PHYWE Cobra4 Mobile-Link 2 Operating Instructions Manual

  • Hello! I am an AI chatbot trained to assist you with the PHYWE Cobra4 Mobile-Link 2 Operating Instructions Manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
Operating instructions
Fig. 1: Cobra4 Mobile-Link 2
Cobra4 Mobile-Link 2
(Cobra4 Mobile-Link 2 incl. accessories)
12620-09
(12620-10)
PHYWE Systeme GmbH & Co. KG
Robert-Bosch-Breite 10
37079 Göttingen
Germany
Phone +49(0)551 604-0
Fax +49(0)551 604-107
Internet www.phywe.com
The unit complies
with the applicable
EC-guidelines
TABLE OF CONTENTS
1 SAFETY PRECAUTIONS
2 PURPOSE AND CHARACTERISTICS
3 FUNCTIONAL AND OPERATING ELEMENTS
3.1 Operating elements on the top
3.2 Sensor interface
3.3 Recording function of the memory card
4 NOTES ON OPERATION
5 MOBILE HANDLING
5.1 Start-up
5.2 Starting the unit
5.3 Measurement data recording process
5.4.1 Sensor/Range (1.1)
5.4.2 Sensor > Setup (1.2)
5.4.3 Datalogger/Setup (2.1)
5.4.4 Datalogger/Files (2.2)
5.4.5 Zero (3)
5.4.6 Calibration (4)
5.4.7 Setup/Date/Time (5.1)
5.4.8 Setup/Battery status (5.2)
5.4.10 Setup/Information (5.4)
5.4.11 Setup/Large Display (5.5)
5.4.12 Utilities/Stopwatch (6.1)
5.4.13 Utilities/Water level (6.2)
5.4.14 Extras/ GPS (6.3)
6 HANDLING VIA A PC
6.1 Start-up
6.2 Starting the unit
7 CHANGING THE BATTERY
8 TECHNICAL DATA
9 PARTS SUPPLIED
10 ACCESSORIES
11 WARRANTY
12 WASTE DISPOSAL
1 SAFETY PRECAUTIONS
- Read the operating instructions thoroughly and com-
pletely prior to starting the unit. This is important for your
own protection and for avoiding damage to the unit.
- Use the unit solely for its intended purpose.
- The unit is intended solely for use in dry rooms where
there is no risk of explosion.
- Protect the device against dust, moisture, and vapours.
Clean the unit with a slightly moist, lint-free cloth. Do not
use any aggressive cleaning agents or solvents.
Attention!
2
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
- Use the experimental set-up solely for the intended pur-
pose.
- Do not open the unit. It is not necessary to replace the
rechargeable battery.
- Based on today's state of software development, it is
impossible to guarantee that a product is completely free
from errors or defects. This is why PHYWE Systeme
GmbH & Co. KG does not accept any liability for damage
that results from the installation or use of this unit.
2 PURPOSE AND CHARACTERISTICS
The Cobra4 Mobile-Link 2 is used for the mobile acquisition
of measurement values that are generated by way of a Co-
bra4 Sensor-Unit.
For this purpose, the unit is equipped with an SD/MMC
memory card. The control menu of the unit enables all rele-
vant adaptations, e.g. the calibration and taring of the unit,
without the need for a PC.
In order to transfer the data to the PC after the completion of
the measurement, the Cobra4 Mobile-Link 2 is equipped with
a USB socket. As an alternative, the memory card can be
removed and the data can be transferred to a PC by way of a
reading device. Ensure that no other Cobra4 device is con-
nected to the PC during the data transfer.
The power supply of the Mobile-Link 2 is ensured by an inte-
grated, rechargeable lithium-ion battery with a charge of
1950mAh. Due to its technology, it exhibits a certain, mini-
mum level of self-discharge. This is why we recommend
checking the charge of the battery prior to using the device,
and to recharge it if necessary.
The battery can be charged via a computer by way of a USB
cable or by way of the supplied power supply unit.
3 FUNCTIONAL AND OPERATING ELEMENTS
3.1 Operating elements on the top
The Cobra4 Mobile-Link 2 is equipped with various elements
for the measurement data display, operation, and status
indication. These elements are explained in detail in the
following sections (Fig. 2).
