Extech Instruments MM750W Wireless Datalogging CAT IV True RMS Multimeter User manual

Category
Measuring, testing & control
Type
User manual

This manual is also suitable for

USER MANUAL
Wireless Datalogging CAT IV True RMS Multimeter
Model MM750W
2 MM750W-en-GB_v1.2 2/20
Table of contents
Introduction ................................................................................................... 3
FEATURES .......................................................................................................................................... 3
Safety ............................................................................................................. 3
Meter Description ........................................................................................... 6
Control Buttons .............................................................................................. 7
Display Symbols .............................................................................................. 7
Multimeter Operating Instructions .................................................................. 8
AUTORANGING/MANUAL RANGE SELECTION ......................................................................................... 8
MAXIMUM (MAX), MINIMUM (MIN), and MAX minus MIN modes ........................................................... 8
RELATIVE ZERO MODE ......................................................................................................................... 8
LCD BACKLIGHT .................................................................................................................................. 9
DATA HOLD ........................................................................................................................................ 9
AUTO POWER OFF (APO) ..................................................................................................................... 9
LOW BATTERY INDICATION ................................................................................................................... 9
DC VOLTAGE MEASUREMENTS ............................................................................................................ 10
AC VOLTAGE, FREQUENCY, DUTY CYCLE MEASUREMENTS ...................................................................... 10
DC CURRENT MEASUREMENTS ........................................................................................................... 11
AC CURRENT, FREQUENCY, DUTY CYCLE MEASUREMENTS ...................................................................... 12
CLAMP ADAPTOR MEASUREMENTS AC/DC ........................................................................................... 13
RESISTANCE MEASUREMENTS ............................................................................................................ 14
CONTINUITY CHECK ........................................................................................................................... 14
DIODE TEST ...................................................................................................................................... 15
CAPACITANCE MEASUREMENTS .......................................................................................................... 15
TEMPERATURE MEASUREMENTS ........................................................................................................ 16
NCV NON-CONTACT (EF) VOLTAGE MEASUREMENTS ............................................................................. 17
Wireless Communication .............................................................................. 17
WIRELESS COMMUNICATIONS OVERVIEW............................................................................................ 17
INSTALLING A WIRELESS MODULE ....................................................................................................... 17
OBTAINING THE APPLICATION (APP) .................................................................................................... 17
USING THE APPLICATION ................................................................................................................... 18
FCC COMPLIANCE ............................................................................................................................. 19
Maintenance ................................................................................................ 19
BATTERY INSTALLATION ..................................................................................................................... 20
FUSE REPLACEMENT.......................................................................................................................... 20
Specifications ............................................................................................... 21
3 MM750W-en-GB_v1.2 2/20
Introduction
Thank you for selecting the Extech Model MM750W Wireless Datalogging CAT IV True
RMS Multimeter. Proper use and care of this meter will provide many years of reliable
service. Please visit our website (www.extech.com) to check for the latest version of this
User Manual and Customer Support.
FEATURES
True RMS responding Auto/Manual Range DMM
6000 count (3-3/4 digit) backlit LCD
Measure AC/DC Voltage and Current, Frequency, Duty Cycle, Capacitance,
Continuity, Diode, and Temperature (Type-K thermocouple)
Non-contact voltage detection
Accepts external clamp adaptor input
Data Hold
MAXIMUM, MINIMUM, and MAX minus MIN recording
Relative Zero mode
Auto Power OFF (APO), can be temporarily disabled
Wireless connectivity with Smart phones and Tablets
Bluetooth® Wireless Datalogger Module (DAT12) stores over 15k readings for
wireless transmission using the ExView® W-Series application
Rated CAT IV 600V, CAT III 1000V
Safety
This symbol adjacent to another symbol, terminal or operating device indicates
that the operator must refer to an explanation in the Operating Instructions to
avoid personal injury or damage to the meter.
WARNING indicates that a potentially hazardous situation, which if not avoided,
could result in death or serious injury.
CAUTION indicates that a potentially hazardous situation, which if not avoided,
may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected
to a circuit point at which the voltage with respect to earth ground exceeds (in this
case) 1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated
with ranges that may, in normal use, be subjected to particularly hazardous
voltages. For maximum safety, the meter and its test leads should not be handled
when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation
or reinforced insulation.
WARNING
CAUTION
MAX
1000V
4 MM750W-en-GB_v1.2 2/20
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures
are taken to limit the transient overvoltages to an appropriate low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the
fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note Examples include switches in the fixed installation and some equipment for industrial use
with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation.
Note – Examples include electricity meters and primary over-current protection equipment
CAUTIONS
Improper use of this meter can cause damage, shock, injury or death. Read and understand
this user manual before operating the meter.
