Steris Cmax X-Ray Image-Guided Surgical Table Operating instructions

Type
Operating instructions
PS00000507
Revision A
CMAX™ X-Ray
Image-Guided Surgical Table
Operator Manual
1 — A Word from STERIS
2
PS00000507_A CMAX™ X-Ray Operator Manual
A Word from STERIS
1.1 Introduction
Thank you for choosing this STERIS product. STERIS is committed to ensuring your continued satisfaction. This manual
contains important information on proper use and maintenance of the STERIS CMAX™ X-Ray table.
IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this equipment can be found in
Section 2, Safety Precautions. Do not operate or service the equipment until you have been trained on this information.
These instructions provide crucial information on how to prevent injuries and property damage that could result from
improper maintenance of the CMAX™ X-Ray table. This manual should be retained in a conveniently accessible area for
quick reference.
Any alteration of the surgical table not authorized or performed by STERIS which could affect operation will void the
warranty, could jeopardize operator and patient safety, could adversely affect efficacy, and could violate national,
federal, state or local regulations.
Intended Use
CMAX™ X-Ray surgical tables are medical devices intended to support and position patients during surgical procedures
carried out in an operating room.
Indications for Use
CMAX™ X-Ray surgical tables have no indication for use, which means that these medical devices are not used to
diagnose, prevent, monitor, treat, alleviate, compensate, replace, modify or control any clinical condition of a patient.
Contraindication – Undesirable Side-Effect – Residual risks
There is no contraindication, undesirable side-effect or residual risks associated with CMAX™ X-Ray surgical tables. All
safety instructions related to the use of these medical devices are listed in Section 2, Safety Precautions.
Patient Target Group
CMAX™ X-Ray surgical tables are not directed to a particular patient population: these medical devices are supposed to
suit any type of patient, regardless of age, anatomy, gender, etc. However, for the safety of the product, weight limits
of patients are indicated in this user manual.
Intended User
CMAX™ X-Ray surgical tables must be used only by qualified personnel who are authorized to practice in the healthcare
facility. These medical devices shall be operated in a controlled environment by fully trained personnel. This product
was developed for surgeons, anesthetists, nurses etc. and is for use in operating rooms, pre- anesthesia rooms, or
recovery rooms.
Essential Performance
CMAX™ X-Ray surgical tables shall support a patient without unwanted movement in a single-fault condition.
1 — A Word from STERIS
3
Operator Manual CMAX™ X-Ray PS00000507_A
Intended Lifetime
The intended lifetime of CMAX™ X-Ray surgical tables is set at 10 years, provided that the conditions of use, the
cleaning and maintenance instructions listed in this user manual are correctly followed.
1.2 Waste Disposal
WARNING
DISPOSAL HAZARD
This product contains materials which may require disposal through appropriately licensed and permitted
hazardous waste management firms.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Use of incorrect hydraulic oil may severely damage the table and/or cause malfunction. Contact STERIS
for proper hydraulic oil.
The following materials may be contained within the table. When disposing of the table or its parts, ensure the proper
disposal of hazardous and other regulated waste in compliance with federal, state and local regulations.
Lead Acid (Pb/H 2SO 4): gelled cell batteries located in the table base at the head-end side of the table. Approximate
weight 21 lbs. (9.5 kg).
Lead (Pb) in Solder: contained in solder on circuit boards and in some miscellaneous wire connections. Minute
amounts.
Electronic and Electrical Parts: not known to require special disposal at date of this manual.
Parts Material: made from aluminum (Al), steel (Fe), cast iron (Fe), copper (Cu), and copper alloys (Cu/x), plastic,
synthetic rubber, plating (Cr, Ni, Zn, Au), Polyvinylchloride (PVC), and adhesives not known to require special disposal
methods at date of this manual.
Hydraulic Oil: contained in hydraulic components located in the base, column, tabletop, and hydraulic system lines and
hoses.
1.3 Service Information
A thorough preventive maintenance program is essential for safe and proper unit operation. This manual contains
maintenance schedules and procedures which should be followed for satisfactory equipment performance.
Only STERIS-trained personnel should attempt to perform maintenance on the table, to avoid personal injury, improper
equipment performance, invalidation of the equipment warranty or other costly damage.
Customers are encouraged to contact STERIS concerning our comprehensive preventive maintenance program. Under
the terms of this program, preventive maintenance, adjustments and replacement of worn parts are provided on a
scheduled basis to help ensure optimal equipment performance and help avoid untimely or costly interruptions. STERIS
maintains a global staff of well equipped, factory-trained technicians to provide these services, as well as expert repair
services. Please contact STERIS for details.
1 — A Word from STERIS
4
PS00000507_A CMAX™ X-Ray Operator Manual
1.4 Contact Information
Manufactured by:
STERIS SAS
116 Avenue de Magudas
33185 Le Haillan
FRANCE
TEL: (+33) 5 56 93 94 94
www.steris.com
Class I Equipment
Type B Applied Part
Splash-Proof Equipment (enclosed equipment protected
against splashing water, IPX4)
Equipment not suitable for use in the presence of a
flammable anesthetic mixed with air or oxygen or nitrous
oxide.
Suitable for intermittent operation 3 min/hr.
The base language of this document is ENGLISH. Any translations must be made from the base
language document.
5
Operator Manual CMAX™ X-Ray PS00000507_A
TABLE OF CONTENTS
Section Number Description Page
1 A Word from STERIS...................................................................................................................................... 2
1.1 Introduction................................................................................................................................. 2
1.2 Waste Disposal ........................................................................................................................... 3
1.3 Service Information .................................................................................................................... 3
1.4 Contact Information ................................................................................................................... 4
2 Safety Precautions ........................................................................................................................................ 6
3 Symbols Used............................................................................................................................................... 14
4 Component Identification ........................................................................................................................... 19
4.1 Overview ................................................................................................................................... 19
5 Operating Instructions ................................................................................................................................ 24
5.1 Operating the Table.................................................................................................................. 24
5.2 Operating the CMAX™ X-Ray Main Hand Control .................................................................. 30
5.3 Joystick Operation..................................................................................................................... 40
5.4 ALS™ (Auto Limit Sensors) ....................................................................................................... 43
5.5 Backup Control (Auxiliary Keypad)........................................................................................... 44
5.6 Examples of Patient Positioning.............................................................................................. 45
5.7 Mounting the Various Adaptable Accessories Onto the Carbon Tabletop ........................... 49
6 Routine Maintenance .................................................................................................................................. 50
6.1 Maintenance General Information ....................................................................................... 50
6.2 Cleaning and Disinfection......................................................................................................... 50
6.3 Maintenance Schedule ............................................................................................................. 54
6.4 CMAX™ X-Ray Spare Parts List ............................................................................................... 55
7 Troubleshooting........................................................................................................................................... 56
7.1 General Troubleshooting Information ..................................................................................... 56
7.2 List of Software Errors ............................................................................................................. 57
7.3 Extension Does Not Lock or Has Too Much Play ................................................................... 57
7.4 The Table Does Not Move........................................................................................................ 58
7.5 Table Remains in Battery Mode When Power Cord is Connected ........................................ 58
7.6 The Table Will Not Switch On .................................................................................................. 58
7.7 The Table Will Not Switch to Unlock Position ........................................................................ 59
8 Electromagnetic Environment.................................................................................................................... 60
9 Commissioning Record ............................................................................................................................... 63
2 — Safety Precautions
6
PS00000507_A CMAX™ X-Ray Operator Manual
Safety Precautions
The following Safety Precautions must be observed when operating or servicing this STERIS equipment. WARNING
indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis,
certain Safety Precautions are repeated throughout the manual. It is important to review ALL Safety Precautions
before operating or servicing the unit.
NOTE: Disregarding this safety information is considered ABNORMAL USE of this product.
NOTE: Any serious incident involving this medical device must be reported to STERIS and the competent authority of
the member state in which the user and/or patient is established.
WARNING
RISK OF INJURY
The modification of this medical device without manufacturer authorization is prohibited.
Pinch Points
WARNING
PINCHING HAZARD
When floating is enabled, moving the floating tabletop can cause pinch points which can injure the patient
or the surgical team.
To avoid serious injury, keep limbs, fingers and other body areas clear of all pinch points when positioning
the table.
When manually positioning table sections, stay clear of potential pinching hazards.
Floor Lock
WARNING
PERSONAL INJURY HAZARD
Tabletop functions are available even if table floor locks are NOT engaged. Lock table before positioning
the patient and starting the surgical procedure.
Do not try to move the table when it is locked.
Make sure the table is stable before engaging the floor locks. If the table is unstable move it to a flatter
surface then engage the floor locks.
2 — Safety Precautions
7
Operator Manual CMAX™ X-Ray PS00000507_A
Patient must be secured to the table in accordance with recommended positioning practices.
Unanticipated patient or table movement could result in death or serious injury
Moving the Table with a Patient Aboard
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
The table can only be unlocked if the floating tabletop is not activated.
The table is intended to be used for moving or adjusting patients in the operating room. It is not to be
used to transport patients outside the operating room. When moving or adjusting the table with a patient
aboard, the following conditions shall be met:
The tabletop must be leveled and centered over the column.
Only lateral slide up to 50mm (2”) is allowed.
The tabletop shall be lowered below 900mm (35”).
The patient shall be in normal orientation and the total weight (patient + accessory) shall not exceed
230kg (507lbs).
The headrest (BF755) may be mounted to the tabletop.
Make sure the battery is properly charged before moving the table.
Before moving the table, ensure its path is clear of obstruction. The table is not intended to cross
threshold or go up or down steps. Always move the table very carefully.
Do not leave objects on the base as they may fall off during motorized manoeuvres and cause a jam
when lowering the column’s sliding covers.
Handling the Surgical Table
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Avoid rough handling of the surgical table during operation. Do not exert excessive lateral loads on it, do
not pull or push the surgical table with the accessories or with the rails.
Handle the remote control and the override system with care. Do not press the keypad with too much
force.
2 — Safety Precautions
8
PS00000507_A CMAX™ X-Ray Operator Manual
Handle the cables with care when connecting or disconnecting the remote control, the joystick, or the
power cord set.
Do not sit on either end of the table, especially when the tabletop is not centered, as this can cause the
table to become unstable and tip.
Do not stand on the base of the surgical table as it can damage the covers.
Pay attention to the power cord when it is connected to the surgical table as it is a tripping hazard. Route
the power cord to the receptacle to minimize the possibility of personnel tripping. Disconnect the cable
before moving the table.
Weight Capacity
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Do not exceed a patient + accessory combined weight of 230 kg (507 lbs) for the standard described
positions and follow the indications and warnings provided.
The use of certain extensions may lower the patient + accessories max weight. When extensions are
installed on the CMAX™ X-Ray surgical table, consult the user manuals for these extensions to find out if
there are weight restrictions related to their use.
Certain positions may cause the tabletop to be significantly off-centered from the column. In these
situations, ensure that there is sufficient stability for the type of surgical procedure requirements.
For all other uses, please contact our network for advice. If you are working intensively on the patient’s
chest, we recommend that you bring the tabletop back to a centered position on the column.
Positioning the Patient
WARNING
RISK OF INJURY
The positions described in this manual are given purely as examples. The accessories required are not
always shown. Under no circumstance should they replace the instructions given by the surgical team in
charge of the patient.
Ensure the stability of the patient and the table when making combined position.
2 — Safety Precautions
9
Operator Manual CMAX™ X-Ray PS00000507_A
Secure the patient properly (ex. using straps), especially when the table is inclined or tilted.
Long surgical procedures can cause tissue compression over bony prominence on patients and result in
tissue ischemia, pressure injury, and post operative problems. Monitor and reposition patients during
surgical procedures, or use patient positioning aids to alleviate areas of high pressure and allow tissue
reperfusion for extended procedures.
Connections
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
The wire terminals of the hand control and joystick are equipped with soft locks that lock into place. Do
not pull on the hand control or Joystick to disconnect them.
The E-Serve connection is only to be used by STERIS service team and with material approved by STERIS.
Never try to connect a device not supplied by STERIS or approved for use with CMAX™ X-Ray to the
surgical table.
Always use the power cord provided by STERIS.
The operator must not make physical contact with the patient while plugging or unplugging the remote
control or joystick, or while handling connectors on these devices.
Potential Equalization Conductor
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Connect together potential equalization conductors to avoid problems with differing potentials across
networks.
Potential equalization conductors protect patients from electric shock during procedures where even
small potentials may cause harm.
When a potential equalization conductor is connected, the table MUST NOT be used with a defibrillator.
When using potential equalization conductors in an electro-medical system, this system must meet IEC
60601-1 requirements.
2 — Safety Precautions
10
PS00000507_A CMAX™ X-Ray Operator Manual
Antistatic Link
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
If antistatic connection is necessary, use only the pads that have been specially developed for this
product. Put the patient in direct contact with these pads and use the table on an antistatic floor, or
connect it to an equipotential cable using the socket designed for it.
Electrical Safety
WARNING
RISK OF PERSONAL INJURY
To prevent any risk of electric shock, the CMAX™ X-Ray should only be connected to a grounded power
supply system.
The CMAX™ X-Ray must always be positioned so that the power cord can be disconnected easily when
connected to the table.
Disconnecting the power cord isolates the CMAX™ X-Ray table from supply mains.
It is preferable to use the batteries instead of the mains if the protective ground in the operating room is
in doubt.
The use of high-frequency surgical equipment, cardiac defibrillators, and cardiac defibrillator monitors in
association with the CMAX™ X-Ray table creates a risk of burns, explosion, or electrical shock.
In case of any damage to the power supply, power cord, plug, or IEC receptacle, do not connect the
equipment and immediately contact STERIS technical support.
Accessories
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
If accessories made by other manufacturers are used, it is the user’s responsibility to ensure they are
compatible with CMAX™ X-Ray and that they do not create a hazard to the patient and the surgical team.
Before connecting accessories to the surgical table, make sure they are not damaged.
2 — Safety Precautions
11
Operator Manual CMAX™ X-Ray PS00000507_A
The end user must ensure that the intended use and the conditions of use (including the safe working
load) of these accessories made by other manufacturers are compatible with the claims of CMAX™ X-Ray
surgical tables.
Emergency Mode
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Use the override system only in emergency mode. When the override system is used, the other control
systems, including the main hand control, are automatically switched off. Since the ALS sensors are
deactivated, ensure you look out for potential section conflicts when moving the tabletop. The pre-
selected patient orientation set as default is the normal orientation; there is no reverse orientation. There
is no immobilization warning LED: you must engage the table locks before moving the tabletop. When
using the override system, always make sure the surgical table is stable and the table locks are engaged
before moving the tabletop, especially when moving the tabletop to greatly off-centered positions.
Cleaning
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Do not use chlorine-based products or other products containing chlorine derivatives; these can cause
excessive corrosion.
Ensure protective covers are properly installed over receptacles to prevent fluids from reaching the
receptacles. After cleaning the table, ensure no residual water remains near the receptacles.
Before cleaning the table, make sure it is switched off and the protective covers are properly installed
over the receptacles.
Do not immerse the hand control or the joystick and do not spray cleaning products directly onto the
receptacles.
Environmental Conditions
Transport / storage: temperature 0°C - 50°C (32°F - 122°F), humidity 10-85% non-condensing, pressure 500-
1060 hPa, table on castors, battery fuses removed.
Use: temperature 10°C-40°C (50°F - 104°F), humidity 10-75% non-condensing, pressure 800 - 1060 hPa (23.6 –
31.3 inHg).
Floor surface punching pressure: 0.13 kg/cm² (13 Pa / 1.85 psi) with a patient + accessories load of 270kg (595
lbs).
Local pressure on the floor: 15.59 kg/cm² (1,529 MPa / 222 psi) with a patient + accessories load of 270kg (595
lbs).
2 — Safety Precautions
12
PS00000507_A CMAX™ X-Ray Operator Manual
WARNING
RISK OF PERSONAL INJURY & EQUIPMENT DAMAGE
Do not use the table in the presence of flammable anesthetic gas, or in areas subject to explosive risks.
CMAX™ X-Ray must not be used in the vicinity of strong magnetic fields (MRI environment).
CMAX™ X-Ray and associated extensions and accessories are not intended to operate in an oxygen rich
environment.
CMAX™ X-Ray and associated extensions and accessories are not intended for outdoor use, and shall not
be exposed to high humidity for an extended time.
X-Ray Radiation
WARNING
RISK OF INJURY
The surgical table's materials, when exposed to X-Ray radiation, can produce and expel harmful radiation
to the environment that is potentially harmful to users. Radiological protection devices should therefore
be worn by users when using X-Ray equipment with CMAX™ X-Ray surgical tables.
Maintenance
WARNING
RISK OF INJURY
The safety, reliability, and proper operation of the surgical table can only be ascertained and granted by
STERIS. Repair and adjustments should only be assigned to STERIS technicians or STERIS-approved
technicians.
Observe the maintenance schedule recommended by STERIS and described in this document.
Never try to adjust or repair the surgical table yourself.
No repair should be undertaken on CMAX™ X-Ray surgical tables until electrical power is switched off,
locked out and tagged out.
During maintenance and repair, only original parts should be used.
No part of the surgical table shall be serviced or maintained while in use with the patient.
2 — Safety Precautions
13
Operator Manual CMAX™ X-Ray PS00000507_A
Hand control storage
CAUTION
EQUIPMENT DAMAGE
Hang the hand control from side rail of the table when not in use to avoid possible damage to the control.
Handle the hand control and the associated cable with care.
Electromagnetic Compatibility
CAUTION
EQUIPMENT DAMAGE
If the table is used close to other electrical equipment, ensure that these comply with the regulations
regarding electromagnetic compatibility.
If high frequency equipment (such as a defibrillator) is being used, make sure to follow the manufacturer’s
instructions carefully.
FCC Part 15: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own expense.
3 — Symbols Used
14
PS00000507_A CMAX™ X-Ray Operator Manual
Symbols Used
Table 3-1. Main Hand Control
Symbol Definition
Floor Locks – Locked / Unlocked
Trendelenburg / Reverse Trendelenburg
Up / Down
Left Tilt / Right Tilt
Return to Level, Origin Position
Patient Orientation – Normal / Reverse
Memory Function
Menu Actions
Menu Selection
Stop All Table Articulation
Table 3-2. TFT Screen Pictograms
Symbol Definition
Floor Lock Engaged
Enable Floating: Tabletop Brake On When Joystick is
Active
Enable Floating: Table Brake on when hand control is
active.
Floor Lock Released
3 — Symbols Used
15
Operator Manual CMAX™ X-Ray PS00000507_A
Table 3-2 TFT Screen Pictograms (Continued)
Symbol Definition
Floating in Progress: Tabletop Free in Transverse +
Longitudinal Direction
Please Wait - Cycle in Progress (e.g. Locking/Unlocking)
Control Interface Active: Joystick / Bluetooth / Hand
Control / E-serve
Bluetooth Joystick Battery Charge Status
Bluetooth Joystick Battery Discharge Status
Alarm Condition – Refer to Troubleshooting Section or
Contact STERIS Technical Department
Table 3-3. Override Control / Joystick
Symbol Definition
Lock / Unlock Tabletop Float
Trendelenburg / Reverse Trendelenburg
Height / Down
Left Tilt / Right Tilt
Select Locked / Unlocked
Powered Backup Control Validation (in Combination with
the Desired Movement)
Start Up Table/Joystick In Bluetooth Mode
3 — Symbols Used
16
PS00000507_A CMAX™ X-Ray Operator Manual
Table 3-3 Override Control / Joystick (Continued)
Symbol Definition
Return to Level
Stop All Table Articulation
Table 3-4. Connection Panel
Symbol Definition
Main Hand Control (Wired)
E-SERVE Serial Port: Reserved for STERIS Use
IMPORTANT: This port is intended for use by STERIS or
STERIS-trained personnel for maintenance procedures only. It
must not be used to move the table during a surgical
procedure.
Joystick (Wired/Bluetooth®)
Accessory port
IMPORTANT: This port must not be used with the CMAX™ X-
Ray or any of its accessories. Do not plug any CMAX™ X-Ray
accessory into it.
Table 3-5. Mobile Base and Packaging Labels
Symbol Definition Symbol Definition
Warning: Read the
Accompanying Documents Hz Frequency Rating of Unit
Attention, Consult Manual
for Further Instructions VA Power Rating of Unit (Volt-
Ampere)
Equipotential
port/outlet/socket
Medical Device
WEEE Directive: the
Product Must Be Recycled
Do Not Wash With
Machine
Name and Address of
Manufacturer
ETL Certification —
Compliance to North
American Safety Standards
3 — Symbols Used
17
Operator Manual CMAX™ X-Ray PS00000507_A
Table 3-5 Mobile Base and Packaging Labels (Continued)
Symbol Definition Symbol Definition
Protection from Electric
Shock Class 1, Type B
Atmosphere Pressure
Limitation
Splash-Proof Equipment Humidity Limit
Date of Manufacture
YYYY/MM/DD
Temperature Limit
/ VDC Direct Current No Stacking for CMAX™ X-
Ray
/ VAC Alternating Current This Way Up
1-4
150830-685 Operator Manual Safety Precautions
CHARGE BATTERY
INTELLI POWER
Symbol Definition
Type B Equipment
Protective Earth Ground
Equipotentiality
Electric Shock Hazard
Control Power ON
Control Power OFF
Powered By AC
Attention, Consult Manual for Further Instructions
AAmperage Rating of the Unit
VVoltage Rating of the Unit
~Alternating Current
Hz Frequency of the Unit
WPower Rating
SNSN
SNSN
SN Serial Number of the Unit
IPX4IPX4
IPX4IPX4
IPX4 Splash-Proof Equipment
Intellipower® Dual Power System Power Panel
(Battery Charge/Discharge Status)
Powered By Battery
Symbols
on Table
Following is a key to symbols on the Cmax Surgical Table.
- OR -- OR -
- OR -- OR -
- OR -
0
BATTERY
Protective Earth (Ground) Fragile, Handle with Care
Product Reference Handle with Care
Serial Number Safety Warning
Batch Number Caution
VVoltage Rating of Unit IPower Supply ON
AAmperage Rating of the
Unit OPower Supply OFF
Ah Energy Unit of the
Batteries (Ampere Hour)
3 — Symbols Used
18
PS00000507_A CMAX™ X-Ray Operator Manual
Table 3-6. Label Definition
Label Definition
Purpose: Indicates a risk of injury from items stored on
the table base
Purpose: Indicates maximum weight capacity
4 — Component Identification
19
Operator Manual CMAX™ X-Ray PS00000507_A
Component Identification
4.1 Overview
4 — Component Identification
20
PS00000507_A CMAX™ X-Ray Operator Manual
1 — Mobile base 7 – 28.5x9.4mm (1.10x0.35”)
Standard accessory rails
13 — Identification label
2 — Column 8 — Connection panel 14 — Battery fuse
3 — Floating tabletop 9 — Carbon tabletop rail accessory 15 — Equipotential plug
4 — Joystick (floating tabletop
control)
10 — Override control 16 — Battery status
5 — Head section interface 11 — Electrical characteristics label 17 — Mains power cable
6 — Main hand control 12 — Mains connection panel
Characteristics
The CMAX™ X-Ray operating table is equipped with a carbon fiber floating tabletop offering exceptional radiological
access.
The CMAX™ X-Ray is suitable for image-guided surgical procedures and has been designed to support patient +
accessory weights of up to 230 kg (507.1 lbs) in multiple position as described in this manual.
The table is controlled by an ergonomic hand control or the joystick in dual-mode (wired/Bluetooth). The innovative
integrated power supply allows the user to choose whether to operate the table on battery or main power. Most table
articulations are powered, whereas the floating tabletop slides manually. E-Serve™ allows for remote diagnostics and
maintenance via computer by approved STERIS technicians (errors, fault log).
The integrated ALS™ linear and angular position sensors prevent collisions between the powered sections of the table
and the floor.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63