8
1. Introduction
1.1 Battery Health (Deterioration) Judgment
To judge Battery Deterioration, measure and manage the internal resistance of the
battery being installed or new replacement battery periodically.
In general, internal resistance value increased by 1.5 ~ 2 times from initial value
(reference value) when battery is deteriorated
[Judgment of Health (Deterioration) ]
In case Lead Storage Cell (Lead-Acid battery), Warning status starts when
internal resistance value increased by 1.5 from initial internal resistance value.
Deterioration standard is little different by manufacturer of Battery
Even though Same Capacity battery, Initial Internal Resistance value of Battery is
different from Type and manufacturer.
“WARNING”, “FAIL” level of Internal Resistance is little different by manufacturer.
[Recommendation of Battery Replacement Time]
Periodically measure Impedance, Voltages and Temperature of the battery.
In management and save the measurement data which recommends the Time of
Battery Replacement.
[Judgment of Battery Replacement Time]
Based on the specifications of the battery in management when installed,
Trend of changing data from every measurement displayed in graphic.
Right Timing of replacement of Battery may be estimated and displayed on LCD
display backed by accumulated measurement data .
1.2 Description
The Analyzer designed for measuring the internal resistance, open circuit voltage
and terminal temperature of secondary batteries including lead storage cells,
nickel-cadmium batteries, lithium–ion batteries and nickel-metal hydride batteries.