Hill-Rom SureTemp Plus 690 Thermometer User manual

Category
Measuring, testing & control
Type
User manual
Welch Allyn
®
SureTemp
®
Plus
690 and 692 Series
Instructions for use
ii Welch Allyn® SureTemp® Plus
© 2020 Welch Allyn, Inc. To support the intended use of the product described in this publication, the purchaser of the product
is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn.
Welch Allyn assumes no responsibility for any injury, or for any illegal or improper use of the product, that may result from
failure to use this product in accordance with the instructions, cautions, warnings, or indications for use published in this
manual.
Welch Allyn is a registered trademark of Welch Allyn, Inc.
PATENT/PATENTS
hillrom.com/patents
May be covered by one or more patents. See above Internet address. The Hill-Rom companies are the proprietors of European,
US, and other patents and pending patent applications.
Hillrom Technical Support
For information about any Hillrom product, contact Hillrom Technical Support: hillrom.com/en-us/about-us/locations
Notice to users and/or patients in EU
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent
authority of the Member State in which the user and/or patient is established.
106701 (CD) 80025967 Ver. B Revision date: 2020-12 901053 ELECTRONIC THERMOMETER
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153 USA
hillrom.com
Welch Allyn, Inc. is a subsidiary of Hill-Rom Holdings, Inc.
and EU IMPORTER Authorized Australian Sponsor
Welch Allyn Australia Pty. Ltd.
Unit 4.01, 2-4 Lyonpark Road
Macquarie Park, NSW 2113
Phone 1800 650 083
Welch Allyn Limited
Navan Business Park, Dublin Road
Navan, Co. Meath C15 AW22
Ireland
Instructions for use iii
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
SureTemp Plus model comparison chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Indications for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbol descriptions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Residual risk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Latex statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Parts and features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Button functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Assembling the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mounting the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Oral temperature measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Axillary temperature measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Rectal temperature measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Monitor mode temperature measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Instrument security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Temperature count . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Instant-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cleaning and disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Cleaning and disinfecting the thermometer and probe . . . . . . . . . . . . . . . . . 15
Cleaning and disinfecting the removable probe well . . . . . . . . . . . . . . . . . . . 15
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Replacing the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Hillrom service policy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Contacting Hillrom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Performance and technical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Standards and compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Lot code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Guidance and manufacturer’s declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
EMC compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Emissions and immunity information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Replacement parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
SureTemp Plus Global Trade Item Numbers (GTIN) . . . . . . . . . . . . . . . . . . . . . . . . . 23
Limited warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
iv Welch Allyn® SureTemp® Plus
Instructions for use 1
Introduction
Follow the operating and maintenance instructions provided in this manual to ensure years of
accurate and reliable service. Read these instructions thoroughly before using the instrument,
and then store these instructions in a safe place for future reference.
The Welch Allyn SureTemp Plus is a portable thermistor thermometer used for accurately
measuring body temperatures at oral, axillary, or rectal sites. The device measures temperature
by default in predictive (Normal) mode, during which it uses a predictive algorithm to calculate
patient temperature. If you cannot correctly measure the patients temperature in Normal
mode, the unit will automatically enter Monitor (Direct) mode.
Normal body temperature is a range, which varies from person to person and is influenced by
factors such as time of day and an individual’s level of activity, ingested medications, age, and
gender. A person’s normal temperature tends to decrease with age.
The following table shows that ranges of normal also vary by site. Do not directly compare
measurements from different sites.
This booklet provides informative tips and instructions for using the SureTemp Plus model 690
and model 692 thermometers. The chart on the following page provides details about the
differences between these SureTemp Plus instruments.
2Welch Allyn® SureTemp® Plus
SureTemp Plus model comparison chart
This manual describes features found in SureTemp Plus model 690 and 692 thermometers.
Certain features are not available to users of the model 690 thermometer.
Intended use
The Welch Allyn SureTemp® Plus thermometer enables the health care professional to make an
accurate prediction of a febrile, afebrile, or hypothermic patient’s oral temperature in
approximately 4-6 seconds (in Normal mode). Pediatric axillary (age 17 and younger)
temperatures can be obtained in approximately 10-13 seconds. Adult axillary temperatures (in
Normal mode) can be obtained in approximately 12-15 seconds. Rectal temperatures (in
Normal mode) can be obtained in approximately 10-13 seconds. Predictive (Normal) mode is
available for oral, rectal, and axillary use.
