Midmark 641 Power Procedures Chair Installation guide

Type
Installation guide
1
© 2007 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | www.midmark.com
Procedures Chair Installation
Applies to Models:
230 (all)
630 (-001 thru -009)
641 (all)
Language or origin: English
003-1406-00 Rev. LA1 10/24/18
Step 1: Cut cable ties.
Remove shipping brace.
Models w/Rotational Base only
Step 2: Remove shipping bracket.
115V
MA644500i
230V
115V
Does the voltage setting that appears in the
fuse holder window match the facility supply
voltage?
If YES, go to Step 4.
If NO, you must perform
Voltage Conversion Procedure.
Step 3: Remove shipping bolts.
Remove chair from skid.
Equipment Alert
Voltage setting that shows thru
window must match supply voltage. Incorrect
setting will result in damage to the chair.
Voltage Conversion Procedure:
A. Open fuse holder cover.
B. Remove fuse holder .
C. Flip fuse holder 180°.
(630 shown)
(630 shown)
NOTE:
Shipping brace may not be present on some models.
Style A
2
© 2007 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | www.midmark.com
MA8330i
MA642802i
Rear Cord Inlet
Front Cord Inlet
Step 4: Connect hand/foot control to cord inlet.
Tighten locking ring.
Note: The cord connectors are “keyed” and must
be oriented properly to connect.
Step 5: Connect power cord.
Equipment Alert
The chair must be level to ensure proper operation.
Step 6: Adjust four leveling feet
until chair is level.
Step 7: Install hand control mounting
discs (2), and holster.
Locking Ring
Attention
Slide narrow portion of cord
connector into bracket as shown.
Midmark models only
Step 8: Install magnetic
base covers (3 pcs.)
(630 shown)
Before connecting the Midmark device power cord into an electrical
power source outlet, it is recommended to contact a local licensed
electrician or a local Authority Having Jurisdiction over the internal
building power distribution to verify that the device when powered
up will comply with all local electrical codes. Use the electrical
power rating marked on the device when determining the
appropriate electrical branch circuit and outlet requirements.
For 115V rated device models with a current rating at or above
12 Amps, a dedicated branch circuit may be required if the branch
circuit protection is provided by a 15 Amp breaker.
  • Page 1 1
  • Page 2 2

Midmark 641 Power Procedures Chair Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI

Other documents