Perma Pure Baldwin-Series 33XP User manual

Type
User manual
P
PE
ER
RM
MA
A
P
PU
UR
RE
E
Perma Pure LLC
Tel: 732-244-0010
8 Executive Dr, PO Box 2105
Tel: 800-337-3762 (toll free US)
Toms River, NJ 08754
Fax: 732-244-8140
www.permapure.com
User Manual
HEATED FILTER PROBES
CLASS I DIVISION 1 SERIES
Model 33XP
SE-MAN-002
Table of Contents 2
Table of Contents
A: Safety Information & Warnings .................................................................................... 3
B: Warranty Information .................................................................................................. 4
C: Specifications ............................................................................................................. 5
D: Overview .................................................................................................................... 6
E: Installation................................................................................................................... 6
F: Maintenance ............................................................................................................... 9
G: Spare/Replacement Parts ......................................................................................... 10
Appendix: ...................................................................................................................... 11
Table of Contents 3
A: SAFETY INFORMATION & WARNINGS
Please be sure to review the following basic safety procedures. These procedures represent the MINIMUM
requirements to operate the equipment safely. It is the ultimate responsibility of the operator to ensure
proper safety practices are utilized at the point of operation.
NEVER attempt to operate this equipment in an explosive, wet, or otherwise hazardous area.
NEVER exceed any specified rating for the equipment. Voltage, temperature and pressure ratings must
be closely observed and not exceeded. Voltage rating of the equipment MUST match the rating on the
data label. Please make sure that it matches before powering up the equipment.
Condensate is potentially dangerous. NEVER handle drain lines, impingers or any other item that may
have come in contact with the gas stream or any hazardous material, without adequate personal
protective equipment. ALWAYS assume that any liquid present is hazardous.
Sample gas is potentially dangerous. A leak test is recommended at initial startup and as often as
necessary to maintain a safe working environment around the equipment. The gas stream exhaust must
exit away from all personnel to prevent dangerous exposure.
NEVER operate the equipment with any part of the enclosure unsecured. All operated doors and covers
must be in place and secured prior to operation. Electrical current may be present behind covers or
doors, even if tools are not necessary to access these components.
NEVER attempt service on this equipment without first disconnecting all energy sources. Repair of this
equipment should only be done by properly trained personnel that are familiar with the potential risks
involved with servicing of the equipment.
NEVER replace fuses with types other than the sample specification of type and current. Do not bypass
this or any other safety device.
NEVER operate this equipment if it is visibly damaged or the possibility exists that it may have been
damaged.
The use of components that have not been purchased through an authorized Perma Pure dealer or
directly from Perma Pure may compromise the safety of the operator. Additionally, use of non-
authorized components may change the operating characteristics of this equipment. Any changes to the
equipment, that modify its operation in any way, are dangerous, and are strictly prohibited.
Read the entire operating manual before attempting to set up or operate the equipment.
Please heed all warning labels on the equipment. They are there to remind you of possible hazardous
conditions.
Verify the integrity of any mechanical and/or electrical connections that are made to the unit.
Verify that the unit is connected to the properly rated power for the system.
Verify that the unit is plumbed properly to operate effectively.
Table of Contents 4
B: WARRANTY INFORMATION
Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform
to the published specifications of Seller and be free from defects in material and workmanship under normal
use and service. Seller’s sole obligation and liability under this warranty is limited to the repair or replacement
at its factory, at Seller’s option, of any such product which proves defective within one year after the date of
start-up (or within 18 months after original shipment at the discretion of Seller) and is found to be defective in
material or workmanship by Seller’s inspection.
Buyer agrees that (1) any technical advice, information, suggestions, or recommendations given to Buyer by
Seller or any representative of Seller with respect to the product or the suitability or desirability of the product
for an particular use or application are based solely on the general knowledge of Seller, are intended for
information guidance only, and do not constitute any representation or warranty by Seller that the product
shall in fact be suitable or desirable for any particular use or application; (2) Buyer takes sole responsibility for
the use and applications to which the product is put and Buyer shall conduct all testing and analysis necessary
to validate the use and application to which Buyer puts the product for which Buyer may recommend the use
or application of the product by others; and (3) the characteristics, specifications, and/or properties of the
product may be affected by the processing, treatment, handling, and/or manufacturing of the product by
Buyer or others and Seller takes no responsibility for he nature or consequence of such operations or as to the
suitability of the product for the purposes intended to be used by Buyer or others after being subjected to
such operations.
SELLER MAKES NO OTHER WARRANTY, EXPLICIT OR IMPLIED, OF THE PRODUCT SUPPLIED HEREUNDER,
INCLUDING, WITHOUT LIMITATION, IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR
PARTICULAR PURPOSE, AND ALL SUCH WARRANTIES ARE HEREBY EXPRESSLY EXCLUDED. SELLER SHALL HAVE
NO LIABILITY FOR LOSS OF PROFITS, OR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES UNDER ANY
CIRCUMSTANCES OR LEGAL THEORY, WHETHER BASED ON NEGLIGENCE, BREACH OF WARRANTY, STRICT
LIABILITY, TORT, CONTRACT, OR OTHERWISE. SELLER SHALL IN NO EVENT BE LIABLE IN RESPECT OF THIS
ORDER AND OR PRODUCT DELIVERED ON ACCOUNT OF THIS ORDER FOR ANY AMOUNT GREATER THAN THAT
PAID TO SELLER ON ACCOUNT OF THIS ORDER.
Section A: Specifications 5
C: SPECIFICATIONS
General Specifications
Probe Mounting
1 ¼” NPT male thread for connection to 150# flange adapter
Stinger Connection
½” NPT female threaded port
Calibration
Integral calibration port with check valve
Temperature Control
Self regulated at 392F (200C), by thermostat. 450F (232C)
over-temp thermostat
Connectors
¼" cal gas, ¼" sample line, ¼" blowback air
Thermocouple
Type K
O-rings
Viton®
Gaskets
Graphoil
Dimensions
14.13” x 10” x 13” in. HWD (From phalange face to probe cap )
(35.89 x 25.4 x 33.02 cm)
Weight
42 lbs
19 kg
Operating Specifications
Calibration Gas
20 psig, 6-10 LPM
Probe Operating Temp
392ºF (200ºC)
Electrical
115/230 VAC, 1 PH, 50/60 Hz, 200 watts
Material Specifications
Enclosure Material
UL® Classified Class I, Division 1 Groups B, C & D Aluminum
Heater Type
Cartridge style
Filter Chamber
Cast 316 stainless steel,
Filter Element
2,5,10,20 micron sintered SS
2 micron ceramic
2 micron SS screen mesh
0.7 micron glass/PTFE coated
Section C: Installation 6
D: OVERVIEW
The Baldwin™-Series Models 33XP Class I Division 1 Heated Filter Probe is designed to be mounted
on a stack or duct in low particulate applications. Its primary function is to provide a heated
environment to maintain sample gas temperatures above dew point and remove particulate
material from the gas sample. The 33XP feature a standard filter element, cartridge heaters, an
integral calibration gas port, and a UL® Classified Class I, Division 1 Groups B, C & D enclosure.
Section C: Installation 7
E: INSTALLATION
Mounting
The 33XP is designed to be mounted directly on a stack or duct with the included 1¼” Schedule 40
male pipe nipple. This pipe nipple can be screwed into a standard ASA flange, either flat or raised
face that has been modified with a 1¼” NPT threaded port in the center. These modified flanges are
available from Perma Pure in kit form. The kit includes the flange of the desired size and the
appropriate bolts and flange gasket.
1. Thread the flange onto the 1¼” filter nipple
2. Install the stinger pipe into the filter mounting nipple using PTFE tape or thread seal paste
that is appropriate for the application.
3. Mount the assembly to the stack flange with the gas connections facing in the desired
direction using the bolts and gasket provided.
4. Makeup the sample gas, calibration gas, and blowback air connections. Typical tightening
for ¼” compression fittings is 1 ¼ turns past finger tight. Do not over-tighten.
5. The sample gas line that runs from the probe to the analyzer area may need to be heated to
prevent condensation of water and acid vapors. Condensation in the sample line is likely to
cause analyzer inaccuracy, unreliable measurements and maintenance problems and must
be avoided.
See Appendix for images and drawings.
Section E: Maintenance 8
Electrical
Connection of the probe to the electrical must be performed by qualified personnel. Connect the
system to an appropriate earth ground. Adhere to all electrical code requirements in effect at the
installation site. This device is intended for Class 1, Div 1 Hazardous installations.