In the Monitor mode, the instrument provides the capability of accurate, long-term monitoring
of actual oral, rectal or axillary temperature, and of following the temperature whether constant,
increasing, or decreasing.
The SureTemp Plus is a clinical grade thermometer intended for use by healthcare practitioners
only; typically, in a hospital, clinic, long-term care, or mobile health care environment. It is not
intended for home use.
Indications for use
The Welch Allyn SureTemp® Plus thermometer is intended to be used by healthcare
professionals, to provide an accurate prediction of patient temperature using the oral, axillary,
SureTemp Plus model
690
SureTemp Plus model
692
Probe cover storage X X
F/C conversion X X
Last temperature recall X X
Monitor mode X X
Detachable probe X X
Oral / axillary / rectal modes on each
thermometer
X X
Color-coded removable probe well X X
Wall-holder standard X
Pulse timer X
Backlight X
Security (electronic) X
Configurable on-screen labeling for
identification of instrument
X
Sold in U.S./Canada X X
Sold internationally X X
Instrument warranty 2 years 3 years
Removable probe well 90 days 90 days
Probe warranty 1 year 1 year
Instructions for use 3
or rectal body sites in 4 to 15 seconds, or to provide an actual temperature reading in the
continuous monitor mode in about 3 minutes.
Symbol descriptions
For information on the origin of these symbols, see the Welch Allyn symbols glossary:
welchallyn.com/symbolsglossary.
Consult instructions for use (IFU). A copy of the IFU is available on this website. A printed copy of the IFU can be
ordered from
Hillrom
for delivery within 7 calendar days.
Separate collection of electrical and electronic equipment. Do not dispose as unsorted municipal
waste.
WARNING The warning statements in this manual identify conditions or practices that could lead
to illness, injury, or death. Warning symbols will appear with a grey background in a black and white
document.
CAUTION The caution statements in this manual identify conditions or practices that could result
in damage to the equipment or other property, or loss of data.
NOTE Presents clarification about an instruction or helpful information about a feature or behavior.
Type BF applied part Do not re-use, single use device
Fragile Recyclable
Temperature limit Product identifier
Humidity limitation Reorder number
Positioning of cell Global Trade Item Number
Battery check Manufacturer
IPXØ Not protected against the ingress of water Serial number
Not made with natural rubber latex Medical device
Keep dry Lot code
hillrom.com
4Welch Allyn® SureTemp® Plus
Warnings
WARNING No modification of this equipment allowed.
WARNING Patient injury risk. Prior to taking a temperature, instruct the patient not to
bite down on the probe as patient injury and damage to the probe may result.
WARNING Patient injury risk. Use only Welch Allyn single-use disposable probe covers.
A new probe cover needs to be used with each new temperature taken. Never take
a temperature measurement without a single-use probe cover securely attached. Failure
to use a probe cover can cause patient discomfort from a heated probe, patient cross-
contamination, and inaccurate temperature readings.
WARNING Patient injury risk. Inaccurate measurement risk. Oral/axillary probes (blue
ejection button at top of probe) and blue removable probe wells are used for taking oral
and axillary temperatures only. Rectal probes (red ejection button) and red removable
probe wells are used for taking rectal temperatures only. Use of the incorrect removable
probe well could result in patient cross-contamination. Use of the probe at the wrong
site will result in temperature errors.
WARNING Patient injury risk. When taking rectal temperatures, insert the probe tip
only 5/8 inch (approximately 1.5 cm) inside the rectum of adults and only 3/8 inch
(approximately 1 cm) inside the rectum of children to avoid the risk of bowel
perforation.
WARNING Inaccurate measurement risk. Never take an axillary temperature through
the patient’s clothing. Carefully place the probe in the axilla, avoiding contact with other
objects or material.
WARNING Patient injury risk. Continuous measurement durations of 3 minutes at the
oral and rectal sites and 5 minutes at the axillary site are recommended for accurate
measurement. Do not continuously measure beyond 10 minutes in any mode.
WARNING To ensure patient safety and temperature measurement accuracy, use only
accessories and supplies recommended or supplied by Welch Allyn.
