Teledyne 2120XL User manual

Category
Carbon monoxide (CO) detectors
Type
User manual
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS FOR
Model 2120XL
Trace Nitrogen in Argon Analyzer
DANGER
This instrument is for analyzing nitrogen in argon only.
Do not introduce any flammable or toxic gases into this instrument.
Hazardous voltages exist on certain components internally which may be
lethal. Disconnect power before servicing.
Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M84744
11-27-13
Model 2120XL
Teledyne Analytical Instruments ii
Copyright © 2013 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any other language or computer language in
whole or in part, in any form or by any means, whether it be electronic, mechanical,
magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne
Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91748.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from
defects of material and of construction, and that our liability shall be limited to replacing or
repairing at our factory (without charge, except for transportation), or at customer plant at
our option, any material or construction in which defects become apparent within one year
from the date of shipment, except in cases where quotations or acknowledgements provide
for a shorter period. Components manufactured by others bear the warranty of their
manufacturer. This warranty does not cover defects caused by wear, accident, misuse,
neglect or repairs other than those performed by Teledyne or an authorized service center.
We assume no liability for direct or indirect damages of any kind and the purchaser by the
acceptance of the equipment will assume all liability for any damage which may result from
its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus,
and to make any alterations in the dimensions, shape or weight of any parts, in so far as
such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which
valuable data can be gathered. The information provided by the instrument may assist the user
in eliminating potential hazards caused by his process; however, it is essential that all
personnel involved in the use of the instrument or its interface, with the process being
measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of Teledyne at the
time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser
is responsible for the training of personnel, for providing hazard warning methods and
instrumentation per the appropriate standards, and for ensuring that hazard warning devices
and instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept
responsibility for conditions beyond its knowledge and control. No statement expressed or
implied by this document or any information disseminated by the manufacturer or its
agents, is to be construed as a warranty of adequate safety control under the user’s process
conditions.
Trace Nitrogen in Argon Analyzer
Teledyne Analytical Instruments iii
Specific Model Information
This instrument relies on the known spectrum emitted from a
plasma discharge from a distinct gas mixture of variable composition at
or near atmospheric pressure. This instrument cannot be used for
analysis on any gas or gas mixture other than nitrogen in argon or a
mixture specified at the time of purchase. Specific filters carefully
chosen and tested at the factory have been installed for the particular gas
mixture. Using this instrument to analyze any other gas mixture will
result in serious error. Consult the factory for additional information for
gas analysis not specified at the time of purchase.
Instrument Serial Number: _______________________
Instrument Range: _______________
Calibrated for: _______________
Background Gas: _______________
Zero Gas: _______________
Span Gas: _______________
Model 2120XL
Teledyne Analytical Instruments iv
Safety Messages
Your safety and the safety of others is very important. We have
provided many important safety messages in this manual. Please read
these messages carefully.
A safety message alerts you to potential hazards that could hurt you
or others. Each safety message is associated with a safety alert symbol.
These symbols are found in the manual and inside the instrument. The
definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions for
details on the specific danger. These cautions warn
of specific procedures which if not followed could
cause bodily Injury and/or damage the instrument.
CAUTION: HOT SURFACE WARNING: This warning is specific to
heated components within the instrument. Failure to
heed the warning could result in serious burns to
skin and underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous voltages appear
within this instrument. This warning is specific to
an electrical hazard existing at or nearby the
component or procedure under discussion. Failure
to heed this warning could result in injury and/or
death from electrocution.
Technician Symbol: All operations marked with this symbol are to be
performed by qualified maintenance personnel only.
NOTE: Additional information and comments regarding a
specific component or procedure are highlighted in
the form of a note.
No
Symbol
Trace Nitrogen in Argon Analyzer
Teledyne Analytical Instruments v
CAUTION: THE ANALYZER SHOULD ONLY BE USED ONLY
FOR THE PURPOSE AND IN THE MANNER
DESCRIBED IN THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER
THAN THAT FOR WHICH IT WAS INTENDED,
UNPREDICTABLE BEHAVIOR COULD RESULT
POSSIBLY ACCOMPANIED WITH HAZARDOUS
CONSEQUENCES.
This manual provides information designed to guide you through
the installation, calibration operation and maintenance of your new
analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular
application or features and/or options added per customer requests.
Please check the front of this manual for any additional information in
the form of an Addendum which discusses specific information,
procedures, cautions and warnings that may be peculiar to your
instrument.
Manuals do get lost. Additional manuals can be obtained from
Teledyne at the address given in the Appendix. Some of our manuals are
available in electronic form via the internet. Please visit our website at:
www.teledyne-ai.com.
Model 2120XL
Teledyne Analytical Instruments vi
Table of Contents
List of Figures ................................................................ ix
List of Tables .................................................................. x
Introduction ..................................................................... 1
1.1 Overview 1
1.2 Typical Applications 1
1.3 Features 2
1.4 Front Panel 3
1.5 Rear Panel 5
1.6 Internal Components 6
1.7 Additional Safety Information 7
1.7.1 Detector Cautions 8
1.7.2 Basic Safety Requirements 9
1.7.3 Precautionary Labels 9
1.7.4 Summary of Known Hazards 10
1.7.4.1 Electrocution 10
1.7.4.2 Pressure 10
1.7.4.3 Purging 11
1.7.4.4 Safe Repair Procedures 11
1.7.4.5 General Precautions for Handling and Storing
High Pressure Gas Cylinders 11
Operational Theory ....................................................... 13
2.1 The Analyzer 13
2.2 Sample System 14
Installation ..................................................................... 17
3.1 Unpacking the Instrument 17
Trace Nitrogen in Argon Analyzer
Teledyne Analytical Instruments vii
3.2 Choosing a Location 18
3.3 Mounting 18
3.4 Rear Panel Connections 19
3.5 Electrical Connections 20
3.5.1 Primary Input Power 20
3.5.2 50-Pin Equipment Interface Connector 20
3.5.3 RS-232 Port 26
3.6 Gas Connections 27
3.6.1 Typical Sample System 29
3.6.2 Gas Connections to the Instrument 31
3.7 Purging 31
3.8 Calibration 33
3.9 Installation Checklist 33
Operation....................................................................... 35
4.1 Powering Up the Analyzer 36
4.2 Zero and Span Calibration 36
4.3 Setup and Operation General Information 37
4.4 Using the Data Entry and Function Buttons 37
4.5 Analyzer Functions 38
4.5.1 System Menu 39
4.5.1.1 Digital Filter 40
4.5.1.2 Self-Test 41
4.5.1.3 Password Protection 42
4.5.1.4 Logout 46
4.5.1.5 ERR 46
4.5.1.6 Auto Calibration (Auto Cal) 47
4.5.1.7 Flow 49
4.5.1.8 Tracking During Calibration 49
4.5.1.9 Calibration Hold Timer (Calholdtimer) 50
4.5.1.10 The Model Screen 50
4.5.1.11 Analog Output Calibration (Out_Cal) 51
4.5.1.12 Linearize (Lin) 52
Model 2120XL
Teledyne Analytical Instruments viii
4.5.2 The Zero and Span Functions 54
4.5.2.1 Before Calibration 54
4.5.2.2 Recommended Calibration Gases 55
4.5.2.3 Zero Cal 55
4.5.2.4 Span Calibration 57
4.5.3 The Range Function 59
4.5.4 The Alarms Function 60
4.6 The Analyze Function 62
4.7 Signal Output 62
Maintenance .................................................................. 65
5.1 Routine Maintenance 65
5.2 Routine Maintenance Schedule 65
5.3 Fuse Replacement 66
5.4 System Self Diagnostic Test 67
5.5 Error Screens 68
5.6 Sensor Replacement 69
Appendix ....................................................................... 71
A.1 Specifications 71
A.2 Recommended Replacement Parts List 72
A.3 Drawing List 73
Trace Nitrogen in Argon Analyzer
Teledyne Analytical Instruments ix
List of Figures
Figure 1-1 Model 2120XL Front Panel 3
Figure 1-2: Model 2120XL Rear Panel 6
Figure 1-3: Internal Component Identification 7
Figure 3-1: Required Front Door Clearance 19
Figure 3-2: Rear Panel of the Model 2120XL 19
Figure 3-3: Equipment Interface Connector Pin Arrangement 21
Figure 3-4: Remote Valve Connections 25
Figure 3-5: FET Series Resistance 26
Figure 3-6: Suggested Sample System 29