1. Refer to drawing 4P-33XP-04-02 in the appendix for connection to AC power and
thermocouple.
2. The K type thermocouple can be used to monitor the temperature of the probe or provide a
signal to an external temperature controller if desired.
Plumbing
Section E: Maintenance 9
F: MAINTENANCE
General
The 33XP filter assembly and heating system do not require routine maintenance.
The filter element requires periodic replacement, depending upon application and dust loading. See
the attached spare parts list for replacement elements.
If the 33XP is used in conjunction with the Baldwin-Series Flow Control Drawer, monitoring the
sample vacuum will allow the operator to determine when to change the filter element. The
operator can log the baseline vacuum reading when the system is first started with a new filter
element.
If the sample vacuum is consistently 25% higher than at start-up, the operator should replace the
filter element with a new filter. Visual inspection will also confirm the condition of the filter
element.
Element Replacement
The filter incorporates a “jack screw” that helps to remove the filter cap in case it becomes stuck
due to buildup of debris inside the housing. Turning the thumbwheel counterclockwise will force the
cap away from the housing and allow for easy removal of the element.
Section G: Spare Parts 10
G: SPARE/REPLACEMENT PARTS
Model 33XP (Part Number 4P-33XP)
Part No.
Description
3FES-015PK
Filter Element Seals: Silicone, Used w/ Screen Mesh 3FES-010 (10 pack)
3FES-010
Filter Element: 316L SS Screen Mesh, 2.0 Micron
3FES-004
Filter Element: 316L SS, 1.25” x 2.975”, 10 Micron
3FES-003
Filter Element: 316L SS, 1.25” x 2.975”, 20 Micron
3FES-005
Filter Element: 316L SS, 1.25” x 2.975”, 5 Micron
3FES-002
Filter Element: 316L SS, 1.25” x 2.975”, 2 Micron
3FEC-002
Filter Element: Ceramic 2 Micron
3FEG-001
Filter Element: Glass, 0.1 Micron
3FEG-003
Filter Element: Glass/TFE Coated, 0.7 Micron
3FET-002
Filter Element: 2 micron Teflon®
4P-FLANGE2
Flange: 2”, 150# with Gasket & Bolts
4P-FLANGE3
Flange: 3”, 150# with Gasket & Bolts
4P-FLANGE4
Flange: 4”, 150# with Gasket & Bolts
4P-FLANGE6
Flange: 6”, 150# with Gasket & Bolts
4P-GCS-212
Gas Cooling Spool Piece: w/ 2” Flanges & 12” Spool
4P-GCS-312
Gas Cooling Spool Piece: w/ 3” Flanges & 12” Spool
4P-GCS-412
Gas Cooling Spool Piece: w/ 4” Flanges & 12” Spool
4P-GCS-206
Gas Cooling Spool Piece: w/ 2” Flanges & 6” Spool
4P-GCS-306
Gas Cooling Spool Piece: w/ 3” Flanges & 6” Spool
4P-GCS-406
Gas Cooling Spool Piece: w/ 4” Flanges & 6” Spool
3PAM-006PK
Filter Element Gasket: Grafoil 1.25” (10 pack)
3PAM-031PK
O-Ring: Pack, Viton, “C” series probes only, 5 ea 1⅞” OD, 2 ¼” OD
3PHS-504
Thermostat, High Limit N.C.
3PHS-503
Thermostat, Control N.C.
3PHS-501
Heater, Cartridge, ⅜” x 3” 40
1TTC-015
Type K, 24” lead, SS
Stingers are available for order. Call factory for details.
Perma Pure LLC
Tel: 732-244-0010
8 Executive Dr, PO Box 2105
Tel: 800-337-3762 (toll free US)
Toms River, NJ 08754
Fax: 732-244-8140
www.permapure.com
Appendix 11
APPENDIX:
Drawings
4P-33XP-01-01 Stack Installation Diagram
4P-33BXP-01-01 Dimensional Drawing
4P-33XP-03-02 Assembly
4P-33XP-05-01 Tubing Drawing
4P-33XP-04-03 Electrical Wiring Diagram
(Attach drawings here)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11

Perma Pure Baldwin-Series 33XP User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI