User Guide
Publication version 1.0
Software version 01-01
2
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Publication information
Publication version Software version Revision date Change description
1.0 01-01 2020-11 First version
y Revision history
Edition notice
This publication is intended for operators of
cobas
®
pure integrated solutions.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version of this publication.
Where to find information
The User Assistance contains all information about the
product, including the following:
• Routine operation
• Maintenance
• Safety
• Troubleshooting information
• Configuration information
• Background information, for example, about analytical
principles
The Safety Guide contains important safety information.
You must read the Safety Guide before operating the
system.
The User Guide focuses on routine operation and
maintenance. The chapters are organized according to
the normal operation workflow.
The Quick Reference Guide gives a brief introduction
to important routine tasks and daily maintenance.
cobas
®
e‑library provides access to important updates,
Method Sheets, Value Sheets, and other important
documents from Roche.
The PC manufacturer’s manual contains all information
about the control unit hardware.
3
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
General attention
To avoid serious or fatal injury, ensure that you are
familiar with the system and safety information before
you use the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the system in a way that is not described
in this publication.
r Store all publications in a safe and easily retrievable
place.
Training
Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave tasks that are not described in the
user documentation to trained Roche Service
representatives.
Screenshots
The screenshots in this publication have been added
exclusively for illustration purposes. Configurable and
variable data, such as tests, results, or path names visible
therein must not be used for laboratory purposes.
Warranty
Any customer modification to the system renders the
warranty or service agreement null and void.
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Copyright
© 202
1, Roche Diagnostics GmbH.
All rights reserved.
License information
cobas
®
pure integrated solutions is protected by
contract law, copyright laws, and international treaties.
cobas
®
pure integrated solutions is licensed for use
between Roche Diagnostics GmbH and a licensee, and
only authorized users are permitted to access and use
the software. Unauthorized use and distribution may
result in civil and criminal penalties.
License agreement for UltraVNC software
UltraVNC is a piece of free software for all commercial
uses, which is installed on the control unit.
You can redistribute the software and/or modify it under
the terms of the GNU General Public License (version 2
or later), as published by the Free Software Foundation.
4
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
A copy of the GNU General Public License (version 2) is
stored on the control unit. The path for the license is C:
\Program Files\uvnc bvba\UltraVNC.
The software is distributed without warranty. There is no
implied warranty of merchantability or fitness for a
particular purpose. For more information, see the GNU
General Public License at http://www.gnu.org/licenses.
The source code for the software is stored on the control
unit. The path for the source code is C:\DriversAndTools
\UltraVNC
Open source and commercial software
cobas
®
pure integrated solutions might contain
components or systems that are open source or
commercial software. For copyright and other notices
and licensing information regarding such software
programs included with cobas
®
pure integrated
solutions please see the following license notifications
and licenses.
The product and cobas
®
pure integrated solutions as a
whole may form a regulated device in accordance with
the applicable laws – please refer to the User Guide and
the labeling for details. Please note that the respective
authorization in accordance with the applicable laws
lapse in case of any unauthorized changes to
cobas
®
pure integrated solutions.
Trademarks
cobas
®
pure is a trademark and the cobas
®
pure logo
is a trademark of Roche.
Other product names and trademarks are the property of
their respective owners.
Feedback
Every effort has been made to ensure that this
publication fulfills the intended purpose. All feedback on
any aspect of this publication is welcome and is
considered during updates. Contact your Roche Service
representative, should you have any such feedback.
Approvals
cobas
®
pure integrated solutions meets the
requirements laid down in:
Directive 98/79/EC of the European Parliament and of the
Council of 27 October 1998 on in vitro diagnostic medical
devices.
5
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Directive 2011/65/EU of the European Parliament and of
the Council of 8 June 2011 on the restriction of the use
of certain hazardous substances in electrical and
electronic equipment.
Directive 2014/53/EU of the European Parliament and of
the Council of 16 April 2014 on the harmonization of the
laws of the Member States relating to the making
available on the market of radio equipment and repealing
Directive 1999/5/EC.
To view the full text of the 2014/53/EU declaration of
conformity, go to the Roche DiaLog Global Web Site
(https://dialog1.roche.com/) and choose the eLabDoc
link.
If you are unable to access Roche DiaLog, contact your
Roche Service representative.
Compliance is provided by means of the Declaration of
Conformity.
The following marks demonstrate compliance:
For in vitro diagnostic use.
Complies with the IVD directive 98/79/EC on in vitro
diagnostic medical devices
Issued by Intertek for Canada and the US.
Instrument approvals
Furthermore, the instrument is manufactured and tested
according to the following international safety standards:
• IEC 61010-1:2010
• IEC 61010-2-101:2015
• UL 61010-1
• UL 61010-2-101
The instrument complies with the emission and immunity
requirements described in standard IEC61326-2-6/
EN61326-2-6.
6
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Contact addresses
Inside the European Union and
EFTA member states
Manufacturer of the
instrument
Hitachi High-Tech Corporation
1-17-1 Toranomon
Minato-ku Tokyo 105-6409
Japan
Authorized
representative
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Japan
Outside the European Union and
EFTA member states
Manufactured by: Hitachi High-Tech Corporation
Manufactured for: Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Distributed in USA by: Roche Diagnostics
9115 Hague Road
Indianapolis, Indiana, USA
7Table of contents
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Table of contents
Publication information 2
Contact addresses 6
Intended use 10
Symbols and abbreviations 10
Safety
1 Safety information
Safety classifications 19
Safety precautions 20
Warning messages 27
Caution messages 38
Notices 40
Product safety labels 43
Safety information for barcode readers 61
Safety interlock 64
Safety information for disposal 67
System overview
2 Overview of the system
About the system 73
Overview of laboratory IT setup 75
About racks and sample containers 80
Sample handling 87
Barcodes on samples 92
3 Overview of the instrument
About the control unit 99
About the sample supply unit 101
About the c303 analytical unit
107
About the e402 analytical unit
124
About rack flow 142
About rack priority 145
Optional components 148
4 Overview of the user interface
About the screen areas 153
About the user interface 162
About the status colors 163
Navigating in the user interface 164
About the Alarms dialog box 165
About reports 167
Taking a screenshot 170
System modes 171
Finding help 174
Searching for QC Value Sheets in the
e‑library 176
5 Overview of reagents and consumables
About reagents and consumables– ISE and
c303
181
About reagents and consumables– e402
187
About expiry of reagents and electrodes 199
Quick reference: Reagents and
consumables– ISE 203
Quick reference: Reagents and
consumables– c303
205
Quick reference: Reagents and
consumables– e402
209
Quick reference: cobaseflow HIV Duo
212
6 Specifications
Specifications– system 215
Specifications – control unit 220
Specifications – sample supply unit 222
Specifications – c303
223
Specifications – e402
228
Operation
7 Before operation
Starting up the system 235
Checking the system performance 239
Checking for system alarms 241
About pre-routine 242
Updating software components 244
Checking the liquid waste containers 245
Performing recommended maintenance 246
Replacing reagents and consumables 248
Downloading required parameters 272
Performing calibration and QC 274
Quick reference: Start up the system 281
Quick reference: Pre-routine workflow 282
Quick reference: Replacing reagents 285
8 During operation
Running patient samples 291
Prioritizing a sample 306
Checking test results 307
Unloading a sample ahead of schedule 313
Unloading racks from the sample supply unit 314
Replacing reagents and consumables during
operation 316
Checking the reagent status 319
Preventive sampling stop - e402
326
Manually masking and unmasking reagents,
tests, and analytical units 327
Performing calibration 333
Performing QC 339
Quick reference: Running patient samples 344
Quick reference: Checking test results 347
Quick reference: Performing calibration 348
Quick reference: Performing QC 350
9 After operation
About stopping the system 355
Stopping operation 356
Stopping operation after the current run 357
8 Table of contents
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Shutting down the system 358
Quick reference: Post-routine workflow 364
10 Non-routine operation
Orders and results 371
Calibration 392
QC 422
Switching off and starting up the system 438
Saving and backing up data 452
Overview of masking functions 466
Maintenance
11 Overview of maintenance
Maintenance definitions 475
About self-operating maintenance 477
About system modes for maintenance 481
Canceling maintenance actions 488
About maintenance pipes and maintenance
pipe functions 489
Creating a maintenance type 495
Checking the maintenance due time 498
Checking the maintenance status 499
Creating a maintenance report 500
List of software maintenance actions 502
List of cleaning solutions 508
List of spare parts and replacement intervals 509
List of maintenance videos 510
Identifying the installed software version 512
12 Maintenance
List of maintenance intervals 515
Daily maintenance 517
Weekly maintenance 523
Every 2 weeks maintenance 525
Monthly maintenance 548
Every 2 months maintenance 565
Every 3 months maintenance 570
Every 6 months maintenance 575
Maintenance as required 580
Troubleshooting
13 System troubleshooting
General troubleshooting information 615
About data alarms and system alarms 617
Data problems without alarm 619
14 Data alarms
List of data alarms 629
Data alarms for tests 633
Calibration data alarms 680
QC data alarms 691
List of data alarms that generate an
automatic rerun test 695
Configuration
15 System settings
Creating a user 703
List of access rights 705
Rack ranges 711
Test keys and test profiles 719
Reagent warning levels 723
Configuring filters for results 725
About non-24/7 operation 726
About the diluted liquid waste container 728
Setting up a host 729
16 Application-related settings
Setting up a test 733
Setting up special reagents, diluents, and
wash solutions 753
Updating an application 755
Deleting an application 759
Analytical principles
17 Measurement technology
About ISE technology 765
About photometric technology 770
About the ECL technology 775
18 Test principles
About test principles - ISE 785
About photometric test principles - c303
787
About sample index principles 803
Overview of immunology principles - e402
807
About limit of detection and limit of
quantitation 814
19 Calibration principles
About calibration of ISE tests 817
About calibration of photometric tests 821
Calibration of immunology tests 846
20 Rules for attaching data alarms
About data alarm rules for photometric tests 857
Index
Glossary
9
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
10
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Intended use
cobaspure integrated solutions is an in-vitro (IVD)
device used for the quantitation of clinical chemistry,
immunochemistry, and ion selective electrolyte
parameters from various biological fluids.
• The system is not intended to be used for the
screening of blood donations.
Intended users
cobaspure integrated solutions is intended to be used
by trained laboratory technicians.
Symbols and abbreviations
Product names
Except where the context clearly indicates otherwise, the
following product names and descriptors are used.
Product name Descriptor
cobaspure integrated solutions
system
cobasc303 analytical unit c303 analytical unit
cobase402 analytical unit e402 analytical unit
cobaseflow cobaseflow test
cobas
®
e‑library
e‑library
cobas
®
e‑services
electronic services
cobas
®
link cobaslink
Sample Cup standard cup
Micro-Sample Cup micro cup
Sarstedt Micro Tube cryo cup
Sarstedt False Bottom Tube Sarstedt FBT
AssayTip tip
AssayCup cup
AssayTip/AssayCup tray tip and cup tray
WasteLiner solid waste box
y Product names
Symbols used in the publication
Symbol Explanation
o
List item
u
Related topics containing further information
y Symbols used in the publication
11
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Symbol Explanation
q
Tip: extra information on correct use or useful
hints
r
Start of a task
I
Extra information within a task
f
Result of an action within a task
c
Frequency of a task
n
Duration of a task
d
Materials that are required for a task
j
Prerequisites of a task
u
Topic (used in cross-references to topics)
p
Task (used in cross-references to tasks)
w
Figure (used in figure titles and cross-
references to figures)
y
Table (used in table titles and cross-references
to tables)
z
Equation (used in cross-references to
equations)
Status button — the blue LED is flashing, the
green LED is lit
Status button — the blue LED is flashing, the
green LED is off
Status button — the blue LED is lit, the green
LED is flashing
Status button — the blue LED is lit, the green
LED is off
Status button — both LEDs are off
y Symbols used in the publication
Home tab: View the home screen of the User
Assistance
Search tab: Find matching topics in the User
Assistance
System explorer tab: Navigate visually through
the system
Interactive help tab: Use the guided, interactive
help to resolve specific problems
Maintenance tab: Perform maintenance
actions
History tab: Show previously viewed topics
again
Favorites tab: Select topics that have been
added to the favorites list
12
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Zoom in button: Enlarge images and videos in
the User Assistance
Play button: Play videos in the User Assistance
Symbols used on products
Symbol Explanation
Global Trade Item Number
Quantity contained in the package
Quantity contained in the package
Orientation of the package during
transportation
Complies with the directive 2011/65/EU and
2015/863/EU on RoHS
y Symbols used on products
Abbreviations
The following abbreviations are used:
Abbreviation Definition
Abs. Absorbance units
ACN Application Code Number
ANSI American National Standards Institute
CDC Closed Development Channel
CFAS Calibrator for automated systems (also
C.f.a.s.)
CSV Comma-separated values
CV Coefficient of variation
EC European Community
ECL Electrochemiluminescence
Eco-D EcoTergent, additive to the incubation
bath water
EFTA European Free Trade Association
EN European standard
EU European Union
y Abbreviations
13
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Abbreviation Definition
IEC International Electrical Commission
IS ISE Internal Standard
ISE Ion-selective electrode
ISEhigh ISEStandardHigh, used as calibrator2
ISElow ISEStandardLow, used as calibrator1
ISES3 ISEStandardHigh, used as calibrator3
according to ISE package insert
ISE-DIL ISE Diluent
IVD In vitro diagnostic
LAS Laboratory automation system
LED Light-emitting diode
LIS Laboratory information system
LLD Liquid level detection
n/a Not applicable
NACL NaCl solution, used as diluent
NaOHD (D1) Wash solution for reagent probes and
reaction cells
OBS Onboard stability
QC Quality control
RCM Reaction calculation mode
REF ISE Reference Electrolyte solution
RFID Radio-frequency identification
SCCS (D3) Additive to prevent reaction cell
carryover in long HbA1c batches
SD Standard deviation
SMS (D2) Selective Mode Solution, a wash
solution for reagent probes and reaction
cells
SSU Sample supply unit
STAT Short turnaround time
SysClean ISE Cleaning Solution/SysClean
WEEE Waste Electrical and Electronic
Equipment
y Abbreviations
14
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Safety
1 Safety information.........................................................................................17
17
Table of contents
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Safety information
In this chapter
1
Safety classifications............................................................. 19
Safety precautions................................................................. 20
About operator qualification................................... 20
About safe and proper use of the system ......... 21
About environmental conditions........................... 22
About installation and deinstallation................... 22
About the protection of personal data and
software security......................................................... 23
Warning messages................................................................ 27
Biohazardous materials ............................................ 27
Electrical safety............................................................ 29
Flammable materials.................................................. 30
Waste............................................................................... 30
Results............................................................................. 31
Caution messages ................................................................. 38
Mechanical safety....................................................... 38
About reagents and working solutions .............. 39
Notices ....................................................................................... 40
Spillage ........................................................................... 40
Tip and cup tray drawer – e402...........................
40
Solid waste drawer – e402.....................................
41
Emergency stop........................................................... 41
Reagents and wash solutions loading
position ........................................................................... 41
Microbeads mixer....................................................... 42
Water container........................................................... 42
Product safety labels............................................................. 43
List of product safety labels.................................... 43
Product safety labels on the sample supply
unit.................................................................................... 45
1 Safety information
18
Table of contents
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Product safety labels on the c303 analytical
unit.................................................................................... 47
Product safety labels on the e402 analytical
unit.................................................................................... 51
Safety information for barcode readers......................... 61
About the laser barcode reader ............................ 61
About the LED barcode readers ........................... 62
Safety interlock ....................................................................... 64
About safety interlock of covers............................ 64
Using maintenance mode - c303........................
65
Safety information for disposal......................................... 67
Disposal information.................................................. 67
1 Safety information
19Safety information
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Safety classifications
The safety precautions and important user notes are
classified according to the ANSIZ535.6-2011 standard.
Familiarize yourself with the following meanings and
icons:
Safety alert
r The safety alert symbol is used to alert you to
potential physical injury hazards. Obey all safety
messages that follow this symbol to avoid possible
damage to the system, injury, or death.
These symbols and signal words are used for specific
hazards:
WARNING!
Warning…
r …indicates a hazardous situation that, if not avoided,
could result in death or serious injury.
CAUTION!
Caution…
r …indicates a hazardous situation that, if not avoided,
could result in minor or moderate injury.
NOTICE!
Notice...
...indicates a hazardous situation that, if not avoided,
may result in damage to the system.
Important information that is not safety relevant is
indicated with the following icon:
i
Tip...
...indicates additional information on correct use
or useful tips.
1 Safety information
20 Safety precautions
Roche Diagnostics
cobas
®
pure integrated solutions · Software version 01-01 · User Guide · Publication version 1.0
Safety precautions
Non-observance of safety precautions
r To avoid serious or fatal injury, read and comply with
the following safety precautions.
The hazard warnings in the user documentation and on
the system cannot cover every possible case, as it is
impossible to predict and evaluate all circumstances
beforehand.
Just following the given directions may, therefore, be
inadequate for operation. Always be alert and use your
common sense.
In this section
About operator qualification(20)
About safe and proper use of the system(21)
About environmental conditions(22)
About installation and deinstallation(22)
About the protection of personal data and software
security(23)
About operator qualification
Insufficient knowledge and skills
As an operator, ensure that you know the relevant safety
precaution guidelines and standards, and the
information and procedures contained in these
instructions.
r Do not carry out operation and maintenance unless a
trainer certified by Roche Diagnostics has trained you
to do so.
r Leave maintenance, installation, or service that is not
described in this document to the trained Roche
Service representatives.
r Carefully follow the procedures specified in the
instructions for operation and maintenance.
r Follow laboratory best practices, especially when
working with biohazardous material.
1 Safety information
/