CEM PETwave Operating instructions

Type
Operating instructions
PETwaveTM
Microwave-Assisted Radiosynthesis
Manual
Contents
Instrument Overview
PETwave 1
Microwave 1
Stirring 1
Temperature 1
Cooling 1
Pressure 1
Safety instructions
Safety Notations 2
Operating Precautions 2
Instrument Parts & Accessories
Standard Parts 3
Optional Parts 4
Instrument Components
Microwave Controller 5
Remote Microwave Cavity 6
Instrument Installation
Location 7
Unpacking 8
Attach Cables/tubing 8
Turn instrument On/Off 10
System Verication 10
Instrument Operation
PETwave display and Keypad with onboard software 11
Synergy Software (Optional) 12
Synergy Software Connection (Optional)
Download Synergy Software 14
IP Address 14
Synergy 18
Perform a Reaction
Prepare Sample 19
Microwave 20
Remove Sample 21
General Settings
Integrated PETwave Software 22
Synergy 23
Instrument Calibration and Verification
Temperature Calibration 28
Temperature Verication 29
Stirring Verication 29
Methods
Method Types 31
Integrated PETwave software 32
Synergy Software 37
Data Review
Integrated PETwave software 41
Synergy Software 41
Routine Maintenance & Cleaning
Infrared Sensor 43
Microwave cable 43
Service Utilities
Firmware Update 44
PETwave Information 45
Technical Data
Dimensions (Overall) 46
Troubleshooting
Troubleshooting 48
Warranty
Limited Warranty Information 49
Technical Assistance
Requested Information 50
Contact Information 50
Instrument overview / 1
Instrument Overview
PETwave
The CEM PETwave™ is designed to enhance the ability to perform chemical reactions under controlled conditions
on a small scale inside a ‘hot’ cell or glove box. The system facilitates solution phase chemistry, solvent
evaporation to concentrate reactants and hydrolysis to remove protecting groups. It accommodates conical shaped
vessels ranging in working volume from 50 µL to 2.5 mL. Primary uses of the PETwave™ are for radiolabeling
chemistry in a ‘hot’ cell or chemistry in the controlled environment of a glove box.
Microwave
Microwave energy is applied to the vessel contents (reactants, catalysts, salts, solvents) to accelerate the
chemical reaction. The microwave absorption properties of some liquid and solid materials, due to their polar and
ionic characteristics, have the capability to signicantly enhance chemical reactions relative to traditional energy
application (heating) techniques. The microwave interaction properties with the reactants, intermediates, catalysts,
and salts provide unique opportunities for the synthetic chemist. Power is programmable in 1 watt increments from
0 to 100 watts.
Stirring
Stirring is accomplished using electromagnetic stirring plates located beneath the cavity in conjunction with stir
bars placed in the reaction vessel; ensuring a homogeneous sample. The stir bars must be Teon® coated iron
oxide bars. When placed in the microwave eld, the iron oxide bars will reect the microwave energy without
damaging the instrument. Stirring can be programed as off, low medium or high.
WARNING
If the instrument is suspected of interfering with the operation of a pacemaker, the instrument should be
turned off or the pacemaker wearer should move away from the instrument.
Temperature
Temperature is measured by an Infrared Control System. This temperature control system uses a non-contact,
infrared sensor located below the microwave cavity oor. The sensor is vessel volume independent and is used in
a feedback loop with the on-board computer to control the temperature rise rate and control point of the vessel
contents. Temperature is programmable from room temperature to 300oC.
Cooling
Cooling – The cooling feature consists of necessary solenoid valves and ports to direct a cooling gas (either
nitrogen or “clean” air) onto the outside wall of the reaction vessel. This provides the ability to rapidly cool
(quench) a reaction after the application of microwave energy. Cooling is programmable as on or off and may be
adjusted using the regulator up to 60 psi.
Pressure
Reactions are performed at or below boiling point of the solvent; therefore, reactions may only performed at
ambient/atmospheric pressure.