Always remove the test leads before replacing the battery or fuses.
Inspect the condition of the test leads and the meter itself for any damage before operating
the meter.
Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages constitute a shock hazard.
Warning! This is a class A’ device. This equipment can cause residential radio signal
interference.
Always discharge capacitors and remove power from the device under test before
performing Diode, Resistance or Continuity tests.
Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of the connection to the recessed electrical contacts. Use other means to
ensure that the terminals are not ‘live’.
Do not operate this device in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Do not allow children or pets access to this device, its batteries and other accessories, or
the packaging materials.
If this device is going to be unused for an extended period, remove the batteries.
Expired or damaged batteries can cause cauterization on contact with the skin. Always use
suitable hand protection.
Do not short-circuit the batteries and do not dispose of batteries in fire.
5 MM750W-en-GB_v1.2 2/20
We designed this device for safe use but please operate with caution. Follow the safety rules,
below, for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
Maximum Input
V DC or V AC
1000VDC / AC rms
uA AC/DC
600mA 1000V fast acting fuse
mA AC/DC
600mA 1000V fast acting fuse
A AC/DC
10A 1000V fast acting fuse (15 seconds max.;
always turn the meter OFF for 15 minutes
between 10A measurement cycles)
Frequency/Duty Cycle, Resistance,
Diode Test, Capacitance, Continuity
1000V DC / AC rms
2. USE EXTREME CAUTION when working with high voltages.
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above earth
ground.
4. NEVER connect the meter leads across a voltage source while the function switch is in the
current, resistance, or diode position; doing so can damage the meter.
5. ALWAYS discharge filter capacitors in power supplies and disconnect the power when making
resistance or diode tests.
6. ALWAYS turn off the power and disconnect the test leads before opening the covers to replace
the fuse or batteries.
7. NEVER operate the meter unless the back cover and the battery and fuse covers are in place
and fastened securely.
8. Do not use the device immediately after moving it from a cold to a warm environment;
condensation may form and damage the meter. Leave the device switched off until it has
reached room temperature before use.
9. Do not use the meter immediately before, during, or after an electrical storm (thunder,
lightning, etc.).
10. If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
6 MM750W-en-GB_v1.2 2/20
Meter Description
METER FRONT
1. Non-Contact Voltage Detector
2. LCD
3. Control buttons
4. Function switch
5. Positive 10A input terminal
6. COM negative input terminal
7. Positive terminal for all inputs except 10A
METER BACK
1. Wireless module compartment
2. Test lead holders
3. Battery compartment
4. 600mA/1000V Fuse compartment
5. Tilt stand
6. 10A/1000V Fuse compartment
7 MM750W-en-GB_v1.2 2/20
Control Buttons
Short press to enter Manual range and to select range manually
Long press to return to Auto range
Short presses step through Maximum/Minimum readings, and
MAX minus MIN; Long press to exit. Note that Auto Range is not
operational in MAX/MIN mode. Refer to the MAX/MIN section of
this manual for details
Short press for Relative mode ON/OFF
Long press for data transmission ON/OFF
Short press to select one of the functions available for a given
rotary switch position; Long press while powering up to defeat
Auto Power OFF (APO)
Short press for Data Hold ON/OFF
Long press backlight ON/OFF
Short press to step throughFrequency/Duty Cycle and to exit
Display Symbols
1. Automatic range
2. Diode test
3. Continuity test
4. Data hold
5. Relative zero
6. Maximum-Minimum mode
7. Temperature units
8. Non-contact voltage detector
9. Units of measure
10. Measurement display digits
11. Low battery icon
12. Minus sign
13. AC and DC symbols
14. Wireless Transmission
8 MM750W-en-GB_v1.2 2/20
Multimeter Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous; please measure with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. When Manual ranging, if OL appears in the display during a measurement, the
value exceeds the selected range. Change to a higher range.
AUTORANGING/MANUAL RANGE SELECTION
The meter defaults to the Auto Range mode. This automatically selects the best range
for measurements and is generally the best mode for most measurements. For Manual
Range operation:
1. Press the R (Range) button. The “AUTO” display indicator will disappear.
2. Press the R key to step through available ranges until you select the desired range.
3. Long press R to exit to return to Auto Range mode.
Note: Manual ranging does not apply for the Capacitance and Frequency functions.
MAXIMUM (MAX), MINIMUM (MIN), and MAX minus MIN modes
Note: The Auto Range function is not available in MAX/MIN mode. Please select the desired
range manually before entering MAX/MIN mode. If the reading exceeds that range, OL
will appear. Auto Range mode will re-activate when you exit MAX/MIN mode.
1. Short press MAX/MIN to activate the MAX/MIN recording mode. The display icon
"MAX" will appear. The meter will display and hold the maximum reading and will
update only when a new “maxoccurs.
2. Press MAX/MIN again and the display icon "MIN" will appear. The meter will display
and hold the minimum reading and will update only when a new “minoccurs.
3. Press MAX/MIN again and the display icon "MAX-MIN" will appear. The meter will
display and hold the MAX minus the MIN reading.
4. Long press MAX/MIN to exit.
RELATIVE ZERO MODE
The relative measurement feature allows you to make measurements relative to a
stored reference. In this mode, the displayed value represents the difference between
the reference and the measured value. Also, use this button to zero the display by
removing stray signals before taking a reading.
1. Take a measurement.
2. Press the REL button to store the reading; the relative icon will appear.
3. The display will then indicate the difference between the stored value and
subsequent measurements.
4. Press the REL button to exit the relative mode.
9 MM750W-en-GB_v1.2 2/20
LCD BACKLIGHT
Long press the backlight key
to switch the backlight ON/OFF.
DATA HOLD
The Data Hold function freezes the reading in the display. Short press the H (Hold) key
to activate/deactivate the Data Hold function.
AUTO POWER OFF (APO)
APO will turn the meter off after 5 minutes of inactivity. The meter will emit five
audible beeps approximately 1 minute before automatically powering off, after which
the meter loudly beeps once as it switches off. To wake the meter, press the M button;
no need to turn the rotary switch.
To disable APO, press and hold the M button while turning the function switch from
the OFF position to any other position. Five (beeps) will sound indicating that the APO
feature has been disabled. Note that the APO will be re-enabled on the next cycle of
power to the meter.
LOW BATTERY INDICATION
The battery icon will appear when the battery voltage is critical. Replace the battery
before continuing to use the meter.
10 MM750W-en-GB_v1.2 2/20
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is switching ON/OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the position.
2. Use the M (Mode) button to select (DC).
3. Insert the black test lead banana plug into the negative
COM jack. Insert the red test lead banana plug into the
positive V jack.
4. Touch the black test probe tip to the negative side of the
circuit. Touch the red test probe tip to the positive side
of the circuit.
5. Read the voltage in the display.
AC VOLTAGE, FREQUENCY, DUTY CYCLE
MEASUREMENTS
WARNING: Risk of Electrocution. The probe tips may not reach the live parts inside
some 240V outlets. As a result, the reading may show 0 volts when the outlet actually
has voltage. Make sure the probe tips are touching the metal contacts inside the
outlet before assuming that no voltage is present.
CAUTION: Do not measure AC voltages if a motor on the circuit is switching ON/OFF.
Large voltage surges may occur that can damage the meter.
1. Set the function switch to the position.
2. Use the M (Mode) button to select (AC).
3. Insert the black test lead banana plug into the negative
COM jack. Insert red test lead banana plug into the
positive V jack.
4. Touch the black test probe tip to the neutral side of the
circuit. Touch the red test probe tip to the “hot” side of
the circuit. Read the voltage in the display.
5. Press the HZ/% button to indicate Hzand read the
frequency in the display.
6. Press the Hz/% button again to indicate ‘%’. Read the
duty cycle % in the display.
7. Press Hz/% again to exit.
11 MM750W-en-GB_v1.2 2/20
DC CURRENT MEASUREMENTS
CAUTION: Do not make 10A current measurements for longer than 15 seconds;
Exceeding 15 seconds may cause damage to the meter and/or the test leads. Always
turn the meter OFF for 15 minutes before measuring again.
WARNING: Never measure current above 600mA in the µA/mA range and 10A in the
10A range. The voltage in the measurement circuit must not exceed 1000VDC
1. Insert black test lead banana plug into the negative COM jack.
2. For current measurements up to 6000µA DC, set the function switch to the µA
position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 600mA DC, set the function switch to the mA
position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 10A DC, set the function switch to the 10A position
and insert the red test lead banana plug into the 10A jack.
5. Press the M button to indicate (DC) on the display.
6. Remove power from the circuit under test, then open up the circuit at the point
where you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit. Touch the red test
probe tip to the positive side of the circuit.
8. Apply power to the circuit and read the current on the display.
12 MM750W-en-GB_v1.2 2/20
AC CURRENT, FREQUENCY, DUTY CYCLE MEASUREMENTS
CAUTION: Do not make 10A current measurements for longer than 15 seconds;
Exceeding 15 seconds may cause damage to the meter and/or the test leads. Always
turn the meter OFF for 15 minutes before measuring again.
