WARNING Defective batteries can damage the device. If the battery shows any
signs of damage or cracking, it must be replaced immediately and only with a
battery approved by Welch Allyn.
WARNING Personal injury risk. Improper handling of the battery can lead to heat
generation, smoke, explosion or fire. Do not short-circuit, crush, incinerate, or
disassemble the battery. Never dispose of batteries in refuse containers. Always
recycle batteries according to national or local regulations.
WARNING Use only Welch Allyn approved accessories, and use them according
to the manufacturer’s directions for use. Using unapproved accessories with the
device can affect patient and operator safety and can compromise product
performance and accuracy, and void the product warranty.
WARNING Welch Allyn is not responsible for the integrity of a facility's power. If
the integrity of a facility's power or protective earth conductor is in doubt, always
operate the device on battery power alone when it is attached to a patient.
WARNING The use of the Spot Vital Signs 4400 device adjacent to or stacked with
other equipment or medical electrical systems should be avoided because it could
result in improper operation. If such use is necessary, the Spot Vital Signs 4400 and
other equipment should be observed to verify that they are operating normally.
WARNING Use only accessories and cables Welch Allyn recommends for use with
the Spot Vital Signs 4400 device. Accessories and cables not recommended by
Welch Allyn may affect the EMC emissions or immunity.
WARNING Maintain minimum separation distance of 12 inches (30 cm) between
any part of the Spot Vital Signs 4400 device and portable RF communication
equipment (including peripherals such as antenna cables and external antennas).
Performance of the Spot Vital Signs 4400 device might degrade if proper distance
is not maintained.
WARNING Use of accessories, transducers, and cables other than those specified
may result in degraded electromagnetic compatibility performance of this device.
WARNING Use of accessories, transducers, and cables other than those specified
may result in increased emissions or decreased immunity of the device.
WARNING Patient harm and equipment damage risk. Carefully route patient
cords and cables to reduce the possibility of patient entanglement. When
transporting the Spot Vital Signs device on a mobile stand, properly secure all
patient cords and cables to keep them clear of the wheels and to minimize trip
hazards.
WARNING Strangulation risk. The cords and cables can wrap around the patient’s
neck. When used on children or vulnerable patient populations, the Spot Vital
Signs 4400 device accessories must only be applied with special care and under
permanent supervision. When used on adults, caution should be taken.
WARNING Choking risk. An oral probe cover enters the patient’s mouth when
taking oral temperatures. When inserting the probe tip inside the mouth of a
patient, ensure that the probe cover remains on the probe tip to avoid the risk of
the patient choking on the probe cover. When using on children or vulnerable
patient populations, the Spot Vital Signs 4400 device must only be used with
special care and under permanent supervision. When used on adults, caution
should be taken.