2 / Safety instructions
Safety instructions
Safety Notations
Warnings, cautions and notes are included throughout this manual and should be read thoroughly and strictly
followed. The following conventions for safety messages are used in this Operations Guide:
WARNING
A warning is inserted for essential information used to emphasize dangerous or hazardous conditions to
the operation, cleaning and maintenance of the instrument which may result in personal injury.
CAUTION
A caution is inserted for essential information use to emphasize procedures which, if not strictly followed,
may result in damage or destruction to the instrument or improper instrument operation.
NOTE
A note is inserted for emphasis of procedures or conditions that may otherwise be misinterpreted or
overlooked and to clarify possible confusing situations.
Operating Precautions
The PETwave™ system must be grounded. In the event of an electrical short circuit, grounding reduces the risk
of electric shock by providing an escape wire for electric current. This instrument is equipped with a cord having
a grounding wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and
grounded. Consult a qualied electrician or service technician if the grounding instructions are not completely
understood or if doubt exists as to whether the instrument is properly grounded. If it is necessary to use an
extension cord, use only a 3-wire extension cord that has a 3-blade grounding plug and a 3-slot receptacle that will
accept the plug from the instrument. The marked rating of the extension cord must be equal to or greater than the
electrical rating of the instrument.
The possibility of instrument-induced electromagnetic interference (EMI) is minimal if the instrument is operated as
outlined in this manual. The instrument should not be placed close to any electrical device susceptible to EMI.
If the instrument is suspected of interfering with the operation of a cardiac pacemaker, the instrument should be
turned off or the pacemaker wearer should move away from the instrument.
This instrument utilizes high voltages and microwave radiation. Only those trained in repair and maintenance of
high voltage and microwave power systems should perform instrument service and repair.
Use of the PETwave in any manner not specied by CEM Corporation could render the instrument operation unsafe
for the operator.
Instrument parts & accessories / 3
Instrument Parts & Accessories
• Retain all packing material for use if returning the system to the manufacturer, subsidiary or distributor for
service.
• If a part is missing, contact your CEM representative.
• Inspect the instrument for shipping damage such as cracks, dents or warping.
• If the instrument has been damaged in shipping, contact the freight carrier to report the damage and to le a
damage report. Contact the CEM Service Department or the local subsidiary or distributor to report damage and
to request service information.
WARNING
If damage to the instrument is noted upon receipt, do not attempt to operate the instrument.
Standard Parts
Image Qty Item Part Number
1 Microwave Controller with Microwave transmission cable connected N/A
1 Remote Microwave Cavity N/A
2 4 mL vial (50 µL-0.5 mL working volume) 186400
2 2 mL vial (100 µL-1.0 mL working volume) 186410
2 1 mL vial (250 µL-2.5 mL working volume) 186405
1
Multi-port Cap (7 ports for reagent addition, vacuum drying, gas
purging, and sample removal)
bottom piece
Silicone o-ring
Top head
186420
186435
186425
4 / Instrument parts & accessories
20 Cap SP-1547
1Regulator
1/4" brass tting adapter
541455-M
BR163380
1
Cooling Line assembly
Tubing, PFA, 3/16"
Nut 3/16"
Ferrule, 3/16”
BR199117
167315
167320
1 Temperature Cable, 8 pin N/A
1 Stirring Cable, 6 pin N/A
1
Power cord
USA (120 V)
Export (230 V)
119635
119645
1 Operations Guide 600864
10' Tubing 1/4" BR22122
Optional Parts
Item Description Part Number
Synergy Software Option
The Synergy Software Option is a PC-based software package that allows the
user to program, monitor, and control all PETwave system functions. The option
includes a laptop computer with Synergy software loaded and both Ethernet
and serial cables to communicate directly to the PETwave. The package creates,
and allows for the full management of a database for all system reaction data.
When placed on a network, the current instrument status is visible from a
remote location.
