Teledyne 2000A-EU User manual

Category
Oxygen Equipment
Type
User manual
i
Thermal Conductivity Analyzer
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS FOR
Model 2000A-EU
Thermal Conductivity Analyzer
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
DANGER
P/N M66182
07/22/05
ECO # 05-0131
ii
Model 2000A-EU
Copyright © 1999 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, tran-
scribed, stored in a retrieval system, or translated into any other language or computer
language in whole or in part, in any form or by any means, whether it be electronic,
mechanical, magnetic, optical, manual, or otherwise, without the prior written consent of
Teledyne Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free
from defects of material and of construction, and that our liability shall be limited to
replacing or repairing at our factory (without charge, except for transportation), or at
customer plant at our option, any material or construction in which defects become
apparent within one year from the date of shipment, except in cases where quotations or
acknowledgements provide for a shorter period. Components manufactured by others bear
the warranty of their manufacturer. This warranty does not cover defects caused by wear,
accident, misuse, neglect or repairs other than those performed by Teledyne or an autho-
rized service center. We assume no liability for direct or indirect damages of any kind and
the purchaser by the acceptance of the equipment will assume all liability for any damage
which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our
apparatus, and to make any alterations in the dimensions, shape or weight of any parts, in
so far as such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by
which valuable data can be gathered. The information provided by the instrument may
assist the user in eliminating potential hazards caused by his process; however, it is
essential that all personnel involved in the use of the instrument or its interface, with the
process being measured, be properly trained in the process itself, as well as all instrumenta-
tion related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent danger,
it has no control over process conditions, and it can be misused. In particular, any alarm or
control systems installed must be tested and understood, both as to how they operate and
as to how they can be defeated. Any safeguards required such as locks, labels, or redun-
dancy, must be provided by the user or specifically requested of Teledyne at the time the
order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The
purchaser is responsible for the training of personnel, for providing hazard warning
methods and instrumentation per the appropriate standards, and for ensuring that hazard
warning devices and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the manufactur-
er or its agents, is to be construed as a warranty of adequate safety control under the
user’s process conditions.
iii
Thermal Conductivity Analyzer
Specific Model Information
The instrument for which this manual was supplied may incorporate one or more
options not supplied in the standard instrument. Commonly available options are listed
below, with check boxes. Any that are incorporated in the instrument for which this
manual is supplied are indicated by a check mark in the box.
Instrument Serial Number: _______________________
Options Available with Order:
2000A-C: Auto Calibration valves (zero/span) built-in gas selector panel
and control valves are electronically controlled to provide
synchronization with the analyzer’s operations.
2000A-G: Stainless steel cell block with nickel filaments and Stainless
Steel fittings and tubing.
2000A-H: Stainless steel cell block with gold filaments for corrosive gas
streams and Stainless Steel fittings and tubing.
2000A-K: 19" Rack Mount available with either one or two analyzers
Control Units installed and ready to mount in a standard rack
2000A-L: Gas selector panel consisting of sample/ref flow meters and
control valves for metering input of sample/calibrations
support gases
2000A-R: Sealed reference cell (application dependent, contact factory).
2000A-N: 220 VAC operation.
iv
Model 2000A-EU
Model 2000A-EU complies with all of the requirements of the
Commonwealth of Europe (CE) for Radio Frequency Interference,
Electromagnetic Interference (RFI/EMI), and Low Voltage Directive
(LVD).
The following International Symbols are used throughout the In-
struction Manual for your visual and immediate warnings and when
you have to attend CAUTION while operating the instrument:
STAND-BY, Instrument is on Stand-by,
but circuit is active
GROUND
Protective Earth
CAUTION, The operator needs to refer to the manual
for further information. Failure to do so may
compromise the safe operation of the equipment.
CAUTION, Risk of Electric Shock
COMBUSTIBLE GAS USAGE WARNING
DANGER
This is a general purpose instrument designed for usage in a
nonhazardous area. It is the customer's responsibility to en-
sure safety especially when combustible gases are being ana-
lyzed since the potential of gas leaks always exist.
