5BLADE Setup Guide
Software Setup
1. With the instrument facing forward, position the power switch to ON. Make sure the power switch can easily be ac-
cessed. DO NOT REMOVE any packing material until instructed to do so.
2. Remove the blue tape from the door and interlock cover.
3. Follow the instructions on the BLADE display once the system is powered on to remove the packing foam within the
fume cabinet. DO NOT REMOVE any packing material until instructed to do so. Retain all packing material for use if
returning the instrument.
4. Remove the blue tape from the exhaust cover.
5. Perform an “Exhaust Validation” test to determine if adequate exhaust is achievable.
5.1. Press the System Menu icon located in the top right corner of the screen.
5.2. Press “Tools” and the “Utilities” tab.
5.3. Locate “Exhaust Validation” and press “Start.”
5.4. Follow instructions located on the screen to complete verication.
5.5. If any of the validation points result in a red “X,” verify that the exhaust requirements have been met and view
the BLADE manual (P/N 601477).
WARNING
DO NOT attempt unit operation if the “Exhaust Validation” test results in a red “X.”
6. Perform an “Autosampler Validation” test to verify pre-determined rack positions and the attenuator.
6.1. Press the System Menu icon located in the top right corner of the screen.
6.2. Press “Tools” and the “Utilities” tab.
6.3. Locate Autosampler and press “Validate.”
6.4. Follow instructions located on the screen to complete calibration.
6.5. If the validation fails, perform an Autosampler Calibration as outlined in the “Calibrations” section of the manu-
al (P/N 601477).
7. Scan the barcode located on the Cap Assembly box which includes the caps and septa as outlined below.
7.1. Ensure that the barcode scanner is connected to the BLADE.
7.2. Locate the box containing the cap and septa.
7.3. Press the System Menu icon in the top right corner of the display.
7.4. Press “Settings” and then press the Run tab.
7.5. Press the barcode icon.
7.6. Use the barcode scanner to scan the barcode on the cap and septa box or the numerical number located be-
neath the barcode can be manually entered. This code can only be used once.
7.7. Press “OK.” The cap count in the settings will increase by 250.
8. Perform a “QC Test”. View the “QC Test” method notes for details regarding sample preparation and the “Perform a
Digestion” section of the manual (P/N 601477) for details about vessel assembly, loading and performing the method.
9. Watch instructional videos found within the BLADE software as outlined below to customize system and user settings.
9.1. Press the System Menu icon in the top right corner of the screen.
9.2. Press “Video.”
10. Read and fully understand the BLADE manual (P/N 601477) before operating the instrument.
11. System setup complete. The BLADE is ready for operation.