1. ON/OFF switch
For switching the unit on and off. The process of
switching the unit on or off is confirmed by way of a
sound signal (brief beeping sound). The unit has an
automatic switch-off function in order to prevent the
battery from being discharged: If no measurement is
performed, the unit will be switched off ten minutes af-
ter the last actuation of a button. If the voltage falls be-
low a critical level during the measurement data re-
cording process, an audible alarm will be issued and
the unit will save the measurement values that have
been recorded up to this moment. Then, the unit will
switch off.
2. Display
Colour representation of the measurement values and
selection elements for the menu navigation.
3. LED Battery
For indicating the battery charging process. The LED
lights when a power supply unit or USB cable is con-
nected.
4. Navigation arrows
For selecting a specific measurement
value in the case of multi-channel sen-
sors and for the menu navigation.
5. Confirmation key
For starting and stopping the measurement data re-
cording process and for the menu navigation.
3.2 Sensor interface
The front of the unit is equipped with an interface via which
the connection to a corresponding Cobra4 Sensor-Unit is
realised. This mechanically secure connection is made by
way of lock knobs and holes on/in both units (Fig. 3).
Fig. 3: Cobra4 Mobile-Link 2 front with sensor interface. The
lock knob can be inserted into the hole of the cor-
responding counterpart.
Fig. 2: Cobra4 Mobile-Link 2 operating elements
3
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
3.3 USB interface for the PC
A USB mini-socket is integrated into the sealing cap at the
front of the device (Fig. 4). It is used for the data transfer to
the PC (after the measurement or simultaneously) and for
charging the integrated battery.
3.3 Recording function of the memory card
The recording of measurement values requires a memory
card (e.g. part no. 12620-01; SD/MMC format). Push the card
into the slot in the housing until it locks into place. On the
display, the symbol will be displayed in the header.
In order to unlock the card, press against it and remove it
(Fig. 5).
If no memory has been inserted, a warning sound will be
issued and a warning symbol will be displayed in the header
of the display: .
4 NOTES ON OPERATION
The device fulfils the technical requirements that are summa-
rised in the current directives of the European Community.
The product characteristics entitle the unit to bear the CE
mark. Compliance with the following instructions must be
ensured:
- The unit must be used under the supervision of an ex-
pert and in an electromagnetically controlled environ-
ment at research, teaching, and training facilities
(schools, universities, institutes, and laboratories).
This means that, in such an environment, radio trans-
mission devices, e.g. mobile phones, should not be used
in the direct vicinity of the unit. The connected cables
must not be longer than 2m.
- The measurement data recording process as well as the
general operation of the device can be influenced by ra-
diofrequency fields.
- Electrostatic charges or similar electromagnetic phe-
nomena (HF, bursts, indirect lightning discharge, etc.)
may affect the unit so that it will not work within the
specified data range.
The following measures reduce or eliminate the poten-
tial interferences: Ensure equipotential bonding and use
shielding when using the device in combination with a
PC (in particular with a laptop computer). Do not use
radiofrequency transmitters (e.g. radio sets or mobile
phones) in the direct vicinity of the unit. After a complete
shut-down of the unit, it must be switched off and on
again by way of the ON/OFF switch (operating ele-
ment ). If the unit does not react, keep the ON/OFF
switch pressed for more than 3 seconds so that the unit
will restart.
5 MOBILE HANDLING
This section describes the start-up of the Cobra4 Mobile-
Link 2 and provides an overview of its handling. Please read
this section thoroughly in order to avoid failures or operating
errors.
5.1 Start-up
If the unit is used for the first time, charge the integrated
battery via the USB interface. This can be realised either with
the supplied power supply unit or with the aid of the USB
cable that must be connected to a computer. The charging
process is indicated by the battery LED on the front of the
unit and, if the unit is switched on, in the header of the dis-
play.
In order to record measurement data, the memory card must
be inserted beforehand as described under 3.3.
If the stored data are to be displayed and evaluated on the
PC by way of the measurement evaluation software "meas-
ure", the necessary drivers and programs must be installed
on the PC beforehand via the supplied CD. To do so, insert
the supplied "measure" CD into the drive of your PC and
follow the instructions.
5.2 Starting the unit
The unit is switched on by way of the ON/OFF switch as
described in 3.1.1.
After the unit has been switched on, the start screen will be
displayed. The header of the display includes various sym-
bols that describe the status of the unit.
If no Cobra4 Sensor-Unit is connected to the device, the
Cobra4 Mobile-Link 2 automatically uses the three measuring
channels of the integrated acceleration sensor. This is con-
firmed in the header of the display by way of the word "Ac-
celeration".
In the upper half of the display, i.e. in the main field, one of
the three channels of the acceleration sensor and the associ-
ated measured values are displayed in a large form. The
other two measuring channels are displayed in smaller form
in the field below.
The footer of the display shows the existing navigation op-
tions of the user (see Fig. 6).
Fig. 4: Cobra4 Mobile-Link 2 front with USB interface
Fig. 5: Cobra4 Mobile-Link 2 view from the side with the
slot for the memory card.
4
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
If a sensor is connected, the name of the Cobra4 Sensor-Unit
will be displayed in the header of the display (see Fig. 7).
The navigation keys ▲▼ can be used to select the value that
is to be displayed in a large form in the main field.
The navigation key opens the analogue display.
The navigation keys ▲▼ can be used to select the value that
is to be displayed in the analogue display.
The scaling values that are displayed must be multiplied by
the value that is indicated in the lower left-hand corner. In the
example above: value x1.
The left arrow button can be used to “zoom” the scaling of
the analogue display. In doing so, the values of the display
will be divided in half. Depending on the sensor, this can be
performed up to three times.
5.3 Measurement data recording process
A measurement can be started and stopped by pressing the
confirmation key . The measurement data recording proc-
ess is indicated by a red, flashing dot in the upper right-hand
corner of the main field of the display. The recording sym-
bol in the lower display line turns into stop symbol . The
measurement values are saved to the memory card at the set
measurement interval (Fig. 9).
Fig. 7: Display with a connected Cobra4 Sensor-Unit in the
recording mode of the main display
Fig. 6: Display when no sensor is connected to the unit.
Fig. 8: Display without a connected Cobra4 Sensor-Unit
Fig. 9: Display without a connected Cobra4 Sensor-Unit in
the recording mode of the main display
Electricity
Voltage U / V
0.00
I −0.01 A
Header
Main field
Minor field
Footer
Acceleration
Acceleration ax / g
0.00
ay −0.01 g
az −1.00 g
Acceleration
ax 0.01g
x1
-2
-1
0
1
2
Acceleration
Acceleration ax / g
0.00
ay −0.01 g
az −1.00 g
5
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
During the measurement, the measurement curve can be
displayed and observed in the data display window by way of
the right arrow key (Fig. 9). The left arrow key leads
you back to the main screen. The data display window can
be opened only during a running measurement.
The measuring channel can be selected and viewed sepa-
rately in the data display window by way of the arrow
keys ▲▼.
The right arrow button can be used to zoom the Y-scaling.
In doing so, the values of the Y-axis will be divided in half.
Depending on the sensor, this can be performed up to three
times.
Depending on the selected measuring channel, the associ-
ated graph, which is currently being recorded, is displayed in
yellow in the main field and a red dot flashes in the upper
right-hand corner. The dot disappears when the measure-
ment is stopped with .
In the minor field, the upper value indicates the current
measuring channel . The value for indicates the sam-
pling rate (logging) that can be set in the "Datalogger" menu.
indicates the duration of the measurement and shows
the name of the measurement.
For every series of measurements, a separate file will be
stored on the memory card. The file name consists of the
date and a running counter: YYMMDDxx.mli (xx = 0099).
Ensure that the date and time are set in the Cobra4 Mobile-
Link. This can be checked and, if needed, changed in the
"Setup" menu.
After the completion of the measurement data recording
process, the data can be transferred to the evaluation soft-
ware "measure" where they can be represented and proc-
essed in graphical form. Either the USB interface or a card
reader and the SD card can be used for the data import. The
format of the measurement value file enables it to be im-
ported into a spreadsheet program, e.g. Excel.
If the device is rotated to the left by 90° (i.e. anti-clockwise)
while the data display window is active, the graph will be
enlarged.
Rotating the device to the right (i.e. clockwise), the original
display will be restored as shown in Fig. 9.
5.4 Control menu
The control menu of the Cora4 Mobile-Link 2 is structured as
shown in Fig. 14.
It can be called up by way of the arrow key in the main
screen and it is subdivided into the following blocks (Fig. 12).
Fig. 11: Graph in the landscape format
Fig. 10: Display without a connected Cobra4 Sensor-Unit in
the recording mode of the data display
Fig. 12: Display of the control menu
Acceleration
t/s
Menu
Sensor
Datalogger
Zero
Calibration
Setup
Utilities
Sensor
Acceleration
2
1
0
-1
-2
ax / g
t / s
1 2 4 5 6 7 8 9 10 11
Acceleration
ax −0.01 g
50 ms
00:01:30
13010101 .mli
ax/g
2
1
0
- 1
- 2
6
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
The arrow keys  of the navigator are used for the
navigation within the menu. Selections must be confirmed
with the confirmation key .
The following subchapters explain the various menu items of
the Cobra4 Mobile-Link 2 based on the menu structure in
Fig. 11.
5.4.1 Sensor/Range (1.1)
In the case of sensors with more than one measuring chan-
nel, the measuring channel that is to be used can be selected
by way of ◄► (e.g. in the case of acceleration: , , or
). For every channel, a different measuring range (22,
44 or 88) can be set. Then, the unit (e.g. g or m/s2) can
be specified (Fig. 13). Confirm the settings with the confirma-
tion key . This will lead you back to the "Sensor" submenu.
5.4.2 Sensor > Setup (1.2)
In the case of some sensors, the "Sensor" menu includes the
additional menu item "Setup".
The following section provides an overview of the individual
changes that are possible for these special sensors:
Temperature compensation
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Chemistry (12630-00)
- Cobra4 Sensor-Unit pH (12631-00)
- Cobra4 Sensor-Unit Conductivity+ (12632-00)
It enables the automatic temperature compensation of a pH
or conductivity measuring channel, e.g. T(1), with regard to a
temperature input, or the manual input of a known tempera-
ture with the purpose of compensation (Fig. 15, first and
second point in the menu).
Apart from the manual adjustment of the temperature com-
pensation, it is also possible to enter the temperature of the
solution. As far as the Cobra4 Sensor-Unit pH (12631-00) is
concerned, the manual control can be switched on or
off via a switch and by way of ◄►.
Temperature coefficient
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Conductivity+ (12632-00)
Input of a temperature coefficient for application to a conduc-
tivity measuring channel (Fig. 15, third point).
Fig. 14: Menu structure of the Cobra4 Mobile-Link
Fig. 13: Display of the "Range" menu for the "Accelera-
tion" sensor in the Cobra4 Mobile-Link
Range
Menu
Channel ax
Range -2…2
Unit g
Menu > Sensor
!
Zero
3
Setup
Date/Time
Battery status
Device
Information
Large Display
5
5.1
5.2
5.3
5.4
5.5
Datalogger
Setup
Files
2
2.1
2.2
Calibration
One Point
Two Point
Reset
4
4.1
4.2
4.3
Sensor
Range
Setup
1
1.1
1.2
1
Utilities
Stopwatch
Water level
6
6.1
6.2
GPS
6.3
2
1
Only in case of sensors with setting options
2
Only in case of Mobile-Link GPS (12620-20)
7
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
Kind of Field
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Tesla (12652-00)
It offers the selection of the type of magnetic field, i.e. either a
homogeneous or an alternating field (Fig. 16).
Acoustic Pulses
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Radioactivity (12665-00)
It is used for adjusting an acoustic feedback that will be is-
sued when a certain event occurs. It can be switched
on and off via a switch (Fig. 17).
GM Voltage
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Radioactivity (12665-00)
It is used to change the counter tube voltage in order to
measure the counter tube characteristics.
Gate time
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Radioactivity (12665-00)
It is used to change the counter tube voltage in order to
measure the counter tube characteristics (Fig. 17).
Fig. 15: Display of the submenu "Setup" for the Cobra4 Sensor-
Unit Conductivity+ (12632-00)
Fig. 16: Display of the submenu "Setup" for the Cobra4 Sen-
sor-Unit Tesla (12652-00)
Fig. 17: Display of the submenu "Setup" for the Cobra4 Sen-
sor-Unit Radioactivity (12665-00)
Setup
Menu
Setup
Menu
Setup
Menu
Temp. Comp. Manual
Fluid temp. 23.0 °C
Coeff. 1.00 %/K
Menu > Sensor
Kind of Field
Menu > Sensor
Acoustic Pulse
GM Voltage 100 V
Menu > Sensor
8
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
Mode
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Timer / Counter (12651-00)
The following operating modes of the Timer / Counter can be
selected (Fig. 17):
Measurement of shading times
Measurement of runtimes from the start until the
crossing of the light barrier
Start-stop when crossing the light barrier
Counter mode
Frequency measurement
Wind measurement (with sensor 12124-00)
Start
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Timer / Counter (12651-00)
It is used to select the type of start, i.e. a manual start with ,
an automatic start with a rising signal or with a falling
signal , or with the external triggers or (Fig. 18).
Gate time
This setting option is available for the following Cobra4 Sen-
sor-Units:
- Cobra4 Sensor-Unit Timer / Counter (12651-00)
5.4.3 Datalogger/Setup (2.1)
The following changes are possible:
- Selection of the single-value measurement mode by way
of a switch. This ensures that only one value will be re-
corded when a measurement is started. (2.1.1)
- Selection of the time interval between the individual
measurements and of the unit of measurement (2.1.2
and 2.1.3).
- Toggling between time and the measurement value
number as an X-data set (2.1.4).
5.4.4 Datalogger/Files (2.2)
In the "Datalogger" menu, already existing measurements
can be displayed and deleted (Fig. 20).
Fig. 19: Display of the submenu "Datalogger" for the Cobra4
Mobile-Link 2
Fig. 18: Display of the submenu "Setup" for the Cobra4 Sensor-
Unit Timer / Counter (12651-00)
Fig. 20: Display of the submenu "Datalogger > Files" for the
Cobra4 Mobile-Link 2
Files
Menu
OK
Setup
Menu
Mode Timer
Start
Gate time 1 s
Menu > Sensor
Setup
Menu
Single Values
Logging every 50
Measure Unit ms
X-Axis data Time
Menu > Datalogger
Menu > Datalogger
13010101.MLI
13010501.MLI
13010502.MLI
13010503.MLI
13010504.MLI
Date: 01/01/2013
Sensor: Acceleration
X-axis data Time
9
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
Deleting files:
Select the file with . Delete the file with the arrow key .
A warning asking whether you really want to delete the file
will be displayed. Press the confirmation key in order to
confirm. The arrow key leads you back to the "Files"
menu.
Opening a file and displaying it as a graph:
Select the file with .
Confirm your selection with . The file will be opened and
displayed as a graph in the landscape format.
The graph can be retraced by way of the arrow keys ◄►.
The green crosshairs on the graph indicate the current posi-
tion. The pair of values of the graph is also displayed in
green. When the last measuring point is reached, a new
segment of the graph will be displayed automatically. You
can exit the graph with and return to the file selection
window (Fig. 20). ▲▼ can be used to change the measuring
channel.
5.4.5 Zero (3)
This menu item enables the measurement values of the
measuring channels to be manually reset to zero.
To do so, the switches of the desired channels are set from
off to on by way of the arrow keys ◄►
(Fig. 20).
You can exit the submenu and return to the menu by way
of .
5.4.6 Calibration (4)
In the case of the one-point calibration (4.1), the value that is
currently measured via a channel (actual) is shifted to an-
other value (set) by way of a constant (Fig. 23).
In the case of the two-point calibration (4.2), each of the two
values that are measured is assigned to a new value.
Before the new calibration data are stored in the sensor, an
"attention" message is displayed. Confirming the message
with stores the values, whereas ◄ cancels the storage.
Fig. 21: Representation of a file in the form of a graph
Fig. 22: Display of the submenu "Zero" for the Cobra4 Mobile-
Link 2
Fig. 23: Display of the submenu "Calibration" for the Cobra4
Mobile-Link 2
Zero
Menu
ax 0.00 g
ay 0.34 g
az 0.00 g
Menu
One Point
Menu
Channel G
Ref. -0.1 μS/cm
Actual 0.3 μS/cm
Menu > Calibration
2
1
0
-1
-2
ax / g
t / s
1 2 4 5 6 7 8 9 10 11
Acceleration
0.95 g
0.8 s
10
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
After an unsuccessful calibration, the values can be reset to
the factory settings by way of "Reset" (4.3). An "attention"
symbol will be displayed requiring the user to confirm the
entry.
Note: If the Cobra4 Sensor-Unit Conductivity+ (12632-00) is
used, the menu item under 4.3 is the item Cell constant.
Here, the cell constant of the conductivity sensor that is used
can be entered and saved.
5.4.7 Setup/Date/Time (5.1)
The date and time can be selected by way of the navigation
keys ▲▼ and they can be adjusted with ◄►.
5.4.8 Setup/Battery status (5.2)
The current consumption of the activated Cobra4 Mobile-
Links leads to a continuous decrease of the supply voltage.
This menu can be used to enquire the battery status. In addi-
tion, the total capacity of the battery, the charging time, and
the remaining capacity are displayed (Fig. 24).
If the capacity of the battery falls below a critical threshold
(10% of the remaining battery capacity) the battery symbol
will be displayed in red in the header of the display. Prior to
the automatic shut-down of the device, any running meas-
urements will be stored automatically.
5.4.9 Setup/Device (5. 3)
Selection of the device properties by way of the arrow
keys ▼▲ and confirmation by way of the confirmation key .
The following settings can be realised by way of ◄►:
- Language of the device menu. You can choose between
German , English , and Russian .
- Sound: The sound of the keystrokes of the device can be
switched on or off .
- GPS: (option) In order to localise the device, a satellite
connection can be set up. This function is switched off as
standard. After it has been switched on by way of the
switch, the GPS symbol will be displayed in the
header.
- The automatic rotation of the device display can be
switched on and off .
Use the confirmation key in order to confirm a new setting.
5.4.10 Setup/Information (5.4)
Here, you can find general information concerning your de-
vice, e.g. the number of the firmware version, the number of
devices, and the last update of the Mobile-Link.
5.4.11 Setup/Large Display (5.5)
Here, the settings options for communication with up to two
large-scale display units (07152-93) can be selected if a
Display-Connect (12623-00) is attached.
The uppermost channel of the measurement display will be
transferred automatically to the first large-scale display unit. If
a second large-scale display unit is connected, the second
measuring channel that is to be displayed can be selected for
five radio channels.
5.4.12 Utilities/Stopwatch (6.1)
This feature enables you to record a time interval in the range
of a hundredth of a second.
After the start of the measurement with , it is possible to
pause the measurement with and to resume it afterwards.
For every pause, one measurement value will be saved to a
list. Up to three measuring points can be displayed simulta-
neously. Then, the first one will be overwritten. Pressing
again stops the measurement and reset the stopwatch to
00:00:00.
Pressing while there is no running measurement will de-
lete all of the measurement values.
The various activities are visualised in the footer of the dis-
play.
Fig. 25: Display of the "Stopwatch" utility of the Cobra4 Mo-
bile-Link 2 after the second measurement
Fig. 24: Display of the submenu "Battery status" for the Co-
bra4 Mobile-Link 2
The setup data will be maintained in a bat-
tery-buffered manner for up to 8 years. How-
ever, this requires regular recharging of the
built-in battery.
Battery status
Menu
Stopwatch
10:20:67
00:01:38
10:20:67
00:00:00
Start
Capacity 1905 mAh
Remaining 1543 mAh
Rem. Time 191 min
81 %
Start
Menu > Setup
11
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
5.4.13 Utilities/Water level (6.2)
Here, you can select a water level that can be used to align
experiments horizontally and vertically on a plane or on a
wall.
Alignment on a plane:
Lay the instrument flat on its back and align the surface until
the bubble is located precisely in the centre of the crosshairs.
Horizontal and vertical alignment:
Set up the instrument on its side and hold it against a rail or
similar that is to be aligned. The rail is aligned when the bub-
ble is located between the lines.
In order to tare the water level to 0, press the right arrow
key ► and confirm with or reject with ◄.
5.4.14 Extras/ GPS (6.3)
Here, the geographical coordinates of the current location are
indicated.
The indication is in degrees, minutes, and seconds.
This menu item can only be used in combination with a Mo-
bile-Link 2 GPS (12620-20) and GPS activated.
See also chapter 5.4.9.
Coordinates will not be displayed unless satellites have been
found. In addition, the successful detection of satellites is
indicated by an icon in the header.
Satellites found
No satellites found
6 HANDLING VIA A PC
6.1 Start-up
First, all of the necessary drivers and operating programs that
can be found on the CD must be installed on the PC. This
requires the full version of the "measure" software, part no.
14550-61.
Fig. 26: Display of the "Water level" of the Cobra4 Mobile-
Link 2 on a plane (level).
Fig. 27: Display of the "Water level" of the Cobra4 Mobile-
Link 2 in a vertical/horizontal position.
Fig. 29: Display of the Cobra4 Mobile Link Mobile-Link 2 when
it is connected to a PC via the USB port.
Fig. 28: GPS coordinates in case of detected satellites
Weather 01
Water Level
Water Level
GPS
Menü
Menu > Extras
Latitude:
Longitude:
51°38‘12“
51°38‘12“
12
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
6.2 Starting the unit
Prior to connecting the unit to the PC, attach the desired
Cobra4 Sensor-Unit to it. Following the connection of the
device to the PC via the USB port, the connection is con-
firmed visually on the display of the Cobra4 Mobile-Link 2
(Fig. 29).
The upper area of the display shows the sensor that has
been identified as well as the device number (if several Co-
bra4 devices are connected via the USB interface, they are
numbered). In the upper right-hand corner, a measurement
that is started via the PC is indicated by the recording sym-
bol .
Once the connection has been set up, Cobra4 Mobile-Link 2
and the connected Sensor-Unit are automatically displayed in
the "measure" software together with all of the measuring
channels. All further steps, such as the recording of meas-
urement values or changing of the measurement parameters,
are software-based and, therefore, not described in this
document.
7 CHANGING THE BATTERY
If the capacity of the battery decreases strongly, it is possible
to change the battery. Please note that damage resulting
from a battery change are not covered by the warranty. You
can send the Mobile-Link to the repairs department of
PHYWE in order to have the battery changed.
Step 1: Open the battery cover by pressing a pen or another
object onto the holder of the battery cover.
Attention: Open the left-hand side first, followed by the right-
hand side, in order to ensure that the flat cable connector that
is located on the left of the battery is not damaged.
Step 2: Disconnect the battery connector from the circuit
board. Remove the old battery. Insert the new battery into the
holder.
Step 3: Connect the connector of the new battery.
Attention: Ensure that the cable is located above the USB
connector or, otherwise, the cover cannot be closed.
8 TECHNICAL DATA
typical for 25°C:
Operating temperature range 540°C
Relative humidity < 80%
Device for mobile data acquisition and storage with a USB
interface for the data transfer to a PC
Power supply integrated lithium-ion
battery 1950 mAh
Data memory SD/MMC
Data rate (max.) 2 kHz in combination with
a PC
Dimensions (LxWx H) 155 mm x65 mm x35 mm
Mass 200 g
Battery life up to 6 h
1
2
13
www.phywe.com, © All rights reserved 12620-09 and 12620-10 / 2715
9 PARTS SUPPLIED
Cobra4 Mobile-Link 2 is supplied with the following compo-
nents:
- Operating Instructions
- Lithium-ion battery
In addition, the set “Cobra4 Mobile-Link 2 incl. accessories”
(art. no. 12620-10) contains:
- USB cable
- Power supply unit
- SD memory card
- "measure" CD for the installation of the drivers and op-
erating software for importing the data from the SD
memory card
10 ACCESSORIES
- USB power supply for Cobra4 devices 07932-99
- Replacement battery for Cobra4 Mobile-
Link 2 and Wireless-Link 2 07932-01
The real-time data recording with a PC requires the full ver-
sion of the "measure" software:
- "measure" DVD, single- and multi-user licence 14550-61
Cobra4 Mobile-Link 2 can be used with all instruments from
the Cobra4 product line:
- Cobra4 Sensor-Unit Hall Effect 11801-01
- Cobra4 Sensor-Unit Chemistry:
pH and 2x Temperature NiCrNi 12630-00
- Cobra4 Sensor-Unit pH, BNC connector 12631-00
- Cobra4 Sensor-Unit Conductivity+
Conductivity / Temperature (Pt1000) 12632-00
- Cobra4 Sensor-Unit Conductivity with
stainless steel electrodes 12633-00
- Cobra4 Sensor-Unit Colorimeter 12634-00
- Cobra4 Sensor-Unit Drop Counter 12636-00
- Cobra4 Sensor-Unit Thermodynamics:
Pressure and 2x Temperature NiCrNi 12638-01
- Cobra4 Sensor-Unit Temperature,
Semiconductor −20…+110 °C 12640-00
- Cobra4 Sensor-Unit 2x Temperature, NiCrNi 12641-00
- Cobra4 Sensor-Unit Force ±40 N 12643-00
- Cobra4 Sensor-Unit Electricity:
Current / Voltage 12644-00
- Cobra4 Sensor-Unit Pressure 12647-00
- Cobra4 Sensor-Unit Motion 12649-00
- Cobra4 Sensor-Unit 3D-Acceleration 12650-01
- Cobra4 Sensor-Unit Timer / Counter 12651-00
- Cobra4 Sensor-Unit Tesla 12652-00
- Cobra4 Sensor-Unit Energy 12656-00
- Cobra4 Sensor-Unit Forceplate 12661-00
- Cobra4 Sensor-Unit Radioactivity 12665-00
- Cobra4 Sensor-Unit Sound Level 12669-00
- Cobra4 Sensor-Unit Weather: air pressure,
humidity, temperature, light intensity, elevation 12670-00
- Cobra4 Sensor-Unit CO2 12671-00
- Cobra4 Sensor-Unit Pulse, heart rate 12672-00
- Cobra4 Sensor-Unit Electrophysiology:
ECG, EMG, EOG 12673-00
- Cobra4 Sensor-Unit Spirometry 12675-00
- Cobra4 Sensor-Unit Oxygen 12676-00
- Cobra4 Sensor-Unit Skin Resistance 12677-00
For fixing Cobra4 Mobile-Link 2 in position with a stand rod in
experimental set-ups, we recommend the following acces-
sory:
- Support for Cobra4 with stand rod 12680-00
For table-top set-ups with Cobra4 Mobile-Link 2, we rec-
ommend as accessory:
- Holder for Cobra4 12681-00
We recommend the following accessory for reading out the
SD memory card of the Cobra4 Mobile-Link 2:
- SD card reader 12620-03
11 WARRANTY
We give a warranty of 24 months for units supplied by us
inside the EU, and a warranty of 12 months outside the EU.
The following is excluded from the warranty: Damage that is
due to non-compliance with the operating instructions, im-
proper use, or natural wear.
The manufacturer can only be held liable for the function and
safety-relevant properties of the unit, if the maintenance,
service, and modifications of the unit are performed by the
manufacturer or by an institution that is expressly authorised
by the manufacturer.
12 WASTE DISPOSAL
The packaging mainly consists of environmentally-friendly
materials that should be returned to the local recycling sta-
tions.
PHYWE Systeme GmbH & Co. KG
Customer Service
Robert-Bosch-Breite 10
37079 Göttingen
Germany
Telephone +49 (0) 551 604-274
Fax +49 (0) 551 604-246
Do not dispose of this product with normal
household waste. If this unit needs to be
disposed of, please return it to the address
that is stated below for proper disposal.
/