10
Roche Diagnostics
cobas
®
vivoDx System · Software version 1.0 · Quick Reference Guide · Publication version 1.0
Steps User action
5 Create a sample test order for
each patient sample.
i
If your system is connected
to a LIS, when you enter the
sample ID, the relevant
fields are filled out
automatically.
1. On the Current and recent tests work area,
choose the New button.
2. On the Patient tab, enter the required sample
and patient information.
3. Scan the barcode on the reagent cap.
4. Choose the Save button.
6 Identify the primary tube, the
reagent cap, and the test tube.
1. From the label printer, get the labels that are
printed automatically.
2. Attach the labels to the tubes and to the
reagent cap.
i
Ensure that the labels are completely
adhered to the surface. All edges of the
labels must be completely flat.
7 Prepare a sample test
cartridge for each patient
sample.
i
Check the Instructions for
Use of the corresponding
control for other test-
specific steps.
1. Follow the test-specific Instructions for Use to
transfer the collected sample from the primary
tube to the test tube.
2. Check that the labels on the reagent cap and
on the test tube match.
3. Screw the reagent cap on the test tube and
tighten it.
8 Load the test cartridges onto
the system.
1. Remove a rack from an input rack slot.
2. Insert the test cartridges in a rack position.
If you are loading control test cartridges,
insert the positive and negative control test
cartridges in adjacent rack positions.
3. Insert the rack into an available input rack
slot.
9 Release test results.
i
Check the Instructions for
Use of the corresponding
test for detailed information
about test results.
1. Wait until sample processing is complete.
2. In the Unreleased results work area, select
the test whose result you want to release.
3. To accept the result, choose the Accept
results button.
4. To invalidate the result, choose the Invalidate
results button.
5. If your system is connected to a LIS, the test
result is sent to the LIS.
10 Unload the test cartridges.
1. Remove the output rack completely.
2. Dispose of the test cartridges from the rack
into a biohazardous safety container.
3. Insert the rack again into an available output
rack slot.
y Routine workflow