909105
Synergy Laptop Option The laptop option includes the Synergy software along with a laptop computer . 909110
Instrument components / 5
Instrument Components
Microwave Controller
Front
Display - A 256 x 128 graphics display and on-board computer for programming and operational control of
the system. The memory will store and recall methods.
Keypad - allows the operator to create, edit, store and recall methods, start and stop methods and change
programmed parameters during operation.
Right
USB Ports (2) - Permit installation of USB memory stick to import, export, backup, and/or save instrument data.
Ethernet port - The PETwave does not utilize this feature.
7 Pin Connector - The PETwave does not utilize this feature.
Rear
Fuses/Fuse holder - Prevent electrical power overload.
Power inlet - Receives the female end of the power cord.
Magnetron Cooling Fan - Draws room air past the magnetron for cooling purposes.
Power Supply Cooling Fan - Draws room air past the power supply for cooling purposes.
Electrical Connector - The PETwave does not utilize this feature.
RS232 Port - Allows additional connection and communication.
Ethernet Port (2) - Allows communication and connection to an external computer for data collection, a local
area network (LAN) or the Internet. (Optional).
Nameplate - Lists the model, serial number, Product Tag, operating voltage, frequency and current of the
instrument.
6 pin Stirrer Connector - Connects the electromagnetic plate located in the Remote Microwave Cavity to the
control board located in the Microwave Controller.
8 pin Temperature Connector - Connects the infrared temperature sensor located in the bottom of the Remote
Microwave Cavity to the Microwave Controller.
Cooling Gas Line - Provides a hose connection for the cooling gas source.
6 / Instrument components
Left
Power switch - Turns AC power on and off to the instrument.
Cooling Gas Line – Provides a hose connection for the cooling gas source. The cooling feature consists of
necessary solenoid valves and ports to direct a cooling gas (either nitrogen or “clean” air) onto the vessel in the
system cavity.
Microwave Transmission Cable - Connects the microwave generation component within the Microwave Controller
to the Remote Microwave Cavity. This is the larger blue cable extending from the left side of the microwave cavity.
Use caution when bending/moving this cable.
Remote Microwave Cavity
Microwave Transmission Port - Connects the Remote Microwave Cavity to the Microwave Controller. Safety
interlock and an interlock monitoring system to prevent microwave emission when the microwave transmission
cable is not properly installed.
6 Pin Stirrer Port - Connects the electromagnetic plate located in the Remote Microwave Cavity to the control
board located in the Microwave Controller.
8 Pin Temperature Port – Connects the infrared temperature sensor located in the bottom of the Remote
Microwave Cavity to the Microwave Controller.
Cooling Inlet – Supplies compressed gas from Microwave Controller to the Remote Microwave Cavity.
Multi-port Cap Holder Pin – Keeps multi-port cap in position during reaction.
Instrument installation / 7
Instrument Installation
WARNING
If damage to the instrument is noted upon receipt, do not attempt to operate the instrument.
This instrument utilizes high voltages and microwave radiation. Only technicians trained in repair and
maintenance of high voltage and microwave power systems should perform instrument service and repair.
Never modify the attenuator access port or insert metallic objects such as wire into the port. Serious
microwave leakage and/or electrical shock may result. The access port of the PETwave is electrically
grounded to the microwave cavity and is designed to prevent leakage of microwave energy.
Cardiac pacemakers require magnets to control operation during checkout. Because the PETwave is
equipped with a variable-speed, electromagnetic sample stirrer, some danger exists if a pacemaker is
positioned in close proximity to the instrument cavity. If the instrument is suspected of interfering with the
operation of a pacemaker, the instrument should be turned off or the pacemaker wearer should move away
from the instrument.
Disconnect the instrument from the AC power source prior to performing any service procedure.
CAUTION
Before connecting power cord to PETwave, ensure proper line voltage is available (120V vs 240V). If large
voltage uctuations are expected, a constant voltage source (UPS) is required.
Location
• Choose a location for the Microwave Controller that is in close proximity, less than 10 feet, to the ‘hot’ cell
or glove box.
