Invacare Solo2 Quick start guide

  • Hello! I am an AI chatbot trained to assist you with the Invacare Solo2 Quick start guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Operang Your Transportable Oxygen Concentrator
1. ReadyourOperator’sManual.
2. ToturntheSOLO
2
concentratoron,pressandholdthe
On/Obuonunlthethreewarninglightsdisplayand
theunitbeeps.
3. Toadjusttheliterow,pressandholdthe
Mode/Return/Highlightbuon(learrowsymbol)to
scrollthroughthethreemenuscreens.
4. WiththeMode&Flowscreendisplayed,pressandhold
theUp/Increasebuon(plussign)ortheDown/Decrease
buon(minussign)unldesiredsengisachieved.
5. WaitunltheMode&Flowscreenashesthreemes.
Invacare SOLO
2
Quick Start Guide
FromyourHomeCareMedicalRespiratoryTeam
Cleaning the Air Filter
Checktheairlterdailyandcleanitatleastonceaweek.
1. Squeezethethumbtabsontheinletltergrilland
removefromtheunit.
2. Lioutthelter.
3. Useavacuumcleanerorwashwithamildliquiddish
detergent(suchasDawn)andwater.Rinsethoroughly.
4. Thoroughlydrythelterandinspectforfraying,
crumbling,tearsandholes.Replacelterifdamage
isfound.
5. Reinstalltheairintakelterandsnaptheltercover
backinplace.
User Maintenance
Home Care Medical, Inc.
5665SWestridgeDr.Suite100/NewBerlin,WI53151
Ph262.786.9870/800.369.6939
www.homecaremedical.com
Geng on with life... Home Care Medical makes it possible.
Please call Home Care Medical
at 1.800.369.6939 extension 207
with any quesons you may
have related to your SOLO
2
.
Corporate Headquarters
5665 South Westridge Drive Suite 100
New Berlin, WI 53151
Ph 414.423.8800
www.homecaremedical.com
Prolonging Baery Life: Do’s and Don’ts
Do’s
• Fullychargethebaerymoduleovernightwhenyourst
receiveyourtransportableoxygenconcentrator.
• Alwaysensurethebaerymoduleisrechargedassoonas
possibleaeritbecomesfullydischarged.Thebaerymay
permanentlydegradeiflefullydischargedforanextended
lengthofme.
• Heatistheworstenemyofabaery.Allowplentyofairto
circulatearoundthetransportableoxygenconcentratorso
thatthebaeryiskeptascoolaspossible.
Don’ts
• DONOTuseorleavethebaerymoduleinexcessive
heatorcold.
• DONOTstoreorleavethebaerymoduleinthecar.
• DONOTstoreyourbaerymodulefullychargedformore
than2weeks.Storingyourbaerywithafullchargemay
degradeitsusefullife.
• DONOTleaveyourbaerymodulepluggedintothetransportable
oxygenconcentratorwhenthetransportableoxygenconcentrator
isnotinuse.Thebaerywilllosecharge,evenwiththeunito.
Indicators and Audible Alarms
Green Indicator: Unitisoperangnormalandprovidinggood
oxygenpurity.
Yellow Indicator: Onorblinkingwhenthereisawarningornon-
opmalcondionpresent.Thetransportableoxygenconcentrator
couldsllbeoperang,butsomeuseraconmayberequired.
Red Indicator: Onorblinkingwhenthereisanalarmthatrequires
theimmediateaconoftheuser.Alarmstypicallyresultintheunit
stoppingwithnodeliveryofoxygen.
Charging the Baery
1. Tochargethebaerymodule,makesurethatthemodule
isfullyseatedontheunit.
2. Performoneofthefollowing:
A. ConnecttheDCpowercable.
B. ConnecttheACpoweradapter.
3. Turnotheunitforafastchargeme.Chargingmes
areextendediftheunitisoperang.
Note: DCpowerinputmaynotbesucienttochargebaeryifunitisoperang.
Note: Ifthebaeryiscompletelydischarged,itwillfullychargeinapproximately
4to5hours.Duringregularuse,itisrecommendedtorechargethebaery,
evenifonlyparallydischarged,asoenaspossible.
Checking the Baery Charge Level
Theinstalledbaerychargelevelcanbecheckedasfollows
withoutoperangtheunit:
• WhenusingthebaeryortheDCpowercable,momentarily
presstheOn/Obuon.Thestandbyscreenwilldisplayfor
ashortperiodofme.
• WhenusingtheACpoweradapterandtheStandbyscreen
isdisplayed.
Note:Whentheunitisoperang,theStandbyscreenisperiodicallydisplayed.
Baery Duraon
Important Baery Informaon
Flow Rate Seng Connuous Flow Mode Pulse Flow Mode
0.5 4.5hours -------
1.0 3.5hours 4.5hours
2.0 2.5hours 3.5hours
3.0 1.5hours 3.0hours
4.0 ------- 2.5hours
5.0 ------- 2.5hours
/