WARNING: Never measure current above 600mA in the µA/mA range and 10A in the
10A range. The voltage in the measurement circuit must not exceed 1000V AC.
1. Insert black test lead banana plug into the negative COM jack.
2. For current measurements up to 6000µA AC, set the function switch to the µA
position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 600mA AC, set the function switch to the mA
position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 10A AC, set the function switch to the 10A position
and insert the red test lead banana plug into the 10A jack.
5. Press the M button to indicate (AC) on the display.
6. Remove power from the circuit under test, then open up the circuit at the point
where you wish to measure current.
7. Touch the black test probe tip to the neutral side of the circuit. Touch the red test
probe tip to the ‘hot’ side of the circuit.
8. Apply power to the circuit. Read the current in the display.
9. Press the Hz/% button to indicate Hz. Read the frequency on the display.
10. Press the Hz/% button again to indicate %. Read the duty cycle on the display.
11. Press the Hz/% button to return to current measurement.
13 MM750W-en-GB_v1.2 2/20
CLAMP ADAPTOR MEASUREMENTS AC/DC
1. Turn the rotary switch to the clamp adaptor position .
2. Press the M button to indicate AC or DC on the display.
3. Connect the external clamp adaptor’s negative lead to the meter’s COM jack.
4. Connect the external clamp adaptor’s positive lead to the meter’s positive jack.
5. The meter displays 1A per each mV input (for example, with a 3mV input, the
meter will display 3A).
6. Take a current measurement with the external clamp adaptor and read the current
on the meter display.
7. For AC, Press the Hz/% button to indicate ‘Hz’. Read the frequency on the display.
8. For AC, Press the Hz/% button again to indicate ‘%’. Read the duty cycle on the
display.
9. Press the Hz/% button to return to current measurement.
14 MM750W-en-GB_v1.2 2/20
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any resistance measurements. Remove the
batteries and unplug the line cords.
1. Set the function switch to the Ω position.
2. Insert the black test lead banana plug into the negative COM
jack. Insert the red test lead banana plug into the positive Ω
jack.
3. Press the M button to indicate Ω on the display.
4. Touch the test probe tips across the circuit or part under test.
It is best to disconnect one side of the part under test so the
rest of the circuit will not interfere with the resistance
reading.
5. Read the resistance in the display.
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that
have voltage on them.
15 MM750W-en-GB_v1.2 2/20
1. Set the function switch to the position.
2. Insert the black lead banana plug into the negative COM
jack. Insert the red test lead banana plug into the positive
jack.
3. Press the M button to indicate and Ω on the display.
4. Short press the REL button to null any stray resistance.
5. Touch the test probe tips to the circuit or wire under test.
6. If the resistance is less than approximately 50 ohms, an
audible tone will sound. If the circuit is open, the display
will indicate OL.
DIODE TEST
WARNING: To avoid electric shock, never measure diodes that have voltage on them.
1. Set the function switch to the position.
2. Insert the black test lead banana plug into the negative COM
jack and the red test lead banana plug into the positive V jack.
3. Press the M button to indicate and V on the display.
4. Touch the test probes to the diode under test. Forward
voltage will typically indicate 0.400 to 0.700V. Reverse
voltage will indicate OL. Shorted devices will indicate near
0V and an open device will indicate OL in both polarities.
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and
discharge all capacitors before taking any measurements. Remove the batteries and
unplug the line cords.
16 MM750W-en-GB_v1.2 2/20
1. Set the function switch to the position.
2. Insert the black test lead banana plug into the negative COM
jack. Insert the red test lead banana plug into the positive
jack.
3. Press the M button to indicate on the display.
4. Short press the REL button to null any stray signal.
5. Touch the test probe tips across the circuit or part under test.
6. Read the capacitance on the display. Note that larger
capacitors can take some time before stabilizing.
TEMPERATURE MEASUREMENTS
WARNING: The supplied thermocouple range is limited and does not measure over
the entire temperature range of the meter.
1. Set the function switch to the TEMP position.
2. Press the M button to indicate
o
C or
o
F on the
display.
3. Connect the Type-K thermocouple to the
thermocouple-to-banana adaptor.
4. Connect the adaptor’s negative lead to the COM
jack.
5. Connect the adaptor’s positive lead to the positive
jack.
6. Touch the Type-K tip to the part under test or leave
in open air.
7. Read the temperature on the display. Use the M
button to select
o
C or
o
F.
17 MM750W-en-GB_v1.2 2/20
NCV NON-CONTACT (EF) VOLTAGE MEASUREMENTS
WARNING: Check the NCV function on a known live circuit before testing on a circuit
of unknown status.
1. Set the function switch to the position. The ‘EF’ and ‘NCV’ icons will appear.
2. The non-contact voltage detection sensor is located at the
top of the meter. Place the sensor near the source of
voltage.
3. When the meter detects voltage it will audibly beep, the
display will blink red, and dashes will appear. The number of
dashes is proportional to the strength of the voltage source
as is the periods between audible beeps. Note that the
meter flashes red with each audible beep.
Wireless Communication
WIRELESS COMMUNICATIONS OVERVIEW
We ship the MM750W with a Bluetooth® Wireless Datalogging Module (DAT12)
installed in the compartment on the back (top) of the meter. The DAT12 transmits real-
time readings and logged readings to paired iOS® or Android™ devices using the free
ExView® W-Series application. The DAT12 can record >15k readings.
INSTALLING A WIRELESS MODULE
The communication modules install in the compartment at the back (top) of the meter;
see the Meter Description section of this user manual for location. To install a
communications module first turn the meter OFF then remove the two Phillips screws
that secure the module compartment. Open the compartment, insert the module with
the arrow correctly oriented at the top of the module, and connect via the 8-pin
connector inside the module compartment. Close the compartment and secure with
the two screws before use.
OBTAINING THE APPLICATION (APP)
For iOS devices, download the ExView® W application from the Apple App Store. For
Android devices, download the ExView® W application from the Google Play
TM
store.
18 MM750W-en-GB_v1.2 2/20
USING THE APPLICATION
1. Turn on the meter and select a measurement function
2. Long press the Tx button to start the wireless transmission (the Tx display icon will
appear when started)
3. On your smart device, tap the ExView® W application icon to start the App.
(Bluetooth® must be enabled on your smart device)
4. Tap the search icon located next to Devices. The App will search for available
devices.
5. When the meter appears in the device list, tap it to connect the meter to the App.
6. For more information please refer to the ExView® W-Series Help Guide by tapping
on the Extech icon and then tapping the Help Guide link, or locate the document
on the extech.com/exvieww web page.
7. Please note that some Android™ devices require that you switch ON the device’s
location setting before the ExView® W app can establish a connection with the
wireless meters.
19 MM750W-en-GB_v1.2 2/20
FCC COMPLIANCE
A wireless module must be installed in the meter before the module can communicate with
iOS®/Android™ phones and tablets via Bluetooth® using the ExView® W application.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference.
2. This device must accept any interference received, including interference that may cause undesired operation.
IC: 1590A-MM750W
FCC ID: IWK-MM750W
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part
15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning the equipment
off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
1. Reorient or relocate the receiving antenna.
2. Increase the separation between the equipment and receiver.
3. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
4. Consult the dealer or an experienced radio/TV technician for help.
WARNING
Changes or modifications not expressly approved by the party responsible for compliance could void the user's
authority to operate the equipment.
Maintenance
WARNING: To avoid electric shock, disconnect the test leads from any source of
voltage before removing the back cover or the battery or fuse covers.
WARNING: To avoid electric shock, do not operate your meter until the battery and
fuse covers are in place and fastened securely.
We designed this meter to provide years of dependable service, please follow these
instructions carefully:
1. KEEP THE METER DRY. If it gets wet, allow it to fully dry before use.
2. USE AND STORE THE METER IN NORMAL TEMPERATURES. Temperature extremes
can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the
electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. Do not use
abrasives, chemicals, cleaning solvents, or detergents.
20 MM750W-en-GB_v1.2 2/20
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old
or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD, remove the batteries to
prevent damage to the unit.
BATTERY INSTALLATION
1. Turn power off and disconnect the test leads from the meter.
2. Open the rear battery cover by removing two screws that secure the cover using a
Phillips head screwdriver. See the Meter Description section of this manual for
location of compartment and screws.
3. Insert the battery into battery holder, observing the correct polarity.
4. Place the battery cover back in place. Secure with the screws.
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or wherever
batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated
to take end-of-life devices to a designated collection point for the disposal of electrical
and electronic equipment.
FUSE REPLACEMENT
1. Turn power off and disconnect the test leads from the meter.
2. The fuse compartments are located on either side of the battery compartment under
the orange protective cover. You will have to pry up on the cover to expose the fuse
compartments.
3. The 10A/1000V fuse is located in the fuse compartment on the right (when facing
the back of the meter), and the 600mA/1000V fuse compartment is located on the
left (see Meter Description section of this manual).
4. Insert the fuse into the fuse holder.
5. Close the fuse compartment and secure with the screws.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23

Extech Instruments MM750W Wireless Datalogging CAT IV True RMS Multimeter User manual

Category
Measuring, testing & control
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI