Site Preparation Guide
Phoenix BLACK™ & Optional Upgrades FM0071 Rev. 3
Page 3 of 3
© 2021 CEM Corporation. All rights reserved.
Electrical Requirements
• Region Specic
• 208/230 V, 60 Hz, 15 A @ 230 V
• Line voltage must be within 207 - 253 V
• Instrument is supplied with a NEMA 6 -15 Plug
• Receptacle, non locking
• 220/240 V, 50 Hz, 15 A @ 240 V
• Line voltage must not vary more than ± 10 % of its specied level
• 200 V, 50 Hz, 15 A
• Line voltage must not vary more than ± 10 % of its specied level
• Dedicated, grounded outlet no more than 8’ (2.5 m) from unit
• Instrument is supplied with a region-specic power cord or an adequately rated
power cord may be used
Compressed Air (Airwave ONLY)
• Dedicated compressed air line providing 90 psi (6 atm) at 360 L/min (13 ft2/min)
• 5∕16" (8 mm) diameter airline ending in an “M” style female 1∕4" coupler
• If the Phoenix AirWave is more than 50 feet from the compressed air supply, the air line should be increased to 5/8" ID.
Compressed air supply should be increased if the exhaust hose is more than 16'.
Environmental Requirements
• Indoor use only
• Altitude up to 6562’ (2000 m)
• A Temperature range of 41 °F (5 °C) to 104 °F (40 °C)
• Relative humidity range of 10 % to 85 %, non-condensing
• Overvoltage Category II
• Pollution degree 2
Plug Receptacle
NEMA 6-15P
NEMA 6-15R
NEMA 6-20R