WARNING Never use a damaged temperature probe. The thermometer consists of
high-quality precision parts and should be protected from severe impact or shock. Do
not use the thermometer if you notice any signs of damage to the probe or instrument.
If the thermometer probe is dropped or damaged, remove it from use and have it
inspected by qualified service personnel.
WARNING Inaccurate measurement risk. For rectal measurements, apply a thin layer of
lubricant to probe cover, if necessary, for patient comfort. Use of excessive lubricant may
affect reading accuracy.
WARNING Inaccurate measurement risk. To ensure optimal accuracy, always confirm
that the correct mode and site are selected.
WARNING Incorrect measurements can result in delayed or incorrect treatment. If you
suspect that the thermometer is delivering inaccurate measurements, first review the
instructions in the sections on temperature measurement in this manual for ways to
avoid inaccurate measurements. If you still believe that the thermometer is inaccurate,
contact a Hillrom Service Center for calibration.
Instructions for use 5
Cautions
Residual risk
This product complies with relevant mechanical safety, performance, electrical and
biocompatibility standards. However, the product cannot completely eliminate potential
patient or user harm from the following:
Harm from mechanical hazards,
Harm from device, function, or unavailability,
Harm from misuse error, and/or
Harm from device exposure to biological triggers that may result in a severe systemic
allergic reaction.
Latex statement
The components of the SureTemp Plus device that could contact the user or patient were not
made with natural rubber latex. This includes all items that could be contacted during normal
operation, and all other functions, such as user maintenance and cleaning, as are defined in the
Instructions for use.
No internal components are known to be made with natural rubber latex.
CAUTION Liquids can damage electronics inside the thermometer. Prevent liquids
from spilling on the thermometer. If liquids are spilled on the thermometer, dry off the
thermometer with a clean cloth. Check for proper operation and accuracy. If liquids
possibly entered the thermometer, remove the thermometer from use until it has been
properly dried, inspected, and tested by qualified service personnel.
CAUTION Welch Allyn recommends that batteries are removed for long term storage
since alkaline batteries can deteriorate and leak over extended periods of time, and
possibly cause damage to the thermometer.
CAUTION Do not autoclaveany component of this product. Please note the cleaning
procedures described in this manual.
CAUTION Do not use this instrument for any purpose other than that specified in this
booklet. Doing so will invalidate the instrument’s warranty.
CAUTION This thermometer complies with current required standards for
electromagnetic interference and should not present a problem to other equipment
nor is it effected by other devices. As a precaution, avoid using this device in close
proximity to other equipment.
CAUTION Inaccurate measurement risk. Patient activities such as strenuous exercise,
ingesting hot or cold liquids, eating, chewing gum or mints, brushing teeth, or smoking
may effect oral temperature measurements for up to 20 minutes.
CAUTION Probe covers are disposable, nonsterilized, and single-use. Probes are also
nonsterilized. Do not autoclave probes and probe covers. Ensure that probe covers are
disposed of according to facility requirements or local regulations.
CAUTION Replace the battery when the battery status is low. Use only AA 1.5 Vdc
alkaline batteries.
6Welch Allyn® SureTemp® Plus
Parts and features
You can find the serial number of your device on a label affixed to the rear housing inside the
unit’s probe cover storage compartment. Please note the parts of your Welch Allyn SureTemp
Plus thermometer as shown below:
Button functions
Thermometer
LCD display
Probe
Ejection button (blue or red)
Removable probe well
(blue or red)
Latching probe
connector
Battery door/probe
cover storage
compartment
C/F (Celsius/
Fahrenheit)
button
Pulse timer
button/Monitor
mode (692 only)
Mode selection
button
Recall button
Box of probe
covers
Button User interaction
Press and release the C/F (Celsius/Fahrenheit) button to toggle between
temperature scales any time a temperature is displayed on the LCD.
Press and release the Pulse timer/Monitor mode (692 only) button to
activate the timer. Audible beeps sound at 0, 15, 30, 45, and 60 seconds. The
timer turns off automatically after 60 seconds. You can stop the timer at any
time by pressing the Pulse timer/Monitor mode button or by removing the
probe from the probe well.
For Monitor mode, press and release the Pulse timer button after you have
withdrawn the probe from the probe well to switch the device into Monitor
mode.
