CEM Phoenix BLACK Nitrogen Purge Operating instructions

Type
Operating instructions
Nitrogen Purge Instructions
Phoenix BLACK
601291 Rev.1
© 2020 CEM Corporation. All rights reserved.
Process.suppor[email protected] | 800-726-3331
Introduction
The Nitrogen Purge Kit allows for an inert airow within the furnace and is compatible with the Phoenix BLACK, Phoenix
BLACK AirWave and Phoenix BLACK Sulfated Ashing systems.
Depending on the Phoenix BLACK conguration, this document should be used in conjunction with the Phoenix BLACK
Manual (P/N 600977) or Phoenix BLACK Sulfate Ash Manual (P/N 600978). Read and fully understand all documentation
before operating the instrument.
Installation
1. Carefully insert the quartz glass tube through the hole in the thermo-
couple access door. Place the tube down into the furnace until it
stops.
2. Position the elbow towards the back of the system.
3. Carefully place the ¼” connector onto the quartz glass tube and
tighten the tting until resistance is felt.
4. Use the 1/8" connector to connect to a nitrogen source.
5. Set the nitrogen pressure at 30 psi. Adjust pressure as needed to
meet desired ashing conditions.
  • Page 1 1