FLIR IM75 User manual

  • Hello! I've reviewed the user manual for the FLIR IM75 Insulation Tester/MultiMeter. This document provides instructions on device operation, safety, and specifications. It covers various measurement types, including voltage, resistance, capacitance, and insulation testing. I'm here to help answer any specific questions you may have about this device.
  • How do I select AC or DC voltage measurements?
    How do I start an insulation resistance test?
    How can I save measurement data?
    How does the 'Compare' feature work?
USERMANUAL
FLIRMODELIM75
InsulationTester/MultiMeter
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
2
TableofContents
1.DISCLAIMERS4
1.1Copyright4
1.2QualityAssurance4
1.3Documentation4
1.4DisposalofElectronicWaste4
2.SAFETY5
2.1FCCCompliance6
2.2IndustryCanadaCompliance7
3.INTRODUCTION8
3.1KeyFeatures8
4.METERDESCRIPTION9
4.1MeterSections9
4.2FunctionSwitchPositions10
4.3FunctionButtons,SelectorPad,andRotarySwitch10
4.3.1SelectorPad11
4.3.2RotarySwitch11
4.4DisplayDescription12
4.5DisplayIconsandIndicators13
5.OPERATION14
5.1PoweringtheMeter14
5.1.1AutoPowerOff14
5.2ManualAC/DCSelection14
5.3AutoRange14
5.4‘Smart’OutofRangeWarnings14
5.5DataHoldandAutoHoldModes14
5.5.1DataHoldMode14
5.6VoltageMeasurements15
5.7InsulationResistanceMeasurements15
5.7.1PI/DARTest15
5.8EarthBondResistanceMeasurements(ZEROΩ)16
5.9ContinuityTest16
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
3
5.10DiodeTest17
5.11CapacitanceMeasurements17
5.12ExtendedFunctionality17
5.12.1SelectingModes17
5.12.2VFDMode(ACVonly)18
5.12.3FrequencyMode(ACVonly)18
5.12.4Pass/FailMode(COMPARE)18
5.12.5SetupUtility18
5.12.6SilentMode19
5.12.7DataRecordingMode(99point)19
5.13KeypadLockout19
5.14StreamingMeasurementDatausingBluetooth®19
5.14.1General19
5.14.2Procedures19
6.MAINTENANCE20
6.1CleaningandStorage20
6.2BatteryReplacement20
6.3FuseReplacement20
6.4DisposalofElectronicWaste20
7.SPECIFICATIONS21
7.1Generalspecifications21
7.2ElectricalRangeSpecifications22
7.3InputSpecifications24
8.TECHNICALSUPPORT25
9.WARRANTIES25
9.1FLIRGlobalLimitedLifetimeWarranty25
9.2FLIRTestandMeasurementLimited2YearWarranty26
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
4
1. Disclaimers
1.1Copyright
©2014,FLIRSystems,Inc.Allrightsreservedworldwide.Nopartsofthesoftwareincludingsource
codemaybereproduced,transmitted,transcribedortranslatedintoanylanguageorcomputer
languageinanyformorbyanymeans,electronic,magnetic,optical,manualorotherwise,without
thepriorwrittenpermissionofFLIRSystems.
Thedocumentationmustnot,inwholeorpart,becopied,photocopied,reproduced,translatedor
transmittedtoanyelectronicmediumormachinereadableformwithoutpriorconsent,inwriting,
fromFLIRSystems.
Namesandmarksappearingontheproductshereinareeitherregisteredtrademarksor
trademarksofFLIRSystemsand/oritssubsidiaries.Allothertrademarks,tradenamesorcompany
namesreferencedhereinareusedforidentificationonlyandarethepropertyoftheirrespective
owners.
1.2QualityAssurance
TheQualityManagementSystemunderwhichtheseproductsaredevelopedandmanufactured
hasbeencertifiedinaccordancewiththeISO9001standard.
FLIRSystemsiscommittedtoapolicyofcontinuousdevelopment;thereforewereservetheright
tomakechangesandimprovementsonanyoftheproductswithoutpriornotice.
1.3Documentation
Toaccessthelatestmanualsandnotifications,gototheDownloadtabat:http://support.flir.com.
Itonlytakesafewminutestoregisteronline.Inthedownloadareayouwillalsofindthelatest
releasesofmanualsforourotherproducts,aswellasmanualsforourhistoricalandobsolete
products.
1.4DisposalofElectronicWaste
Aswithmostelectronicproducts,thisequipmentmustbedisposedofinan
environmentallyfriendlyway,andinaccordancewithexistingregulationsfor
electronicwaste.
PleasecontactyourFLIRSystemsrepresentativeformoredetails.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
5
2. Safety
SafetyNotes
Beforeoperatingthedevice,youmustread,understand,andfollowallinstructions,dangers,
warnings,cautions,andnotes.
FLIRSystemsreservestherighttodiscontinuemodels,partsoraccessories,andotheritems,orto
changespecificationsatanytimewithoutpriornotice.
Removethebatteriesifthedeviceisnotusedforanextendedperiodoftime.
WarningStatements
Donotoperatethedeviceifyoudonothavethecorrectknowledge.Formalqualificationsand/or
nationallegislationfortheelectricalinspectionscanapply.Incorrectoperationofthedevicecan
causedamage,shock,injuryordeathtopersons.
Donotstartthemeasuringprocedurebeforeyouhavesetthefunctionswitchtothecorrect
position.Thiscancausedamagetotheinstrumentandcancauseinjurytopersons.
Donotchangetoresistancewhenyoumeasurethevoltage.Thiscancausedamagetothe
instrumentandcancauseinjurytopersons.
Youmustdisconnectthetestleadsfromthecircuitthatyoudidatestonbeforeyouchangethe
range.Ifyoudonotdothis,damagetotheinstrumentandinjurytopersonscanoccur.
Donotreplacethebatteriesbeforeyouremovethetestleads.Thiscancausedamagetothe
instrumentandcancauseinjurytopersons.
Donotusethedeviceifthetestleadsand/orthedeviceshowsignsofdamage.Injurytopersons
canoccur.
Becarefulwhenyoudothemeasurementsifthevoltagesaremorethan25VACrmsor35VDC.
Thereisariskofshockfromthesevoltages.Injurytopersonscanoccur.
Donotdodiode,resistanceorcontinuitytestsbeforeyouhaveremovedthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
Donotusethedeviceasatooltoidentifyliveterminals.Youmustusethecorrecttools.Injuryto
personscanoccurifyoudonotusethecorrecttools.
Makesurethatchildrencannottouchthedevice.Thedevicecontainsdangerousobjectsandsmall
partsthatchildrencanswallow.Ifachildswallowsanobjectorapart,speakwithaphysician
immediately.Injurytopersonscanoccur.
Donotletchildrenplaywiththebatteriesand/orthepackingmaterial.Thesecanbedangerous
forchildreniftheyusethemastoys.
Donottouchexpiredordamagedbatterieswithoutgloves.Injurytopersonscanoccur.
Donotcauseashortcircuitofthebatteries.Thiscancausedamagetotheinstrumentandcan
causeinjurytopersons.
Donotputthebatteriesintoafire.Injurytopersonscanoccur.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
6
Cautions
Donotusethedeviceforaprocedurethatitisnotmadefor.Thiscancausedamagetothe
protection.
Thissymbol,adjacenttoanothersymbolorterminal,indicatesthattheuser
mustrefertothemanualforfurtherinformation.
Thissymbol,adjacenttoaterminal,indicatesthat,undernormaluse,
hazardousvoltagesmaybepresent.
Doubleinsulation.
ULlistingisnotanindicationoraverificationoftheaccuracyofthemeter
2.1FCCCompliance
Thisdevicecomplieswithpart15oftheFCCRules.Operationissubjecttothefollowingtwo
conditions:
1. Thisdevicemaynotcauseharmfulinterference.
2. Thisdevicemustacceptanyinterferencereceived,includinginterferencethatmaycause
undesiredoperation.
ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassBdigitaldevice,
pursuanttopart15oftheFCCRules.Theselimitsaredesignedtoprovidereasonableprotection
againstharmfulinterferenceinaresidentialinstallation.Thisequipmentgenerates,uses,andcan
radiateradiofrequencyenergyand,ifnotinstalledandusedinaccordancewiththeinstructions,
maycauseharmfulinterferencetoradiocommunications.However,thereisnoguaranteethat
interferencewillnotoccurinaparticularinstallation.Ifthisequipmentdoescauseharmful
interferencetoradioortelevisionreception,whichcanbedeterminedbyturningtheequipment
offandon,theuserisencouragedtotrytocorrecttheinterferencebyoneormoreofthe
followingmeasures:
1. Reorientorrelocatethereceivingantenna.
2. Increasetheseparationbetweentheequipmentandreceiver.
3. Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichthereceiveris
connected.
4. Consultthedealeroranexperiencedradio/TVtechnicianforhelp.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
7
WARNING
Changesormodificationsnotexpresslyapprovedbythepartyresponsibleforcompliance
couldvoidtheuser'sauthoritytooperatetheequipment.
2.2IndustryCanadaCompliance
Thisdevicecomplieswithpart15oftheFCCRules.Operationissubjecttothefollowingtwo
conditions:
ThisdevicecomplieswithIndustryCanadalicenseexemptRSSstandard(s).Operationissubjectto
thefollowingtwoconditions:(1)thisdevicemaynotcauseinterference,and(2)thisdevicemust
acceptanyinterference,includinginterferencethatmaycauseundesiredoperationofthedevice.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
8
3. Introduction
ThankyouforselectingtheFLIRIM75InsulationMultiMeter.Thisdeviceisshippedfullytestedand
calibratedand,withproperuse,willprovideyearsofreliableservice.
3.1KeyFeatures
4000countextralargedigitaldualdisplaywithbacklighting
AutoRangewithsmartoverrangeindication
AutoHoldandStandardDataHoldfeatures
Onscreenmenuselectionandnavigatorkeydrive
Variablefrequencydrivemode(lowpassfilter)
Frontpanelinsulationresistancetestvoltageselector
COMPAREfeaturewithfrontpanelcolorcodedPASS/FAILLEDindication
0.1%DCVaccuracy
99recorddatarecord/recallmemory
Highpowerbuiltinworklights
Bluetooth®interface
SafetyCategoryRating:CATIV600V,CATIII1000V
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
9
1
24
3
11
2
3
4
5
6
8
7
4. MeterDescription
4.1MeterSections
1. CourtesyWorklights
2. LCDDisplay
3. PassFail(green/red)LED
4. RotaryFunctionSwitch
5. FunctionButtons(detailedbelow)
6. Positive(+)probeinputjackforVoltage,
Capacitance,Diode,&Continuity
7. COM(negative‐)probeinputjack
8. Positive(+)probeinputjackforinsulation
resistanceandearthbondresistance


1. CourtesyWorkLights
2. TestProbeHolders
3. TiltStand
4. BatteryandFuseAccess
Fig41FrontView
Fig42RearView
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
10
4.2FunctionSwitchPositions
OFF ThemeterisswitchedOFFandinfullpowersavingmode.
Themetercanmeasurevoltage(V)throughtheprobeinputs.
Themetercanmeasurecapacitance,diodepolarityorcontinuitythrough
theprobeinputs.ThetypeofmeasurementischosenbytheMODEbutton.
ZEROΩ EarthBondResistancetestingmode.
50V,100V,250V,500V,1KV:Insulationresistancetestvoltageselections
4.3FunctionButtons,SelectorPad,andRotarySwitch
MODE
Usethisbuttontochangetheoperatingmodeofthemeter(ex:ACvs.DC)
AlsoselectsInsulationResistancevaluestocompareforPASS/FAILtesting
HOLD
PresstoHOLDreadingsonthedisplayduringnormalDMMmeasuringmodes
(seeSection5.5DataHoldandAutoHoldModes).PressandholdforKeypad
lockoutmode(seeSection5.13KeypadLockout).
InINSULATIONTESTINGMODEthisbuttonwillservetoenablethetestlock
functionforcontinuousinsulationresistancetesting.SeeSection5.7Insulation
ResistanceMeasurementsforfurtherinformation.
UsethisSelectorPadtostart/stopInsulationResistanceTests,toenable
extendedfunctionalitymodes,andtonavigatemodemenuoptions.Further
detailsprovidedinSection4.3.1SelectorPad
TheRotarySwitchisusedtoselectameasurementfunction.Detailsare
providedinSection4.3.2RotarySwitch
Pressthisbuttontoexitanextendedfunctionalitymode.
Pressthisbuttontoenable/disablethedisplaybacklight.
Pressandholdthebuttonfor2secondstoenable/disablethe
worklight.
Presstoenable/disableMETERLiNK®(Bluetooth)communication,see
section5.14StreamingmeasurementdatausingBluetooth®
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
11
4.3.1SelectorPad
Therearefive(5)‘soft’functionbuttonsarrangedinasquare,asshownbelow.Thefunctions
changedependingonthesubfunctionchosenatthetime.Generallythearrowbuttonsare
usedfornavigationandtheTESTOKbuttonisusedforselectinganobjectorstartingatest.
Detailedinformationbelow:
IM75 Selector Pad
PressingtheTESTOKbuttonselectsamenudrivenoption.TheTESTOKbuttonisalsousedto
enterthesubmenusforselectionscreens.
NOTE:WheninINSULATIONTESTINGMODEpressingandholdingTESTOKoutputsatest
voltage(untilthebuttonismanuallyreleased).InTestLockmodethisbutton(usedin
conjunctionwiththeHOLD/LOCKbutton)isusedtoStartandStopcontinuousInsulation
Resistancetests.SeeSection5.InsulationResistanceMeasurementsforfurtherdetails.Donot
confusetheInsulationResistancetestlockfeaturewiththeKeypadLockoutfeature(section
5.13).
TheLEFT()buttonscrollstopreviouschoicesinthecurrentlevelofthemenu.When
reachingthestartofagroupofselections,furtherpressesoftheLEFTbuttonwrapstothe
endofthelist.
TheRIGHT()buttonscrollstonextchoicesinthecurrentlevelofthemenu.Whenreaching
theendofagroupofselections,furtherpressesoftheRIGHTbuttonwrapstothestartofthe
list.
TheUP()buttonscrollstopreviousselectableoptionsinthecurrentlevelofthemenu.
Whenreachingthestartofagroupofselections,furtherpressesoftheUPbuttonareignored.
TheDOWN()buttonscrollstopreviousselectableoptionsinthecurrentlevelofthemenu.
Whenreachingtheendofagroupofselections,furtherpressesoftheDOWNbuttonare
ignored.
4.3.2RotarySwitch
Therearenine(9)dedicatedRotarySwitchpositions:
OFF:ThemeterisswitchedOFF.
AC/DCVOLTAGE:Themetercanmeasurevoltagethroughtheprobeinputs.Thetypeof
measurement(AC/DC)ischosenbytheMODEbutton.ThedefaultmeasurementisAC
voltage.
Fig43SelectorPad
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
12
CAPACITANCE/DIODE/CONTINUITY:Themetercanmeasurecapacitance,diodepolarity
orcontinuitythroughtheprobeinputs.ThetypeofmeasurementischosenbytheMODE
button.
ZeroΩ:Themetercanmeasureearthbondresistance,throughtheprobeinputs.The
measurementunitisOhms().
50V:Themetercansourcea50VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
100V:Themetercansourcea100VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
250V:Themetercansourcea250VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
500V:Themetercansource500VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
1000V:Themetercansource1000VDCTESTVoltageandmeasureInsulationResistance
throughtheZeroΩ/InsulationandCOMprobeinputs.
4.4DisplayDescription
1
5
6
34
2
5
1. Maindisplayarea
2. Secondarydisplayarea
3. Bargraph(matchesthereadingonthemaindisplay)
4. Modesofoperation(detailedlaterinthisguide)
5. Unitsofmeasureforprimaryandsecondarydisplays
6. Alertsandstatusicons(detailedbelow)
Fig44Displaysnapshot
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
13
4.5DisplayIconsandIndicators
IndicatesthatBluetooth®communicationisactive
TESTIndicatesanInsulationTestinprogressandtestvoltageispresent
Indicatesthatthemeasuredvoltageisgreaterthan30V(ACorDC)
HIndicatesthatthemeterisinHoldmode(solidH)orautohold(flashingH)
Indicatesthatthemeterisinlockedmode
Memoryiconshownwiththeactivememorylocation(199)
Indicatesthebatteryvoltagestatus
APOIndicatesthattheautopowerofffunctionisenabled
IndicatesthatthemeterismeasuringACvoltage
IndicatesthatthemeterismeasuringDCvoltage
Analogbargraph
VFD(variablefrequencydrive)modeicon
Frequencymodeicon
EarthBondResistanceicon
INSULInsulationResistanceicon
COMPCompare(Pass/Fail)icon
PIPolarizationIndexicon
DARDielectricAbsorptionRatioicon
Settingsmodeicon
Silentmodeicon
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
14
5. Operation
Note:Beforeoperatingthedevice,youmustread,understand,andfollowallinstructions,dangers,
warnings,cautions,andnotes.
Note:Whenthemeterisnotinuse,thefunctionswitchshouldbesettotheOFFposition.
Note:Whenconnectingtheprobeleadstothedeviceundertest,connecttheCOM(negative)lead
beforeconnectingthepositivelead.Whenremovingtheprobeleads,removethepositivelead
beforeremovingtheCOM(negative)lead.
5.1PoweringtheMeter
1.Setthefunctionswitchtoanypositiontoswitchonthemeter.
2.Ifthebatteryindicator showsthatthebatteryvoltageisloworifthemeterdoesnotpower
on,replacethebattery.Seesection6.2BatteryReplacement.
5.1.1AutoPowerOff
Themeterenterssleepmodeafter30minutesofinactivity.Themeterbeepsthreetimes20
secondsbeforepoweringoff.PressanybuttonorturntheRotarySwitchtopreventthemeter
frompoweringoff.Theautopowerofftimeoutisthenreset.Todisableautopoweroff(APO)
presstheMODEbuttonwhileturningonthemeter.TochangetheAPOtime,readsection
5.12.6SetupUtility
5.2ManualAC/DCSelection
Whenmeasuringvoltage,presstheMODEbuttontoselectACorDCvoltage.
5.3AutoRange
AutoRangeistheonlymodeofoperationforCapacitance;rangesareautomaticallyselectedby
themeter.InsulationandEarthBondResistancetestsallowforvoltagetestselectionusingthe
rotaryswitch.
5.4‘Smart’OutofRangeWarnings
Iftheinputisover/underthefullscalerange,thedisplaywillshow>xor<x,wherexisthehighor
lowendofthefullscalerange.
5.5DataHoldandAutoHoldModes
Thesetwomodesarediscussedindetailbelow.IntheSettingsmenu,ifAutoHoldissettoOFF,the
meterwilloperateinthestandardHoldmode.IfAutoHoldissettoONinthesettingsmenu,the
meterwilloperateintheAutoHoldmode.
5.5.1DataHoldMode
InDataHoldmode,presstheHOLDbuttontocapturethemeasurementthatisshownonthe
maindisplay.Thecapturedreadingwillbedisplayedinthesecondary,upperdisplayarea(the
maindisplaywillthenfunctionnormally).The‘H’displayiconwillappearsolidontheLCDin
theHoldmode.ToexittheHoldmode,presstheHOLDbuttonagain.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
15
5.5.2AutoHoldmode
IfAutoHoldmodeisswitchedONinthesettingsmenu,presstheHOLDbuttontoactivate
AutoHoldmode(the‘H’displayiconwillappearflashing).InAutoHoldmodereadingsare
capturedinthesaywayasinstandardHoldmode.Thedifferenceisthatthecaptured
reading,displayedonthesmaller,secondarydisplaydigits,willupdatewheneverareading50
unitshigherorlowerthanthecapturedreadingismeasured.ToexittheAutoHoldmode
presstheHOLDbutton.
5.6VoltageMeasurements
1.Setthefunctionswitchtothe position.
2.InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3.UsetheMODEbuttontoselectACorDCvoltagemeasurements.
The indicatorwillbedisplayedforACmeasurements.
The indicatorwillbedisplayedforDCmeasurements.
4.Connecttheprobeleadsinparalleltothepartundertest
5.Readthevoltagevalueonthedisplay
5.7InsulationResistanceMeasurements
1. Usetherotaryswitchtoselecttheinsulationvoltage(50V,100V,250V,500V,or1kV).
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheInsulation
terminal.
3. SelectINSULtoenabletheInsulationResistanceFunctionmodeasdescribedinsection5.12.1
Selectingmodes.
4. DuringtheTest,theinsulationvoltagewillappearinthesecondary(upper)readingarea.The
resistancewillbeshowninthemainreadingarea.
5. PressTESTOKtostarttest.
6. Tolockthetestforcontinuoustesting,presstheHOLD/LOCKbutton(lockiconwillappear)and
thenpressTESTOKtotestcontinuously.
7. TheTESTiconontheLCDwillbeshownwhilethevoltageisoutputted.
8. PressTESTOKagaintostopthetest
NOTE:The99pointDataRecordmemoryfunctioncannotbeactivatedintheusualwaywhile
usingtheInsulationResistancefunction(sincetheTESTOKbuttonisusedfortestinginthis
caseandnotforsavingreadings).Thelastreading,however,isheldonthedisplayandcanthen
besavedinthenormalwayasdescribedintheExtendedFunctionalitysectionbelowunder
Section5.12.8DataRecordMode.
5.7.1PI/DARTest
Note:PI(PolarizationIndex);DAR(DielectricAbsorptionRatio)
1. SelectthePIorDARtestasdescribedinsection5.12.1Selectingmodes.
2. PresstheTESTOKbuttontotest.
3. PresstheMODEbuttontoshowthetesttimeattheupperreadingarea.(LCDwillshow
CLOCKiconandtheelapsedtime).PIisatenminutetestandDARisaoneminutetest.
4. Thetestwillstopautomaticallyandtheresult(ratio)willbeshownonthemaindisplay
area.ThePASS/FAILLEDwillindicatethetestresult(GreenforPASS,andRedforFAIL).
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
16
5.8EarthBondResistanceMeasurements(ZERO)
1. MovethefunctiondialtotheEarthbondresistanceposition(ZEROΩ);theEarthbond
resistanceiconwillappear.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheZEROOhm
terminal.
3. Beforetesting,shorttheendsoftheprobesandpresstheMODEbutton.TheZEROiconwill
appearonthedisplayandtheoffsetvaluewillbestored.Ifthetestleadsarenotshortedorif
thereisresidualresistance,thedisplaywillshowthe>2.00ohms(thisindicationcouldbedue
tofaultyleads,impropertestleadinsertion,ormeterproblem.Testsshouldnotbeattempted
untilrootcauseofthehighresistanceisdetermined.
4. PressTESTOKtorunatest.
5. TolockthetestON,presstheHOLD/LOCKbutton(theLOCKicon willappear)andthen
pressTESTOKtotestcontinuously.
6. PressTESTOKagaintostopthetest.
7. NotethattheEarthBondResistanceMeasurementmodecanalsobeusedtomeasure
standardresistanceto40kohms
NOTE:The99pointDataRecordmemoryfunctioncannotbeactivatedintheusualwaywhile
usingtheEarthBondResistancefunction(sincetheTESTOKbuttonisusedfortestinginthis
caseandnotforsavingreadings).Thelastreading,however,isheldonthedisplayandcanthen
besavedasdescribedintheExtendedFunctionalitysectionbelowunderSection5.12.8Data
RecordingMode.
5.9ContinuityTest
Warning:Donotdodiode,resistanceorcontinuitytestsbeforeremovingthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3. UsetheMODEbuttontoselectcontinuitymeasurement.Theindicatorwillbedisplayed.
4. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
5. Touchthetipsoftheprobeacrossthecircuitorcomponentundertest.
6. Iftheresistanceis30±5Ω(nominal)orless,themeterwillbeep.
Note:ThisthresholdisuserselectableintheSETUPmenuundertheCntin(continuity)setting:
Range:1050Ω;Increment:1;Default:30Ω.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
17
5.10DiodeTest
Warning:Donotdodiode,resistanceorcontinuitytestsbeforeyouhaveremovedthepowerfrom
capacitorsandotherdevicesundertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. InserttheblackprobeleadintotheCOMterminalandredprobeleadintotheVterminal.
3. UsetheMODEbuttontoselectthediodetestfunction.The indicatorwillbedisplayed.
4. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
5. Touchthetipsoftheprobeacrossthediodeorsemiconductorjunctionundertest.Makea
noteofthevalueonthedisplay.
6. Reversethepolarityoftheprobes,byinterchangingtheprobetestlocations.
7. Touchthetipsoftheprobeacrossthediodeorsemiconductorjunctionundertest.Makea
noteofthenewvalueonthedisplay
5.11CapacitanceMeasurements
Warning:Donotdocapacitancetestsbeforeremovingthepowertothecapacitororotherdevices
undertestduringameasurement.Injurytopersonscanoccur.
1. Setthefunctionswitchtothe position.
2. UsetheMODEbuttontoselectthecapacitancemeasurement.TheF(Farad)unitwillbe
displayed.
3. InserttheblackprobeleadintotheCOM(negative)terminalandtheredprobeleadintothe
positiveterminal.
