Roche cobas t 511 User manual

Type
User manual

Roche cobas t 511 is a coagulation analyzer with fully automated, random continuous-access, and software controlled system for clotting, chromogenic, and immunoturbidimetric analysis. It's intended for qualitative and quantitative in vitro coagulation determinations to aid in diagnosing coagulation abnormalities and monitoring anticoagulant therapy.

Roche cobas t 511 is a coagulation analyzer with fully automated, random continuous-access, and software controlled system for clotting, chromogenic, and immunoturbidimetric analysis. It's intended for qualitative and quantitative in vitro coagulation determinations to aid in diagnosing coagulation abnormalities and monitoring anticoagulant therapy.

cobas t 511 coagulation analyzer
Safety Guide Version 1.1
Software version 1.0
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
2
Publication information
Edition notice This publication is for users of the cobas t 511
coagulation analyzer.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of printing. However, Roche Diagnostics
International Ltd. reserves the right to make any changes
necessary without notice as part of ongoing product
development.
Where to find information The User Assistance contains all information about the
product, including the following:
Routine operation
Maintenance
Safety
Troubleshooting information
A software reference
Configuration information
Background information
The Safety Guide contains important safety information.
You must read the Safety Guide before operating the
instrument.
The User Guide focuses on routine operation and
maintenance. The chapters are organized according to
the normal operation workflow.
Copyright © 2017, F. Hoffmann-La Roche Ltd. All rights reserved.
Contact addresses
Publication version Software version Revision date Change description
1.0 1.0 March 2017 Initial version
1.1 1.0 July 2017 Updated safety messages
y Revision history
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Made in Switzerland
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
3
T
able of contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Conventions used in this publication. . . . . . . . . . . 5
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
About operator qualification . . . . . . . . . . . . . . . . . . 8
About safe and proper use of the system . . . . . . . 8
Miscellaneous safety precautions overview . . . . . 10
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Biohazardous materials . . . . . . . . . . . . . . . . . . . . . . 13
Electromagnetic interference . . . . . . . . . . . . . . . . . 15
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Reagents and other working solutions . . . . . . . . . 16
Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Data security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Caution messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Reagents and other working solutions . . . . . . . . . 23
Fatigue due to long hours of operation . . . . . . . . . 24
Data security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Circuit breakers and fuses . . . . . . . . . . . . . . . . . . . 26
Electromagnetic compatibility. . . . . . . . . . . . . . . . . 26
Mechanical stress. . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Temperature. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Spillage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Safety labels on the system. . . . . . . . . . . . . . . . . . . . . . 29
List of safety labels on the instrument. . . . . . . . . . 29
Location of safety labels on the system. . . . . . . . . 30
Other labels on the system . . . . . . . . . . . . . . . . . . . 34
Safety information for disposal. . . . . . . . . . . . . . . . . . . 36
Disposal information . . . . . . . . . . . . . . . . . . . . . . . . 36
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
4 Preface
Preface
Use this publication together with the cobas t 511
coagulation analyzer User Guide or User Assistance.
Operation and maintenance actions are described in the
User Guide and User Assistance.
Intended use
The cobas t 511 coagulation analyzer is a fully
automated, random continuous-access, software
controlled system for clotting, chromogenic and immuno-
turbidimetric analysis intended for qualitative and
quantitative in vitro coagulation determinations using a
wide variety of coagulation tests, the results of which aid
in the diagnosis of coagulation abnormalities and in
monitoring anticoagulant therapy.
The cobas t 511 coagulation analyzer is a system that
uses human citrated plasma for the determination of
coagulation analytes.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
5
Conventions used in this publication
Product names Except where the context clearly indicates otherwise, the
following product names and abbreviations are used:
Acronyms The following acronyms are used:
Product name Abbreviation
Instrument Software for
cobas t 511 coagulation
analyzer
software
cobas t 511 coagulation
analyzer
analyzer
System Cleaner cobas t System Cleaner
y Product names
Acronym Definition
ANSI American National Standards
Institute
CFR Code of Federal Regulations
CISPR Comité International Spécial des
Perturbations Radioélectriques
FCC Federal Communications
Commission
IEC International Electrical
Commission
ISO International Organization for
Standardization
SOP Standard Operating Procedure
y Acronyms
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
6 Introduction
Introduction
!
