Error Code Help text Problem Solution
2.2.5.001.8
Tips unexpected During Load Check the warning
appears
“Waste Rack Position 2: Tips un-
expected at xx”
1. Make sure the tip trays are properly
inserted
2. Make sure the waste rack is clean and
no parts (tips etc.) are present under-
neath the empty tip trays
3. Make sure the waste rack is pushed
completely to the back
4. If all of the above does not help, re-
place the waste rack (accessory)
2.5.004.1
The Tip Tray on the Waste Rack is
not loaded.
Restart Instrument and Control Unit.
If error persists, generate a problem
report and contact your Roche service
partner.
2.5.004.18
Waste Rack Position 3: Tip Tray
missing
Restart Instrument and Control Unit.
If error persists, generate a problem
report and contact your Roche service
partner.
2.11.5.0.103001.103.0
No empty tip tray in
waste rack to drop
tips.
Possible Causes
• No empty waste tray
• Waste tray not detected by the
LPS -> Dirty LPS / Damaged LPS
Restart Instrument and Control Unit.
If error persists, generate a problem
report and contact your Roche service
partner.
2.4.022.1.2.0
Instrument Access
Error 2
Possible Causes
• Message Channel blocked.
Restart Instrument and Control Unit.
If error persists, generate a problem
report and contact your Roche service
partner.