  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo portable DO analyzers InTap4000e and InTap4004e Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Instruction Manual
Portable Oxygen Analyzer InTap4000e
Portable Oxygen Analyzer InTap4004e
Order number: 52 201 080
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
2 InTap 4000e/InTap 4004e
Mettler-Toledo GmbH
Process Analytics
P.O. Box
8902 Urdorf
Switzerland
Tel.: +41 44 729 62 11
Fax: +41 44 729 66 36
Warranty
METTLER TOLEDO warrants this product to be free from significant deviations in material and workmanship for a period
of one year from the date of purchase. If repair is necessary and not the result of abuse or misuse within the warranty
period, please return by freight pre-paid and amendment will be made without any charge. METTLER TOLEDO's Customer
Service Dept. will determine if the product problem is due to deviations or customer abuse. Out-of-warranty products will
be repaired on an exchange basis at cost.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 3
Instruction Manual
Portable Oxygen Analyzer InTap4000e
Portable Oxygen Analyzer InTap4004e
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
4 InTap 4000e/InTap 4004e
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.1 At the very beginning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.2 Important notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2 Overview of the InTap 4000 e / InTap 4004 e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.1 Instrument guide InTap 4000 e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.2 Instrument guide InTap 4004 e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
2.3 Features and functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3 Unpacking and inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
4 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.1 InTap 4000 e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.2 InTap 4004 e . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5 The Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
5.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
5.2 Preparation for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
5.3 Preconditioning of the sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
5.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
5.5 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
5.6 Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
5.7 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
5.8 Exchange of electrolyte and membrane body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Contents
Page
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 5
6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
6.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
6.2 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
6.3 Monitoring and test functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
6.4 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
6.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
6.6 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
6.7 Data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
6.8 Clock mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
6.9 Serial Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
6.10 Printing of measured values and protocols/transfer to a PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
7 Diagnostics, maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
7.1 Sensoface
®
, Sensocheck
®
functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
7.2 Battery replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
7.3 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
7.4 Maintenance and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
8 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
9 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
10 EU Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
Page
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
6 InTap 4000e/InTap 4004e
1 Introduction
1.1 At the very beginning
Thank you for your decision to purchase our O
2
Analyzer InTap4000e/InTap4004e. This portable instrument is
intended for the measurement of dissolved oxygen (DO) in liquid media in food and beverage processing.
Please read through this manual before operation of the InTap4000e/InTap4004e and carefully follow the in-
structions given. This will ensure that you obtain correct measurement results and gain overall benefit from the
complete range of features of the instrument.
If you have any questions concerning the use of this O
2
analyzer, please do not hesitate to contact your METTLER
TOLEDO Authorized Distributor, who will be pleased to be of assistance.
1.2 Important notes
Proper use
The portable oxygen analyzer InTap4000e/InTap4004e is intended solely for the measurement of dissolved
oxygen in liquid media in the field of food and beverage processing. The InTap4000e/InTap4004e may
only be operated in conjunction with original accessories from METTLER TOLEDO.
All other uses are considered outside the intention of the manufacturer and can lead to erroneous measurements
and/or damage to the instrument.
Proper use is understood to include compliance with the requirements laid down in this instruction manual,
in particular, compliance with the safety instructions.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 7
General safety instructions
The InTap4000e/InTap4004e may be operated and serviced only by personnel familiar with the unit, and who
have read and understood this instruction manual.
The InTap4000e/InTap4004e may only be fitted with alkaline AAA 1.5V batteries (LR03). The use of
rechargeable or other types of batteries with a different specification is not permissible and can lead to in-
strument damage.
The InTap4000e/InTap4004e may not be operated in a hazardous area.
There may be no interference with the unit outside the scope of the work described in this manual relative to
service/maintenance of the InTap4000e/InTap4004e and its accessories. The InTap4000e/InTap4004e may
only be repaired by an authorized METTLER TOLEDO representative.
Without the written consent of METTLER TOLEDO Process Analytics no changes may be made to the
InTap4000e/InTap4004e.
Use only original accessories/spare parts for the InTap4000e/InTap4004e as recommended/supplied by
METTLER TOLEDO. The use of components of other origin can lead to erroneous measurements or to instrument
damage.
Legend for the symbols used in this manual
This symbol denotes a safety or danger warning which, if disregarded, can lead to endangerment of the user of
the instrument, damage to the analyzer generally or to erroneous measurements.