General attention
To avoid serious or fatal injury, read this publication
thoroughly before running the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the instrument in a way that is not
described in this publication.
r Keep this publication in a safe place to ensure that it
is not damaged and remains available for use.
This publication must always be easily accessible.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
7
Safety classifications
The safety precautions and important user notes are
classified according to the ANSI Z535.6 standard.
Familiarize yourself with the following meanings and
icons:
!
Safety alert
r The safety alert symbol is used to alert you to potential
physical injury hazards. Obey all safety messages that
follow this symbol to avoid possible damage to the
system, injury, or death.
These symbols and signal words are used for specific
hazards:
WARNING
!
Warning...
r ...indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
!
Caution...
r ...indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury.
NOTICE
Notice...
r ...indicates a hazardous situation that, if not avoided,
may result in damage to the system.
Important information that is not safety relevant is
indicated with the following icon:
q Tip...
...indicates additional information on correct use or useful
tips.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
8 Safety precautions
Safety precautions
!
To avoid serious or fatal injury, read and
comply with the following safety precautions.
In this section
About operator qualification (8)
About safe and proper use of the system (8)
Miscellaneous safety precautions overview (10)
About operator qualification
Insufficient knowledge and skills As a user, ensure that you know the relevant safety
precaution guidelines and standards and the information
and procedures contained in these instructions.
r Do not carry out operation and maintenance unless
Roche Diagnostics has trained you to do so.
r Leave maintenance, installation, or service that is not
described to trained Roche Service representatives.
r Carefully follow the procedures specified in the
instructions for operation and maintenance.
r Observe good laboratory practices, especially when
you work with biohazardous material.
About safe and proper use of the system
Correct usage
Risk-ID Requirement of Regulatory Affairs
checklist
Using the instrument in a way not described by the
manufacturer, can impair the protection provided.
r Do not use the analyzer in a way that is not described
in the user instructions.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
9
Missing personal protective equipment Working without personal protective equipment means
danger to life or health.
r Wear appropriate personal protective equipment,
including, but not limited to, the following items:
o Eye protection with side shields
o Fluid-resistant laboratory coat
o Approved lab gloves
o Face shield if there is a chance of splashing or
splattering
r Follow good laboratory practices and regularly change
lab gloves to minimize the risk of infection and
contamination (especially after contact with waste or
sample material).
Regular cleaning To prevent inaccurate results and unsafe operation of the
system:
r Regularly clean and/or decontaminate the instrument
as required. Follow good laboratory practices for
cleaning and decontamination.
r Use only approved cleaning solutions for cleaning.
r Ensure that the laboratory is regularly cleaned and is
maintained in an orderly manner.
u See section Cleaning and decontamination in the user
documentation.
Errors in installation Only trained Roche Service representatives may install
the system.
r Leave installation that is not described to trained
Roche Service representatives.
Exchange or removal of parts Unauthorized exchange or removal of system parts can
damage the system or stop it from functioning correctly.
r Do not exchange or remove any part of the instrument
unless instructed to do so.
r Leave replacement of other instrument parts to
trained Roche Service representatives.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
10 Safety precautions
Unsuitable environmental conditions Operation outside of the specified ranges may lead to
incorrect results or malfunction of the system.
r Use the system indoors only, and avoid heat and
humidity outside of the specified range.
r Make sure that the system's ventilation openings
always remain unobstructed.
r To maintain the environmental conditions of the
system, perform maintenance in accordance with the
specified intervals.
r Keep the operating instructions undamaged and
available for use. Operating instructions must be
easily accessible for all users.
u See Environmental conditions in the user
documentation.
Non-approved spare parts Use of non-approved spare parts or devices may result in
malfunction of the system and may render the warranty
null and void.
r Use only spare parts and devices approved by Roche
Diagnostics.
Non-specified third-party software Installation of third-party software is not approved by
Roche Diagnostics and may result in malfunction.
r Do not install third-party software.
Non-specified consumables Use of non-specified consumables can lead to incorrect
results.
r Do not use consumables that are not intended for use
with the analyzer.
u For a list of supported materials, see the user
documentation.