• Provides enough space for Microwave Controller: At least 8 in. (20 cm) of open space on each side and 6 in. (15
cm) of open space in the rear of the instrument for ventilation. The space should be at least 31 in. (77 cm) wide
by 23 in. (59 cm) deep with a height clearance of at least 10 in. (89 cm).
• Provides enough space in hot cell for Remote Microwave Cavity: At least 10 in. (25 cm) wide by 10 in. (25 cm)
deep with a height clearance of at least 15 in (38.1 cm).
• Is free from vibration of large equipment and/or excessive walk-through trafc.
• Provides a temperature range of 41 °F (5 °C) to 104 °F (40 °C) and a humidity range of ±10-85%
relative humidity.
• Provides adequate space for sample handling (and computer placement if applicable), and
• Permits the system to be connected to a dedicated, grounded 120 or 240 VAC outlet. The PETwave System
should be operated on a stabilized, constant voltage AC power supply.
8 / Instrument installation
Unpacking
1. Verify all parts and accessories are available. View “Instrument Parts and Accessories” for the list.
2. Carefully remove the Microwave Controller from its shipping carton and place on an appropriate location near
the ‘hot’ cell or glove box.
CAUTION
Do NOT use force to bend microwave transmission cable (blue cable extending from left side of
instrument). Extreme bending of the microwave transmission cable will render the PETwave inoperative.
3. Place the Remote Microwave Cavity inside the hot cell or glove box.
Attach Cables/tubing
1. Insert the power cord into the power inlet receptacle located on the back of the instrument and the opposite
end into the electrical outlet.
2. Connect the 8 pin temperature cable and 6 pin stirrer cable to the Microwave Controller.
3. Connect the cooling line tube to the tube extending from the back of the Microwave Controller.
Instrument installation / 9
4. Connect the cooling line tube to the Remote Microwave Cavity.
5. Connect the 8 pin temperature cable to the Remote Microwave Cavity.
6. Connect the 6 pin stirrer cable to the Remote Microwave Cavity.
7. Install blue microwave transmission cable extending from the left side of the Microwave Controller to the Re-
mote Microwave Cavity. Turn nut until nger tight.
CAUTION
Do NOT use force to bend microwave transmission cable. Extreme bending of the microwave transmission
cable will render the PETwave inoperative.
Over tightening the nut can damage the microwave transmission cable. Under tightening the microwave
transmission cable will cause a “door open” error on the Microwave Controller.
8. Connect the pressure regulator to either a nitrogen or clean air supply.
8.1. Cut the ¼" tubing in half.
8.2. Install one length of the tubing into the left side of the Microwave Controller by pushing in on the
tubing until resistance is felt.
8.3. Place the opposite end of the tubing into the regulator.
10 / Instrument installation
NOTE
The regulator has an arrow printed on it to show the direction that the air must ow. Set up the regulator
according to the arrow.
8.4. Place the second length of tubing into the opposite end of the regulator.
8.5. Connect the opposite end of the tubing to the air source.
8.6. To change the amount of air ow from the regulator
8.6.1. Pull the knob out.
8.6.2. Rotate the knob on the regulator clockwise to increase the amount of air ow.
8.6.3. Rotate the knob on the regulator counter clockwise to decrease the amount of air ow.
8.6.4. The minimum compressed gas requirement is 25 psi (20 L/min ow).
8.6.5. The maximum ow rate is 60 psi.
Turn instrument On/Off
1. Position Microwave Controller on bench top and Remote Microwave Cavity in hot cell or glove bag.
2. Turn the instrument on using the power switch on the left side of the Microwave Controller. The software will
take up to a minute to load. While in the reaction programming screen, If a round bottom ask does not appear
in the lower right corner of the software, then the blue microwave transmission cable has not been securely
fastened and a “Door Open” message will appear. See section “Attach cables/tubing” for instructions on how
to install the blue microwave transmission cable.
