  • Hello! I've reviewed the installation instructions for the HK Instruments DPG series differential pressure gauges. This document provides details on installing, calibrating, and connecting the pressure tubes for the devices. The instructions cover essential aspects of the series, which is designed for use in HVAC/R applications, and includes models with various measurement ranges.
  • What is the DPG series designed to measure?
    Where are DPG series devices commonly used?
    What is the overpressure protection value for DPG series?
Copyright HK Instruments 2023 www.hkinstruments. Installaon version 5.0 2023
INSTALLATION
INSTRUCTIONS
DIFFERENTIAL PRESSURE GAUGES
& MANOMETERS - DPG series
READ THESE INSTRUCTIONS CAREFULLY BEFORE
ATTEMPTING TO INSTALL, OPERATE OR SERVICE THIS
DEVICE.
• Failuretoobservesafetyinformaonandcomplywith
instruconscanresultinPERSONALINJURY,DEATHAND/OR
PROPERTY DAMAGE.
• Retaintheseinstruconsforfuturereference.
• Thisproduct,wheninstalled,willbepartofanengineered
systemwhosespecicaonsandperformancecharacteriscs
arenotdesignedorcontrolledbyHKInstruments.Review
applicaonsandnaonalandlocalcodestoassurethatthe
installaonwillbefunconalandsafe.Useonlyexperiencedand
knowledgeabletechnicianstoinstallthisdevice.
SPECIFICATIONS
ThankyouforchoosinganHKInstrumentsDPGseriesdierenal
pressuregauge.TheDPGseriesisintendedforuseincommercial
environments in HVAC/R applicaons. It is designed to detect air
and non-corrosive gas pressures (dierenal and posive) and to
provideaclearanalogdisplayofthemeasuredvalues.
TheDPGseriesincludeseldadjustablezeropointcalibraonand
overpressureprotecon.
DPGseriesdevicesarecommonlyusedinHVAC/Rsystemsfor:
•fan,blowerandltermonitoring
•pressureandowmonitoring
•pressuremonitoringincleanrooms
SPECIFICATIONS
Performance
Accuracy:
<±3%F.S(20°C),(DPG60<±5%;DPG100<±4%)
Overpressure:
Burstpressure:20kPa
Technical Specicaons
Media compability:
Dryairornon-aggressivegases
Measuring units:
Pa
Measurement ranges:
0–60Pa
0–100Ps
0–120Pa
0–200Pa
0–250Pa
0–300Pa
0–400Pa
0–500Pa
0–600Pa
0–800Pa
0–1kPa
0–1,5kPa
0–2kPa
0–3kPa
0–5kPa
Environment:
Operangtemperature:-5...+60°C
Storagetemperature:-40...+85°C
Physical
Case:
ABS
Cover:
PC
Membrane:
Silicone
Machinery:
Aluminumandsteelspring
Unit mounng diameter
110mm
Weight:
234g
Protecon standard:
IP65
Conformance
Meetstherequirementsfor:
CE:  UKCA:
RoHS:  2011/65/EU S.I.2012/3032
INTRODUCTION WARNING
APPLICATIONS
RoHS
COMPLIANT
Copyright HK Instruments 2023 www.hkinstruments. Installaon version 5.0 2023
INSTALLATION
Figure 1 - Mounng
1)Mountthedeviceinthedesiredlocaon.
2)Adjustthezeropointbyturningthezerosetscrewontopof
thelid.
3)Connectthepressuretubes.Connectposivepressuretoport
labeled"+"andnegavepressuretoport"-".
Surfacemount
Flushmount
DIMENSIONAL DRAWINGS
106.5
0 600
P.max.100 kPa
DPG600
Made in Finland
500
200
100
400
300
Pa
ZERO SET
93.5
120.0
94.0
106.9
4.2
15.244.9
110.0
60.0
0 600
P.max.100 kPa
DPG600
Made in Finland
500
200
100
400
300
Pa
ZERO SET
Copyright HK Instruments 2023 www.hkinstruments. Installaon version 5.0 2023
Thesellerisobligatedtoprovideawarrantyofveyearsforthedeliv-
eredgoodsregardingmaterialandmanufacturing.Thewarrantyperiod
isconsideredtostart onthe delivery date ofthe product. If adefect
in raw materials or a producon awis found, theseller is obligated,
whentheproductissenttothesellerwithoutdelayorbeforeexpira-
onofthewarranty,toamendthemistakeathis/herdiscreoneither
by repairing the defecve product or by delivering free of charge to
thebuyeranewawlessproductandsendingittothebuyer.Delivery
costsfortherepairunderwarranty willbepaidbythebuyerandthe
return costs by the seller. The warranty does not comprise damages
causedbyaccident,lightning,oodorothernaturalphenomenon,nor-
malwearandtear,improperorcarelesshandling,abnormaluse,over-
loading,improperstorage,incorrectcareorreconstrucon,orchanges
andinstallaonworknotdonebytheseller.Theseleconofmaterials
fordevicespronetocorrosionisthebuyer’sresponsibility,unlessother-
wiseislegallyagreedupon.Shouldthemanufactureralterthestructure
ofthedevice,thesellerisnotobligatedtomakecomparablechanges
todevicesalreadypurchased.Appealingforwarrantyrequiresthatthe
buyerhascorrectlyfullledhis/herduesarisenfromthedeliveryand
stated in the contract. The seller will give a new warranty for goods
thathavebeenreplacedorrepairedwithinthewarranty,howeveronly
totheexpiraonoftheoriginalproduct’swarrantyme.Thewarranty
includestherepairofadefecvepartordevice,orifneeded,anewpart
ordevice,butnotinstallaonorexchangecosts.Undernocircumstance
isthesellerliablefordamagescompensaonforindirectdamage.
WARRANTY POLICY
RECYCLING/DISPOSAL
Thepartsleoverfrom installaonshouldberecycledaccordingto
your localinstrucons. Decommissioned devicesshouldbedisposed
ofasmixedwasteaccordingtoyourlocalinstrucons.
/