Miscellaneous safety precautions overview
Electromagnetic compatibility The cobas t 511 coagulation analyzer complies with the
emission and immunity requirements described in this
part of the IEC 61326 series.
The analyzer complies with the emission requirements
described in this part of the FCC CFR 47, Part 15 Class A.
u Electromagnetic compatibility (26)
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
11
System not used for an extended period r Follow the procedure for shutting down the analyzer
for an extended period.
r Remove and refrigerate any remaining reagents and
QC materials.
r For further information, contact your Roche Service
representative.
u See Shutting down the analyzer for an extended period
in the user documentation.
Damage in transit r Do not attempt to relocate or transport the system.
r Leave relocation and transportation to Roche Service
representatives.
u Related topics
Disposal information (36)
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
12 Warning messages
Warning messages
!
List of warning messages
Failure to observe warning messages may result in death
or serious injury.
r Before operating the system, read the warning
messages carefully.
In this section
Electrical safety (12)
Biohazardous materials (13)
Electromagnetic interference (15)
Mechanical safety (15)
Reagents and other working solutions (16)
Waste (19)
Data security (20)
Electrical safety
Power interruption
Risk-ID DRQ-38183 A power failure or momentary drop in voltage may
damage the system or lead to data loss.
r Operate only with an uninterruptible power supply
(UPS). Operation outside of the specified ranges may
lead to incorrect results.
r Ensure periodic maintenance of the UPS.
r Regularly backup results.
Electric shock
Risk-ID DRQ-38485 Risk of personal injury due to electric shock if covers are
left open or when touching power supply components
during operation.
r Do not remove any cover of the system except those
covers specified in the instructions.
r Do not touch power supply components during
operation.
r Do not attempt to work on any electronic equipment.
r Only Roche Service representatives are permitted to
install, service, and repair the system.
u Related topics
List of safety labels on the instrument (29)
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
13
Biohazardous materials
Infectious samples Contact with samples containing material of human origin
may result in infection. All materials and mechanical
components associated with samples containing material
of human origin are potentially biohazardous.
r Follow good laboratory practices, especially when
working with biohazardous material.
r Keep all covers closed while the system is operating.
r Wear appropriate personal protective equipment.
r If any biohazardous material is spilled, wipe it up
immediately and apply a disinfectant.
r If sample or waste comes into contact with your skin,
wash the affected area immediately with soap and
water and apply a disinfectant.
Consult a physician.
User infection and injury Contact with analyzer mechanisms or with the chassis or
cover may result in personal injury and infection.
r Whenever possible, keep the main cover closed.
r Be careful not to hit your head when the cover is
open.
r Pay attention to the main cover during automatic
movement of analyzer parts, check for obstructions
and keep out of reach.
r Do not touch any parts of the system other than those
specified.
r Never reach into the instrument while parts are
moving.
r Carefully observe all instructions given in this
publication.
Contact with probes
Risk-ID DRQ-36606 Contact with probes may result in injury and infection.
r Avoid touching the end of the reagent probes and
sample probes.
r Avoid contacting the reagent probes and sample
probes during operation.
r When cleaning near the probes or replacing probes,
take care not to puncture yourself.
r Wear appropriate personal protective equipment.
Take extra care when working with lab gloves. They
can easily be pierced or cut, leading to infection.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
14 Warning messages
Contaminated sample tubes, racks, and
surfaces
Risk-ID DRQ-36608 | DRQ-36614 Contact with contaminated sample tubes, sample racks,
loading positions, and internal surfaces may result in
infection.
r Use caution when replacing sample tubes.
r Use caution when handling sample racks, loading
positions, and internal surfaces.
r Wear appropriate personal protective equipment.
Splashed sample
Risk-ID DRQ-36609 Contact with splashed sample may result in infection.
r Use caution when cleaning the analyzer surfaces in
the sample pipetting area.
r Wear appropriate personal protective equipment.
Contaminated incubator and analyzer area
Risk-ID DRQ-36610 Contact with contaminated incubator and analyzer area
may result in infection.
r Use caution when cleaning the incubator and analyzer
area.
r Wear appropriate personal protective equipment.