3. To switch the instrument off, use the power switch. The PETwave system can be turned off or left on when not
in use. The system is ready to run once the software has completely loaded.
System Verification
See section “Instrument Calibration and Verication.
Instrument operation / 11
Instrument Operation
There are two interfaces for operating the PETwave; the integrated software using the keypad and display or
Synergy software. Synergy is an optional data management software.
PETwave display and Keypad with onboard software
Keypad
HOME
STIRRING
POWER
START/PAUSE
COOLING
ENTER DISCOVER
TEMPERATURE
+/- STOP
TIME
PRESSURE
ARROWS
HOME – Press this key to return to the previous menu or home screen.
START/PAUSE – Press to start the current method or pause during method operation.
ENTER – Press to accept entries while editing or creating a method or in System Setup.
DISCOVER – Press to load, save and create methods and review data.
STOP – Press to stop any process. Pressing this key when performing a method will stop microwave radiation
and launch the cooling function. Pressing this key during the cooling cycle will abort the cooling cycle and return
the system to the home screen.
ARROW – Press to navigate through method programming or System Setup steps.
+/- – Press to increase or decrease highlighted method parameters or to navigate through System Setup steps.
POWER – Press this key to change the maximum applied microwave power setting prior to or during a method.
If this key is pressed during a method, this does not stop or pause the method but updates the setting in real
time.
STIRRING – Press this key to change the stirring setting prior to or during a method. Use the <+/-> keys to
toggle to the appropriate setting (High, Medium, Low, Off). If this key is pressed during a method, this does not
stop or pause the method but permits changing the setting in real time.
COOLING – Press this during a method to perform the cooling function.
TEMPERATURE - Press this key to change the temperature setpoint before or during a method. Use the <+/->
keys to increase or decrease the setting. If this key is pressed during a method, this does not stop or pause a
method but updates the setting in real time.
PRESSURE – The PETwave does not utilize this feature.
TIME – Press this key to change the time value during or after a method. Use the <+/-> keys to increase or
decrease the time setting. Pressing this key allows for real time changes to this value.
12 / Instrument operation
Display
Shows menus, method parameters and instrument status on a 256 x 128 graphics display.
Rxn Programming Discover Home Load, Save, Create Method
Synergy Software (Optional)
The Synergy Software is a PC-based software package that allows the user to program, monitor, and control all
PETwave system functions. The package creates and allows for the full management of a database for all system
reaction data.
Menu Bar
File
New Folder - The “New Folder” function allows a user to create a new folder to store methods under. To create
a new folder, select “New Folder” from the “File” drop-down menu, enter the name you wish to use for the new
folder, and then click “OK”. Click “Cancel” to exit without creating a new folder. A new folder can only be created
within an existing folder.
Rename Folder - The “Rename Folder” function allows the user to change the name of the currently highlighted
folder. After selecting the “Rename Folder” option from the “File” drop-down menu, the user must enter the new
folder name, then click “OK”. Clicking “Cancel” will exit without renaming the folder.
Delete Folder - The “Delete Folder” function allows the user to permanently erase the highlighted folder. After
selecting the “Delete Folder” option from the “File” drop-down menu, the user must click “OK” to conrm the
deletion. Clicking “Cancel” will exit without deleting the folder.
Log Off - The “Log Off” function allows the current user to be logged out and takes the user to the log-in screen.
Change Password - The “Change Password” function allows the user to change the password used to log into
the current user’s prole. The user must enter their current password, then enter the new password twice for
conrmation.
Exit - The “Exit” function closes the Synergy software and returns the user to the desktop.
Instrument operation / 13
Method
• See “Methods” section for details.
Mode
• Explorer -The PETwave does not utilize this feature.
Instrument Keys
Start - The “Start” function initiates the method operation. To run a method, the user must select a method from
your user prole, then select “Start” from the “Instrument Keys” drop-down menu. A method can also be started
by selecting the START button.
Pause - The “Pause” function interrupts the currently processing method. To pause a method, the user may
select “Pause” from the “Instrument Keys” drop-down menu. A method can also be paused by using the PAUSE
button. To restart the method, press the START button.
