  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo 30 425 550 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InTap Portable oDO Analyzer
30 461 735 B
cs Návodkpoužití 3
da Brugervejledning 17
de Benutzerhandbuch 31
en User Manual 45
es Manualdeusuario 59
 Käyttöohje 73
fr Guidedel‘utilisateur 87
hu Felhasználóiútmutató 101
it Manualeperl‘utente 115
ja ユーザマニ 129
ko
사용자 매뉴얼 143
nl Handleiding 157
pl Podręcznikużytkownika 171
pt Manualdousuário 185
ru Руководство пользователя 199
sv Användarmanual 213
th
คู่มือผู้ใช ้ 227
zh 简明用户手册 245
User Manual
InTap Portable oDO Analyzer
2 InTap Portable oDO Analyzer • 30 461 735 B
©Mettler-ToledoGmbH01/2021
Subjecttotechnicalchanges.PrintedinSwitzerland.B
10 InTap Portable oDO Analyzer • 30 461 735 B
5 Struktura menu
cs
24 InTap Portable oDO Analyzer • 30 461 735 B
5 Menustruktur
da
38 InTap Portable oDO Analyzer • 30 461 735 B
5 Menüstruktur
de
45InTap Portable oDO Analyzer • 30 461 735 B
User Manual
InTap Portable oDO Analyzer
Content
1 Safety and Intended Use 46
2 TechnicalDataandFurtherInformation 47
3 DesignandConnections 48
4 DisplayandOperatingElements 50
5 Menu Structure 52
6 Commissioning 53
7 Calibration 54
8 MaintenanceandRepair 56
9 Disposal 57
10 Certicatesandapprovals 58
en
InProisaregisteredtrademarkoftheMETTLERTOLEDOGroup
inSwitzerlandandafurthertwelvecountries.
ISMisaregisteredtrademarkoftheMETTLERTOLEDOGroup
inSwitzerland,USA,theEuropeanUnionandafurtherve
countries.
InTap,iSenseandOptoCaparetrademarksoftheMETTLER
TOLEDOGroup.
46 InTap Portable oDO Analyzer • 30 461 735 B
1 Safety and Intended Use
TheInTap™portableoDOanalyzermustonlybeconnected,commissioned,andmaintained
byqualifiedspecialistsinfullcompliancewiththeinstructionsinthisUserManual,theappli-
cablenormsandlegalregulations.
ThespecialistmusthavereadandunderstoodthisUserManualandmustfollowtheinstruc-
tionsitcontains.
TheReferenceManualprovidesfurtherinformationonthedeviceandcanbedownloadedvia
Internet“www.mt.com/InTap“.
TheInTapportableoDOanalyzershouldbeoperatedonlybypersonnelfamiliarwiththedevice
andwhoarequalifiedforsuchwork.
Intended Use
The InTap portable oDO analyzer is a portable measurement device for dissolved oxygen
measurementandforcalibrationanddataloggingofopticaldissolvedoxygenISM
®
sensors
(InPro
®
6970i)installedintheprocess.TheInTapandthesensorscommunicateviaBlue-
tooth.Forthis,theseparateBluetoothinterfaceT100(ordernumber30432819)hastobe
mountedontoeachsensor.
TheInTapisdesignedforuseinthebeverageindustries,innon-hazardousareas.TheInTap
isintendedtobeusedindoor.
ThefollowingInTapvariantsareavailable:
InTapportableoDOanalyzer(metricconnectors):ordernumber30425550
InTapportableoDOanalyzer(imperialinch/USconnectors):ordernumber30457912
en
47InTap Portable oDO Analyzer • 30 461 735 B
2 Technical Data and Further Information
Themostimportanttechnicaldatasuchassupplyvoltagearegivenonthenameplate.The
nameplateisplacedonthebottomofthedevice.Payparticularattentiontothetechnicaldata
onthenameplate.
Measurementparameter OpticalDissolvedOxygenSaturation
Measuringrange 0to2000ppb
Accuracy
1)
(sensor) ≤±(1%+2ppb)
Responsetimet
98
at+25°C(+77°F)
(airtoN
2
)
<20s
Powersupply
1)
External:12VDC,2.5A
Internal:Li-ionbattery45.4Wh
Batterylifetime Up to 24 hours
DatastorageUSB2.0 8GB
Operatingtemperature –5to+45°C(+23to+113°F)withmedia
Upto+80°C(+176°F)forcleaningwith2%NaOH
Withpoweradapteronlyupto+40°C(+104°F)
Storagetemperature –20to50°C(–4to+122°F)
Operatingpressure 0to6bar(0to87psi)
Designpressure 10bar(145psi)
Protectionrating IP 67
Hoseconnection Variantwithmetricconnectors:Swagelok6mm
Variantwithimperialconnectors:Swagelok1/4"
Weight 3.5kg(7.7lbs)
Dimensions(height3width3depth)
280mm3170mm3150mm(11.0"36.7"35.9")
Altitude Max.2000m
Relativehumidity 0to95%non-condensing
Pollution degree 2
1) Measurementspecificationbasedonsensorspecification
en
48 InTap Portable oDO Analyzer • 30 461 735 B
3 Design and Connections
1
4
5
6
10
2
7 89
3
1 Monochrome4"touchdisplay
2 On/Offbutton
3 Housingwithmeasurementcell,O
2
InTapsensor,valve,battery,HWboard
4 Connectionforhose“Mediumin”
5 Connectionforhose“Mediumout”
6 Rotaryknobforadjustingmediumflowrate
7 USBstickfordatalogging
8 USBtypeBconnectionforsoftwareupdates
9 USBtypeAconnection
10 ConnectionforAC/DCadapterforloadingbattery
en
49InTap Portable oDO Analyzer • 30 461 735 B
Scope of delivery
Powersupply30W:AC/DCadapterwithACadaptersetEU,US,AU,UK
USBstick,plugged-intheUSBconnection
Supply voltage (Power)
Youhavethefollowingpossibilitiestopowerthedevice:
eitherviatheinternalbattery
orviathesuppliedAC/DCadapterconnectedto100to240VAC/50to60Hz.
Hoses “Medium In” and “Medium Out”
Youcanconnectthefollowinghosesto“MediumIn”and“MediumOut”:
InTap portable oDO analyzer Hoses
Description Order number Description Order number
Withmetricconnectors 30425550 6mm 30422575
Withimperialinch/
US connectors
30457912 1/4" 30432818
Adjusting medium flow rate
Theflowrateisadjustedwiththerotaryknob.Theflowratedependsonthemediumpressure
Theflowrateincreaseswithincreasingmediumpressure.
1. Turnrotaryknobclockwisetotheendstop.
2. Turnrotaryknobanti-clockwisetoadjusttheflowrate.Therotaryknobhasaclick-stop
positionevery15°.E.g.foraflowrateofapproximately200ml/minat2baryouhave
toturntherotaryknobtothe30°position.Thatmeansyouhear2clicks.
en
50 InTap Portable oDO Analyzer • 30 461 735 B
4 Display and Operating Elements
\ CAL
Calibrate InTap Sensor
Calibrate In-Line Sensor
InTap
200
A
ppbO2
25.0
°C
DLI
ACT
0 d
0.0 d
\ CAL \ Calibrate InTap Sensor
Chan
Unit
Methods
Options
InTap
% air
2-Point
Options
Verify Cal
InTap
200
A
ppbO2
25.0
°C
DLI
ACT
0 d
0.0 d
1 2
34
6 7
55
1 Homescreen
2 Measurementscreen
3 Menuscreen,heree.g.“Calibrationmenu”
4 Parametrizationscreen,heree.g.“CalibrationInTap”
5 NavigationtoHomescreen
6 Navigationtonext-highermenulevel,hereMeasurementscreen
7 Navigationtonext-highermenulevel,hereCalibrationmenuscreen
en
51InTap Portable oDO Analyzer • 30 461 735 B
Display or operating
element
Description Display or operating
element
Description
USBstickisconnected. EnterTrenddisplayscreen.
Displayofthecurrentbattery
status
EnterISMmenu.
TheInTapispoweredviathe
AC/DCadapter.
c
EnterCalibrationmenu.
Failure:Nobatteryorbattery
isempty.
C
EnterConfigurationmenu.
Failure:Batteryislow
(<5%).TheInTapisnot
poweredviatheAC/DC
adapter.
EnterMeasuringPoint(MP)
menu.
Failure:USBstickisnot
connectedordamaged.
H
ReturntoMenuscreen.
Failure:Leakage,high
temperature,batteryfaulty
etc.RefertotheReference
Manual.
Enternext-lowermenulevel.
EnterMessagesscreen.
p
Returntonext-highermenu
level.
F
Entermenuscreen.
< >
Changebetweenpages
withinonemenulevel.
s
Enterstartscreen.
en
52 InTap Portable oDO Analyzer • 30 461 735 B
5 Menu Structure
en
53InTap Portable oDO Analyzer • 30 461 735 B
6 Commissioning
TheInTapissuppliedreadyforuse.Afterswitchingon,youcanimmediatelyusetheInTapas
ameasuringdeviceorasameasuringdevicewithdataloggingfunction.
Atthefactorythedataloggingissettoautomaticmode.Usetheparameter“LoggingMode”to
changetomanualmode(CAL\CONFIG\MeasurementMode).
Notethatwithfactorysettings,thedisplayisswitchedoffaftersomeminutes,butthemeasure-
mentwillcontinue.Usethemenu“DisplaySetup”tochangethesettings(CAL\CONFIG\Dis-
playSetup).
ForcommunicationwithanIn-Linesensor(processcalibration)aseparateBluetoothinterface
T100(ordernumber30432819)hastobemountedontoeachsensorandthesensorhas
tobepairedwiththeInTap.
Pairing InTap with In-Line Sensors
Prerequisite
TheIn-LinesensoriswithintheBluetoothrangeoftheInTap.
MP – Select
New MP
New BT device recognized 00
MP_01
MP_02
MP_03
Edit
Edit
Edit
MP_01 is selected.
1
2
3
4
1 A new Bluetooth device is within the Bluetooth
rangeandwasrecognized.Thesensorneedstobe
parametrized.
2 TheIn-Linesensor“MP_01”isselected.The
sensoriswithintheBluetoothrangeandalready
parameterized.
3 TheIn-Linesensor“MP_02”isnotselected.The
sensoriswithintheBluetoothrangeandalready
parameterized.
4 TheIn-Linesensor“MP_03”isnotselected.The
sensorisalreadyparameterizedbutnotwithinthe
Bluetoothrange.
en
54 InTap Portable oDO Analyzer • 30 461 735 B
1. Enter“MeasuringPoint(MP)”menu.SeeChapter 4 on Page 50.
Thescreen“MPSelect”withtheitem“NewBTdevicerecognized00”isshown.
2. Tap“+”fortheitem“NewBTdevicerecognized00”.
3. Configurethenewmeasuringpoint(MP).
4. Tap“OK”
Thescreen“MPSelect”isshown.
5. Activatetheradiobutton.
6. Returntothenext-highermenulevel.
7. Answerthequestion“Savechanges?”with“Yes”.
TheMeasurementScreenfortheselectedmeasuringpointisshown.Aftersomeseconds
themeasurementresultsareshown.Tap“Save”tosavetheresultsontheUSBstick.Tap
“Repeat”togetnewvaluesfromthesensor.Tap“Cancel”tocanceltheprocedure.
7 Calibration
Youcanalso calibratethedismounted InTapsensororthe dismountedIn-LineSensorvia
iSense™software.SeeOperatingInstructionsoftheiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Calibration Menu c
Aftereverysuccessfulcalibrationdifferentoptionsareavailable.If”Adjust”or”Calibrate”is
chosen,themessage”Calibrationsavedsuccessfully!”isdisplayed.Press”Done”.
Option ISM sensors
Adjust
Calibrationvaluesarestoredinthesensorandusedforthemeasurement.
Additionally,thecalibrationvaluesarestoredinthecalibrationhistory.
Calibrate Calibrationvaluesarestoredinthecalibrationhistoryfordocumentation,butnot
beusedforthemeasurement.Thecalibrationvaluesfromthelastvalid
adjustmentarefurtherusedforthemeasurement.
Cancel Calibrationvaluesarediscarded.
en
55InTap Portable oDO Analyzer • 30 461 735 B
Calibrating an In-Line Sensor
Prerequisite
TheIn-LinesensorispairedwiththeInTapandiswithintheBluetoothrangeoftheInTap.
1. Enter“CalibrateIn-LineSensor”menu.Menupath:CAL>CalibrateIn-LineSensor
Thescreen“MPSelect”isshown.
2. SelecttheMP(In-Linesensor)youwanttocalibrate.Thereforeactivatethecorrespond-
ingradiobutton.
3. Returntothenext-highermenulevel.
Thescreen“CalibrateIn-LineSensor”isshown.
4. Settheoptions“CalPressure”,“Rel.Humidity”and“Stability”.
5. Tap“Cal”.
6. Confirmthemessage“Thereisa“autologging”procedurerunning”with“Yes”.
7. Answerthequestion“FirstcalibrationafterOptoCap™change”.
ThecurrentvaluesfortheInTapsensorandtheIn-Linesensorareshown.
8. Tap“Next”.
ThevaluesforP100andP0areshown.
9. Tap“Adjust”or“Calibrate”.
10.Tap“Done”tofinishcalibrationoftheIn-Linesensor.
Calibrating the InTap Sensor
h
Note:Useasairgasdry,cleanandoil-freecompressedair.UseaszerogasN
2
or CO
2
witha
purityofatleast99.9%.Forbestresultsuseazerogaswithapurityof99.995%.
1. Connectthecompressedairtotheconnection“MediumIn”.
2. PurgetheInTapfor5minuteswithcompressedairtoremovetheexcesshumidityoutof
themeasurementcell.Ensureasteadymediumflowwithoutpressurebuild-upinthe
measurementcell.Forthis,performtwoturnswiththerotaryknob.SeeChapter 3 on
Page48.
3.
Turnoffcompressedair.
4. Enter“CalibrateInTapSensor”menu.Menupath:CAL>CalibrateInTapSensor
5. Settheunits.
en
56 InTap Portable oDO Analyzer • 30 461 735 B
6. Settheoptions“CalPressure”,“Rel.Humidity”and“Stability”.
7. Tap“Cal”.
8. Confirmthemessage“Thereisa“autologging”procedurerunning”with“Yes”.
9. Answerthequestion“FirstcalibrationafterOptoCapchange”correctly.
Themessage“ConnectthegastoInTapinlet.”isshown.
10.Turnonthecompressedair.
11.Ensureasteadymediumflowwithoutpressurebuild-upinthemeasurementcell.
12.Tap“Next”.
Thecurrentvalueisshown.
Ifthecriteriaaremet,themessage“Pleasechangegas.”isshown.
13.Connectthesecondreferencegas(zerogas)totheconnection“MediumIn”.
14.Tap“Next”.
Thecurrentvalueisshown.
Ifthestabilitycriteriaaremet,thevaluesforP100andP0areshown.
15.Tap“Adjust”or“Calibrate”.
16.Tap“Done”tofinishcalibrationoftheInTapsensor.
8 Maintenance and Repair
Cleanthesurfaceswithasoftdampclothanddrythesurfaceswithaclothcarefully.
Replacing the OptoCap of the InTap Sensor (Sensors Sensing Element)
Typically,theOptoCaphastobereplacedafter6to12months.Thisintervaldependsonthe
aggressiveness(CIP/SIP)oftheapplicationconditions.
Fordismantlingthesensorrefertothe“ExchangingtheInTapSensor“section.Forreplacingthe
OptoCaprefertotheUserManual“InPro6000OpticalO
2
Sensors“.
Exchanging the InTap Sensor
1. Ifnecessary,exchangethesensorasshowninthefollowingfigure.
2. Tightencoverscrewswithamaximumtorqueof5Nm.
3. Calibratesensor.SeeChapter 7 on Page 54.
en
57InTap Portable oDO Analyzer • 30 461 735 B
Replacing the Battery
ThebatterymayonlybereplacedbyMETTLERTOLEDO.Looseningthescrewsoftheinner
coverwillleadtotheexpirationofanywarrantyclaims.
9 Disposal
Environmental Protection
Wasteelectricalproductsshouldnotbedisposedofwithhouseholdwaste.Pleaserecycle
wherefacilitiesexist.CheckwithyourLocalAuthorityorretailerforrecyclingadvice.
en
58 InTap Portable oDO Analyzer • 30 461 735 B
10 Certicates and approvals
Europeandirectives TheEUDeclarationofconformityispartofthedelivery.
Safety standards CAN/CSA–C22.2No.61010-1-12
ANSI/ULStd.No.61010-1(3rdEdition)
Overvoltage Category II
FCC(USA) CFR47FCCPart15
ANSIC63.4-2014
IC(Canada) ICES-003,Issue6
ANSIC63.4-2014
en
66 InTap Portable oDO Analyzer • 30 461 735 B