Rinse mechanism malfunction
Risk-ID DRQ-36611 Malfunction of the rinse mechanism may result in direct
contact with sample and infection when replacing a
probe.
r Use caution when replacing probes.
r Wear appropriate personal protective equipment.
Smoke due to electrical malfunction Electrical malfunction can result in the emission of
hazardous smoke. Inhaling smoke emitting from the
instrument can lead to personal injury.
r If you see smoke coming from the instrument:
o Avoid inhaling
o Disconnect from power supply
o Contact Roche Service immediately
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
15
Troubleshooting procedures Corrective measures in troubleshooting procedures can
result in exposure to biohazardous materials.
r Always follow the troubleshooting procedures given in
software wizards and/or in the user documentation.
r Wear appropriate personal protective equipment
when implementing corrective measures.
Electromagnetic interference
Malfunction of system and incorrect results
due to interfering electromagnetic fields
Risk-ID DRQ-38185 This system is designed and tested to CISPR 11 Class A
standard. In a domestic environment it may cause radio
interference, in which case, you may need to take
measures to mitigate the interference.
r The electromagnetic environment should be evaluated
prior to operation of the device.
r Do not operate this system in close proximity to
sources of strong electromagnetic fields (for example
unshielded intentional RF sources), as they may
interfere with proper operations.
Mechanical safety
Main cover
Risk-ID DRQ-38484 Risk of personal injury when closing the main cover. Your
fingers can be pinched between the main cover and the
housing.
r Use caution when closing the main cover.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
16 Warning messages
Moving parts
Risk-ID DRQ-31531 | DRQ31532 Contact with moving parts of the incubator, analyzer,
transfer arms, sample probes, or reagent probes may
result in personal injury.
r Keep all covers closed and in place while the system is
operating.
r Do not touch any parts of the system except those
parts specified. Keep away from moving parts during
operation.
r During operation and maintenance, carefully follow
the instructions.
r Observe all warning labels on the analyzer.
Reagents and other working solutions
Incorrect results due to calibration with mixed
calibrator material lots
Risk-ID DRQ60003 | DRQ60004 |
DRQ60005
Using different calibrator lots in a single calibration can
lead to incorrect results.
r Do not mix different calibrator lots for a single
calibration, also for repeat point measurement.
r Visually control the calibration result and graph prior
to release. Check the list of used materials.
r Use caution when preparing a calibrator material for
measurement and pay particular attention when
attaching e-barcodes to secondary tubes for
calibrators.
Foam, clots, films, or bubbles
Risk-ID DRQ-16937 | DRQ36619 Incorrect results may occur due to foam, fibrin clots, films,
or bubbles in reagents or samples.
r Ensure good sample preparation and reagent
handling techniques to avoid the formation of foam,
clots, and air bubbles in all reagents, samples, and QC
materials.
r Check that there is no foam in calibrator tubes, QC
tubes, or sample tubes.
r Make sure that the samples contain no insoluble
contaminants, such as fibrin or dust.
r Ensure that no air bubbles are aspirated into the
sample pipetting mechanism.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
17
Incorrect results due to insufficient sample
volumes
Risk-ID DRQ-60105 | DRQ-60106 | Risk-
60136
Pipetting samples with insufficient volumes can lead to
coagulum being aspirated and incorrect results.
r Ensure that there is enough plasma in the sample
tube.
r When performing high numbers of measurements
from the same sample material (e.g., for precision
studies), use a secondary cup.
Evaporation of samples or reagents
Risk-ID DRQ-36616 Evaporation of samples or reagents may lead to incorrect
or invalid results.
r Sample material may evaporate if left open. Do not
leave samples open for any length of time.
r Do not use improperly stored reagents. Ensure that
reagents are stored according to the Instructions for
Use.
r The system does not allow the use of expired
reagents, calibrators, or QCs.