Press and release the Mode selection button after you have withdrawn the
blue probe from the probe well to change the current measurement site to
oral, adult axillary, or pediatric axillary. The selected measurement site is
displayed by a flashing body site icon. Note that no other measurement sites
are available when using the red probe in the Rectal mode.
Press and release the Recall button when the display is blank to recall the last
completed predicted temperature. The LCD will display the recalled
temperature for five seconds along with the battery icon, temperature scale,
mode selection, and probe-position icon (if applicable).
Instructions for use 7
Display indicators
Instrument malfunction
The device is not working properly. Contact Hillrom Technical Support
for assistance.
Broken probe
The probe is broken. Replace probe or contact Hillrom Technical
Support for assistance.
Loss of tissue contact
The probe has lost contact with the patient’s tissue. The icon will
disappear once proper contact is achieved. The icon will later reappear
at the end of the temperature measurement. It is recommended that
you take a new temperature reading.
ID field
This is a seven-character field that allows customized identification of
the instrument via user-selected letters and numbers. See the model
690/692 Service manual for setup instructions.
Walking segments
The dark bar will move around the display when the thermometer is in
the process of taking a temperature.
Oral mode
This flashing icon and/or word are displayed to show that the oral mode
is selected.
Adult axillary mode
This flashing icon and/or words are displayed to show that the adult
axillary mode is selected.
Ped axillary mode
This flashing icon and/or words are displayed to show that the pediatric
axillary mode is selected.
8Welch Allyn® SureTemp® Plus
Setup
Assembling the unit
Your SureTemp Plus thermometer is shipped to you with three 1.5V AA batteries installed.
Rectal mode
This flashing icon and/or word are displayed to show that the rectal
mode is selected.
Monitor mode
This icon is displayed while in Monitor mode. Monitor mode is used for
continuous temperature monitoring when difficult situations prevent
accurate temperatures in the normal mode. For oral and rectal sites,
three minutes of continuous monitoring is recommended for an
accurate temperature reading. For axillary sites, five minutes of
continuous monitoring is recommended.
Patient temperature exceeds upper measurement range limit of
110.0 °F (43.3 °C)
Patient temperature is lower than measurement range limit of 80.0
°F (26.7 °C)
Ambient temperature exceeds upper measurement range limit of
104.0 °F (40.0 °C)
Ambient temperature is lower than measurement range limit of 50.0
°F (10.0 °C)
Pulse timer
This icon indicates that the device is in the pulse timer mode.
Low battery
A single flashing segment appears on the battery symbol. Normal
operation is not affected. Replace the batteries.
Dead battery
An open battery symbol remains on the LCD and the device will not
operate. Replace the batteries.
Security mode
The instrument is in security mode. Return the thermometer to the wall
holder to resume normal operation See page 14 to set the instrument in
Security mode (model 692 only).
Instructions for use 9
1. Install the oral/axillary probe (blue ejection button) or optional rectal probe (red ejection
button) by inserting the latching probe connector on the end of the coiled cord into the
unit’s plug receptacle. Ensure that the connector snaps into place. The LCD will illuminate
for approximately 7 seconds and will then turn off.
2. Connect the appropriate removable probe well to the thermometer. Note that the cut-out
in the probe well fits over the probe cord strain relief. Ensure that the removable probe well
“snaps” into position and is completely seated to the thermometer. Place the probe into the
well.
3. Insert the box of Hillrom disposable probe covers into the probe cover storage holder. To
open the box of probe covers, remove and discard the tear-away corner. You will see the
ends of the probe covers in the box.
Mounting the unit
Mount the wall-mount bracket in a location such as an interior wall or wall divider so that the
thermometer remains at ambient room temperature. Mounting the unit on an uninsulated
exterior wall, in an area directly hit by sunlight, above heat radiators or powered equipment
(including computers and lights), or below air conditioning outlets is not recommended.
Proper mounting of the unit will prevent tangling of the probe cord with other equipment at
the patient’s bedside.
Oral temperature measurement
Using probe with blue ejection button and blue probe well
When used correctly, the SureTemp Plus thermometer accurately measures an oral temperature
in approximately 4–6 seconds. The ability of the SureTemp Plus thermometer to take an
accurate oral temperature requires correct user technique.
1. Mount the instrument so
that all other objects are
no closer than 4 inches
(10.25 cm) from the sides
and bottom of the unit.
2. Mount the wall mount a
maximum of 48 inches
(122 cm) from the floor.
3. Install the wall mount to
allow at least 12 inches
30.5 cm) clearance on top
of the instrument. This
provides ease of
removing and installing
the probe and probe well
or installing probe covers
onto the probe.
10 Welch Allyn® SureTemp® Plus
Taking an oral temperature
1. Ensure that the oral probe (blue ejection button) and the blue probe well are installed.
2. Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.
3. Verify that the oral model icon is selected by observing the flashing head icon on the
instrument’s display. If this icon is not flashing, press the Mode selection button until
the head icon appears.
4. Load a probe cover by inserting the probe into a probe cover and pressing the probe
handle down firmly. The probe handle will move slightly to engage the probe cover.
5. With the Oral mode indicator flashing, quickly place the probe
tip under the patient’s tongue on either side of the mouth to
reach the rear sublingual pocket. Have the patient close his/her
lips around the probe.
6. Hold the probe in place, keeping the tip of the probe in contact
with the oral tissue throughout the measurement process.
Rotating “walking” segments appear on the display, indicating
that measurement is in progress.
7. The unit will beep three times when the final temperature is
reached. The measurement site, temperature scale, and patient
temperature will display on the LCD. The final temperature will remain on the display for 30
seconds.
8. If you cannot correctly measure the patient’s temperature in Normal mode,
the unit will automatically enter Monitor mode. In this mode, measurement
time is extended. Either repeat the temperature measurement in Normal
mode in the opposite sublingual pocket or keep the probe in place for three
minutes in Monitor mode. The thermometer will not beep to indicate a final
temperature. Record the temperature before removing the probe from the site, as the
temperature reading is not maintained in memory.
9. After the temperature measurement is complete, remove the probe from the patients
mouth. Eject the probe cover by firmly pressing the ejection button on the top of the probe.
WARNING Do not take a patient’s temperature without using a Hillrom disposable
probe cover. Doing so can cause patient discomfort, patient cross contamination, and
erroneous temperature readings.
WARNING Verify that the probe cover does not come off prior to use. A probe cover that
comes off during use can cause asphyxiation. If a probe cover fails to stay on, then the
retention mechanism may be damaged. Remove the device from service for inspection
by qualified service personnel.
WARNING Use only Hillrom probe covers. The use of other manufacturer’s probe covers
or no probe cover may produce temperature measurement errors and/or inaccuracy.
WARNING Patient injury risk. Do not exceed the recommended temperature
measurement durations in Direct mode. Continuous measurement durations of 3 minutes
at the oral and rectal sites and 5 minutes at the axillary site are recommended for accurate
measurement. Do not continuously measure beyond 10 minutes in any mode.
Instructions for use 11
10. Return the probe to the probe well. The LCD display will go blank.
Axillary temperature measurement
Using probe with blue ejection button and blue probe well
When used correctly, the SureTemp Plus thermometer accurately measures an axillary
temperature for pediatric patients (ages 17 and younger) in approximately 10–13 seconds and
for adult patients (ages 18 and older) in approximately 12–15 seconds.
Placing the thermometer into axillary mode
1. Ensure that the axillary probe (blue ejection button) and the blue probe well are installed.
2. Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.
3. Verify that the axillary mode is selected by observing the
correct flashing axillary icon on the instrument’s display. If
this icon is not flashing, press the Mode selection button
to select the Adult axillary or Pediatric axillary mode.
Taking an axillary temperature
1. Load a probe cover by inserting the probe into a probe cover and pressing the probe han-
dle down firmly. The probe handle will move slightly to engage the probe cover.
2. With the correct axillary mode indicator flashing, lift the patient’s arm so that the entire
axilla is easily seen. Place the probe as high as possible in the axilla. Do not allow the probe
tip to come into contact with the patient until the probe is placed in the measurement site.
Before this, any contact between the probe tip and the tissue or other material may cause
inaccurate readings.
3. Verify that the probe tip is completely surrounded by axillary tissue and place the arm
snugly at the patient’s side. Hold the patient’s arm in this position and do not allow
CAUTION Patient actions may interfere with accurate oral temperature readings.
Ingesting hot or cold liquids, eating food, chewing gum or mints, brushing teeth,
smoking, or performing strenuous activity may affect temperature readings for up to
20 minutes after activity has ended.
WARNING Do not take a patient’s temperature without using a Hillrom disposable
probe cover. Doing so can cause patient discomfort, patient cross contamination, and
erroneous temperature readings.
CAUTION To ensure optimal accuracy, always confirm that the correct axillary mode is
selected.
CAUTION After a temperature is taken and the probe is returned to the probe well, the
instrument reverts to the original measurement site mode.
Adult axillary mode
icon
Pediatric axillary
mode icon
WARNING Do not take an axillary temperature through patient’s clothing. Direct
contact between patient’s skin and the probe is required.
WARNING Use only Hillrom probe covers. The use of other manufacturer’s probe covers
or no probe cover may produce temperature measurement errors and/or inaccuracy.
12 Welch Allyn® SureTemp® Plus
movement of the arm or probe during the measurement cycle. Rotating “walking”
segments appear on the display, indicating that measurement is in progress.
4. The unit will beep three times when the final temperature is reached. The measurement
site, temperature scale, and patient temperature will display on the LCD. The final
temperature will remain on the display for 30 seconds.
5. If you cannot correctly measure the patient’s temperature in Normal mode,
the unit will automatically enter Monitor mode. In this mode, measurement
time is extended. Either repeat the temperature measurement in Normal
mode in the opposite axilla or keep the probe in place for five minutes in
Monitor mode. The thermometer will not beep to indicate a final temperature. Record the
temperature before removing the probe from the site, as the temperature reading is not
maintained in memory.
6. After the temperature measurement is complete, remove the probe from the patients
axilla. Eject the probe cover by firmly pressing the ejection button on the top of the probe.
7. Return the probe to the probe well. The LCD display will go blank.
Rectal temperature measurement
Using probe with red ejection button and red probe well
When used correctly, the SureTemp Plus thermometer accurately measures rectal temperature
in approximately 10–13 seconds.
Taking a rectal temperature
1. Ensure that the rectal probe (red ejection button) and the red probe well are installed.
The instrument will only operate in Rectal mode when the red rectal probe and probe well
are installed.
2. Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.
3. Observe the flashing lower-body icon on the unit’s display. Load a probe cover by
inserting the probe into a probe cover and pressing the probe handle down firmly.
The probe handle will move slightly to engage the probe cover.
WARNING Long-term continuous monitoring beyond five minutes is not recommended
in the Axillary mode.
WARNING Probe contact with electrodes, bandages, etc., poor tissue contact, taking a
temperature over clothing, or prolonged exposure of axilla to ambient air can cause
inaccurate temperature readings.
WARNING Cross contamination or nosocomial infection risk. Thorough hand-washing
greatly reduces the risk of cross-contamination and nosocomial infection.
WARNING Do not take a patient’s temperature without using a Hillrom disposable
probe cover. Doing so can cause patient discomfort, patient cross contamination, and
erroneous temperature readings.
WARNING Use only Hillrom probe covers. The use of other manufacturer’s probe
covers or no probe cover may produce temperature measurement errors and/or
inaccuracy.
Instructions for use 13
4. With the Rectal mode indicator flashing, separate the patient’s buttocks with one hand.
Using the other hand, gently insert the probe only 1.5 cm (5/8 in.) inside the rectum (less for
infants and children). The use of a lubricant is optional.
5. Tilt the probe so that the tip of the probe is in contact with tissue. Keep the hand separating
the buttocks in place, and hold the probe in place throughout the measurement process.
Rotating “walking” segments appear on the display, indicating that measurement is in
progress.
6. The unit will beep three times when the final temperature is reached. The measurement
site, temperature scale, and patient temperature will display on the LCD. The final
temperature will remain on the display for 30 seconds.
7.
If you cannot correctly measure the patient’s temperature in Normal mode, the
unit will automatically enter Monitor mode. In this mode, measurement time is
extended. Either repeat the temperature measurement in Normal mode or keep
the probe in place for three minutes in Monitor mode. The thermometer will not
beep to indicate a final temperature. Record the temperature before removing
the probe from the site, as the temperature reading is not maintained in memory.
