WARNING Liquids can damage electronics inside the device. Prevent
liquids from spilling on the device.
If liquids are spilled on the device:
1. Power down the device.
2. Disconnect the power plug.
3. Remove battery pack from the device.
4. Dry off excess liquid from the device.
Note If liquids possibly entered the device, remove the device
from use until it has been properly dried, inspected, and
tested by qualified service personnel.
5. Reinstall battery pack.
6. Reconnect the power plug.
7. Power on the device and verify that the device functions normally before
using it.
WARNING The device may not function properly if dropped or damaged.
Protect it from severe impact and shock. Do not use the device if you
notice any signs of damage. Qualified service personnel must check any
device that is dropped or damaged for proper operation before putting the
device back into use.
WARNING Defective batteries can damage the device. If the battery
shows any signs of damage or cracking, it must be replaced immediately
and only with a battery approved by Welch Allyn.
WARNING Personal injury risk. Improper handling of the battery can lead
to heat generation, smoke, explosion or fire. Do not short-circuit, crush,
incinerate, or disassemble the battery. Never dispose of batteries in refuse
containers. Always recycle batteries according to national or local
regulations.
WARNING Use only Welch Allyn approved accessories, and use them
according to the manufacturer’s directions for use. Using unapproved
accessories with the device can affect patient and operator safety and can
compromise product performance and accuracy, and void the product
warranty.
WARNING Welch Allyn is not responsible for the integrity of a facility's
power. If the integrity of a facility's power or protective earth conductor is
in doubt, always operate the device on battery power alone when it is
attached to a patient.
WARNING Safety risk. Make frequent electrical and visual checks on
cables, sensors, and electrode wires. All cables, sensors, and electrode
wires must be inspected and properly maintained and in proper working
order to allow the equipment to function properly and to protect patients.
WARNING The use of the Spot Vital Signs 4400 device adjacent to or
stacked with other equipment or medical electrical systems should be
avoided because it could result in improper operation. If such use is
necessary, the Spot Vital Signs 4400 and other equipment should be
observed to verify that they are operating normally.
10 About warnings and cautions Welch Allyn
®
Spot Vital Signs 4400