1. ShutdowntheBDAccuriC6.
2. UnscrewtheBDAccuriC6SIPcollar,removethesamplestageandinstalltheLow-ProleBDCSamplerSIPcollar(Figures
1-3).Thelow-proleBDCSamplerSIPcollarensuresthatcollisionsdonotoccurduringoperationoftheBDCSampler.
Figure1
Figure2 Figure3
BD CSampler™ Installation Guide
BD CSampler Accessory Kit
Shipping Contents
• BDCSamplerModule
• BDCSamplerMat
• MountingBolts(3)
• InstallationandRemovalTool
• BDCSamplerSIPCollar
• BDCSamplerSoftwareCD
(notshown)
• 24-tuberack(notshown)
Install and validate the BD Accuri
C6 Flow Cytometer
with 6- and 8-peak beads prior to installing the BD
CSampler.InspecttheBDCSamplerAccessoryKitcontentstoensuredeliveryofallcomponents.
BDCSamplerAccessoryKit
BDCSampler
Module
Mat
Tool
BDCSampler
SIPCollar
Bolts
WARNING Failuretoshut
downtheBDAccuriC6duringBD
CSamplerinstallationcouldresultin
damagetoeitherorbothelectronic
modules.
Low-ProleBDCSampler
SIPcollar
BDC6SIPcollar
www.AccuriCytometers.comForResearchUseOnly.Notforuseindiagnosticprocedures.Class1LaserProduct7820020Rev-3
BD CSampler™ Installation Guide
5. AlignthefrontoftheBDCSamplersothattheholesonthemodulearelinedupwiththethreethreadedholesintheBD
AccuriC6.HoldthemodulewithonehandandscrewinthefrontrightmountingbolttosecuretheBDCSamplertotheBD
AccuriC6,tighteningwiththetoolprovided(Figure6).
6. Screwintheothertwomountingbolts.WARNINGDonotovertighten(Figure7).
Figure7
Figure8
Figure9
Figure10
Figure4 Figure5
Figure6
Tab
Holes
8. Placethematinposition(Figure10).WARNINGFailuretokeepthematclearof
plates,tubesoranyothermaterialsmayresultindamagetotheBDCSampler.
9. InstallBDCSamplersoftware.
10. TurnontheBDAccuriC6.
11. Runvalidationbeadstoensurecorrectoperation.RefertoBDAccuriC6Software
UserGuide,Chapter1.
3. OpenthelidoftheBDAccuriC6andlocatetheholesusedtomounttheBDCSampler(Figure4).
4. HoldtheBDCSamplermodulewithbothhandsandlocatethetabintheslotonthebackoftheBDAccuriC6(Figure5).
7. ConnectthecabletothesocketatthebackoftheBDAccuriC6andswitchontheBDAccuriC6.TheBDCSamplerwillauto-
maticallyalignitselfandthenremaininthehomeposition(Figure8-9).
www.AccuriCytometers.comForResearchUseOnly.Notforuseindiagnosticprocedures.Class1LaserProduct7820020Rev-3
/