The customer should ensure that the principles of operating of
this equipment is well understood by the user. Misuse of this
product in any manner, tampering with its components, or
unauthorized substitution of any component may adversely
affect the safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne, no responsibility by Teledyne, its affiliates, and
agents for damage or injury from misuse or neglect of this
equipment is implied or assumed.
v
Thermal Conductivity Analyzer
Table of Contents
1 Introduction
1.1 Overview........................................................................ 1-1
1.2 Typical Applications....................................................... 1-2
1.3 Main Features of the Analyzer ....................................... 1-2
1.4 Model Designations....................................................... 1-3
1.5 Front Panel (Operator Interface) ..................................... 1-3
1.6 Recognizing Difference Between LCD & VFD............... 1-5
1.7 Rear Panel (Equipment Interface) .................................. 1-5
1.8 Gas Connections ........................................................... 1-7
2 Operational Theory
2.1 Introduction .................................................................... 2-1
2.2 Sensor Theory ............................................................... 2-1
2.2.1 Sensor Configuration............................................... 2-1
2.2.2 Calibration ............................................................... 2-2
2.2.3 Effects of Flowrate and Gas Density........................ 2-3
2.2.4 Measurement Results.............................................. 2-3
2.3 Electronics and Signal Processing ................................ 2-3
2.4 Temperature Control ...................................................... 2-5
3 Installation
3.1 Unpacking the Analyzer................................................. 3-1
3.2 Mounting the Analyzer ................................................... 3-1
3.3 Electrical Connections (Rear Panel).............................. 3-3
3.3.1 Primary Input Power............................................... 3-4
3.3.2 Primary Input Power............................................... 3-4
3.3.3 50-Pin Equipment Interface Connector .................. 3-4
3.3.3.1 Analog Outputs .............................................. 3-5
3.3.3.2 Alarm Relays ................................................. 3-6
3.3.3.3 Digital Remote Cal Inputs.............................. 3-7
3.3.3.4 Range ID Relays ........................................... 3-9
3.3.3.5 Network I/O .................................................... 3-9
3.3.3.6 Remote Valve Connector ............................... 3-9
3.3.4 RS-232 Port ........................................................... 3-10
3.4 Gas Connections ........................................................... 3-11
3.4.1 Sample System Design .........................................3-13
3.4.2 Pressure and Flow Rate Regulation......................3-13
3.4.3 VENT Exhaust .......................................................3-14
3.4.4 SAMPLE Gas.........................................................3-14
vi
Model 2000A-EU
3.4.5 REFERENCE Gas ................................................3-15
3.4.6 ZERO Gas .............................................................3-15
3.4.7 SPAN Gas..............................................................3-15
3.5 Testing the System.........................................................3-16
3.6 Warm Up at Power Up....................................................3-16
4 Operation
4.1 Introduction .................................................................... 4-1
4.2 Using the Data Entry and Function Buttons ................... 4-1
4.3 The
System
Function ..................................................... 4-4
4.3.1 Setting the Display................................................. 4-5
4.3.2 Setting up an Auto-Cal........................................... 4-5
4.3.3 Password Protection .............................................. 4-6
4.3.3.1 Entering the Password................................... 4-7
4.3.3.2 Installing or Changing the Password ............. 4-7
4.3.4 Logging Out ........................................................... 4-9
4.3.5 System Self-Diagnostic Test .................................. 4-9
4.3.6 The Model Screen ................................................. 4-10
4.3.7 Checking Linearity with ALGORITHM ................... 4-10
4.4 The
Zero
and
Span
Functions ....................................... 4-11
4.4.1 Zero Cal................................................................. 4-12
4.4.1.1 Auto Mode Zeroing ........................................ 4-12
4.4.1.2 Manual Mode Zeroing.................................... 4-13
4.4.1.3 Cell Failure .................................................... 4-14
4.4.2 Span Cal................................................................ 4-14
4.4.2.1 Auto Mode Spanning ..................................... 4-15
4.4.2.2 Manual Mode Spanning................................. 4-15
4.5 The
Alarms
Function...................................................... 4-16
4.6 The
Range
Function ...................................................... 4-18
4.6.1 Manual (Select/Define Range) Screen .................. 4-19
4.6.2 Auto (Single Application) Screen........................... 4-19
4.6.3 Precautions............................................................ 4-21
4.7 The
Analyze
Function.................................................... 4-22
4.8 Programming ................................................................. 4-22
4.8.1 The Set Range Screen .......................................... 4.23
4.8.2 The Curve Algorithm Screen ................................. 4-25
4.8.2.1 Checking the Linearization ............................ 4-25
4.8.2.2 Manual Mode Linearization ........................... 4-26
4.8.2.3 Auto Mode Linearization................................ 4-27
4.9 Special Function Setup.................................................. 4-28
4.9.1 Output Signal Reversal.......................................... 4.28
vii
Thermal Conductivity Analyzer
4.9.2 Special - Inverting Output ...................................... 4-29
4.9.3 Special - Polarity Coding ....................................... 4.29
4.9.4 Special - Nonlinear Application Gain Preset.......... 4-29
Maintenance
5.1 Routine Maintenance..................................................... 5-1
5.2 System Self Diagnostic Test........................................... 5-1
5.3 VFD Display .................................................................. 5-2
5.4 Fuse Replacement......................................................... 5-2
5.5 Major Internal Components............................................ 5-3
5.6 Cell, Heater, and/or Thermistor Replacement ................ 5-5
5.6.1 Removing the Cell Compartment........................... 5-5
5.6.2 Removing and Replacing the Cell Block ............... 5-6
5.6.3 Removing the Heater and/or Thermocouple .......... 5-7
5.7 Cleaning ........................................................................ 5-4
5.8 Phone Numbers............................................................. 5-5
Appendix
A-1 Specifications ................................................................ A-1
A-2 Recommended 2-Year Spare Parts List ......................... A-3
A-3 Drawing List................................................................... A-4
A-4 19-Inch Relay Rack Panel Mount................................... A-4
A-5 Calibration Procedure for TG Application........................... A-5
viii
Model 2000A-EU
1-1
Thermal Conductivity Analyzer Introduction 1
Teledyne Analytical Instruments
Introduction
1.1 Overview
The Analytical Instruments Model 2000 Thermal Conductivity Ana-
lyzer is a versatile microprocessor-based instrument for measuring a com-
ponent gas in a background gas, or in a specific mixture of background
gases. 2000A-EU Analyzer complies with all of the requirements of the
Comonwealth of Europe (CE) for Radio Frequency Interference and Elec-
tromagnetic Interfaces (RFI/EMI) protection. It compares the thermal
conductivity of a sample stream with that of a reference gas of known
composition. The 2000 can—
measure the concentration of one gas in a mixture of two gases.
measure the concentration of a gas in a specific mixture of
background gases.
measure the purity of a sample stream containing a single
impurity or a mixture of impurities.
The standard 2000 is preprogrammed with automatic linearization
algorithms for a large number of gases and gas mixtures. The factory can
add to this data base for custom applications, or the sophisticated user can
add his own unique application.
This manual covers the Model 2000A-EU General Purpose flush-
panel and rack-mount units only. These units are for indoor use in a
nonhazardous environment.
Many of the Model 2000 features covered in this manual are optional,
selected according to the customers specific application. Therefore, the user
will find much here that does not apply to his instrument. This is unavoid-
able due to the number of possible combinations of features available. We
have endeavored to make the manual as usable and convenient as possible,
in light of this flexibility.
1-2
1 Introduction Model 2000A-EU
Teledyne Analytical Instruments
1.2 Typical Applications
A few typical applications of the Model 2000 are:
Power Generation
Air liquefaction
Chemical reaction monitoring
Steel manufacturing and heat treating
Petrochemical process control
Quality assurance
Refrigeration and storage
Gas proportioning control.
1.3 Main Features of the Analyzer
The main features of the Model 2000 Thermal Conductivity Analyzer
include:
Three independent, user definable, analysis ranges allow up to
three different gas applications with one concentration range
each, or up to three concentration ranges for a single gas appli-
cation, or any combination.
Special recalibration range for multiple applications. Recalibrat-
ing one, recalibrates all.
Automatic, independent linearization for each range.
Auto Ranging allows analyzer to automatically select the proper
preset range for a given single application. Manual override
allows the user to lock onto a specific range of interest.
RS-232 serial digital port for use with a computer or other
digital communications device.
Two adjustable concentration alarms and a system failure alarm.
Extensive self-diagnostic testing, at startup and on demand.
A 2-line alphanumeric display screen, driven by microprocessor
electronics, that continuously prompts and informs the operator.
High resolution, accurate indication of target or impurity gas
concentration from large, bright, meter readout. (0-9999 ppm
through 0-100 % depending on types of gas involved.)
Standard, proven sensor cell design.
Wide range of custom applications, ranges, and linearization.
1-3
Thermal Conductivity Analyzer Introduction 1
Teledyne Analytical Instruments
Microprocessor based electronics: 8-bit CMOS microprocessor
with 32 kB RAM and 128 kB ROM.
Auto and remote calibration capabilities.
CE Mark Certified.
Four analog outputs: two for measurement (0–1 V dc and
Isolated 4–20 mA dc) and two for range identification.
Compact and versatile design: Small footprint, yet internal
components are accessible.
1.4 Model Designations
The Model 2000A-EU is ordinarily custom programmed at the factory
to fit the customer’s application. Many parameters, including the number of
channels, the gas application, the materials specification of the sampling
system, and others, are options. The most common options, are covered in
this manual. See the Specific Model Information checklist in the front pages
of this manual for those that apply to your Model 2000A analyzer. Some
standard models that are not covered in this manual are listed here.
Models 2000B: NEMA-4, bulkhead mounted enclosure for general
purpose, nonhazardous environments.
Models 2010: Split architecture models using a sealed explosion-proof
enclosure for the Analysis Unit and a general purpose
remote Control Unit for installation in a safe area.
Models 2020: Both the analysis section and control unit are in a single
explosion proof enclosure.
1.5 Front Panel (Operator Interface)
The 2000A is housed in a rugged metal case with all controls and
displays accessible from the front panel. See Figure 1-1. The front panel
has thirteen buttons for operating the analyzer, a digital meter, and an
alphanumeric display. They are described briefly here and in detail in the
Operations chapter of this manual.
Function Keys: Six touch-sensitive membrane switches are used to
change the specific function performed by the
analyzer:
Analyze Perform analysis for target-gas content of a sample
gas.
1-4
1 Introduction Model 2000A-EU
Teledyne Analytical Instruments
System Perform system-related tasks (described in detail in
chapter 4, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the user definable ranges for the instrument.
Data Entry Keys: Six touch-sensitive membrane switches are used to
input data to the instrument via the alphanumeric VFD display:
Left & Right Arrows Select between functions currently
displayed on the VFD screen.
Up & Down Arrows Increment or decrement values of
functions currently displayed.
Enter Moves VFD on to the next screen in a series. If none
remains, returns to the
Analyze
screen.
Figure 1-1: Model 2000A Front Panel
1-5
Thermal Conductivity Analyzer Introduction 1
Teledyne Analytical Instruments
Escape Moves VFD back to the previous screen in a series. If
none remains, returns to the
Analyze
screen.
Digital Meter Display: The meter display is a LED device that
produces large, bright, 7-segment numbers that are legible in any lighting.
It produces a continuous trace readout from 0-9999 ppm or a continuous
percent readout from 1-100 %. It is accurate across all analysis ranges.
Alphanumeric Interface Screen: The VFD screen is an easy-to-use
interface between operator and analyzer. It displays values, options, and
messages that give the operator immediate feedback.
