FLIR IM75 User manual

Category
Multimeters
Type
User manual
USERMANUAL
FLIRMODELIM75
InsulationTester/MultiMeter
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
2
TableofContents
1.DISCLAIMERS4
1.1Copyright4
1.2QualityAssurance4
1.3Documentation4
1.4DisposalofElectronicWaste4
2.SAFETY5
2.1FCCCompliance6
2.2IndustryCanadaCompliance7
3.INTRODUCTION8
3.1KeyFeatures8
4.METERDESCRIPTION9
4.1MeterSections9
4.2FunctionSwitchPositions10
4.3FunctionButtons,SelectorPad,andRotarySwitch10
4.3.1SelectorPad11
4.3.2RotarySwitch11
4.4DisplayDescription12
4.5DisplayIconsandIndicators13
5.OPERATION14
5.1PoweringtheMeter14
5.1.1AutoPowerOff14
5.2ManualAC/DCSelection14
5.3AutoRange14
5.4‘Smart’OutofRangeWarnings14
5.5DataHoldandAutoHoldModes14
5.5.1DataHoldMode14
5.6VoltageMeasurements15
5.7InsulationResistanceMeasurements15
5.7.1PI/DARTest15
5.8EarthBondResistanceMeasurements(ZEROΩ)16
5.9ContinuityTest16
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
3
5.10DiodeTest17
5.11CapacitanceMeasurements17
5.12ExtendedFunctionality17
5.12.1SelectingModes17
5.12.2VFDMode(ACVonly)18
5.12.3FrequencyMode(ACVonly)18
5.12.4Pass/FailMode(COMPARE)18
5.12.5SetupUtility18
5.12.6SilentMode19
5.12.7DataRecordingMode(99point)19
5.13KeypadLockout19
5.14StreamingMeasurementDatausingBluetooth®19
5.14.1General19
5.14.2Procedures19
6.MAINTENANCE20
6.1CleaningandStorage20
6.2BatteryReplacement20
6.3FuseReplacement20
6.4DisposalofElectronicWaste20
7.SPECIFICATIONS21
7.1Generalspecifications21
7.2ElectricalRangeSpecifications22
7.3InputSpecifications24
8.TECHNICALSUPPORT25
9.WARRANTIES25
9.1FLIRGlobalLimitedLifetimeWarranty25
9.2FLIRTestandMeasurementLimited2YearWarranty26
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
4
1. Disclaimers
1.1Copyright
©2014,FLIRSystems,Inc.Allrightsreservedworldwide.Nopartsofthesoftwareincludingsource
codemaybereproduced,transmitted,transcribedortranslatedintoanylanguageorcomputer
languageinanyformorbyanymeans,electronic,magnetic,optical,manualorotherwise,without
thepriorwrittenpermissionofFLIRSystems.
Thedocumentationmustnot,inwholeorpart,becopied,photocopied,reproduced,translatedor
transmittedtoanyelectronicmediumormachinereadableformwithoutpriorconsent,inwriting,
fromFLIRSystems.
Namesandmarksappearingontheproductshereinareeitherregisteredtrademarksor
trademarksofFLIRSystemsand/oritssubsidiaries.Allothertrademarks,tradenamesorcompany
namesreferencedhereinareusedforidentificationonlyandarethepropertyoftheirrespective
owners.
1.2QualityAssurance
TheQualityManagementSystemunderwhichtheseproductsaredevelopedandmanufactured
hasbeencertifiedinaccordancewiththeISO9001standard.
FLIRSystemsiscommittedtoapolicyofcontinuousdevelopment;thereforewereservetheright
tomakechangesandimprovementsonanyoftheproductswithoutpriornotice.
1.3Documentation
Toaccessthelatestmanualsandnotifications,gototheDownloadtabat:http://support.flir.com.
Itonlytakesafewminutestoregisteronline.Inthedownloadareayouwillalsofindthelatest
releasesofmanualsforourotherproducts,aswellasmanualsforourhistoricalandobsolete
products.
1.4DisposalofElectronicWaste
Aswithmostelectronicproducts,thisequipmentmustbedisposedofinan
environmentallyfriendlyway,andinaccordancewithexistingregulationsfor
electronicwaste.
PleasecontactyourFLIRSystemsrepresentativeformoredetails.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
5
2. Safety
SafetyNotes
Beforeoperatingthedevice,youmustread,understand,andfollowallinstructions,dangers,
warnings,cautions,andnotes.
FLIRSystemsreservestherighttodiscontinuemodels,partsoraccessories,andotheritems,orto
changespecificationsatanytimewithoutpriornotice.
Removethebatteriesifthedeviceisnotusedforanextendedperiodoftime.
WarningStatements
Donotoperatethedeviceifyoudonothavethecorrectknowledge.Formalqualificationsand/or
nationallegislationfortheelectricalinspectionscanapply.Incorrectoperationofthedevicecan
causedamage,shock,injuryordeathtopersons.
Donotstartthemeasuringprocedurebeforeyouhavesetthefunctionswitchtothecorrect
position.Thiscancausedamagetotheinstrumentandcancauseinjurytopersons.
Donotchangetoresistancewhenyoumeasurethevoltage.Thiscancausedamagetothe
instrumentandcancauseinjurytopersons.
Youmustdisconnectthetestleadsfromthecircuitthatyoudidatestonbeforeyouchangethe
range.Ifyoudonotdothis,damagetotheinstrumentandinjurytopersonscanoccur.
Donotreplacethebatteriesbeforeyouremovethetestleads.Thiscancausedamagetothe
instrumentandcancauseinjurytopersons.
Donotusethedeviceifthetestleadsand/orthedeviceshowsignsofdamage.Injurytopersons
canoccur.
Becarefulwhenyoudothemeasurementsifthevoltagesaremorethan25VACrmsor35VDC.
Thereisariskofshockfromthesevoltages.Injurytopersonscanoccur.
Donotdodiode,resistanceorcontinuitytestsbeforeyouhaveremovedthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
Donotusethedeviceasatooltoidentifyliveterminals.Youmustusethecorrecttools.Injuryto
personscanoccurifyoudonotusethecorrecttools.
Makesurethatchildrencannottouchthedevice.Thedevicecontainsdangerousobjectsandsmall
partsthatchildrencanswallow.Ifachildswallowsanobjectorapart,speakwithaphysician
immediately.Injurytopersonscanoccur.
Donotletchildrenplaywiththebatteriesand/orthepackingmaterial.Thesecanbedangerous
forchildreniftheyusethemastoys.
Donottouchexpiredordamagedbatterieswithoutgloves.Injurytopersonscanoccur.
Donotcauseashortcircuitofthebatteries.Thiscancausedamagetotheinstrumentandcan
causeinjurytopersons.
Donotputthebatteriesintoafire.Injurytopersonscanoccur.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
6
Cautions
Donotusethedeviceforaprocedurethatitisnotmadefor.Thiscancausedamagetothe
protection.
Thissymbol,adjacenttoanothersymbolorterminal,indicatesthattheuser
mustrefertothemanualforfurtherinformation.
Thissymbol,adjacenttoaterminal,indicatesthat,undernormaluse,
hazardousvoltagesmaybepresent.
Doubleinsulation.
ULlistingisnotanindicationoraverificationoftheaccuracyofthemeter
2.1FCCCompliance
Thisdevicecomplieswithpart15oftheFCCRules.Operationissubjecttothefollowingtwo
conditions:
1. Thisdevicemaynotcauseharmfulinterference.
2. Thisdevicemustacceptanyinterferencereceived,includinginterferencethatmaycause
undesiredoperation.
ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassBdigitaldevice,
pursuanttopart15oftheFCCRules.Theselimitsaredesignedtoprovidereasonableprotection
againstharmfulinterferenceinaresidentialinstallation.Thisequipmentgenerates,uses,andcan
radiateradiofrequencyenergyand,ifnotinstalledandusedinaccordancewiththeinstructions,
maycauseharmfulinterferencetoradiocommunications.However,thereisnoguaranteethat
interferencewillnotoccurinaparticularinstallation.Ifthisequipmentdoescauseharmful
interferencetoradioortelevisionreception,whichcanbedeterminedbyturningtheequipment
offandon,theuserisencouragedtotrytocorrecttheinterferencebyoneormoreofthe
followingmeasures:
1. Reorientorrelocatethereceivingantenna.
2. Increasetheseparationbetweentheequipmentandreceiver.
3. Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichthereceiveris
connected.
4. Consultthedealeroranexperiencedradio/TVtechnicianforhelp.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
7
WARNING
Changesormodificationsnotexpresslyapprovedbythepartyresponsibleforcompliance
couldvoidtheuser'sauthoritytooperatetheequipment.
2.2IndustryCanadaCompliance
Thisdevicecomplieswithpart15oftheFCCRules.Operationissubjecttothefollowingtwo
conditions:
ThisdevicecomplieswithIndustryCanadalicenseexemptRSSstandard(s).Operationissubjectto
thefollowingtwoconditions:(1)thisdevicemaynotcauseinterference,and(2)thisdevicemust
acceptanyinterference,includinginterferencethatmaycauseundesiredoperationofthedevice.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
8
3. Introduction
ThankyouforselectingtheFLIRIM75InsulationMultiMeter.Thisdeviceisshippedfullytestedand
calibratedand,withproperuse,willprovideyearsofreliableservice.
3.1KeyFeatures
4000countextralargedigitaldualdisplaywithbacklighting
AutoRangewithsmartoverrangeindication
AutoHoldandStandardDataHoldfeatures
Onscreenmenuselectionandnavigatorkeydrive
Variablefrequencydrivemode(lowpassfilter)
Frontpanelinsulationresistancetestvoltageselector
COMPAREfeaturewithfrontpanelcolorcodedPASS/FAILLEDindication
0.1%DCVaccuracy
99recorddatarecord/recallmemory
Highpowerbuiltinworklights
Bluetooth®interface
SafetyCategoryRating:CATIV600V,CATIII1000V
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
9
1
24
3
11
2
3
4
5
6
8
7
4. MeterDescription
4.1MeterSections
1. CourtesyWorklights
2. LCDDisplay
3. PassFail(green/red)LED
4. RotaryFunctionSwitch
5. FunctionButtons(detailedbelow)
6. Positive(+)probeinputjackforVoltage,
Capacitance,Diode,&Continuity
7. COM(negative‐)probeinputjack
8. Positive(+)probeinputjackforinsulation
resistanceandearthbondresistance


1. CourtesyWorkLights
2. TestProbeHolders
3. TiltStand
4. BatteryandFuseAccess
Fig41FrontView
Fig42RearView
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
10
4.2FunctionSwitchPositions
OFF ThemeterisswitchedOFFandinfullpowersavingmode.
Themetercanmeasurevoltage(V)throughtheprobeinputs.
Themetercanmeasurecapacitance,diodepolarityorcontinuitythrough
theprobeinputs.ThetypeofmeasurementischosenbytheMODEbutton.
ZEROΩ EarthBondResistancetestingmode.
50V,100V,250V,500V,1KV:Insulationresistancetestvoltageselections
4.3FunctionButtons,SelectorPad,andRotarySwitch
MODE
Usethisbuttontochangetheoperatingmodeofthemeter(ex:ACvs.DC)
AlsoselectsInsulationResistancevaluestocompareforPASS/FAILtesting
HOLD
PresstoHOLDreadingsonthedisplayduringnormalDMMmeasuringmodes
(seeSection5.5DataHoldandAutoHoldModes).PressandholdforKeypad
lockoutmode(seeSection5.13KeypadLockout).
InINSULATIONTESTINGMODEthisbuttonwillservetoenablethetestlock
functionforcontinuousinsulationresistancetesting.SeeSection5.7Insulation
ResistanceMeasurementsforfurtherinformation.
UsethisSelectorPadtostart/stopInsulationResistanceTests,toenable
extendedfunctionalitymodes,andtonavigatemodemenuoptions.Further
detailsprovidedinSection4.3.1SelectorPad
TheRotarySwitchisusedtoselectameasurementfunction.Detailsare
providedinSection4.3.2RotarySwitch
Pressthisbuttontoexitanextendedfunctionalitymode.
Pressthisbuttontoenable/disablethedisplaybacklight.
Pressandholdthebuttonfor2secondstoenable/disablethe
worklight.
