Extech Instruments MA160 User manual

Category
Multimeters
Type
User manual
USERGUIDE
TrueRMS200AAC/DCOpenJawClampMeter
ModelMA160
MA160enGB_V1.010/15
2
TableofContents
1. INTRODUCTION 3
2. SAFETYINFORMATION 4
3. DESCRIPTIONS 6
4. OPERATION 9
PoweringtheMeter 9
DisplayBacklight 9
WorkLight 9
DataHold 9
TestLeadConsiderations 9
VoltageMeasurements 10
ACVoltageMeasurements 10
DCVoltageMeasurements 11
CurrentMeasurementsUsingtheClamp 12
NonContactVoltageDetector 13
ResistanceMeasurements 14
ContinuityMeasurements 15
CapacitanceMeasurements 16
DiodeTest 17
5. MAINTENANCE 18
6. SPECIFICATIONS 19
MA160enGB_V1.010/15
3
1. Introduction
ThankyouforselectingtheExtechMA160ClampMeter.
TheMA160isafeaturepackedOpenJawTrueRMSdigitalClampmeter.
MeasureAC/DCvoltage,ACcurrent,DCcurrent,Resistance,Diode,Continuity,andCapacitance.
TheMA160alsoincludesaNonContactVoltageDetectionfeature.
OtherfunctionsincludeDataHold,Relativemode,Worklight,Backlight,andAutopoweroff.
Thisdeviceisshippedfullytestedandcalibratedand,withproperuse,willprovideyearsof
reliableservice.Pleasevisitourwebsite(www.extech.com)tocheckforthelatestversionofthis
UserGuide,ProductUpdates,ProductRegistration,andCustomerSupport.
Features
6000countdigitaldisplay
LargebacklitLEDdisplay
TrueRMSACmeasurements
0.8%DCVaccuracy
DataHold
Relativemode
AutoPowerOFF(APO)
200AACcurrentmeasurements
200ADCCurrentMeasurements
NonContactVoltageDetector
Visualandaudiblecontinuitymeasurement
alert
Lowbatteryindicator
CATIII600V/CATII1000V
MA160enGB_V1.010/15
4
2. SafetyInformation
Toensurethesafeoperationandserviceofthemeter,followtheseinstructionsclosely.Failureto
observewarningscanresultinsevereinjury.
WARNINGS
WARNINGSidentifyhazardousconditionsandactionsthatcouldcauseBODILYHARMorDEATH.
Whenhandlingtestleadsorprobes,keephandsandfingersbehindthefingerguardsatalltimes.To
avoidelectricalshockdonottouchexposedelectricalwire,connectors,unusedinputterminals,or
circuitsundertest.
Removetestleadsfromthemeterbeforeopeningthebatterycompartmentormeterhousing.
UsethemeteronlyasspecifiedinthisUserGuideoraccompanyingQuickStarttoavoidcompromising
theprotectionsprovidedbythemeter.
Besuretousetheproperterminals,switchpositions,andrangeswhentakingmeasurements.
Verifythemeter’soperationbymeasuringaknownvoltage.Have
themeterservicedifthemeter
respondsunusuallyoriftherearequestionsregardingthemeter’sfunctionalintegrity.
Donotapplymorethantheratedvoltage,asmarkedonthemeter,betweenterminalsorbetweenany
terminalandearthground.
Donotmeasurevoltagesabove1000VDCor600VAC
betweenterminalandgroundtopreventelectrical
shockanddamagetotheClampmeter.
Usecautionworkingwithvoltagesabove30VACRMS,42VACpeak,or60VDC.Thesevoltagesposea
shockhazard.
Toavoidmisleadingreadingsthatcouldleadtoelectricshockandinjury,replace
thebatteriesassoonas
thelowbatteryindicatorisdisplayed.
Disconnectpowertothecircuitunde rtestanddischargeallhighvoltagecapacitorsbeforetesting
resistance,continuity,diodes,orcapacitance.
Donotusethemeterinthepresenceofexplosivegasorvapor.
Toreduceriskof
fireorelectricshock,donotusethemeterifitiswetanddonotexposethemeterto
moisture.