Standby Button: The Standby turns off the display and outputs,
but circuitry is still operating.
CAUTION: The power cable must be unplugged to fully
disconnect power from the instrument. When
chassis is exposed or when access door is open
and power cable is connected, use extra care to
avoid contact with live electrical circuits.
Access Door: For access to the thermal conductivity sensor or the
front panel electronics, the front panel swings open when the latch in the
upper right corner of the panel is pressed all the way in with a narrow
gauge tool. Accessing the main electronics circuit board requires unfasten-
ing rear panel screws and sliding the electronics drawer out of the case.
(See chapter 5.)
CAUTION: The Access door must be closed and latched for
CE mark compliance to be in effect.
1.6 Recognizing Difference Between LCD & VFD
LCD has GREEN background with BLACK characters. VFD has
DARK background with GREEN characters. In the case of VFD - NO
CONTRAST ADJUSTMENT IS NEEDED.
1.7 Rear Panel (Equipment Interface)
All electrical inputs and outputs to the 2000A are made through rear-
panel connectors. The connectors are described briefly here and in detail in
chapter 3, Installation.
1-6
1 Introduction Model 2000A-EU
Teledyne Analytical Instruments
Figure 1-2: Model 2000A-EU Rear Panel
Power Connection Universal AC power source.
9-Pin RS-232 Port Serial digital concentration signal output
and control input.
50-Pin Equipment Interface Port
Analog Outputs 0-1 V dc concentration plus 0-1 V dc
range ID, and isolated 4-20 mA dc
plus 4-20 mA dc range ID.
Alarm Connections 2 concentration alarms and 1 system
alarm.
Remote Valve Used in the 2000 for controlling
external solenoid valves only.
Remote Span/Zero Digital inputs allow external control
of analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.
Range ID Contacts Four separate, dedicated,
range-identification relay contacts
(00, 01, 02, 03).
1-7
Thermal Conductivity Analyzer Introduction 1
Teledyne Analytical Instruments
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 2000 is accurate to 2-3 %. Accordingly, internally sched-
uled calibrations can vary 2-3 % per day.
1.8 Gas Connections
The gas connectors are on the bottom of the Model 2000A chassis
near the front panel. There are no gas control valves inside the main chas-
sis. Electronic input/output ports are provided on the rear panel for the
operation of solenoid valves under the complete control of the Model 2000
electronics. See section 3.3.
A sample system must be provided for introduction of zero and span
gas, as well as sample gas, into the sample path, and for controlling the
flowrates through the sample and reference paths of the analyzer. Appropri-
ate pressure reducing regulators must be installed at all gas supply sources.
Gas Connector-and-Control Panels for specific applications are avail-
able as extra cost additions. These panels are usually designed around a
standard manifold that attaches to the Model 2000 series analyzer below
the front panel.
For those customers wishing to incorporate their own sample controls,
the recommended system piping schematic is included among the drawings
at the rear of the manual.
1-8
1 Introduction Model 2000A-EU
Teledyne Analytical Instruments
2-1
Thermal Conductivity Analyzer Operational Theory 2
Teledyne Analytical Instruments
Operational Theory
2.1 Introduction
The analyzer is composed of two subsystems:
1. Thermal Conductivity Sensor
2. Electronic Signal Processing, Display and Control.
The sensor is a thermal conductivity comparator that continuously
compares the thermal conductivity of the sample gas with that of a refer-
ence gas having a known conductivity.
The electronic signal processing, display and control subsystem
simplifies operation of the analyzer and accurately processes the sampled
data. A microprocessor controls all signal processing, input/output, and
display functions for the analyzer.
2.2 Sensor Theory
For greater clarity, Figure 2-1 presents two different illustrations, (a)
and (b), of the operating principle of the thermal conductivity cell.
2.2.1 Sensor Configuration
The thermal conductivity sensor contains two chambers, one for the
reference gas of known conductivity and one for the sample gas. Each
chamber contains a pair of heated filaments. Depending on its thermal
conductivity, each of the gases conducts a quantity of heat away from the
filaments in its chamber. See Figure 2-1(a).
The resistance of the filaments depends on their temperature. These
filaments are parts of the two legs of a Wheatstone bridge circuit that
unbalances if the resistances of its two legs do not match. See Figure
2-1(b).
2-2
2 Operational Theory Model 2000A-EU
Teledyne Analytical Instruments
Figure 2-1: Thermal Conductivity Cell Operating Principle
If the thermal conductivities of the gases in the two chambers are
different, the Wheatstone bridge circuit unbalances, causing a current to
flow in its detector circuit. The amount of this current can be an indication
of the amount of impurity in the sample gas, or even an indication of the
type of gas, depending on the known properties of the reference and
sample gases.
The temperature of the measuring cell is regulated to within 0.1 °C by
a sophisticated control circuit. Temperature control is precise enough to
compensate for diurnal effects in the output over the operating ranges of
the analyzer. (See Specifications in the Appendix for details.)
2.2.2 Calibration
Because analysis by thermal conductivity is not an absolute measure-
ment, calibration gases of known composition are required to fix the
upper and lower parameters (“zero” and “span”) of the range, or ranges, of
analysis. These gases must be used periodically, to check the accuracy of
the analyzer.
During calibration, the bridge circuit is balanced, with zero gas
against the reference gas, at one end of the measurement range; and it is
sensitized with span gas against the reference gas at the other end of the
measurement range. The resulting electrical signals are processed by the
analyzer electronics to produce a standard 0-1V, or an isolated 4–20 mA
dc, output signal, as described in the next section.
2-3
Thermal Conductivity Analyzer Operational Theory 2
Teledyne Analytical Instruments
2.2.3 Effects of Flowrate and Gas Density
Because the flowrate of the gases in the chambers affects their cooling
of the heated filaments, the flowrate in the chambers must be kept as equal,
constant, and low as possible.
When setting the sample and reference flowrate, note that gases
lighter than air will have an actual flowrate higher than indicated on the
flowmeter, while gases heavier than air will have an actual flowrate lower
than indicated. Due to the wide range of gases that are measured with the
Thermal Conductivity Analyzer, the densities of the gases being handled
may vary considerably.
Then, there are limited applications where the reference gas is in a
sealed chamber and does not flow at all. These effects must be taken in
consideration by the user when setting up an analysis.
2.2.4 Measurement Results
Thermal conductivity measurements are nonspecific by nature. This
fact imposes certain limitations and requirements. If the user intends to
employ the analyzer to detect a specific component in a sample stream, the
sample must be composed of the component of interest and one other gas
(or specific, and constant, mixture of gases) in order for the measured
heat-transfer differences to be nonambiguous.
If, on the other hand, the user is primarily interested in the purity of a
process stream, and does not require specific identification of the impurity,
the analyzer can be used on more complex mixtures.
2.3 Electronics and Signal Processing
The Model 2000 Thermal Conductivity Analyzer uses an 8031 micro-
controller (Central Processing Unit—CPU) with 32 kB of RAM and 128
kB of ROM to control all signal processing, input/output, and display
functions for the analyzer. System power is supplied from a universal
power supply module designed to be compatible with any international
power source. (See Major Internal Components in chapter 5 Maintenance
for the location of the power supply and the main electronic PC boards.)
The Temperature Control board is mounted on the inner face of the
rear panel, under the power input receptacle. The signal processing elec-
tronics including the microprocessor, analog to digital, and digital to
analog converters are located on the Motherboard at the bottom of the case.
2-4
2 Operational Theory Model 2000A-EU
Teledyne Analytical Instruments
The Preamplifier board is mounted on top of the Motherboard as shown in
the figure (in chapter 5). These boards are accessible after removing the
back panel. Figure 2-2 is a block diagram of the Analyzer electronics.
Figure 2-2: Block Diagram of the Model 2000 Electronics
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85