4. Touchthetipsoftheprobeacrossthepartundertest.
5. Readthecapacitancevalueonthedisplay.
Note:Forverylargecapacitancevalues,itmaytakeseveralsecondsforthemeasurementto
settleandthefinalreadingtostabilize.
5.12ExtendedFunctionality
Avarietyofmodesareavailablethatprovideextendedfunctionalityasdetailedbelow.
5.12.1SelectingModes
Themodeiconsapplicablefortheselectedmeasurementtypearedisplayedinthelowerpart
ofthedisplay.Whenamodeisenabled,theiconisframed.
Usetheleft/rightarrowbuttonstonavigatetothedesiredmodeicon.Thecurrentlyselected
iconwillflash.
1. PresstheTESTOKbuttontoenabletheselected(flashing)mode.
2. Usetheup/downarrowbuttonstostepthroughthemodeoptions(ifany).Referto
thesectionrelatedtothespecificmodefordetailedinstructions.
3. PresstheTESTOKbuttontodisabletheselected(flashing)mode.
Fig51ModeIcons
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
18
5.12.2VFDMode(ACVonly)
InVFD(variablefrequencydrive)mode,highfrequencynoiseiseliminatedfromthevoltage
measurementthroughtheuseofalowpassfilter.VFDmodeisavailablewhenmeasuring
ACV.
1. Select andthenpresstheTESTOKbuttontoenabletheselected(flashing)mode.
5.12.3FrequencyMode(ACVonly)
InFrequencymode,thefrequencyisdisplayedinthemaindisplayandtheperiodisdisplayed
inthesecondarydisplay.FrequencymodeisavailablewhenmeasuringACvoltage.
1. Select andenableFrequencymodeasdescribedinsection5.12.1,Selectingmodes.
5.12.4Pass/FailMode(COMPARE)
1. Selectthedesiredinsulationvoltagevalueusingtherotaryswitch.
2. SelectCOMPandenablethePass/FailFunctionmodeasdescribedinsection5.12.1
Selectingmodes.
3. UsetheMODEbuttontoselecttheresistancevaluetocompare.(Thereareresistance
valuesegmentsandaCOMPAREiconontheLCD,Theresistancevalueoptionsare
100kohm,200kohm,500kohm,1Mohm,2Mohm,5Mohm,10Mohm,20Mohm,50Mohm)
4. PressTESTtostarttheprocess.
5. Observethegreenlight/redlightforPASS/FAILindication.
5.12.5SetupUtility
InSetup,definethesettingsforvariousmeteroptions:
Autopoweroff(indicatedbythetextAPO):Amodewherethetimeperiodafter
whichthemeterenterssleepmodecanbeset.Therangeis1to30minutes,orOFF.
Thefactorydefaultis10minutes.
Autobacklightoff(indicatedbythetextb.Lit):Amodewherethetimeperiodafter
whichthebacklightturnsoffcanbeset.Therangeis1to30minutes,orOFF.The
factorydefaultis5minutes.
Continuitythreshold(indicatedbythetextCntin):Amodewherethethresholdfor
continuitytestscanbeset.
Autohold(indicatedbythetextA.Hold):SelectAutoholdmodeON(Autoholdmode
active)orOFF(Dataholdmodeactive).Formoreinformation,seesection5.5Data
holdmodeandAutoholdmode.
1. Select andenableSetupmodeasdescribedinsection5.12.1Selectingmodes.
2. Usetheup/downarrowsbuttontocyclethroughthemodeoptionsAPO,b.Lit,Cntin,
A.Hold,andRESETshownonthesecondarydisplay.
3. PresstheTESTOKbuttontoactivatethedisplayedoption.
APO:Usetheleft/rightarrowbuttonstochangetheautopowerofftime.
b.Lit:Usetheleft/rightarrowbuttonstochangetheautobacklightofftime.
A.Hold:Usetheleft/rightarrowbuttonstoconfigureautohold/datahold.Onindicates
thattheautoholdmodeisactive.Offindicatesthatthedataholdmodeisactive.
Cntin:Usetheleft/rightarrowbuttonstochangethecontinuitythreshold.
RESET:PresstheTESTOKbuttontoresetthesettingstothefactorydefault.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
19
5.12.6SilentMode
InSilentmode,thealertbeeperisdisabled.Silentmodedoesnotaffectthecontinuitybeeper.
SelectandenableSilentmodeasdescribedinsection5.12.1Selectingmodes.
5.12.7DataRecordingMode(99point)
Themeterhas99memorylocationsforthestorageofmeasurementdata.
SelectandenableMEMmode asdescribedinsection5.12.1Selectingmodes.
Usethearrowkeystocyclethroughthemodeoptions:SAVE,LOAD,andCLEARshownonthe
secondarydisplay.
SAVE:Thedataonthemaindisplayissavedtoamemorylocationindicatedbythe‘88’inthe
upperdisplayarea.
LOAD:Thedatastoredinthememorylocation(identifiedbythe‘88’indicator)isdisplayed.
Usearrowkeystoscrollthememorylocations.Usethe buttontoexittheLOADmode.
CLEAR:Thedatainallmemorylocationsiserased.
5.13KeypadLockout
InLockoutmodethemeterignoresallbuttonpressesexceptHOLD.Theautopowerofffunction,
seesection5.1.1AutoPoweroff,isdisabledinLockoutmode.
PressandholdtheHOLD/LOCKbuttonfor3secondstoenter/exitthelockoutmode.Inlockout
mode,thelockindicatorisdisplayed.DonotconfuseLockoutmodewiththetestlockfeatureused
forcontinuousInsulationResistancetesting(seeSection5.7InsulationResistanceMeasurements).
5.14StreamingMeasurementDatausingBluetooth®
5.14.1General
SomeIRcamerasfromFLIRSystemssupportBluetoothcommunication,andtothosecameras
youcanstreammeasurementdatafromthemeter.Thedataisthenmergedintotheresultant
tableintheIRimage.ThisisfurtherfacilitatedthroughaFLIRsoftwaretoolknownas
METERLiNK®.TostreamdatatoaniOS/Androiddevice,downloadtheapplicationFLIR
TOOLSTMandfollowtheapplicationonscreenprompts.
StreamingmeasurementdataisaconvenientwaytoaddimportantinformationtoanIRimage.
Forexample,whenidentifyinganoverheatedcableconnection,youmaywanttoknowits
voltage.TheBluetoothrangeis10m(32ft)maximum.
5.14.2Procedures
1. PairtheIRcamerawiththeinstrument.Refertothecameramanualforinformationon
howtopairBluetoothdevices.
2. Turnonthecameraandthenturnonthemeter.
3. PressthebuttononthemetertoenableBluetooth.TheBluetoothiconwillappear
flashingintheupperlefthandcornerofthemeterdisplay.
4. Choosethevariablethatyouwanttouse(voltage,bondresistance,etc.).Resultsfromthe
meterwillnowautomaticallybedisplayedintheresultanttableinthetopleftcornerof
theIRcamerascreen.
FLIRIM75USERGUIDEDocumentIdentifier:IM75enUS_AA
20
6. Maintenance
6.1CleaningandStorage
Cleanthemeterwithadampclothandmilddetergent;donotuseabrasivesorsolvents.
Ifthemeterisnottobeusedforanextendedperiod,removethebatteriesandstorethem
separately.
6.2BatteryReplacement
TheBatterysymbolflasheswithno‘bars’whenthebatterieshavereachedacriticallevelof7.0V.
ThesymbolstaysactiveandvisiblewhiletheLCDispowered.
Themeterdisplaysreadingswithinspecificationswhilethelowbatteryindicatorison.Whenthisis
nolongerpossible,thedisplaymustbeblanked.Themeterpowersoffbeforeitdisplaysanoutof
tolerancevoltage.
1. Toavoidelectricalshock,disconnectthemeterifconnectedtoacircuit,removetheprobe
fromtheterminals,andsetthefunctionswitchtotheOFFpositionbeforeattemptingto
replacethebatteries.
2. Unscrewandremovethebatterycompartmentcover.
3. ReplacethesixstandardAAAbatteries,observingcorrectpolarity.
4. Securethebatterycompartmentcover.
6.3FuseReplacement
Thefuseaccessedviathebatterycompartmentcover.Thefuseisrated440mA/1000V,ceramic
fastblowwithaminimumInterruptRatingof10kA.
6.4DisposalofElectronicWaste
Aswithmostelectronicproducts,thisequipmentmustbedisposedofinan
environmentallyfriendlyway,andinaccordancewithexistingregulationsfor
electronicwaste.PleasecontactyourFLIRSystemsrepresentativeformoredetails.
/