System water leakage in sample pipetting
mechanism
Risk-ID DRQ-36615 System water leakage in the sample pipetting mechanism
can lead to incorrect results.
r Regularly perform QCs and visual checks for system
water leakage.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
18 Warning messages
Incorrect results due to incorrect handling of
reagents, QCs, calibrators, or consumables
Risk-ID DRQ-36627 | DRQ-48608 Incorrect handling of reagents, QCs, calibrators, or
consumables may lead to incorrect results.
r Opening and closing a sample tube, QC tube, or
calibrator tube can influence the inside pressure. Use
caution when opening and closing tubes.
r Do not use reagents, QCs, calibrators, or other
consumables that were exposed to heat or to light for
an extended time.
r The system does not support the use of expired
reagents, calibrators, or QCs.
r Adhere to the storage conditions defined in the
Instructions for Use.
r Do not use reagents or consumables that have been
dropped on the floor or compromised in any other
way.
r Do not manipulate supplies in any way not specified in
the user documentation or Instructions for Use.
Incorrect results due to using expired
reagents, QC materials and calibrators
Risk-ID DRQ-36626 | DRQ-36627 | DRQ-
25755
Using expired reagents, QC materials and calibrators can
lead to incorrect results.
r Using expired reagents, QC materials and calibrators
is not supported by the system.
r Do not change QC expiry dates. Changing the expiry
date can lead to incorrect QC evaluation and is fully
the responsibility of the operator.
Exchanging reagent cassettes between
systems
Risk-ID DRQ-16939 | DRQ-36631 Exchanging reagent cassettes between systems can lead
to incorrect results and is not supported.
r Do not exchange reagent cassettes between systems.
r If using multiple analyzers in the laboratory, mark the
labels of any unloaded reagent cassettes to avoid
reloading the reagent cassette onto another
instrument.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
19
Incorrect results due to hemolysis, icterus
and lipemia
Risk-ID DRQ13513 Result quality can be affected by the presence of
hemoglobin, bilirubin and lipids in the sample.
r Ensure good sample preparation and handling
techniques.
Reagent probe/flow path
Risk-ID DRQ36620 Incorrect cleaning or rinsing of the probe/flow path of the
reagent pipetting mechanism can lead to incorrect
results.
r Perform QCs regularly.
Sample carryover
Risk-ID DRQ36617 Sample or rinse water carryover in the pipetting
mechanism can lead to incorrect results.
r Regularly check instrument performance to prevent
incorrect results due to carryover.
Waste
Infectious waste
Risk-ID DRQ-36612 Contact with liquid waste or solid waste may result in
infection. All materials and mechanical components
associated with the waste systems are potentially
biohazardous.
r Wear appropriate personal protective equipment.
Take extra care when working with lab gloves. They
can easily be pierced or cut, leading to infection.
r If any biohazardous material is spilled, wipe it up
immediately and apply a disinfectant.
r If waste comes into contact with your skin, wash the
affected area immediately with soap and water and
apply a disinfectant.
Consult a physician.
Roche Diagnostics
cobas t 511 · Software version 1.0 · Safety Guide · Version 1.1
20 Warning messages
Liquid waste tubing clogging
Risk-ID DRQ36613 Risk of infection due to contact with liquid waste when
removing clogging in the liquid waste tubing.
r Use caution when unclogging the liquid waste tubing.
r Wear appropriate personal protective equipment.
Solid waste bags Attempting to empty and reuse solid waste bags can
result in spillage. Spilled solid waste can lead to
contamination, risk of slipping, and risk of personal injury
during instrument cleanup.
r Do not attempt to empty and reuse solid waste bags.
r When disposing of solid waste, dispose of the entire
bag.
Environmental harm The system generates liquid and solid waste which is
potentially biohazardous. Improper disposal may
contaminate the environment.
r Treat solid waste as infectious waste.
r Dispose of waste in accordance with the local
regulations.
u Related topics
List of safety labels on the instrument (29)
Disposal information (36)
Data security
Data loss or unauthorized access to system
data
Risk-ID DRQ-39467 Unsecured system backup and archive files can result in
data loss or unauthorized access to system data.
r Make sure that system backup and archive files
exported by the system are physically secure and are
protected from any unauthorized access.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Roche cobas t 511 User manual

Type
User manual

Roche cobas t 511 is a coagulation analyzer with fully automated, random continuous-access, and software controlled system for clotting, chromogenic, and immunoturbidimetric analysis. It's intended for qualitative and quantitative in vitro coagulation determinations to aid in diagnosing coagulation abnormalities and monitoring anticoagulant therapy.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI