Note
Use this kit to replace the capacitor on Midmark Ultra-Series Dental Chairs. J-Box component locations vary
depending on model and the options attached.
003-10205-00 Rev. AA1 (7/18)
Contents:
Capacitor ................1
Double-sided Tape .1
Special Tools:
none
Capacitor Replacement Kit Install [Kit# 002-10356-00]
Style G
Hydraulic Fluid Reservoir
Capacitor
Kit capacitor
Remove
existing
capacitor
Hydraulic Cover
Install Kit Capacitor...
A) Peel off tape film, rotate the kit capacitor to fit and
press it onto the chair frame near the hydraulic fluid
reservoir.
B) Connect the two wires you removed from the existing
capacitor to the kit capacitor terminals. Either wire
can be connected to either terminal.
C) Install the hydraulic cover.
D) Reconnect power to the chair.
WARNING
Always disconnect power from the chair
before removing any covers. Failure to do so
may result in personal injury.
Remove Existing Capacitor...
A) Rotate chair seat to one side.
B) Remove three screws and the hydraulic cover; locate
the existing capacitor.
C) Discharge the existing capacitor.
D) Disconnect the two wires from the existing capacitor.
E) Remove existing capacitor (it may be mounted with
a single screw or with double-sided tape).
WARNING
Use an insulated handle screwdriver to
jumper across both terminals to discharge
the capacitor before touching it. Failure to do so could
result in an electrical shock.