Figure 4-1: Hierarchy of Functions and Sub Functions 39
Figure 5-1: Removing Fuse Block from Housing 66
Figure 5-2: Installing Fuses 67
Model 2120XL
Teledyne Analytical Instruments x
List of Tables
Table 3-1: Analog Output Connections 21
Table 3-2: Alarm Relay Contact Pins 22
Table 3-3: Remote Calibration Connections 23
Table 3-4: Range ID Relay Connections 24
Table 3-5: Commands via RS-232 Input 26
Table 3-6: Required RS-232 Options 27
Table 5-1: Routine Maintenance Schedule 66
Trace Nitrogen in Argon Analyzer
Teledyne Analytical Instruments xi
DANGER
COMBUSTIBLE GAS USAGE
WARNING
This is a general purpose instrument designed for use in a
non-hazardous area. It is the customer's responsibility to
ensure safety especially when combustible gases are
present since the potential of gas leaks always exist.
Never introduce gases other than argon into the analyzer. If
explosive, flammable or corrosive gases or gas mixtures are
allowed to flow into the analyzer, fire or explosion can result.
Sample gas introduced must be at or very close to
atmospheric pressure or damage to the detector will result.
To avoid serious injury, read all precautionary labels
attached to equipment, cylinders, containers, and boxes
prior to start-up.
Labels attached in appropriate areas of the analyzer warn
you of inherent hazards associated with the system. For
personal safety, read the labels and perform directed
precautions before handling the equipment.
The customer should ensure that the principles of operating
this equipment are well understood by the user. Misuse of
this product in any manner, tampering with its components,
or unauthorized substitution of any component may
adversely affect the safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne Analytical Instruments, referred as TAI, no
responsibility by TAI, its affiliates, and agents for damage or
injury from misuse or neglect of this equipment is implied or
assumed.
Model 2120XL
Teledyne Analytical Instruments xii
Trace Nitrogen in Argon Analyzer Introduction
Teledyne Analytical Instruments 1
Introduction
1.1 Overview
This manual describes installation, operation and maintenance for
the Model 2120XL Trace Nitrogen in Argon Gas Analyzer. Section 1
describes the analyzer in general terms and provides additional safety
information pertinent to the proper operation of the instrument.
The Teledyne Model 2120XL Trace Nitrogen in Argon Gas
Analyzer is a robust analytical tool for measuring trace amounts of
nitrogen in argon. Using precise optical filtering, single line emission
characteristic of nitrogen is produced with an intensity proportional to
the nitrogen concentration.
The Model 2120XL has three user programmable analysis ranges
extending from 0-1 ppm to 0-100 ppm with corresponding analog output
signals that are proportional to the concentration on the selected range.
The nitrogen concentration is displayed on the front panel and the
analog output signals are available at the rear panel. Digital
communication is achieved with a standard RS-232 serial port.
The system is easy to operate with all controls and indicators
accessible from the front panel. Digital flow control, two adjustable
concentration alarms, system alarm, self-diagnostics are just a few of the
many features included on the standard Model 2120XL instrument.
1.2 Typical Applications
The Model 2120XL Trace Nitrogen in Argon Gas Analyzer is used
in a wide range of applications including:
Air separation plants
Argon Purification Plants
Specialty Gas Laboratories
Specialty Steel Manufacturing
Gas Management/Monitoring Systems
Introduction Model 2120XL
Teledyne Analytical Instruments 2
Quality Control for Truck Fills & Gas Cylinders
Process Control
New Line Certification
Chemical Plants
Welding Gas Management
Semiconductor Manufacturing
1.3 Features
The Model 2120XL comes equipped with the following standard
features:
A 2-line alphanumeric vacuum fluorescent display (VFD)
screen, driven by microprocessor electronics that
continuously prompts and informs the operator.
High resolution, accurate readings of nitrogen content from
low ppm levels through 100 ppm in argon. Large, bright,
meter readout.
Advanced high frequency plasma generator produces stable
electroluminescent discharge with minimal heating yielding a
spectral emission characteristic of the gas mixture.
Custom low-noise optical filter for precise narrow band
selection and low optical interference.
Amplified optics circuit for stable nitrogen detection at low
ppb level.
Microprocessor based electronics: 8-bit CMOS
microprocessor with 32 kB RAM and 128 kB ROM.
Three user definable output ranges (from 0-1 ppm through 0-
100 ppm) allow best match to users process and equipment.
Auto Ranging allows analyzer to automatically select the
proper preset range for a given measurement. Manual
override allows the user to lock onto a specific range of
interest.
Two adjustable concentration alarms and a system failure
alarm.
Trace Nitrogen in Argon Analyzer Introduction
Teledyne Analytical Instruments 3
Extensive self-diagnostic testing, at startup and on demand,
with continuous power-supply monitoring.
Two way RFI protection.
RS-232 serial digital port for use with a computer or other
digital communication device.
Four analog outputs: two for measurement (0–1 VDC and
isolated 4–20 mA DC) and two for range identification.
Convenient and versatile, steel, flush-panel or rack-
mountable case with slide-out electronics drawer.
1.4 Front Panel
Operator controls and displays are located on the front panel as
shown in Figure 1-1.
Figure 1-1 Model 2120XL Front Panel
The standard 2120XL Nitrogen in Argon Analyzer is panel
mounted designed for easy installation in a standard 19” instrument
rack. All user controls and displays accessible from the front panel.
The front panel has thirteen buttons for operating the analyzer, a
digital meter, and an alphanumeric vacuum fluorescent (VFD) display.
Introduction Model 2120XL
Teledyne Analytical Instruments 4
Function Keys:
Six touch-sensitive membrane switches are used to change the specific
function performed by the analyzer:
Analyze Perform analysis for nitrogen content in an argon
gas mixture.
System Perform system-related tasks (described in detail
in Chapter 4, Operation.).
Span Span calibrate the analyzer.
Zero Zero calibrate the analyzer.
Alarms Set the alarm setpoints and attributes.
Range Set up the 3 user definable ranges for the
instrument.
Data Entry Keys:
Six touch-sensitive membrane switches are used to input data to the
instrument via the alphanumeric VFD display:
◄► Select between functions currently displayed on
the VFD screen.
▲▼ Increment or decrement values of functions
currently displayed.
Enter Advances VFD display to the next screen in a
series or returns to the Analyze screen if none
remain.
Escape Backs VFD display to the previous screen in a
series or returns to the Analyze screen if none
remain also used to abort an entry.
Digital Meter Display:
The meter display is a Light Emitting Diode (LED) device that produces
large, bright, 7-segment numbers that are legible in any lighting. It
produces a continuous readout from 0-10,000 ppm. It is accurate across
all analysis ranges without the discontinuity inherent in analog range
switching.
Alphanumeric Interface Screen:
The VFD screen is an easy-to-use interface from operator to analyzer. It
displays values, options, and messages that give the operator immediate
feedback.
Trace Nitrogen in Argon Analyzer Introduction
Teledyne Analytical Instruments 5
Standby Button:
The Standby turns off the display and outputs, but circuitry is still
operating.
CAUTION: THE POWER CABLE MUST BE UNPLUGGED TO
FULLY DISCONNECT POWER FROM THE
INSTRUMENT. WHEN CHASSIS IS EXPOSED OR
WHEN ACCESS DOOR IS OPEN AND POWER
CABLE IS CONNECTED, USE EXTRA CARE TO
AVOID CONTACT WITH LIVE ELECTRICAL CIRCUITS
.
Access Door:
For access to internal components of the analyzer, the front panel swings
open when the latch in the upper right corner of the panel is pressed all
the way in with a narrow gauge tool. Accessing the main circuit board
requires unfastening rear panel screws and sliding the unit out of the
case.
1.5 Rear Panel
The rear panel, shown in Figure 1-2, contains the gas and electrical
connectors for external inlets and outlets. Some of those depicted are
optional and may not appear on your instrument. The connectors are
described briefly here and in detail in Chapter 3 Installation.
Power Connection Universal AC power source.
Digital I/O Analog Outputs:
0–1 VDC nitrogen concentration
plus 0-1 VDC range ID, and
isolated 4–20 mA DC nitrogen
concentration plus 4-20 mA DC
range ID.
Alarm Signals:
2 concentration alarms and 1
system alarm.
Remote Span/Zero
Digital inputs allow external
control of analyzer calibration.
Introduction Model 2120XL
Teledyne Analytical Instruments 6
Calibration Contact
Notifies external equipment that
instrument is being calibrated and
readings are not monitoring
sample.
Range ID Contacts
Four separate, dedicated, range
relay contacts. Low, Medium,
High, Cal (not used).
RS-232 Port Serial digital concentration signal
output and control input.
Gas Inlet and Outlet One inlet and one vent.
Purge Connections Inlet and outlet connections for
purging the analyzer.
Instrument Air Used for driving optional auto
calibration valves.
Figure 1-2: Model 2120XL Rear Panel
1.6 Internal Components
The internal components can be accessed by removing the top
cover of the analyzer. Figure 1-3 shows an inside view of the analyzer
Trace Nitrogen in Argon Analyzer Introduction
Teledyne Analytical Instruments 7
and identifies specific components of the analyzer. See also Figure 1-4
for internal components inside the card cage.
Figure 1-3: Internal Component Identification
WARNING: HIGH VOLTAGE. ELECTROCUTION HAZARD.
UNPLUG THE ANALYZER BEFORE REMOVING THE
COVER. THE OUTPUT OF THE HIGH-VOLTAGE
TRANSFORMER AND THE ANALYTICAL CELL
ELECTRODES CAN APPROACH VOLTAGES OF
10,000 VAC. THE ANALYZER SHOULD BE SERVICED
ONLY BY A QUALIFIED SERVICE TECHNICIAN.
1.7 Additional Safety Information
Note: The material provided in this section contains information
to promote safety in the operation and maintenance of this
equipment. It is not intended to supersede, replicate, or
replace any safety documentation or procedures provided
from or established by official safety sources.
Introduction Model 2120XL
Teledyne Analytical Instruments 8
Do NOT operate the Model 2120XL Trace Nitrogen in Argon
Gas Analyzer until you read and understand the operating,
maintenance, and safety instructions included in this
manual.
Anyone involved with the operation of this equipment
including plant engineering, operations, and management, must
understand the potential hazards involved, and know and observe all
required safety precautions.
Your safety and the safety of equipment, nearby facilities, and
personnel require a proper safety attitude and emphasis on safe work
procedures. This is the essence of any good safety program. If at any
time you identify safety deficiencies, immediately correct them and
bring them to the attention of management.
Before an accident can be prevented, it must be anticipated. Use
pre-job discussions with your coworkers and supervisors to identify
hazards and the means to avoid them. At your facility, various gases
may exist in liquid and/or gaseous states. Familiarize yourself with the
hazards associated with each gas found at your facility.
Read and understand the Material Safety Data Sheets (MSDS) for
the materials used with this equipment. All personnel who work in the
vicinity of this equipment should read, understand, and follow all safety
information contained in the MSDSs, in addition to following all
government and facility safety regulations.
CAUTION: NEVER INTRODUCE GASES OTHER THAN ARGON
INTO THE ANALYZER. IF EXPLOSIVE, FLAMMABLE,
OR CORROSIVE GASES OR GAS MIXTURES ARE
ALLOWED TO FLOW INTO THE ANALYZER, FIRE OR
EXPLOSION CAN RESULT. THIS ANALYZER IS NOT
DESIGNED TO BE USED IN HAZARDOUS AREAS.
1.7.1 Detector Cautions
The Model 2120XL uses a detection technique based on
spectroscopic emission. The detector is a thin-walled pure quartz cell
located in an electromagnetic field created by a specific high intensity
plasma generator. This electromagnetic field creates a plasma that emits
light at different wavelengths. A filter for the nitrogen is used to avoid
any interference and get the best performance.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85