Presstoenable/disableMETERLiNK®(Bluetooth)communication,see
section5.14StreamingmeasurementdatausingBluetooth®
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
11
4.3.1SelectorPad
Therearefive(5)‘soft’functionbuttonsarrangedinasquare,asshownbelow.Thefunctions
changedependingonthesubfunctionchosenatthetime.Generallythearrowbuttonsare
usedfornavigationandtheTESTOKbuttonisusedforselectinganobjectorstartingatest.
Detailedinformationbelow:
IM75 Selector Pad
PressingtheTESTOKbuttonselectsamenudrivenoption.TheTESTOKbuttonisalsousedto
enterthesubmenusforselectionscreens.
NOTE:WheninINSULATIONTESTINGMODEpressingandholdingTESTOKoutputsatest
voltage(untilthebuttonismanuallyreleased).InTestLockmodethisbutton(usedin
conjunctionwiththeHOLD/LOCKbutton)isusedtoStartandStopcontinuousInsulation
Resistancetests.SeeSection5.InsulationResistanceMeasurementsforfurtherdetails.Donot
confusetheInsulationResistancetestlockfeaturewiththeKeypadLockoutfeature(section
5.13).
TheLEFT()buttonscrollstopreviouschoicesinthecurrentlevelofthemenu.When
reachingthestartofagroupofselections,furtherpressesoftheLEFTbuttonwrapstothe
endofthelist.
TheRIGHT()buttonscrollstonextchoicesinthecurrentlevelofthemenu.Whenreaching
theendofagroupofselections,furtherpressesoftheRIGHTbuttonwrapstothestartofthe
list.
TheUP()buttonscrollstopreviousselectableoptionsinthecurrentlevelofthemenu.
Whenreachingthestartofagroupofselections,furtherpressesoftheUPbuttonareignored.
TheDOWN()buttonscrollstopreviousselectableoptionsinthecurrentlevelofthemenu.
Whenreachingtheendofagroupofselections,furtherpressesoftheDOWNbuttonare
ignored.
4.3.2RotarySwitch
Therearenine(9)dedicatedRotarySwitchpositions:
OFF:ThemeterisswitchedOFF.
AC/DCVOLTAGE:Themetercanmeasurevoltagethroughtheprobeinputs.Thetypeof
measurement(AC/DC)ischosenbytheMODEbutton.ThedefaultmeasurementisAC
voltage.
Fig43SelectorPad
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
12
CAPACITANCE/DIODE/CONTINUITY:Themetercanmeasurecapacitance,diodepolarity
orcontinuitythroughtheprobeinputs.ThetypeofmeasurementischosenbytheMODE
button.
ZeroΩ:Themetercanmeasureearthbondresistance,throughtheprobeinputs.The
measurementunitisOhms().
50V:Themetercansourcea50VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
100V:Themetercansourcea100VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
250V:Themetercansourcea250VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
500V:Themetercansource500VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
1000V:Themetercansource1000VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
4.4DisplayDescription
1
5
6
34
2
5
1. Maindisplayarea
2. Secondarydisplayarea
3. Bargraph(matchesthereadingonthemaindisplay)
4. Modesofoperation(detailedlaterinthisguide)
5. Unitsofmeasureforprimaryandsecondarydisplays
6. Alertsandstatusicons(detailedbelow)
Fig44Displaysnapshot
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
13
4.5DisplayIconsandIndicators
IndicatesthatBluetooth®communicationisactive
TESTIndicatesanInsulationTestinprogressandtestvoltageispresent
Indicatesthatthemeasuredvoltageisgreaterthan30V(ACorDC)
HIndicatesthatthemeterisinHoldmode(solidH)orautohold(flashingH)
Indicatesthatthemeterisinlockedmode
Memoryiconshownwiththeactivememorylocation(199)
Indicatesthebatteryvoltagestatus
APOIndicatesthattheautopowerofffunctionisenabled
IndicatesthatthemeterismeasuringACvoltage
IndicatesthatthemeterismeasuringDCvoltage
Analogbargraph
VFD(variablefrequencydrive)modeicon
Frequencymodeicon
EarthBondResistanceicon
INSULInsulationResistanceicon
COMPCompare(Pass/Fail)icon
PIPolarizationIndexicon
DARDielectricAbsorptionRatioicon
Settingsmodeicon
Silentmodeicon
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
14
5. Operation
Note:Beforeoperatingthedevice,youmustread,understand,andfollowallinstructions,dangers,
warnings,cautions,andnotes.
Note:Whenthemeterisnotinuse,thefunctionswitchshouldbesettotheOFFposition.
Note:Whenconnectingtheprobeleadstothedeviceundertest,connecttheCOM(negative)lead
beforeconnectingthepositivelead.Whenremovingtheprobeleads,removethepositivelead
beforeremovingtheCOM(negative)lead.
5.1PoweringtheMeter
1.Setthefunctionswitchtoanypositiontoswitchonthemeter.
2.Ifthebatteryindicator showsthatthebatteryvoltageisloworifthemeterdoesnotpower
on,replacethebattery.Seesection6.2BatteryReplacement.
5.1.1AutoPowerOff
Themeterenterssleepmodeafter30minutesofinactivity.Themeterbeepsthreetimes20
secondsbeforepoweringoff.PressanybuttonorturntheRotarySwitchtopreventthemeter
frompoweringoff.Theautopowerofftimeoutisthenreset.Todisableautopoweroff(APO)
presstheMODEbuttonwhileturningonthemeter.TochangetheAPOtime,readsection
5.12.6SetupUtility
5.2ManualAC/DCSelection
Whenmeasuringvoltage,presstheMODEbuttontoselectACorDCvoltage.
5.3AutoRange
AutoRangeistheonlymodeofoperationforCapacitance;rangesareautomaticallyselectedby
themeter.InsulationandEarthBondResistancetestsallowforvoltagetestselectionusingthe
rotaryswitch.
5.4‘Smart’OutofRangeWarnings
Iftheinputisover/underthefullscalerange,thedisplaywillshow>xor<x,wherexisthehighor
lowendofthefullscalerange.
5.5DataHoldandAutoHoldModes
Thesetwomodesarediscussedindetailbelow.IntheSettingsmenu,ifAutoHoldissettoOFF,the
meterwilloperateinthestandardHoldmode.IfAutoHoldissettoONinthesettingsmenu,the
meterwilloperateintheAutoHoldmode.
5.5.1DataHoldMode
InDataHoldmode,presstheHOLDbuttontocapturethemeasurementthatisshownonthe
maindisplay.Thecapturedreadingwillbedisplayedinthesecondary,upperdisplayarea(the
maindisplaywillthenfunctionnormally).The‘H’displayiconwillappearsolidontheLCDin
theHoldmode.ToexittheHoldmode,presstheHOLDbuttonagain.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
15
5.5.2AutoHoldmode
IfAutoHoldmodeisswitchedONinthesettingsmenu,presstheHOLDbuttontoactivate
AutoHoldmode(the‘H’displayiconwillappearflashing).InAutoHoldmodereadingsare
capturedinthesaywayasinstandardHoldmode.Thedifferenceisthatthecaptured
reading,displayedonthesmaller,secondarydisplaydigits,willupdatewheneverareading50
unitshigherorlowerthanthecapturedreadingismeasured.ToexittheAutoHoldmode
presstheHOLDbutton.
5.6VoltageMeasurements
1.Setthefunctionswitchtothe position.
2.InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3.UsetheMODEbuttontoselectACorDCvoltagemeasurements.
The indicatorwillbedisplayedforACmeasurements.
The indicatorwillbedisplayedforDCmeasurements.
4.Connecttheprobeleadsinparalleltothepartundertest
5.Readthevoltagevalueonthedisplay
5.7InsulationResistanceMeasurements
1. Usetherotaryswitchtoselecttheinsulationvoltage(50V,100V,250V,500V,or1kV).
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheInsulation
terminal.
3. SelectINSULtoenabletheInsulationResistanceFunctionmodeasdescribedinsection5.12.1
Selectingmodes.
4. DuringtheTest,theinsulationvoltagewillappearinthesecondary(upper)readingarea.The
resistancewillbeshowninthemainreadingarea.
5. PressTESTOKtostarttest.
6. Tolockthetestforcontinuoustesting,presstheHOLD/LOCKbutton(lockiconwillappear)and
thenpressTESTOKtotestcontinuously.
7. TheTESTiconontheLCDwillbeshownwhilethevoltageisoutputted.
8. PressTESTOKagaintostopthetest
NOTE:The99pointDataRecordmemoryfunctioncannotbeactivatedintheusualwaywhile
usingtheInsulationResistancefunction(sincetheTESTOKbuttonisusedfortestinginthis
caseandnotforsavingreadings).Thelastreading,however,isheldonthedisplayandcanthen
besavedinthenormalwayasdescribedintheExtendedFunctionalitysectionbelowunder
Section5.12.8DataRecordMode.
5.7.1PI/DARTest
Note:PI(PolarizationIndex);DAR(DielectricAbsorptionRatio)
1. SelectthePIorDARtestasdescribedinsection5.12.1Selectingmodes.
2. PresstheTESTOKbuttontotest.
3. PresstheMODEbuttontoshowthetesttimeattheupperreadingarea.(LCDwillshow
CLOCKiconandtheelapsedtime).PIisatenminutetestandDARisaoneminutetest.
4. Thetestwillstopautomaticallyandtheresult(ratio)willbeshownonthemaindisplay
area.ThePASS/FAILLEDwillindicatethetestresult(GreenforPASS,andRedforFAIL).
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
16
5.8EarthBondResistanceMeasurements(ZERO)
1. MovethefunctiondialtotheEarthbondresistanceposition(ZEROΩ);theEarthbond
resistanceiconwillappear.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheZEROOhm
terminal.
3. Beforetesting,shorttheendsoftheprobesandpresstheMODEbutton.TheZEROiconwill
appearonthedisplayandtheoffsetvaluewillbestored.Ifthetestleadsarenotshortedorif
thereisresidualresistance,thedisplaywillshowthe>2.00ohms(thisindicationcouldbedue
tofaultyleads,impropertestleadinsertion,ormeterproblem.Testsshouldnotbeattempted
untilrootcauseofthehighresistanceisdetermined.
4. PressTESTOKtorunatest.
5. TolockthetestON,presstheHOLD/LOCKbutton(theLOCKicon willappear)andthen
pressTESTOKtotestcontinuously.
6. PressTESTOKagaintostopthetest.
7. NotethattheEarthBondResistanceMeasurementmodecanalsobeusedtomeasure
standardresistanceto40kohms
NOTE:The99pointDataRecordmemoryfunctioncannotbeactivatedintheusualwaywhile
usingtheEarthBondResistancefunction(sincetheTESTOKbuttonisusedfortestinginthis
caseandnotforsavingreadings).Thelastreading,however,isheldonthedisplayandcanthen
besavedasdescribedintheExtendedFunctionalitysectionbelowunderSection5.12.8Data
RecordingMode.
5.9ContinuityTest
Warning:Donotdodiode,resistanceorcontinuitytestsbeforeremovingthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3. UsetheMODEbuttontoselectcontinuitymeasurement.Theindicatorwillbedisplayed.
4. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
5. Touchthetipsoftheprobeacrossthecircuitorcomponentundertest.
6. Iftheresistanceis30±5Ω(nominal)orless,themeterwillbeep.
Note:ThisthresholdisuserselectableintheSETUPmenuundertheCntin(continuity)setting:
Range:1050Ω;Increment:1;Default:30Ω.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
17
5.10DiodeTest
Warning:Donotdodiode,resistanceorcontinuitytestsbeforeyouhaveremovedthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3. UsetheMODEbuttontoselectthediodetestfunction.The indicatorwillbedisplayed.
4. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
5. Touchthetipsoftheprobeacrossthediodeorsemiconductorjunctionundertest.Makea
noteofthevalueonthedisplay.
6. Reversethepolarityoftheprobes,byinterchangingtheprobetestlocations.
7. Touchthetipsoftheprobeacrossthediodeorsemiconductorjunctionundertest.Makea
noteofthenewvalueonthedisplay
5.11CapacitanceMeasurements
Warning:Donotdocapacitancetestsbeforeremovingthepowertothecapacitororotherdevices
undertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. UsetheMODEbuttontoselectthecapacitancemeasurement.TheF(Farad)unitwillbe
displayed.
3. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
4. Touchthetipsoftheprobeacrossthepartundertest.
5. Readthecapacitancevalueonthedisplay.
Note:Forverylargecapacitancevalues,itmaytakeseveralsecondsforthemeasurementto
settleandthefinalreadingtostabilize.
5.12ExtendedFunctionality
Avarietyofmodesareavailablethatprovideextendedfunctionalityasdetailedbelow.
5.12.1SelectingModes
Themodeiconsapplicablefortheselectedmeasurementtypearedisplayedinthelowerpart
ofthedisplay.Whenamodeisenabled,theiconisframed.
Usetheleft/rightarrowbuttonstonavigatetothedesiredmodeicon.Thecurrentlyselected
iconwillflash.
1. PresstheTESTOKbuttontoenabletheselected(flashing)mode.
2. Usetheup/downarrowbuttonstostepthroughthemodeoptions(ifany).Referto
thesectionrelatedtothespecificmodefordetailedinstructions.
3. PresstheTESTOKbuttontodisabletheselected(flashing)mode.
Fig51ModeIcons
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
18
5.12.2VFDMode(ACVonly)
InVFD(variablefrequencydrive)mode,highfrequencynoiseiseliminatedfromthevoltage
measurementthroughtheuseofalowpassfilter.VFDmodeisavailablewhenmeasuring
ACV.
1. Select andthenpresstheTESTOKbuttontoenabletheselected(flashing)mode.
5.12.3FrequencyMode(ACVonly)
InFrequencymode,thefrequencyisdisplayedinthemaindisplayandtheperiodisdisplayed
inthesecondarydisplay.FrequencymodeisavailablewhenmeasuringACvoltage.
1. Select andenableFrequencymodeasdescribedinsection5.12.1,Selectingmodes.
5.12.4Pass/FailMode(COMPARE)
1. Selectthedesiredinsulationvoltagevalueusingtherotaryswitch.
2. SelectCOMPandenablethePass/FailFunctionmodeasdescribedinsection5.12.1
Selectingmodes.
3. UsetheMODEbuttontoselecttheresistancevaluetocompare.(Thereareresistance
valuesegmentsandaCOMPAREiconontheLCD,Theresistancevalueoptionsare
100kohm,200kohm,500kohm,1Mohm,2Mohm,5Mohm,10Mohm,20Mohm,50Mohm)
4. PressTESTtostarttheprocess.
5. Observethegreenlight/redlightforPASS/FAILindication.
5.12.5SetupUtility
InSetup,definethesettingsforvariousmeteroptions:
Autopoweroff(indicatedbythetextAPO):Amodewherethetimeperiodafter
whichthemeterenterssleepmodecanbeset.Therangeis1to30minutes,orOFF.
Thefactorydefaultis10minutes.
Autobacklightoff(indicatedbythetextb.Lit):Amodewherethetimeperiodafter
whichthebacklightturnsoffcanbeset.Therangeis1to30minutes,orOFF.The
factorydefaultis5minutes.
Continuitythreshold(indicatedbythetextCntin):Amodewherethethresholdfor
continuitytestscanbeset.
Autohold(indicatedbythetextA.Hold):SelectAutoholdmodeON(Autoholdmode
active)orOFF(Dataholdmodeactive).Formoreinformation,seesection5.5Data
holdmodeandAutoholdmode.
1. Select andenableSetupmodeasdescribedinsection5.12.1Selectingmodes.
2. Usetheup/downarrowsbuttontocyclethroughthemodeoptionsAPO,b.Lit,Cntin,
A.Hold,andRESETshownonthesecondarydisplay.
3. PresstheTESTOKbuttontoactivatethedisplayedoption.
APO:Usetheleft/rightarrowbuttonstochangetheautopowerofftime.
b.Lit:Usetheleft/rightarrowbuttonstochangetheautobacklightofftime.
A.Hold:Usetheleft/rightarrowbuttonstoconfigureautohold/datahold.Onindicates
thattheautoholdmodeisactive.Offindicatesthatthedataholdmodeisactive.
Cntin:Usetheleft/rightarrowbuttonstochangethecontinuitythreshold.
RESET:PresstheTESTOKbuttontoresetthesettingstothefactorydefault.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
19
5.12.6SilentMode
InSilentmode,thealertbeeperisdisabled.Silentmodedoesnotaffectthecontinuitybeeper.
SelectandenableSilentmodeasdescribedinsection5.12.1Selectingmodes.
5.12.7DataRecordingMode(99point)
Themeterhas99memorylocationsforthestorageofmeasurementdata.
SelectandenableMEMmode asdescribedinsection5.12.1Selectingmodes.
Usethearrowkeystocyclethroughthemodeoptions:SAVE,LOAD,andCLEARshownonthe
secondarydisplay.
SAVE:Thedataonthemaindisplayissavedtoamemorylocationindicatedbythe‘88’inthe
upperdisplayarea.
LOAD:Thedatastoredinthememorylocation(identifiedbythe‘88’indicator)isdisplayed.
Usearrowkeystoscrollthememorylocations.Usethe buttontoexittheLOADmode.
CLEAR:Thedatainallmemorylocationsiserased.
5.13KeypadLockout
InLockoutmodethemeterignoresallbuttonpressesexceptHOLD.Theautopowerofffunction,
seesection5.1.1AutoPoweroff,isdisabledinLockoutmode.
PressandholdtheHOLD/LOCKbuttonfor3secondstoenter/exitthelockoutmode.Inlockout
mode,thelockindicatorisdisplayed.DonotconfuseLockoutmodewiththetestlockfeatureused
forcontinuousInsulationResistancetesting(seeSection5.7InsulationResistanceMeasurements).
5.14StreamingMeasurementDatausingBluetooth®
5.14.1General
SomeIRcamerasfromFLIRSystemssupportBluetoothcommunication,andtothosecameras
youcanstreammeasurementdatafromthemeter.Thedataisthenmergedintotheresultant
tableintheIRimage.ThisisfurtherfacilitatedthroughaFLIRsoftwaretoolknownas
METERLiNK®.TostreamdatatoaniOS/Androiddevice,downloadtheapplicationFLIR
TOOLSTMandfollowtheapplicationonscreenprompts.
StreamingmeasurementdataisaconvenientwaytoaddimportantinformationtoanIRimage.
Forexample,whenidentifyinganoverheatedcableconnection,youmaywanttoknowits
voltage.TheBluetoothrangeis10m(32ft)maximum.
5.14.2Procedures
1. PairtheIRcamerawiththeinstrument.Refertothecameramanualforinformationon
howtopairBluetoothdevices.
2. Turnonthecameraandthenturnonthemeter.
3. PressthebuttononthemetertoenableBluetooth.TheBluetoothiconwillappear
flashingintheupperlefthandcornerofthemeterdisplay.
4. Choosethevariablethatyouwanttouse(voltage,bondresistance,etc.).Resultsfromthe
meterwillnowautomaticallybedisplayedintheresultanttableinthetopleftcornerof
theIRcamerascreen.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
20
6. Maintenance
6.1CleaningandStorage
Cleanthemeterwithadampclothandmilddetergent;donotuseabrasivesorsolvents.
Ifthemeterisnottobeusedforanextendedperiod,removethebatteriesandstorethem
separately.
6.2BatteryReplacement
TheBatterysymbolflasheswithno‘bars’whenthebatterieshavereachedacriticallevelof7.0V.
ThesymbolstaysactiveandvisiblewhiletheLCDispowered.
Themeterdisplaysreadingswithinspecificationswhilethelowbatteryindicatorison.Whenthisis
nolongerpossible,thedisplaymustbeblanked.Themeterpowersoffbeforeitdisplaysanoutof
tolerancevoltage.
1. Toavoidelectricalshock,disconnectthemeterifconnectedtoacircuit,removetheprobe
fromtheterminals,andsetthefunctionswitchtotheOFFpositionbeforeattemptingto
replacethebatteries.
2. Unscrewandremovethebatterycompartmentcover.
3. ReplacethesixstandardAAAbatteries,observingcorrectpolarity.
4. Securethebatterycompartmentcover.
6.3FuseReplacement
Thefuseaccessedviathebatterycompartmentcover.Thefuseisrated440mA/1000V,ceramic
fastblowwithaminimumInterruptRatingof10kA.
6.4DisposalofElectronicWaste
Aswithmostelectronicproducts,thisequipmentmustbedisposedofinan
environmentallyfriendlyway,andinaccordancewithexistingregulationsfor
electronicwaste.PleasecontactyourFLIRSystemsrepresentativeformoredetails.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28

FLIR IM75 User manual

Category
Multimeters
Type
User manual

Ask a question and I''ll find the answer in the document