IndividualprotectiveequipmentshouldbeusedifHAZARDOUSLIVEpartsintheinstallationwhere
measurementsaretobecarriedoutcouldbeaccessible.
CAUTIONS
CAUTIONSidentifyconditionsandactionsthatcouldcauseDAMAGEtothemeterorequipment
undertest.Donotexposethemetertoextremesintemperatureorhighhumidity.
Disconnectthetestleadsfromthetestpointsbeforechangingthepositionofthefunction(rotary)
switch.
Donotexposethemetertoextremesintemperatureortohighhumidity.
Neversetthemetertotheresistance,diode,capacitance,microamp,orampfunctionswhen
measuringthevoltageofapowersupplycircuit;thiscouldresultinmeterdamageanddamagetothe
equipmentundertest.
MA160enGB_V1.010/15
5
SafetySymbolsthataretypicallymarkedonmetersandinstructions
Thissymbol,adjacenttoanothersymbol,indicatestheusermustrefertothe
manualoruserguideforfurtherinformation.
Riskofelectricalshock
Equipmentprotectedbydoubleorreinforcedinsulation
LowBatterysymbol
ConformstoEUdirectives
Donotdiscardthisproductinhouseholdtrash.
ACmeasurement
DCmeasurement
Earthground
PERIEC1010OVERVOLTAGEINSTALLATIONCATEGORY
OVERVOLTAGECATEGORYI
EquipmentofOVERVOLTAGECATEGORYIisequipmentforconnectiontocircuitsinwhichmeasuresare
takentolimitthetransientovervoltagestoanappropriatelowlevel.
NoteExamplesincludeprotectedelectroniccircuits.
OVERVOLTAGECATEGORYII
EquipmentofOVERVOLTAGECATEGORYIIisenergyconsumingequipmentto
besuppliedfromthefixed
installation.
NoteExamplesincludehousehold,office,andlaboratoryappliances.
OVERVOLTAGECATEGORYIII
EquipmentofOVERVOLTAGECATEGORYIIIisequipmentinfixedinstallations.
NoteExamplesincludeswitchesinthefixedinstallationandsomeequipmentforindustrialusewith
permanentconnectiontothefixedinstallation.
OVERVOLTAGECATEGORYIV
EquipmentofOVERVOLTAGECATEGORYIVisforuseattheoriginoftheinstallation.
NoteExamplesincludeelectricitymetersandprimaryovercurrentprotectionequipment
MA160enGB_V1.010/15
6
3. Descriptions
MeterDescription
1. NonContactVoltageDetector
2. Openjaw
3. WorkLight
4. NCValertLEDlamp
5. WorkLightswitch
6. Rotaryfunctionswitch
7. Mode(M)button
8. Relative
,DCAZerobutton
9. HOLDandBacklightbutton
10. Display
11. Testleadjacks
Note:Batterycompartmentonbackofmeter
Fig31METERDESCRIPTION
1
3
11
10
9
8
6
5
2
7
4
MA160enGB_V1.010/15
7
DisplayIconDescriptionsforMA160
Fig32MA160METERDISPLAY
1. Unitsofmeasure
2. DCAZero
3. Relativemode
4. Datahold
5. AlternatingCurrent
6. DirectCurrent
7. Automaticrange
8. Continuity
9. Diode
10. Capacitance
11. NonContactVoltageDetector
12. Autopoweroff
13. Batterystatus
14. LCDdisplay
MA160enGB_V1.010/15
8
PushButtonDescriptions
MomentarypressesoftheMode(M)buttonperformthefunctionsshowninthetable
below.
Fig.34Mode(M)ButtonFunctionTable
SwitchPosition Mode(M)ButtonFunction
ACDC
AC DC
Ω


Momentarilypresstoaccess/exittheRelativemode.Thismodeofoperationisonly
availableforAC/DCvoltage,ACCurrent,andCapacitance.
InDCAmode,presstozerothedisplay;pressandholdtoexitthezero.
Pressandholdthisbuttontoactivate/deactivatetheLCDbacklight.Momentary
pressesswitchDATAHOLDONorOFF.
