcode labels ................... A–29
Disposal of device ................................................................................................... A–32
Information on waste of electrical and electronic equipment (WEEE) .... A–32
Roche Diagnostics
Operator’s Manual · Version 2.1 A-5
cobas p 512 pre-analytical system (64x) 1 General information
Preface
Preface
The cobas p 512 (64x) pre-analytical system is a computer controlled, stand-alone
system capable of sorting of barcoded sample tubes. The system operates with an
online connection to the LIS.
It consists of the following components:
o Input sorter with sorting unit
o Tube transport
o Barcode scanner
o Optical tube type identification (options: QSI full version, QSI-LLD-SI, TTI
camera)
o Decapping module
o Recapping module (optional)
o Output sorter with sorting unit (optional: with cobas connection module (CCM)
in the output sorter area)
o Control unit
Using the operator’s manual
This operator’s manual is valid for the cobas p 512 (64x) pre-analytical system. It
contains information for the system operators at the application level. The
information contained in this manual is not intended for administrators or service
technicians.
The operator’s manual is divided into the following sections:
NOTICE
Read the operator’s manual with the utmost care and attention before using the system for
the first time. Particular attention should be paid to descriptions of hazards and
precautions.
Failure to observe notices pertaining to risks and safety contained in this operator’s manual
can have serious implications for health and safety as well as result in material damage.
NOTICE
Keep this manual in a safe place accessible to all operators of the system.
Part Label Contents
A General information Manual description, safety information
B System description System overview, description of the hardware,
description of the software, sample flow, reports
C Operation Working with the system
D Maintenance Maintenance and cleaning instructions
E Troubleshooting Describing and remedying faults
F Appendix Glossary, technical details
Roche Diagnostics
A-6 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Using the operator’s manual
Units and conventions used in this document
Visual cues are used to help locate and interpret information in this manual quickly.
This section explains formatting conventions used in this manual.
The following symbols are used:
Screen graphics, signs and symbols
Screen displays are presented in the operator’s manual as shown in the following
example:
Symbol Used for
a Start of procedure
End of procedure
o List item
e
Cross-reference
f Call-up (software reference)
Tip
Safety alert
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Figure A-1 Screenshot
Roche Diagnostics
Operator’s Manual · Version 2.1 A-7
cobas p 512 pre-analytical system (64x) 1 General information
Using the operator’s manual
Command entries
A 15" touch screen monitor mounted on a swivel arm between the input sorter and
the output sorter is used to program or control the system. This manual depicts
keyboard entries and buttons as follows:
Safety symbols
Throughout the operator’s manual, the safety instructions are labeled as follows:
Enter key
[ENTER]
Page up key
[PAGE UP]
Page down key
[PAGE DN]
Scroll up key
[ï‚­]
Scroll down key
[]
Scroll right key
[ï‚®]
Scroll left key
[]
NOTICE
This symbol indicates advice that users should observe to ensure errors do not occur when
operating the system.
This symbol indicates a potential hazard. The operator or the system may be at risk if the
warning is not heeded.
CAUTION
Caution: Indicates a hazardous situation that, if not avoided, can result in injury to the
operator and/or damage to the system.
WARNING
Warning: Indicates a hazardous situation that, if not avoided, could result in death or
serious injury.
Roche Diagnostics
A-8 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Abbreviations
Abbreviations
The list below contains an explanation of the abbreviations used in this manual:
Abbreviation Meaning
AAmpere
AC Alternating Current
Ah Ampere hour
ASTM American Society for Testing and Materials
BGV Berufsgenossenschaftliche Vorschriften
CCM cobas connection module
CE Communauté Européenne
CPU Central Processing Unit
dB Decibel
DC Direct Current
DIN Deutsche Industrie Norm (German industry standard)
ebox-one Control unit of the CCM
EC European Community
EDP Event Distribution Protocol
EMC Electromagnetic Compatibility
EN European Standard
EPROM Erasable Programmable Read Only Memory
FCC Food Chemicals Codex
GB GigaByte
GUI Graphical User Interface
Hz Hertz
IEC International Electrotechnical Commission
IVD In-Vitro Diagnostics
LASER Light Amplification by Stimulated Emission of Radiation
LCD Liquid Crystal Display
LED Light Emitting Diode
LIS Laboratory Information System
LLD Liquid Level Detection
MB Megabyte
Mbps Megabits per second
MTH Multi-Tube Holder
NN Normal Null (mean sea level)
NW Nominal Width
PE Protective Earth
PIO Parallel I-O Interface
PTT Primary Tube Transport
RAM Random Access Memory
RCM Rack Conveyor Module
RFM Rack Feed Module
RJ45 Registered Jack 45
Roche Diagnostics
Operator’s Manual · Version 2.1 A-9
cobas p 512 pre-analytical system (64x) 1 General information
Safety precautions
Safety precautions
Safety regulations IEC 61010-1 and the applicable national electrical safety guidelines
must be observed! Safe operation and reliability of the system can only be ensured if
o the electrical installation of the system meets appropriate IEC requirements.
o prescribed installation is performed by a trained expert.
o visual inspection, PE conductor test, insulation test and leakage current test are
carried out by a qualified electrician, if necessary (e.g. according to BGV A3 or
other applicable safety guidelines).
o the system is operated in accordance with the operating instructions.
Product IT security
RS232/422 Recommended Standard 232/422
TCP/IP Transmission Control Protocol/ Internet Protocol
VVolt
VA Volt-Ampere
WEEE Directive 2002/96/EG covering Waste of Electrical and Electronic
Equipment
Abbreviation Meaning
CAUTION
Pay attention to the following safety precautions!
If you ignore these precautions, you can suffer minor to moderate injury. Each precaution
is important.
CAUTION
Loss of data integrity or unavailability of the system due to malicious software or
unauthorized system access
To avoid infection of malicious software or the unauthorized misuse of the system, the
following recommendations are essential:
o Make sure other computers and services on the network (for example: the LIS,
archiving share or -service, backup share or -service) are properly secured and
protected against malicious software and unauthorized access.
o Laboratory IT is responsible for the security of their local area network, especially in
protecting it against malicious software and attacks. This protection might include
measures, such as a firewall, to separate the device from uncontrolled networks as well
as measures that ensure that the connected network is free of malicious code.
o Restrict physical access to the system and all attached IT infrastructure (computer,
cables, network equipment etc.).
Roche Diagnostics
A-10 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Safety precautions
Operator qualification
As a user, you are required to have sound knowledge of relevant safety precaution
guidelines and standards and of the information and procedures found in the
operator’s manual.
o Do not carry out any operation or maintenance procedures unless you are trained
to do so by Roche Diagnostics.
o Carefully follow the procedures specified in the operator’s manual for operation
and maintenance.
o Leave maintenance, installation, or service that is not described in the operator’s
manual to trained Roche Service representatives.
o Follow good laboratory practices, especially when working with biohazardous
material.
Roche Diagnostics
Operator’s Manual · Version 2.1 A-11
cobas p 512 pre-analytical system (64x) 1 General information
Safety precautions
Warning labels on the system
Warning labels are used to indicate potential safety hazards and point out any
remaining risk.
All warning labels on the module and in areas that can be accessed by the operator
are documented below.
ß
Figure A-2 Location of warning labels on the front side
Roche Diagnostics
A-12 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Safety precautions
Figure A-3 Location of warning labels on the rear side
Roche Diagnostics
Operator’s Manual · Version 2.1 A-13
cobas p 512 pre-analytical system (64x) 1 General information
Safety precautions
Rating plate
The rating plate is located on the lower rear side of the system besides the interface
block.
A Required voltage, frequency and power
consumption
B Serial number of the system
C Reference number (order number) of the
system
D Type of system
E Warning label F C/N key of the system
G Date of manufacture H Required mains fuse
Figure A-4 Rating plate
E
C
B
A
F
G
D
H
Read the operator’s manual with the utmost care and attention before using the system for
the first time. Particular attention should be paid to descriptions of hazards and
precautions.
Failure to observe notices pertaining to risks and safety contained in this operator’s manual
can have serious implications for health and safety as well as result in material damage.
Roche Diagnostics
A-14 Operator’s Manual · Version 2.1
1 General information cobas p 512 pre-analytical system (64x)
Safety precautions
Protective hoods
A pinching hazard warning label is attached to each of the protective hoods on the
front of the system.
Interlocks of protective hoods
A biohazard warning label is attached besides each of the interlocks for the protective
hoods.
A Protective hood B Warning label
Figure A-5 Pinching hazard warning label
A
B
CAUTION
Pinching hazard!
Contact with moving parts may cause serious injury.
Do not reach into the running system without putting it in standby mode.
A Interlock B Warning label
Figure A-6 Biohazard warning label
A
B
WARNING
Risk of infection!
This warning label warns you of the potential hazards posed by sample material. Always
wear gloves when working on the system and comply with applicable laboratory
regulations.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276

Roche cobas p 512 LCP1 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI