Safety
2Safety
2.1Generalsafetynotes
WARNING!
Riskofinjuryordamageduetoelectricalshorts
Connectorpinsoncablesconnectedtothepower
modulecanstillbeliveevenwhenthesystem
isoff.
–Cableswithlivepinsshouldbeconnected,
restrainedorcovered(withnon-conductive
materials)sothattheyarenotexposedto
humancontactormaterialsthatcouldcause
electricalshorts.
–Whencableswithlivepinshavetobe
disconnected,forexample,whenremovingthe
buscablefromtheremoteforsafetyreasons,
makesuretorestrainorcoverthepins(with
non-conductivematerials).
Riskofdamagetotheconnectorpins
Ifyoutouchtheconnectorpins,theycanbecome
dirtyortheycanbedamagedbyelectrostatic
discharge.
–Donottouchtheconnectorpins.
Riskofdamagetothemobilitydevice
Therearenouser-serviceablepartsinsideany
case.
–Donotopenordisassembleanycase.
2.2SafetyinformationonIDC
CAUTION!
Riskofinjuryordamage
Malfunctioningorunintendedmovementmay
occuriftheIDCwasexposedtoforcefulimpacts
suchaswallcollisionorfalldamage,orifitis
visiblydamaged.Internalpartsmightbebroken.
–Ifunintendedmovementoccurs,stopusingthe
IDCandpowerwheelchairimmediatelyand
contactaqualiedtechnician.
CAUTION!
Riskofinjuryordamage
Ifclampingleverorhandleandhousingscrews
arettedtootightlyorreplacedbyhardwarenot
approvedbyInvacare,IDCinternalpartsmight
getdamaged.Thiscouldleadtoanunintended
drivesignalandtheIDCcouldstartrunningby
itselfwhenIDCiscontrollingthewheelchair.
–DoNOTtightenhandlescrewsatall.Handle
maygetstuckorblocked.
–Tightenclampingleverandhousingscrews
hand-tightonly(max.6Nm).
–Tightenscrewonmonoblockto1Nm.
–Onlyuseclampingleverandscrewsprovided
byInvacareforxingIDC.
–Testmobilitydevicebeforeuse.
1643411-A5