Stop - The “Stop” function immediately stops the currently running method and initiates the quenching function.
To stop a method, the user may select “Stop” from the “Instrument Keys” drop-down menu. A method can also
be stopped by using the STOP button.
Data
• See Section “View Data” for details.
Options
Network Setup - The “Network Setup” function allows the user to connect the Synergy software to a PETwave® or
Explorer™. To connect a unit, the user must select “Network Setup” from the “Options” drop-down menu.
Manage Users - The “Manage Users” function is only available to users with Administrator access to the Synergy
software. Anyone with an Administrator account can add, alter, and delete other users’ accounts. It is advised to
reserve Administrator accounts to a select few users. To manage user accounts, select “Manage Users” from
the “Options” drop-down menu.
Convert ChemDriver Database - The PETwave does not utilize this feature.
Open Vessel - The PETwave does not utilize this feature
Temperature - The PETwave does not utilize this feature.
Preferences - The “Preferences” function allows the user to alter their personal settings within the Synergy
software. To adjust their preferences, the user must select “Preferences” from the “Options” drop-down menu.
To save their adjustments after the program is closed, the user must check “Save As Defaults” at the bottom of
the dialogue box before clicking “OK” or “Apply”.
Help
Contents - Allows the user to view the help topics of the Synergy software.
About - The “About” function allows the user to view the current version of Synergy software, PETwave Firmware
and serial number.
14 / Synergy software connection (optional)
Synergy Software Connection (Optional)
Download Synergy Software
1. Access the CEM website www.cem.com.
2. Select “PETwave” from the “Products” menu.
3. Select “Resource Library.” Scroll to the “Software” section. The current version of Synergy will be available for
download.
4. Log-in or create a CEM user account.
NOTE
If the serial number cannot be located, contact the CEM Corporation Synthesis Department
800-726-3331 or 704-821-7015.
5. Download the Synergy software onto a USB memory stick.
6. Locate a computer designated for the PETwave and place near Microwave Controller. The Synergy Software
requires a computer with Windows XP, Windows 7, Windows 8, Windows 8.1 or Windows 10. Minimum system
requirements are a Pentium 4 class processor running at 2.80 GHz with 1G of RAM and at least 40GB of avail-
able disk space.
7. Turn the computer on.
8. If any applications or dialogue boxes are open, close them.
9. Install the USB memory stick containing the Synergy software and open the Synergy folder.
10. Follow the instructions displayed on the computer screen for installation. A short cut will be automatically cre-
ated with the Synergy icon and placed on the computer desktop when installation is complete.
IP Address
The computer and Synergy software require a Fixed IP address for operation.
Computer
1. Connect an Ethernet cable from the PETwave system to the computer.
2. Turn the PETwave (Microwave Controller) on.
3. Enter the “Local Area Connection Properties screen.” Follow either “Windows XP” or “Windows 7” for IP ad-
dress setup.
Windows XP
3.1. From the computer’s desktop, select “Start” (located in the lower left corner of the screen).
3.2. Select “Control Panel.
3.3. If using the “Classic View,” select “Network Connections.” If using the “Category view,” select
“Network and Internet Connections” and then select “Network Connections.” The following screen
will appear.
Synergy software connection (optional) / 15
3.4. Select “Local Area Connection.” The following screen will appear.
Windows 7, Windows 8, Windows 8.1 and Windows 10
3.1. From the computer’s desktop, select “Start” (located in the lower left hand corner of the screen).
3.2. Select “Control Panel.
3.3. If using “Category View,” select “View network status and tasks” within the “Network and Internet”
section. If using “large icons” or “small icons” view, select “Network and Sharing Center.” The
following screen will appear.
16 / Synergy software connection (optional)
3.4. Select “Change adapter settings” located in the upper left corner of the window. The following
screen will appear.
3.5. Select “Local Area Connection.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

CEM PETwave Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI