  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments Extech SD910 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
USER MANUAL
3ChannelDCVoltageDatalogger
ModelSD910
SD910enGB_V1.29/16
2
TableofContents
1. INTRODUCTION 3
2. DESCRIPTIONS 4
3. OPERATION 5
Power 5
ConnectingVoltageCables 5
Datalogging 5
Time/Date/SampleRateCheck 6
SDCardDataStructure 6
TransferringDatatoaPC 6
AdvancedSettings 7
SystemRESET 8
RS232Interface 8
Backlight 8
4. MAINTENANCE 8
BatteryReplacement 8
CleaningandStorage 8
5. SPECIFICATIONS 9
SD910enGB_V1.29/16
3
1. Introduction
Thank you for selecting the Extech SD910 3-Channel Voltage Datalogger.
TheSD910isathreechannel0to300mVor0to3000mVDCvoltagemonitoranddatalogger.
Dataisstoredataselectablerateandstoredforeasyexporttoaspreadsheet.
Thisdeviceisshippedfullytestedandcalibratedand,withproperuse,willprovideyearsof
reliableservice.Pleasevisitourwebsite(www.extech.com)tocheckforthelate stversionof
thisUserGuide,ProductUpdates,ProductRegistration,andCustomerSupport.
Features
TripleLCDsimultaneouslydisplays(3)300mVor(3)3000mVchannels
Dataloggerdate/timestampsandstoresreadingsonanSDcardinExcel®formatforeasy
transfertoaPC
Selectabledatasamplingrate:1,2,5,10,30,60,120,300,600seconds,auto
SD910enGB_V1.29/16
4
2. Descriptions
1. Channel1Display
2. Channel2Display
3. Channel3Display
4. LOGand
(ENTER)button
5. SETbutton
6. and
(TIME)button
7. and (Power)button
8.
ACadaptorsocket
9.
Resetbutton
10.
RS232output
11.
SDmemorycardsocket
12.
Channel1input
13.
Channel2input
14.
Channel3input
Note:BatteryCompartmentandTiltStandareloc ated onthebackofthemeter.
SD910enGB_V1.29/16
5
3. Operation
Power
1. Installsix(6)1.5VAAAbatteries.Thebatteriesprovidebackuppowerfortheclock.
2. ConnecttheAC/DCpoweradaptortotheDC9Vinputonthemeter.Thepoweradaptoris
requiredfordatalogging.
3. Whenthe
iconappearsinthedisplaythebatteriesareweakandshouldbereplaced.
However,inspec.measurementsmaystillbemadeforseveralhours,afterthelowbattery
indicatorappears.
ConnectingVoltageCables
1. Connectthevoltagewirestotheplugasshownandthen
inserttheplugintothesocketsonthebottomofthe
datalogger(observepolarity)
2. Uptothreecablesmaybeinstalled.
3. Thevoltagefortheinsertedcableswillbedisplayed1,2,
and3fromtopto
bottom.
4. 0.00willbedisplayedforopenorunusedinputs.
Datalogging
1. OpentheleftsidedoorandinsertaformattedSDcard
Notes
TheSDcardshouldbe1GBto4GBincapacity.
Donotusememorycardsformattedbyothermetersorcameras.UsetheSDcard
formattingprocedureundertheadvancedfeaturessectionofthismanualto
correctly
formatthecard.
Theinternalclockmustbesettothecorrecttime.Seetheadvancedfeaturessectionof
thismanualtosettheclock.
Thedefaultdatastructureusesadecimalpoint“.”asthenumericdecimalindicator.See
theadvancedfeaturessectionofthismanualtochangethistoacomma“,”.
IftheSDmemorycardisnotinstalled,EMPTYwillappearinthedisplay.
Ifthevoltagecableisnotinstalled,randomnumbersmayappearinthedatafile.
Ifthevoltagecableisnotinstalled,afterdatadownloadthe
eunitssymbolmaynot
appearinthedisplay.
SD910enGB_V1.29/16
6
Displayederrormessages:
Thememorycardisfullorthereisaproblemwiththecard
Thebatteryislowanddataloggingisdisabled
TheSDcardisnotinserted
2. PresstheLOGbuttonfor>2secondstobeginlogging.“DATALOGGER”willappearinthe
displayandthemeterwillbeepeachtimethedataisrecorded(ifthebeeperisenabled).
3. Tostopdatalogging,presstheLOGbuttonfor>2seconds.“DATALOGGER”willchangeto
“DATA”andthemeterwillcountdownthroughtherecordeddata.
NOTE:Toavoidcorruptingdata,donotremovethememory
cardwithoutproperlyendingthe
recordfunction.
Time/Date/SampleRateCheck
PressandHoldthe buttonfor>2secondsandthedisplaywillcyclethroughthedate,time
andsamplerateinformation.
SDCardDataStructure
1. WhentheSDcardiffirstinsertedintothe dataloggerthefolderMVA01iscreated.
2. ThefirstdataloggingsessionwillthencreateafileMVA01001.XLS.Alldatawillbesaved
tothisfileuntilthenumberofcolumnsreaches30,000.
3. After30,000columnsanewfile,MVA010
02.XLSiscreated.Thisisrepeatedevery30,000
columnsuntilMVA01099.XLS.Atthispointanewfolder,MVA02iscreatedandthe
processisrepeated.MVA10isthefinalfolder.
TransferringDatatoaPC
1. Removethememorycardfromthe dataloggerandplugitintotheSDcardslotonthePC.
2. LaunchExcelandopenthedatafileonthememorycard.Thefilewillappearsimilarto
thefigurebelow.
no
CArd
LobAt
CH-
CArd
SD910enGB_V1.29/16
7
AdvancedSettings
TheSETfunctionisusedto:
FormattheSDmemorycard
Setthedateandtime
Setthesamplingtime
SetthebeepersoundON/OFF
SettheSDcardDecimalcharacter
SettheRS232dataoutputON/OFF
Setthevoltagerangeto300mVor
3000mV
1. PressandHoldtheSETbuttonfor>2secondstoenterthesettingmode.
Thefirstfunction(SdF)willappearinthedisplay.PresstheSETbuttontostepthroughthe
sevenfunctions.Usetheandbuttonstoadjusttheselectedfunction.Usethe“LOG”
buttontostepthroughfieldswithinafunction.IntheSETmode,ifnobuttonispressed
within5seconds,theloggerwillrevertbacktothestandardmode.
2. SdFFormattheSDcard.
PressthebuttontoselectyESorno.ForyES,pressthe
button.WhenyESandEnt
appear,pressthe keyagaintoformatthecardanderaseallexistingdata.Thescreen
willdisplayaflashingyEsandESCwhilethememoryisbeingerasedandformatted.
3. dAtESetthedateandtime.
Presstheorbuttonstoadjusttheselected(blinking)field.Pressthe
buttonto
storethevalueandtostepthroughthevariousfields(Year,Month,Day,Hour,Minute,
Second)
4. SPtSetthesamplerate.
PressthebuttontoselectthedesiredsamplerateandpressEntertostoretheselection.
Theselectionsare:5,10,30,60,
120,300,600secondsandAUTO.InAUTO,thedatawillbe
storedeverytimethereisachangeof>10digits.
5. bEEP‐SetthebeeperONorOFF.
PressthebuttontoselectONorOFFandpress
tostoretheselection.
6. dEC‐SettheSDcardDecimalcharacter.
PressthebuttontoselectUSA(decimal)orEuro(comma)andpress
tostorethe
selection.
7. rS232‐SettheRS232dataoutputON/OFF.
PressthebuttontoselectONoroffandpress
tostoretheselection.
8. RngSetthevoltagerange
Pressthebuttontoselecteitherto3000mVorthe300.0mVrangeandpress
tostore
theselection.
9. ESCExitthesettingmode.
PresstheSETbuttontoreturntonormaloperation.
SD910enGB_V1.29/16
8
SystemRESET
IfaconditionappearswheretheCPUdoe snot respondtokeystrok esorthemeterseemsfrozen,
presstheRESETbutto nonthe sideof thedata logger (useapaper clipor similarpoint e dobject)to
returnthemeter toaworkingstate.
RS232Interface
Forstreamingofda tatoaPCviatheRS232Outputjack,theopt ional407001USBkit(RS232to
USBcableanddr iver CD)alongwiththe407001software(ava ilablefreeatwww.ex tech.com)are
required.
Backlight
Pressthe butt ontoturnonthebacklight.Thebacklightwillautoturnoffafter5se co nds.
4. Maintenance
BatteryReplacement
1. Toreplaceorinstallbatteries,removethePhilipsheadscrewthatsecurestherear
batterycoverandliftoffthecover.
2. ReplacethesixAAAbatteries(usealkalineheavydutytype),observingpolarity.
3. Replaceandsecurethecover.
BatterySafetyNotes:Pleasedisposeofbatteriesresponsibly;never
disposeofbatteriesina
fire,batteriesmayexplodeorleak.Ifthemeterisnottobeusedfor60daysormore,remove
thebatteryandstoreseparately.Donotmixbatterytypesorfreshnesslevels;pleaseuse
batteriesofthesametypeandofthesamefreshnesslevel.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatterieswerepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.The
userisobligatedtotake
endoflifedevicestoadesignatedcollectionpointforthedisposalofelectricalandelectronic
equipment.
CleaningandStorage
Periodicallywipethecasewithadampclothandmilddetergent;donotuseabrasivesor
solvents.
Pleaseremovethebatteriesifthemeterisstoredforalong periodoftime.
SD910enGB_V1.29/16
9
5. Specifications(at23±5°C)
Display 60mmx50mm(2.4x2.0”)LCDwithbacklight
MemoryCard SDmemorycard,1GBto16GB.
InputChannels 3
DCVoltageRange Resolution Accuracy
0to300.0mV 0.1mV ±(0.5%+0.2mV)
0to3000mV 1mV ±(0.5%+2mV)
Inputimpedance 5.6MOhms
Inputs Negative()terminalsconnectedtoanalogground
DataloggerSamplingTime 1/2/5/10/30/60/120/300/600seconds/auto
(withthesamplingratesetto‘1’secondsomedatapointscouldbelost)
Dataerrornumber≤0.1%no.oftotalsaveddatatypically
Displayupdaterate Approx.1second
DataOutput RS232
OperatingTemperature 0to50°C(32to122°F)
OperatingHumidity <85%R.H.
PowerSupply 9VACadaptorrequiredfordatalogging
6AAA(UM4)Alkalineorheavyduty1.5V
batteriesare
usedforclockbackupandmeasurementdisplayonly.
Batterylife Dependentonsamplerate,fornewalkalinebatteriesand
60secondsamplingtime,>onemonthistypical.Fast
samplerateswillsignificantlyreducebatterylife.
ThebatterylifewilldecreaseifthebeeperissetON
Weight 199g(0.44lbs.)
Dimension 132x80x32mm(5.2x3.1x1.3 “)
Note:AbovespecificationtestsperformedunderenvironmentalRFFieldStrength<3V/M&
frequency<30MHzonly.
Copyright©20152016FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
ISO9001Certified
www.extech.com
/