FunctionSwitchDescription
1. MeterPOWEROFFposition
2. AC/DCCurrentmode(useMbuttontochooseACorDC)
3. AC/DCVoltagemode(useMbuttontochooseACorDC)
4. Capacitance,Continuity,Diode,Resistancemodes
(useMbuttontochoosemode)
5. NonContactVoltageDetectpositionwithalertLED
Fig33MA160FunctionSwitch
MA160enGB_V1.010/15
9
4. Operation
CAUTION:ReadandunderstandalloftheSafetystatementslistedinthesafetysection
ofthismanualpriortouse.
PoweringtheMeter
1. Turntherotaryfunctionswitchtoanypositiontopowerthemeter.Checkthebatteriesifthe
unitfailstopowerON.RefertotheMaintenancesectionforbatteryreplacement.
2. TurnthefunctionswitchtotheOFFpositiontopowerOFFthemeter.
3. ThemeterhasanAuto
PowerOFFfeature(APO)wherethemeterswitchesOFFafter15
minutesofinactivity.WhenAPOisenabled,theAPOiconwillshowonthedisplaywhenthe
meterispoweredON,
Thelowbatterysymbol
appearsonthedisplaywhenthebatteryvoltageweakensbelowthe
threshold.
DisplayBacklight
WiththemeterpoweredON,pressandholdthebacklightbutton toswitchthebacklightONor
OFF.Notethatexcessiveuseofthebacklightwillshortenthebatterylife.
WorkLight
WiththemeterpoweredON,presstheWorkLightbuttonontherightsideofthemeter.A
momentarypresswillswitchtheWorkLightoff.TheWorkLightlampislocatedonthebackofthe
unittowardthebottomoftheclampjaw.Notethatexcessiveuseofthe
backlightwillshortenthe
batterylife.
DataHold
TofreezetheLCDmeterreading,presstheDataHold(H)button.Whiledataholdisactive,theH
displayiconappearsontheLCD.Pressthe(H)buttontoreturntonormaloperation.TheHiconwill
switchOFF.
TestLeadConsiderations
TestleadprobecoverscanberemovedforCATII1000Vinstallations.Usethetestleadprobe
coversforCATIII1000Vinstallations.Donotmeasurevoltages>1000VDCor600VAC.Remove
thestoragecapsfromthemeterendoftestleadsbeforeconnectingleadstothemeter.
MA160enGB_V1.010/15
10
VoltageMeasurements
WARNING:
RemovethetestleadprobecoversforCATII1000Vinstallations.Usethetestleadprobecovers
forCATIII600Vinstallations.
Donotmeasurevoltagesgreaterthan600VACor1000VDC.
CAUTION:Whenconnectingthetestleadstothecircuitordeviceundertest,connectthe
blackleadbeforethered;whenremovingthetestleads,removetheredbeforetheblacklead.
Disconnectprobesandcircuitundertestafterallmeasurementsarecompleted.
ACVoltageMeasurements
1. Inserttheblacktestleadbananaplugintothenegative(COM)jackandtheredtestlead
bananaplugintothepositive(V/)jack.
2. Turnthefunctionswitchtothe
position.UsetheMbuttontoselectAC.
3. ReadtheWarningandCautionstatementsabovetodeterminewhetherornottousethe
testleadprobecovers.
4. Touchthetestprobetipstothecircuitundertest.
5. Readthedigitalvalueshownonthedisplay.Themeter
isAutoRangingandtherefore
selectstheproperdecimalpointposition.Themeteralsodisplaysthemeasurement
type,unitsymbols,andotherrelevantmultifunctionicons.
6. ThemeteriscapableofdetectingACvoltagesto600V.
Fig41ACVOLTAGEMEASUREMENTS
MA160enGB_V1.010/15
11
DCVoltageMeasurements
1. Inserttheblacktestleadbananaplugintothenegative(COM)jackandtheredtestlead
bananaplugintothepositive(V/)jack.
2. MovetheFunctionSwitchtothe position.
3. UsetheMbuttontoselectDC.
4. ReadtheWarningandCautionstatementsatthebeginningoftheVoltageMeasurement
sectiontodeterminewhetherornottousethetestleadprobecovers.
5. Touchthetestprobetipstothecircuitundertest.Besureto
observethecorrectpolarity
(redleadtopositive,blackleadtonegative).
6. Readthedigitalvalueofthemeasurementinthedisplay.Thedisplaywillalsoindicate
theproperdecimalpoint(AutoRange)andmeasurementtype/unitssymbols.Ifthe
polarityisreversed,thedisplaywillshow()minus
beforethevalue.
7. ThemeteriscapableofdetectingDCvoltagesto1000V.
Fig42DCVOLTAGEMEASUREMENTS
MA160enGB_V1.010/15
12
CurrentMeasurementsUsingtheClamp
WARNING:Donotmeasurethecurrentonacircuitwhenthevoltageincreasestomore
than750VACor1000VDC.Thiscancausedamagetotheinstrumentandcancauseinjuryto
persons.
Fig.43CorrectandIncorrectClamping
1. Ensurethattheprobeleadsaredisconnectedfromthemeter.
2. Setthefunctionswitchtothe
position.
3. UsetheMbuttontoselectACorDC.ForDCpressZEROtoremoveanyresidual
magnetismandtonullthedisplaybeforeclampingontoaconductor.
4. Placethemetersoonlyoneconductorislocatedatthebaseofthejaws.Foroptimum
results,center
theconductorinthejaws.
5. Readthecurrentmeasurementinthedisplay.Thedisplaywillindicatetheproperdecimal
pointandvalue.InDC,whenthepolarityisreversed,thedisplaywillshow()minus
beforethevalue.
MA160enGB_V1.010/15
13
NonContactVoltageDetector
WARNING:Itispossibleforvoltagetobepresentinacircuitevenifthemeterdoesnot
beeporflashtheNCVLEDlamp.AlwaysverifymeteroperationonaknownliveACcurrent
circuitandverifythatthebatteriesarefreshbeforeuse.
WhenthemetersensesanACVoltageorelectromagneticfield>100VAC,thefollowingoccurs:
TheaudiblebeepersoundsONandOFF
TheLEDlampflashesONandOFF
Thedisplayshows1,2,3,or4dashes
Thegreatertheelectricalfieldstrength,thefastertherate
oftheaudiblebeeper,theflashingof
theLEDlamp,andthenumberofdashesdisplayed.Ifthemeterdoesnotemitatoneorflash
theLEDinthismode,thereisstillthepossibilitythatvoltageispresent;pleaseusecaution.
1. Turnthefunctionswitchtothe
NCVpositiontoselectNonContactVoltageDetect mode.
2. NotethatEFisdisplayedwheninthismode.IftheEFdoesnotdisplaywhenthefunction
switchisturnedtotheNCVposition,checkthebatteriesanddonotusethemeteruntilEF
isdisplayed.
3. To
test,placethemeternearasourceofelectricalenergy.Notethatthetipofthemeter
offersthehighestsensitivity.
4. Notetheaudiblebeeping,theflashingLED,andthedisplayeddasheswhena sourceof
electricalenergyisdetected.
Fig44NONCONTACTVOLTAGEDETECTOR
MA160enGB_V1.010/15
14
ResistanceMeasurements
CAUTION:SwitchOFFpowertothedeviceundertestbeforemeasuring.Donotteston
circuitsordeviceswhere60VDCor30VACispresent.
1. Inserttheblacktestleadbananaplugintothenegative(COM)jack.Inserttheredtest
leadbananaplugintothepositive(V/)jack.
2. TurntheFunctionSwitchtotheposition.
3. UsetheMbuttontoselecttheicononthedisplayindicatingresistanceonly
(without
thecontinuity/diode/capacitanceiconsshowing).
4. Touchthetestprobetipsacrossthecircuitorpartundertest.Itisbesttodisconnectone
sideofthepartundertestsotherestofthecircuitwillnotinterferewiththeresistance
reading.
5. Readtheresistancevalueinthe
display.Thedisplaywillindicatetheproperdecimal
pointandvalue.Ifthereadingisoutofrange,theOLdisplayiconwillappear.
Fig45RESISTANCEMEASUREMENTS
ResistanceMeasurementNotes:
Thedisplaywillshow“OL”whenanopencircuitisdetectedoriftheresistance>maximum
range.
Thetestleadsintroduceanerrorofapprox.0.1Ω0.2Ωforlowresistancemeasurements.
UsetheRelativemodetoobtainaccuratereadings.Shortthetestleadstogether,
pressthe
RELbutton,andthenmeasurealowresistance.Themetersubtractstheshortcircuitvalue
fromthereading.
Ifthetestleadresistanceofprobeis>0.5Ωwhenshorted,inspectthetestleadsandthe
connection.
Itmaytakeseveralsecondsforthereadingto
stabilizewhenmeasuringresistance>1MΩ.
Thisisnormaloperation.
Forpersonalsafety,donotmeasureacircuitwithvoltages>30VDCorAC.
MA160enGB_V1.010/15
15
ContinuityMeasurements
1. InserttheblacktestleadintothenegativeCOMterminalandtheredtestleadintothe
positiveterminal.
2. Setthefunctionswitchtothe position.
3. UsetheMbuttontoselecttheContinuitymode.LookfortheContinuityicononthe
display.
4. Touchthetestprobetipsacrossthewireorcircuitundertest.
5. Iftheresistanceis<10,thebeeperwillsoundcontinuously.Ifthe resistanceisbetween
10and100Ωthebeepermayormaynotsound.Foranopencircuitconditionthemeterwill
displayOL.
Fig46CONTINUITYMEASUREMENTS
ContinuityMeasurementNotes:
Turnoffpowertothecircuitundertestanddischargecapacitorsbeforemeasuring
continuity.
Opencircuitvoltageisapprox.‐3.5V
Disconnecttestleadsandcircuitmeasuredaftermeasurementsarecompleted.
MA160enGB_V1.010/15
16
CapacitanceMeasurements
WARNING:Toavoidelectricshock,removepowertothecircuitundertestanddischarge
thecapacitorundertestbeforemeasuring.Donottestoncircuitsordeviceswhere60VDCor
30VACispresent.
1. Setthefunctionswitchtothe
capacitanceposition.
2. InserttheblacktestleadbananaplugintothenegativeCOMjack
andtheredtestleadbananaplugintothepositive jack.
3. PresstheMbuttontoselecttheunitofmeasuresymbolF.
4. Touchthetestprobetipsacrossthepartundertest.
5. Forsmallcapacitancevalues,usetheRelativemode
toremovetestleadandprobe
capacitance.
6. Readthecapacitancevalueinthedisplay.
7. Thedisplaywillindicatetheproperdecimalpointandvalue.
Fig47CAPACITANCEMEASUREMENTS
CapacitanceMeasurementNotes:
Thedisplaywillshow“OL”ifacapacitorisshortcircuitedorifthemeasuredcapacitance>
maximumrangeoftheinstrument.
Capacitance measurements>600μFmayrequireseveralsecondstoobtainastable
reading.
Inordertoensuremeasurementaccuracy,dischargeresidualcharges
beforemeasuring
capacitance;Usemaximumsafetywhenworkingwithhighvoltagecapacitorstoprevent
damagetotheinstrumentandrisktopersonalsafety.
Disconnecttestleadsandcircuitundertestaftermeasurementsarecompleted.
MA160enGB_V1.010/15
17
DiodeTest
1. InserttheblacktestleadbananaplugintothenegativeCOMjackandtheredtest
leadbananaplugintothepositivejack.
2. Turnthefunctionswitchtoposition.UsetheMbuttontoselectthediodefunction,the
diodeandvoltagesymbolswillappearontheLCD
wheninDiodetestmode.
3. Touchthetestprobetipstothediodeorsemiconductorjunctionundertest.Notethe
meterreading.
4. Reversethetestleadpolaritybyreversingtheredandblackleads.Notethisreading.
5. Thediodeorjunctioncanbeevaluatedasfollows:
If
onereadingdisplaysavalue(typically0.400Vto0.900V)andreversereadingdisplays
OL,thediodeisgood.
