BD FACSCanto II
Flow Cytometer
Reference Manual
© 2006, Becton, Dickinson and Company. All rights reserved. No part of this publication may be reproduced,
transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any
form or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without prior
written permission from BD Biosciences.
The information in this guide is subject to change without notice. BD Biosciences reserves the right to change its
products and services at any time to incorporate the latest technological developments. Although this guide has been
prepared with every precaution to ensure accuracy, BD Biosciences assumes no liability for any errors or omissions,
nor for any damages resulting from the application or use of this information. BD Biosciences welcomes customer
input on corrections and suggestions for improvement.
BD FACSDiva software © Becton, Dickinson and Company. This software is the property of Becton, Dickinson and
Company. Each sale of a stored unit of this software grants the purchaser a nontransferable, nonexclusive, personal
license. This software may not be duplicated, reproduced, or copied in any form or by any means whatsoever, except
as otherwise permitted by law.
BD FACSCanto clinical software © Becton, Dickinson and Company. This software is the property of Becton,
Dickinson and Company. Each sale of a stored unit of this software grants the purchaser a nontransferable,
nonexclusive, personal license. This software may not be duplicated, reproduced, or copied in any form or by any
means whatsoever, except as otherwise permitted by law.
BD, the BD logo, BD Calibrite, BD FACS, BD FACSCalibur, BD FACSCanto, BD FACSDiva, BD FACSFlow,
BD Falcon, BD Multiset, BD Multitest, and BD Trucount are trademarks of Becton, Dickinson and Company.
JDS Uniphase is a trademark of JDS Uniphase, Inc.
Alexa Fluor and Texas Red are registered trademarks, and Pacific Blue is a trademark of Molecular Probes, Inc.
iFLEX is a trademark of Point Source, Ltd.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Sapphire is a trademark and Coherent is a registered trademark of COHERENT, INC.
Teflon is a registered trademark of E.I. du Pont de Nemours and Company.
All other company and product names might be trademarks of the respective companies with which they are
associated.
Patents
PE and APC: US 4,520,110; 4,859,582; 5,055,556; Europe 76,695; Canada 1,179,942
PerCP: US 4,876,190
Cy5.5 and Cy7: US 5,268,486; 5,486,616; 5,569,587; 5,569,766; and 5,627,027
Pe-Cy7: US 4,542,104
APC-Cy7: US 5,714,386
FCC Information
WARNING: Changes or modifications to this unit not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
NOTICE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant
to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction manual, can cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his or her own expense.
Shielded cables must be used with this unit to ensure compliance with the Class A FCC limits.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Réglement sur the matériel brouilleur du
Canada.
Notice
BD Biosciences delivers software and workstations that are intended for running the instruments supplied by
BD Biosciences. It is the responsibility of the buyer/user to ensure that all added electronic files including software
and transport media are virus free. If the workstation is used for Internet access or purposes other than those specified
by BD Biosciences, it is the buyer/user’s responsibility to install and maintain up-to-date virus protection software.
BD Biosciences does not make any warranty with respect to the workstation remaining virus free after installation.
BD Biosciences is not liable for any claims related to or resulting from buyer/user's failure to install and maintain
virus protection.
History
Revision Date Change Made
640806 Rev. A 5/06 Initial release
v
Contents
About This Manual ix
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . xi
Chapter 1: Introduction 13
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Flow Cytometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Fluidics Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Computer Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
BD FACS Loader (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Barcode Reader (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Chapter 2: BD FACS Loader Option 33
Loader Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Using the Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Opening and Closing the Loader Doors . . . . . . . . . . . . . . . . . . . . . . . . . 36
Acquiring with the Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Changing Operational Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Maintaining the Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Spill Containment Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Cleaning External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
vi BD FACSCanto II Flow Cytometer Reference Manual
Chapter 3: Barcode Reader Option 45
Installing and Using the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Cleaning the Barcode Reader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
1D Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
2D Barcode Symbologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Chapter 4: Maintenance 51
Scheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Emptying the Waste Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Purging the Fluidics Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Decontaminating the Fluidics System (Long Clean) . . . . . . . . . . . . . . . . 59
Replacing the Air Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Replacing Fluidics Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Unscheduled Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Changing a Cubitainer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Removing an Air Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Cleaning External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Removing Bubbles from the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Purging the Bubble Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Decontaminating the Fluidics System for Storage . . . . . . . . . . . . . . . . . . 79
Replacing the Bal Seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Resetting the Cytometer Circuit Breaker . . . . . . . . . . . . . . . . . . . . . . . . . 85
Reconnecting the Ethernet and RS232 Cables . . . . . . . . . . . . . . . . . . . . . 86
Reconnecting the Fluidics Cart Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Replacing the Fluidics Level Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Replacing the Fluidics Cart Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Contents vii
Chapter 5: Troubleshooting 95
Instrument Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Fluidics Cart Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Loader Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
Appendix A: Technical Overview 105
Flow Cytometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Fluidics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Optics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Light Scatter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Fluorescence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Optical Filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Detector Arrays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Spillover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Electronics System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Pulses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Pulse Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Appendix B: Optics Configurations 125
4-2 Optics Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
4-2-2 Optics Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
5-3 Optics Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
viii BD FACSCanto II Flow Cytometer Reference Manual
Appendix C: Supplies and Replacement Parts 139
Instrument Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Installation Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
Other Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Bar Code Reader Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
Labware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Appendix D: Technical Specifications 145
Cytometer Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Optics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Fluidics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Signal Processing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Fluidics Cart Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
BD FACS Loader Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Index 153
ix
About This Manual
This manual contains reference information about the BD FACSCanto™ II flow
cytometer and the BD FACS™ Loader. Operating instructions are contained in
the BD FACSCanto II Instructions for Use.
Most instrument functions are controlled by BD FACSCanto™ clinical software
and BD FACSDiva™ software. BD FACSCanto clinical software contains
modules for dedicated clinical applications with automatic gating algorithms,
while BD FACSDiva software is non–application specific. You can use either
software to perform instrument quality control. Each software package has its
own reference manual, provided on the documentation CD.
BD Biosciences recommends that first-time users of this instrument take
advantage of operator training offered with the sale of every new instrument.
The BD FACSCanto II Flow Cytometer Reference Manual assumes you have a
working knowledge of basic Microsoft® Windows® operation. If you are not
familiar with the Windows operating system, refer to the documentation
provided with your computer.
x BD FACSCanto II Flow Cytometer Reference Manual
Conventions
The following tables list conventions used in this manual. Table 1 lists symbols
that are used in this manual or on safety labels to alert you to a potential hazard.
Text and keyboard conventions are shown in Table 2.
Table 1 Hazard symbols
a
a. Although these symbols appear in color on the instrument, they are in black and white throughout this
reference manual; their meaning remains unchanged.
Symbol Meaning
CAUTION: hazard or unsafe practice that could result in material damage,
data loss, minor or severe injury, or death
Electrical danger
Laser radiation
Biological risk
Table 2 Text and keyboard conventions
Convention Use
! Tip
Features or hints that can save time or prevent difficulties
NOTICE
Provides information that supplements the topic material
Italics Book titles and new or unfamiliar terms
> Menu selection
For example, “select File > Print” means select Print from the File
menu.
Ctrl-X Press the indicated keys simultaneously
For example, Ctrl-P means hold down the Control key while
pressing the letter p.
About This Manual xi
Technical Assistance
For technical questions or assistance in solving a problem:
Read the section of the manual specific to the operation you are
performing.
See Chapter 5, Troubleshooting.
If additional assistance is required, contact your local BD Biosciences technical
support representative or supplier.
When contacting BD Biosciences, have the following information available:
product name, part number, and serial number
software version number
•error messages
details of system performance
For instrument support within the US, call (877) 232-8995.
For support within Canada, call (888) 259-0187.
Customers outside the US and Canada, contact your local BD representative or
distributor.
xii BD FACSCanto II Flow Cytometer Reference Manual
13
1
Introduction
The BD FACSCanto II system uses a fixed-optics design and advanced digital
electronics to support multicolor analysis of up to eight fluorescent markers and
two scatter parameters per assay.
The instrument does not require special facilities—it plugs into a standard wall
outlet, uses air-cooled lasers, and provides its own air pressure and vacuum for
the fluidics and waste. You can prepare samples on the BD FACS™ Sample Prep
Assistant II and import the sample prep worklist.
For further automation, use BD FACSCanto clinical software and the BD FACS
Loader for sample acquisition. Alternatively, use BD FACSDiva software for
more flexibility in acquisition and analysis.
See the following for more information:
Intended Use on page 14
System Components on page 15
System Requirements on page 31
14 BD FACSCanto II Flow Cytometer Reference Manual
Intended Use
The BD FACSCanto II system is intended for use as an In Vitro Diagnostic device
for identification and enumeration of lymphocyte subsets in human cells in
suspension for flow cytometry.
Chapter 1: Introduction 15
System Components
The BD FACSCanto II system consists of a benchtop flow cytometer, a self-
contained fluidics cart, and the BD FACSCanto II workstation. System options
include an automated sample loader and a barcode reader.
For a description of system components, see:
Flow Cytometer on page 16
Fluidics Cart on page 26
Computer Workstation on page 29
BD FACS Loader (Optional) on page 29
Barcode Reader (Optional) on page 30
16 BD FACSCanto II Flow Cytometer Reference Manual
Flow Cytometer
Cytometer power is controlled by the power button. All other cytometer and
fluidics cart functions are controlled by BD FACSCanto clinical software and
BD FACSDiva software (provided with the instrument).
Figure 1-1 BD FACSCanto II flow cytometer
The BD FACSCanto II flow cytometer consists of a fluidics subsystem, an optics
subsystem, and an electronics subsystem. For a more in-depth discussion of
fluidics, optics, electronics, and flow cytometry, see the Technical Overview on
page 105.
Fluidics
The fluidics system consists of the sample injection tube (SIT), the aspirator arm,
the flow cell, a pressurized interior reservoir (plenum), and a network of tubing
that provides sheath and cleaning fluids to and removes waste from the flow cell.
data ports
side door
flow cell
access door
optics
access door
sample
injection
tube
fluidics cart
connections
power button
acquisition
indicator lights
Chapter 1: Introduction 17
The following table briefly describes these components.
flow cell where laser beam intercepts particles
tubing brings sheath and cleaning fluids to and aspirates waste from flow cell
SIT hollow metal tube that brings sample to flow cell
aspirator arm movable waste aspiration port
aspirator arm
bar
metal bar used to push aspirator arm away from SIT during manual
loading
tube guide helps guide tube onto the SIT (Loader operation only)
tube eject
cylinder
ejects tube (Loader operation only)
tube sensor detects tube position on SIT
aspirator arm
tubing
aspirator arm bar
flow cell
tube guide
(Loader option only)
sample injection tube (SIT)
tube eject cylinder
(used with
Loader option only)
tube sensor
Bal seal access
18 BD FACSCanto II Flow Cytometer Reference Manual
When you install tubes onto the SIT, a pump within the fluidics cart pressurizes
the plenum, which then provides sheath fluid to the flow cell. At the same time,
the sample tube is pressurized and sample is pushed up the SIT and into the flow
cell.
When you remove tubes from the SIT, the cytometer cleans the SIT by flushing
sheath solution down the inside and outside of the tube. The flushed sheath
solution is aspirated by the aspirator arm.
SIT cleaning between tubes is automatic when you use BD FACSCanto clinical
software. In BD FACSDiva software, SIT cleaning between tubes is automatic
unless you disable it by deselecting the SIT Flush checkbox on the Acquisition
Dashboard.
Do not leave a tube of distilled water on the SIT between sample tubes, or during
or after daily shutdown.
Never install a tube onto the SIT that contains more than 3 mL of fluid. This will
cause the tube sensor to become wet and fail to properly sense tubes.
Flow Cell
Once the sample moves into the flow cell, particles move in single file through the
laser beams. The scattered and emitted light from these particles provides
information about their size, shape, granularity, and fluorescence properties.
The flow cell is behind the flow cell access door. For more information, see
Fluidics System on page 106.
Chapter 1: Introduction 19
Figure 1-2 Flow cell
Optics
The optics system in the BD FACSCanto II cytometer is composed of both
excitation optics and collection optics. Excitation optics bring light to the flow
cell. Collection optics gather the light signals emitted or scattered by the particles.
Optics Configurations
Several optics configurations are available for the BD FACSCanto II cytometer.
The 4-2 configuration can collect four fluorescence signals from a blue laser
source and two fluorescence signals from a red laser source. The 4-2-2
configuration adds a violet laser and can collect two fluorescence signals from the
violet laser source. The 5-3 configuration uses only the blue and red lasers, but
can collect an additional fluorescence signal from each laser source.
The excitation and collection optics components for the 4-2 configuration are
described in this section. The additional optics configurations and their
components are described in Appendix B, Optics Configurations.
FSC diode
where lasers
intercept
sample stream
flow cell
obscuration
bar
20 BD FACSCanto II Flow Cytometer Reference Manual
Excitation Optics
The excitation optics consist of lasers, fiber optic cables, beam-shaping prisms,
and an achromatic focusing lens, as shown in Figure 1-3.
The BD FACSCanto II instrument uses low-powered air-cooled and solid state
lasers that do not have special power and cooling requirements.
Fiber optic cables direct the laser light onto beam-shaping prisms, which in turn
transmit the laser light to a focusing lens. The lens directs the laser light onto the
sample stream within the flow cell (Figure 1-3). See also Figure A-3 on page 109.
Figure 1-3 Optical pathway (figure shows violet laser available with 4-2-2 configuration only)
When the flow cell access door opens, an interlock shutters the laser light and
blocks its pathway for safety reasons.
Table 1-1 4-2 configuration fluorochromes
Laser
Wavelength
(nm)
Commonly Used Fluorochromes
Coherent® Sapphire™
Solid State
488 (blue) FITC, PE, PE-Cy7,
PerCP or PerCP-Cy5.5
JDS Uniphase™ HeNe
Air Cooled
633 (red) APC, APC-Cy7
1
2
3
4
interlock focusing lens
prisms flow cell
1
3
2
4
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158