8.
After the temperature measurement is complete, remove the probe from the patient’s
rectum. Eject the probe cover by firmly pressing the ejection button on the top of the probe.
9. Return the probe to the probe well. The LCD display will go blank.
10. Wash your hands.
Monitor mode temperature measurement
Using probe with blue ejection button (oral/axillary) or red
ejection button (rectal)
Monitor mode continuously displays the temperature of the probe for as long as the probe
remains in place at the measurement site and remains within the operating patient temperature
range. The patient’s oral and rectal temperature will reach final equilibrium in approximately
three minutes in the Monitor mode. Axillary temperature will reach equilibrium in
approximately five minutes in the Monitor mode.
Taking a temperature using Monitor mode
WARNING Patient injury risk. When taking rectal temperatures, insert the probe tip
only 5/8 inch (approximately 1.5 cm) inside the rectum of adults and only 3/8 inch
(approximately 1 cm) inside the rectum of children to avoid the risk of bowel
perforation.
WARNING Long-term continuous monitoring beyond three minutes is not
recommended in Rectal mode.
WARNING Washing hands greatly reduces the risk of cross-contamination and
nosocomial infection.
CAUTION Monitored temperatures are not stored in memory for recall.
WARNING Do not take a patient’s temperature without using a Welch Allyn disposable
probe cover. Doing so can cause patient discomfort, patient cross contamination, and
erroneous temperature readings.
14 Welch Allyn® SureTemp® Plus
1. Ensure that both the appropriately colored probe and matching probe well for the intended
measurement site (oral/axillary or rectal) are installed.
2. Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.
3. Load a probe cover by inserting the probe into a probe cover and pressing the probe
handle down firmly. The probe handle will move slightly to engage the probe cover.
4.
For model 692, press and release the Pulse timer / Monitor mode button
after you have withdrawn the probe from the probe well to switch the device
into Monitor mode. The Monitor mode indicator appears on the LCD
.
5. For model 690, take the patient’s temperature using the normal Oral, Rectal, Pediatric
axillary, or Adult axillary mode as previously described.
a. Continue to hold the probe in position after the temperature is displayed.
b. Press the Mode selection button until the Monitor mode indicator
appears on the LCD.
6. Hold the thermometer in place for the required duration:
7. Record the temperature before removing the probe from the site.
8. Eject the probe cover by firmly pressing the ejection button on the top of the probe.
9. Replace the probe in the probe well to clear the display and reset the thermometer to
Normal mode. The thermometer will not beep to indicate a final temperature.
Instrument security
The SureTemp model 692 thermometer is equipped with electronic security options designed
to reduce the risk of theft. These options include both Temperature count and Instant-on
security features.
Temperature count
The Temperature Count security feature allows you to take a set number of temperature
readings before the instrument goes into the security alarm state. The selectable counts are 25,
50, 100, and 200. As the instrument nears the final reading, it will display SECURITY COUNT as a
warning, along with a number representing the last 5 temperature counts (i.e., 5,4,3,2,1)
remaining on the security counter.
When the instrument enters the alarm state, it beeps for approximately 10 seconds. At this time,
the instrument displays SEC on the LCD along with SECURITY COUNT in the ID field in the
upper left-hand corner of the screen. The instrument is then disabled. Returning the instrument
to the wall holder resets the security count. If the count is set to “OFF” the temperature count
function is disabled and no security alarm will occur.
To set the Temperature count security feature, refer to the Service manual.
WARNING Use only Welch Allyn probe covers. The use of other manufacturer’s probe
covers or no probe cover may produce temperature measurement errors and/or inaccuracy.
WARNING The unit will automatically enter Monitor mode if the probe is withdrawn
from the probe well and is not replaced within 60 seconds of inactivity.
Oral 3 minutes
Axillary 5 minutes
Rectal 3 minutes
WARNING Long-term continuous monitoring beyond three to five minutes is not
recommended in any mode.
Instructions for use 15
Instant-on
When the Instant-on feature is activated, the instrument goes into the security alarm state
within five seconds after removal from the wall holder. When the instrument enters the alarm
state, it beeps continuously and displays SEC on the LCD. Pressing the Mode selection button
while removing the probe from the probe well, within 30 seconds of the instruments removal
from the wall holder, clears the alarm and allows the instrument to operate. Returning the
instrument to the wall holder will reset the instant audio alarm.
The options for the Instant-on feature are ON and OFF. In the OFF setting, no security alarm will
sound.
To set the Instant-on security feature, refer to the Service manual.
Cleaning and disinfection
Clean and disinfect on a routine basis according to each facility's policy or whenever
contamination occurs.
Cleaning and disinfecting the thermometer and probe
Wipe the thermometer and probe with an appropriate health care low- or intermediate-level
cleaner/disinfecting wipe that incorporates either a 1:10 sodium hypochlorite (bleach) solution
or isopropyl alcohol as the active disinfection ingredient. Follow wipe manufacturer's
instructions for appropriate use, contact times, and applicable warnings and precautions.
Cleaning and disinfecting the removable probe well
Remove the probe well from the unit. Unplug the latching probe connector to prevent the
device from consuming battery power while you are cleaning the probe well.
Clean the inner surface of the probe well by swabbing the surface with a cloth dampened
with a 1:10 sodium hypochlorite (bleach) solution or isopropyl alcohol solution as the active
disinfection ingredient. Clean the probe well's outer surface by swabbing or wiping the
surface with one of the solutions mentioned above.
Thoroughly dry all surfaces before re-assembling the instrument.
Re-connect the latching probe connector to the thermometer. Ensure that the connector
snaps into place.
Re-install the probe well in the thermometer and snap the probe well into place.
Insert the probe into the probe well.
WARNING Cleaning solutions, including those that contain bleach, can cause skin
irritation if not completely wiped clean from the probe or thermometer following cleaning.
CAUTION DO NOT immerse or soak the thermometer or probe in any type of fluid.
CAUTION DO NOT use steam, heat, or gas sterilization on the thermometer or probe.
CAUTION DO NOT autoclave the thermometer or the probe.
CAUTION DO NOT use hard or sharp objects to clean the probe well. This could damage
the probe well and cause the unit to not function properly.
CAUTION DO NOT use steam, heat, or gas sterilization on the probe well.
CAUTION DO NOT autoclave the probe well.
16 Welch Allyn® SureTemp® Plus
Maintenance
Replacing the batteries
1. Remove the box of probe covers from the probe cover storage compartment on the back of
the unit.
2. Remove the probe and the probe well.
3. Press inward on the textured grips on each side of the assembly to remove the battery door
from the back of the unit.
4. Remove the three AA 1.5Vdc alkaline batteries and replace all three batteries with AA 1.5
Vdc alkaline batteries. Match the +/- polarity of each battery to the +/- symbols inside the
battery compartment.
5. Snap the probe cover storage assembly back into place.
6. Replace the probe and probe well.
Service
Except for the replaceable AA batteries, this instrument contains no user-serviceable parts.
Return the thermometer to a Hillrom authorized service center for service and repair. Refer to
the Service manual.
Hillrom service policy
Only a Hillrom Service Center can perform or approve all repairs on products under warranty.
Unauthorized repairs will void the warranty. Qualified electronics personnel or a Hillrom Service
Center should repair out-of-warranty products.
Contacting Hillrom
If you encounter an equipment problem that you cannot resolve, contact the Hillrom Service
Center nearest you for assistance. Find a Hillrom Service Center at
welchallyn.com/en/service-support.html.
Before returning a product for repair, you must first obtain authorization from Hillrom. If you are
advised to return a product to Hillrom for service or repair, you should schedule the repair with
the Hillrom Service Center nearest you.
Hillrom service personnel with provide you with a Return Materials Authorization (RMA)
number. Print this number on the outside of your shipping box.
Hillrom does not accept returns that are missing an RMA number.
Disposal
Users must adhere to all federal, state, regional, and/or local laws and regulations as they
pertain to the safe disposal of medical devices and accessories. If in doubt, the user of the
device should first contact Hillrom Technical Support for guidance on safe disposal protocols.
WARNING Use size AA 1.5 Vdc alkaline batteries only. Use of any other battery size or
type could damage the thermometer and cause personal injury.
CAUTION Remove the batteries if the instrument is not used for an extended period
of time to avoid possible damage to the thermometer due to battery leakage.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30

Hill-Rom SureTemp Plus 690 Thermometer User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI