Roche cobas p 671 Short Guide

Type
Short Guide

This manual is also suitable for

Short Guide
Replacement of the Centrifuge Model
Hettich ROTANTA 46 RSC
with
Hettich ROTANTA 460 Robotic
cobas p 471/ 671
Modification kit ID-No. 01090259090
Roche PVT GmbH
Maybachstrasse 30
D-71332 Waiblingen
Germany
cobas p 471/ 671 Installation on Site
Disclaimer
This manual is a product of Roche PVT GmbH.
All rights are reserved. No part of this manual may be reproduced in any form (print, photocopy,
microfilm or another procedure) or processed using electronic systems, copied or distributed without
the written permission of Roche PVT GmbH. Roche PVT GmbH assumes no liability for any errors,
omissions or damages resulting from the use of the information contained in this manual.
Neither Roche PVT GmbH nor its subsidiaries assume any liability to the purchaser of this product or
any third parties for damages, losses, costs or expenses incurred by the purchaser or third parties due
to accident, misuse of the product or unapproved modifications, repairs or improvements to the
instrument.
Roche PVT GmbH assumes no liability for damages or disruptions resulting from the use of options or
accessories that are not original Roche PVT products or that do not have express approval from
Roche PVT GmbH.
Note: All of the product names contained in this manual are trademarks of the respective companies.
If the trademark symbols © or ® are not included, it can not be assumed that the name is an
unregistered product name.
The content of this manual is subject to change without any communicative obligations
by Roche PVT GmbH.
Copyright by Roche PVT GmbH, Waiblingen 2012
Short Guide – V1.1 – 05/2012 2/8
cobas p 471/ 671 Installation on Site
1. Replacement of the Centrifuge Model
The former Hettich laboratory centrifuge ROTANTA 46 RSC and the current ROTANTA 460 R differ in,
among other things, the way of data transfer. ROTANTA 46 RSC is equipped with an opto data
connector box, which converts data from the centrifuge into serial transmission, whereas the
ROTANTA 460 R includes a RS232 interface.
The following describes how to modify an cobas p 471/ 671 centrifuge system to deploy both
centrifuge variants.
NOTICE
For cobas p 471/ 671 systems, the new ROTANTA 460 R centrifuge and the previous
ROTANTA 46 RSC can be used in mixed mode operation.
1.1 Physical installation of the new Hettich ROTANTA 460 Robotic
CAUTION
Before installing a centrifuge, make sure that the voltage indicated on the
rating plate is the same as the mains voltage.
1.1.1 Centrifuge replacement
Open the maintenance door(s) on the rear of the system and disconnect the cables from the
centrifuges.
Detach the cables from the cable holders.
Remove the centrifuge(s).
WARNING
Crushing hazard: A centrifuge is very heavy. Installation must be carried
out by qualified personnel! Improper handling may cause serious injury.
Do not step under the load and keep fingers and hands away from the
bottom when the centrifuge is lifted and moved.
NOTICE
It is strongly recommended to use the special lifting/ moving hoist which can be ordered from
Roche PVT GmbH quoting the order number 52560000002.
The lifting/ moving hoist enables a single person to safely remove and install a centrifuge. Furthermore,
it serves to transport a centrifuge over even surfaces.
Please refer to the flashcards on how to use the lifting/ moving hoist
'27184-hubwagen_de_flashcards.pdf'
Short Guide – V1.1 – 05/2012 3/8
cobas p 471/ 671 Installation on Site
Centrifuge lifting/ moving hoist
Figure 1
To install the new centrifuge proceed as described in
'Service Manual Æ Installation Æ Unpacking and mounting the centrifuge'.
ROTANTA 460 Robotic
Figure 2
Short Guide – V1.1 – 05/2012 4/8
cobas p 471/ 671 Installation on Site
1.1.2 Cable laying and connections
On the rear of the system, unwrap the spiral guard from the cables and discard the data cable to
the opto coupler.
NOTICE
The opto coppler box, which is no longer used, can be left in the system.
Connect the power supply cable(s) to the left and the RS232 cable(s) to the right.
Feed the cable(s) through the cutout and guide it/ them to the computer.
Connect the serial cable(s) with the RS232 ports on the rear of the computer.
Computer; rear side
Figure 3
Short Guide – V1.1 – 05/2012 5/8
cobas p 471/ 671 Installation on Site
Wrap the spiral guards around the cables and fix them in place with the cable holders provided.
ROTANTA 460 R; rear side
Figure 4
Check that the pullout can move freely.
CAUTION
Make sure that the pullout is not obstructed and
the cables are not pinched.
1.2 Centrifuge addressing
After installation, check the centrifuge(s) for correct addressing.
cobas p 471: Centrifuge = Addr. A
cobas p 671: Left centrifuge 1 = Addr. A, right centrifuge 2 = Addr. B
Please refer to the appropriate Hettich manual; section 'Set the address at the centrifuge'.
Short Guide – V1.1 – 05/2012 6/8
cobas p 471/ 671 Installation on Site
1.3 Software installation
Please follow these steps for software installation:
Power the system and the centrifuge(s) on.
Save the User System ('Administration Æ User Backup').
Insert the 'Modification Centrifuge' CD.
Go to menu item 'Administration Æ Data Installation'.
Press/ click the [OK] button.
Applies to cobas p 471
Select 'Centrifuge is ROTANTA 460 R ' by pressing [1].
Figure 5 cobas p 471 system
Applies to cobas p 671: Selecting Centrifuge 1 and/ or Centrifuge 2
Using the numeric keys, select Centrifuge 1 with [1] and/ or Centrifuge 2 with [2],
depending on which of the centrifuges is/ are to be configured.
Figure 6 cobas p 671 system
Short Guide – V1.1 – 05/2012 7/8
cobas p 471/ 671 Installation on Site
NOTICE
Seen from in front, centrifuge 1 is located to the left, centrifuge 2 to the right.
To abort the installation process, select 'Cancel' by pressing the respective numeric key
and press [ENTER].
Select the item 'Proceed' by pressing the respective numeric key and press [ENTER].
A message of successful configuration will be displayed, prompting you to finish configuration.
Once centrifugation configuration is concluded, remove the CD from the drive and restart the
system.
After changing the centrifuge configuration, the new centrifuge(s) will be displayed on screen during
boot up.
1.3.1 Troubleshooting
If any problems occur during or after the configuration procedure please refer to the below
troubleshooting:
Upgrade aborted
If during the upgrade a message appears, informing you that upgrading has been aborted, check the
following and repeat the installation:
Has the appropriate centrifuge been correctly assigned
(Centrifuge is ROTANTA 46 RSC or 460 R)?
Is the respective centrifuge switched on?
Are all cables properly connected?
NOTICE
If other problems occur during configuration, make a note of the message displayed before you
terminate the installation and consult Roche PVT Software Support.
1.4 Inspection log book for Hettich laboratory centrifuges
Please work through the original Hettich inspection log book for the appropriate centrifuge.
1.5 Test run
Perform a test run according to the cobas p 471/ 671 Service Manual/ Checklist.
CAUTION
The system must not be put into routine operation without
successful test run.
Short Guide – V1.1 – 05/2012 8/8