This symbol draws attention to additional information and important advice for correct and economic use of
the instrument.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
8 InTap 4000e/InTap 4004e
2 Overview of the InTap 4000e/InTap 4004e
2.1 Instrument guide InTap4000e
1
6
2 3 4 5
7891011
13
14
15
12
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 9
Legend
1 Casing
2 Transmitter
3 Key pad (key functions see «Section 6.2»)
4 Display (description see «Section 6.1»)
5 Sensor VP connector
6 Oxygen sensor
7 Connections not used (not visible in figure)
8 Serial interface connection (not visible in figure)
9 Black top cover
10 Battery case
11 Measuring chamber
12 Feed tube (1.5 m)
13 Three-way valve
14 Discharge tube (0.3 m) with integrated
reduction nipple
15 Bypass tube (0.3 m)
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
10 InTap 4000e/InTap 4004e
2.2 Instrument guide InTap4004e
1
6
2 3 4 5
7891011
12
15
16
13
13
14
12
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 11
Legend
1 Casing
2 Transmitter
3 Key pad (key functions see «Section 6.2»)
4 Display (description see «Section 6.1»)
5 Sensor VP connector
6 Oxygen sensor
7 Connections not used (not visible in figure)
8 Serial interface connection (not visible in figure)
9 Black top cover
10 Battery case
11 Fixation set
12 Measuring chamber
13 Fluidics enclosure
14 Sampling valve
15 Discharge tube
16 Zwickel adapter
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
12 InTap 4000e/InTap 4004e
2.3 Features and functions
The InTap4000e/InTap4004e include the following features and functions:
Ergonomically optimized enclosure with hose-proof protection IP66
Amperometric measuring principle
Large, well-arranged display
Choice of reading (ppm, ppb, mg/l or %)
Simultaneous readings of oxygen value and temperature
Automatic shut-off
Temperature reading switchable from °C to °F
Programmable clock and calendar format
Data logger for the automatic recording of measurement values
Bidirectional serial RS 232 interface with universal interface cable for printer and PC
Automatic calibration in air
Low battery indicator
Sensor status indicator «Sensoface
®
» and «Sensocheck
®
»
Unit settings and calibration function protected with access code
Enhanced Paraly SW109
Zwickel adapter unit (only InTap4004e)
3 Unpacking and inspection
Upon receipt, inspect the carton for damage; if any is found, notify carrier immediately. Ensure that all items
noted on packing list are present and in good condition. If there is any problem contact your METTLER TOLEDO
Distributor.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 13
Scope of delivery
InTap4000e
Item
1 x Carrying case
1 x Transmitter (measuring instrument)
1 x Oxygen sensor fitted in measuring
chamber
1 x Casing blue (with top cover, black)
1 x Carrying strap
1 x CD-ROM containing transfer software
«Paraly SW109»
1 x RS 232 interface cable with adapter
1 x Cleaning and conditioning set
consisting of:
20 x Cleaning and conditioning tablets
1 x Plastic syringe with tube connection
1 x Plastic measuring beaker
1 x Membrane kit consisting of:
4 x Membrane bodies (InTap 4000e/4004e)
1 x O-ring set
1 x Bottle containing 25 ml electrolyte
1 x Fitting instructions
1 x Liquid handling set consisting of:
1 x supply tube
1 x bypass tube
1 x discharge tube with reduction nipple
1 x 3-way valve (already mounted)
2 x screws for valve (already mounted)
1 x Instruction manual (en)
InTap4004e
Item
1 x Transmitter (measuring instrument)
1 x Oxygen sensor fitted in fluidics enclosure
1 x Casing blue (with top cover, black)
1 x Fluidics enclosure
1 x Carrying strap
1 x CD-ROM containing transfer software
«Paraly SW109»
1 x RS 232 interface cable with adapter
1 x Cleaning and conditioning set
consisting of:
20 x Cleaning and conditioning tablets
1 x Plastic syringe with tube connection
1 x Plastic measuring beaker
1 x Membrane kit consisting of:
4 x Membrane bodies (InTap 4000e/4004e)
1 x O-ring set
1 x Bottle containing 25 ml electrolyte
1 x Fitting instructions
1 x Instruction manual (en)
1 x Tubing adapter kit for conditioning
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
14 InTap 4000e/InTap 4004e
4 Start-up
4.1 InTap4000e
Important! Before using the sensors for the first time, the electrolyte should be replaced (see «Section 5.8»).
Due to possible adverse conditions during transport and storage (e.g. airfreight, pressure and temperature
variations), the quality of the electrolyte may become impaired. Poor electrolyte quality can lead to erroneous
measurement values.
Note: Insert included batteries before start-up (refer to «Section 7.2» for detailed information).
Note: The sensor is delivered ex factory not connected to the transmitter. Therefore, it has to be polarized before
first time operation.
Connect the InTap sensor to the analyzer. After 6 hours, the sensor is fully polarized and ready for operation.
Leave the sensor permanently connected to the transmitter after first time polarization. This ensures that the
sensor is continuously supplied with a polarization voltage (even when the unit is switched off), thus avoiding
a polarization time of about 6 h before being ready to take a measurement. In addition, this protects the sensor
from contamination.
On the following page is the sequence for correct start-up and measurement application. Comply with
the described procedure in order to ensure optimal response time, a correct measurement result and
maximum hygiene.
Start-up
4.1
InTap4000e
4.2
InTap4004e
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 15
1. Connect discharge tube (0.3 m) and bypass tube (0.3 m) to three-way valve (see picture below).
Warning! Push the tubes into the quick couplers to the stop.
Warning! The discharge tube may not be fitted with any form of extension, as this could allow air (avoid
measurement errors) to penetrate into the measuring chamber between measuring cycles.
Legend
1 Measuring chamber
2 Reduction nipple (diaphragm)
3 Discharge tube (0.3 m)
4 Three-way valve
5 Bypass tube (0.3 m)
1
2
3
4
5
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
16 InTap 4000e/InTap 4004e
2. Perform configuration of the measuring unit according to «Section 6.4»
This step is necessary during initial installation/start-up or if new requirements for configuration rise
have arisen.
3. Perform calibration according to «Section 6.5»
Calibration is essential during initial start-up of the instrument. In most instances, automatic calibration
in air is sufficient.
Note: We recommend periodic re-calibration. Frequency of re-calibration and the method used
(see «Section 6.5»), depend upon extent of use of the instrument, the required measurement accuracy
and you own experience.
4. Insert the sensor into the measuring chamber
After calibration, re-insert the sensor into the measuring chamber and tighten the sleeve nut.
Legend
1 Sensor in measuring chamber
2 Sleeve nut
1
2
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 17
5. Prepare the cleaning and conditioning solution
20
ml
15
10
5
20
40
20
40
60
80
100 ml
20
40
60
80
100 ml
5 Min.
Fill the measuring beaker with
40 ml of water.
Add one conditioning tablet.
Let stand for 5 minutes (do not
stir or shake) until the tablet has
completely dissolved.
Draw off 25 ml of cleaning and
conditioning solution using the
syringe (pull plunger up to the
stop). Make sure to remove any
air bubbles.
6. Fill the measuring chamber with cleaning and conditioning solution
Connect the syringe with tube
connection (diam. 6 mm) to
the measuring chamber.
Turn the three-way valve to
«Measuring position», then inject
approx. 15 ml of cleaning and
conditioning solution into measur-
ing chamber.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
18 InTap 4000e/InTap 4004e
Let cleaning and conditioning solution act:
Minimum time of exposure to the cleaning and conditioning solution: 15 minutes
Maximum time of exposure to the cleaning and conditioning solution: 1 week
Note: If no measurements are carried out with the InTap within three days, the cleaning and conditioning
solution should be removed from the measuring chamber, and both the sensor and the measuring chamber
thoroughly dried.
Turn the sampling valve to the
«BYPASS» position, then remove
syringe with tubing connection.
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
Printed in Switzerland 52 201 080
InTap 4000e/InTap 4004e 19
7. Connecting the feed tube
8. Process connection
As the beverage industry does not have any valid worldwide standard for connections to storage tanks or
pipes, no special tube adapter for connection to processes is included in the scope of our supply. Connec-
tion of the feed tube to the process is the responsibility of the customer.
For the Perlick-type connections used on storage tanks in the USA, METTLER TOLEDO offers either the
InTap4004e or an appropriate tank sample adapter port (order no. 52 200 261) as an optional accessory.
Warning! Leaky tube couplings between process and measuring chamber (sensor) are the most
frequent cause of erroneous measurements. All tube couplings and tube extensions between the
process connection and the measuring chamber must therefore be absolutely leakproof.
Warning! Special care has been taken in the selection of a material with low oxygen permeability for
the feed tube connected to the process. Therefore, only use the feed tube supplied by us.
Warning!
Push feed tube into the quick coupler to the stop.
1 Feed tube (1.5 m)
2 Three-way valve in
«BYPASS» position
1
2
© 07/04 Mettler-Toledo GmbH, CH-8606 Greifensee
52 201 080 Printed in Switzerland
20 InTap 4000e/InTap 4004e
Connection to tank/pipe Connection to tank/pipe
with Perlick-type connection
(USA-Standard)
Tank or pipe
To measuring
chamber
Feed tube (1.5 m)
Tank or pipe
To measuring
chamber
Sample port connection
(order no. 52 200 261)
Feed tube (1.5 m)
Trysock sample port (Zwickel)
Feed tube (1.5 m, diam. 6/4 mm) to be fitted directly to
tank or pipe connection, or by means of special adapter
(responsibility of customer).
/