IfbothreadingsdisplayOLthedeviceisopen.
Ifbothreadingsareverysmallor0’,thedeviceisshorted.
Fig48DIODETESTING
MA160enGB_V1.010/15
18
5. Maintenance
WARNING:Toavoidelectricalshock,removethetestleads,disconnectthemeterfromany
circuitandturnOFFthemeterbeforeopeningthecase.Donotoperatewithanopencase.
BatteryReplacement
1. Removethetestleadsfromthemeter.
2. RemovethePhillipsheadscrewthatsecuresthebatterycompartmentcoveronthebackof
themeter.
3. OpenthebatterycompartmentandreplacethetwoAAbatteries,observingcorrect
polarity.Reassemblethemeterbeforeuse
BatterySafetyNotes:Pleasedisposeofbatteriesresponsibly;neverdisposeofbatteriesinafire,
batteriesmayexplodeorleak.Ifthemeterisnottobeusedfor60daysormore,removethe
batteryandstoreseparately.Donotmixbatterytypesorfreshnesslevels;pleaseuse
batteries
ofthesametypeandofthesamefreshnesslevel.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriatecollection
sites,theretailstorewherethebatterieswerepurchased,orwhereverbatteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeendof
lifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronicequipment.
CleaningandStorage
Periodicallywipethecasewithadampclothandmilddetergent;donotuseabrasivesorsolvents.
Pleaseremovethebatteriesifthemeterisstoredforalongperiodoftime.

MA160enGB_V1.010/15
19
6. Specifications
ELECTRICALSPECIFICATIONS
Accuracyisgivenas±(%ofreading+leastsignificantdigits)at23
C±5
Cwithrelativehumidity
<80%.Accuracyisspecifiedforaperiodofoneyearaftercalibration.
Function Range Resolution Accuracy(reading) ‘OL’Protection
ACCurrent
200.0A 0.1A ±(2.5%+5digits) 200A
TRMS,FrequencyResponse50to60Hz
DCCurrent
200.0A 0.1A ±(2.5%+5digits) 200A
DCZerofunctioninuse
ACVoltage
6.000V 0.001V ±(1.2%+5digits)
1000VDC
750VAC
60.00V 0.01V
±(1.2%+3digits)
600.0V 0.1V
750V 1V ±(1.5%+5digits)
TrueRMSvoltageapplicableto10%~100%oftherange
InputImpedance:≥10M;Frequencyresponse40~400Hz
DCVoltage
6.000V 0.001V
±(0.8%+3digits)
1000VDC
750VAC
60.00V 0.01V
600.0V 0.1V
1000V 1V ±(1.0%+5digits)
InputImpedance:≥10MΩ
Resistance
600.0Ω 0.1Ω ±(1.2%+2digits)
1000VDC
750VAC
6.000kΩ 0.001kΩ
±(1.0%+2digits)
60.00kΩ 0.01kΩ
600.0kΩ 0.1kΩ
6.000MΩ 0.001MΩ
±(1.2%+2digits)
60.00MΩ 0.01MΩ
±(1.5%+5digits)
Continuity
600.0Ω 0.1Ω ‐‐‐‐‐ 1000VDC;750VAC
Lessthan10Ωbeepersounds.10Ωto100Ωundetermined.Greaterthan100Ω
beeperoff(OLdisplayed).Opencircuitvoltageapprox.1.2V
Diode
6.000V 0.001V 0.5to0.8V 1000VDC;750VAC
OpenCircuitVoltage:Approx.3.3V
MA160enGB_V1.010/15
20
Function Range Resolution Accuracy(reading) ‘OL’Protection
Capacitance
60.00nF 0.01nF
±(4.0%+20digits)
1000VDC
750VAC
600.0nF 0.1nF
6.000µF 0.001µF
60.00µF 0.01µF
600.0µF 0.1µF
6.000mF 0.001mF ±(10%)
60.00mF 0.01mF Forreferenceonly
NonContact
VoltageDetector
(NCV)
100Vrms;≤10mm(LED/Buzzerindication)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21

Extech Instruments MA160 User manual

Category
Multimeters
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI