  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments SDL730 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
UserManual
HeavyDuty
DifferentialPressureManometer
ModelSDL730
AdditionalUserManualTranslationsavailableatwww.extech.com
2 SDL730-en-GB_V1.0 12/16
Introduction
CongratulationsonyourpurchaseoftheExtechSDL730PressureManometer.Thisdevice
measuresgaugeanddifferentialpressureintherangeof±100psi.Supportedmeasurement
unitsarembar,psi,Kg/cm
2
,mmHg,inchHg,metersofH
2
O,inchesofH
2
O,atmosphere,hPA,
andkPA.LoggeddatareadingsarestoredonanSDcardfortransfertoaPC.Thismeteris
shippedfullytestedandcalibratedand,withproperuse,willprovideyearsofreliableservice.
Pleasevisitourwebsite(www.extech.com)tocheckforthelatestversionofthisUser
Guideand
CustomerSupport.
Safety
InternationalSafetySymbols
Thissymbol,adjacenttoanothersymbolorterminal,indicatestheusermustrefertothe
manualforfurtherinformation.
MeterDescription
1. P1Input(positive+)
2. P2Input(negative‐)
3. LCDDisplay
4. HOLDandBacklight
key
5. MAXMINkey
6. SETandClock
key
7. RS232outputjack
8. Resetbutton
9. PowerAdaptorjack
10. SDcardslot
11. ENTERandLOGkey
12. Downarrowkey
(Press&keyatsametimetozero
meter)
13. Uparrow/UNITkey
(Press&keyatsametime
tozero
meter)
14. PowerONOFF
key
Notes:Items7,8,and9arelocatedbehindthesnapoffcompartmentcoveronmeter’srightside.
Batterycompa rtmen t,tiltstand,andtripodmount arelocatedontherearoftheinstrument.
1
11
12
13
14
2
3
4
5
6
7
8
9
10
3 SDL730-en-GB_V1.0 12/16
GettingStarted
PowerONOFF
Powerthemeterbypressingandholdingthepowerbutton foratleast1.5seconds.
Pressandholdthepowerbuttonforatleast1.5secondstopowerOFFthemeter.
Thismeterispoweredbysix(6)1.5VDC‘AA’batteriesorbyoptionalACadaptor.Ifthe
meterwillnotswitchONpleasecheckthatfreshbatteries
areinstalledintherearbattery
compartmentor,inthecaseoftheACadaptor,checkthattheadaptorisconnected
correctlytothemeterandtoanACsource.
DisplayBacklight
ToturnthedisplaybacklightONorOFF,pressandholdthebacklight buttonforatleast1.5
seconds.ThemeterwillbeepwhenswitchingthebacklightONorOFFunlessthebeeperis
disabled.
UnitsofMeasure
Thecurrentlyselectedunitofmeasureisshownbelowthemeasurementvalueonthemeter’s
LCD.Tochangetheunitofmeasure,pressandholdtheUNITbuttonuntilthedesiredunitof
measureappearsandthenreleasetheUNITbutton.Themeterbeginsscrollingthroughthe
availableunitsofmeasure
(seetablebelow)aftertheUNITbuttonhas beendepressedforat
least1.5seconds.
UNIT DISPLAYINDICATOR
psi PSI
InchHg InHg
InchH
2
O InH2O
hPA hPA
kPA _PA
Bar bAr
Kg/cm
2
_gC2
mmHg‐Hg
MetersH
2
O‐tH2O
Atmospheres AtP
ZEROAdjustment
Tonull,orzero,thedisplayforaconnectedsensor,pressandholdtheupanddownarrow
keys foratleast1.5seconds.ThemeterwillshowCAL0onthebottomleftofthescreen.
DataHold
TofreezeadisplayedreadingontheLCD,momentarilypresstheHOLDbutton(theHOLDicon
willappearabovethereading).ToexitHOLD,presstheHOLDbuttonagain.
4 SDL730-en-GB_V1.0 12/16
SetupMode
Basicsettingsataglance
Toviewthecurrentconfigurationofthemeterwithregardtotime,date,anddatalogging
samplingratepresstheSETbuttonmomentarily.Themeterwillnowdisplaytheconfigurationin
quicksuccession.Iftheinformationismissedonthefirsttry,simplypresstheSETbuttonagain
untilallofthe
informationisnoted.
AccessingtheSetupmode
1. PressandholdtheSETbuttonforatleast1.5secondstoaccesstheSetupmenu.
2. PresstheSETbuttonmomentarilytostepthroughtheavailableparameters.Theparameter
typeisshownonthebottomoftheLCDandthecurrentselectionforthattypeisshown
aboveit.
3. Whenaparameterisdisplayedthatistobechanged,usethearrowkeystochangethe
setting.PresstheENTERbuttontoconfirmachange.
4. PressandholdtheSETbuttonforatleast1.5secondstoexittheSetupmode.Notethatthe
meterautomaticallyswitches
outoftheSetupmodeifnokeyispressedwithin7seconds.
5.
TheavailableSetupparametersarelistedbelow.Additionaldetailedinformationis
providedbelowthislist:
dAtE
Settheclock(Year/Month/Date ;Hours/M inutes/Seconds)
SPt
Setthedataloggersamplingrate (Hours/Minutes/Seco nds)
PoFF
Automaticpoweroffmanagement(Enableordisabletheautopoweroff
function)
bEEP
SetthebeepersoundON/OFF
dEC
Setthenumericalformat;USA(decimal:20.0 0)orEuropean(comma:20,00)
SdF
FormattheSDmemory card
SettingtheClockTime
1. AccessthedAtEparameter.
2. Usethearrowkeystochangeavalue
3. UsetheENTERbuttontostepthroughtheselections
4. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemeter
toautomaticallyswitchtothenormaloperating
mode).
5. Theclockwillkeepaccuratetimeevenwhenthemeterisswitchedoff.However,ifthe
batteryexpirestheclockwillhavetoberesetafterfreshbatteriesareinstalled.
SettingtheDataloggerSamplingTime(Rate)
1. AccesstheSPtparameter.
2. Thesamplingratecanbesetfrom‘0’seconds(manuallogmode)upto8hours,59minutes,
and59seconds.
(Datalosemayoccurwhenloggingata1se co n drate)
3. UsetheENTERbuttontomovethroughtheHours,Minutes,andSecondsdigitgroups and
usethearrowkeystochangethedigitvalues.
4.
PresstheENTERbuttontoconfirmtheentry.
5 SDL730-en-GB_V1.0 12/16
5. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitchtothe normaloperating
mode).
Enabling/DisablingtheAutoPowerOFFFeature
1. AccessthePoFFparameter.
2. UsethearrowbuttonstoselectONorOFF.WiththeAutoPowerOFFfeatureenabled,the
meterwillautomaticallyswitchOFFafter10minutesofinactivity.
3. PressENTERtoconfirmsetting.
4. PressandholdtheSETbuttonforatleast1.5seconds
toexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitchtothe normaloperating
mode).
SettheBeeperSoundONorOFF
1. AccessthebEEPparameter.
2. UsethearrowbuttonstoselectONorOFF.
3. PressENTERtoconfirmsetting.
4. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitch
tothenormaloperating
mode).
NumericalFormat(commaordecimal)
EuropeanandUSAnumericalformatsdiffer.ThemeterdefaultstoUSAmodewhereadecimal
pointisusedtoseparateunitsfromtenths,i.e.20.00;TheEuropean formatusesacomma,i.e.
20,00toseparateunitsfromtenths.Tochangethissetting:
1. AccessthedECparameter.
2. Usethe
arrowbuttonstoselectUSAorEUro.
3. PressENTERtoconfirmsetting.
4. PressandholdtheSETbuttonforatleast1.5secondstoexittothenormaloperationmode
(orsimplywait7secondsforthemetertoautomaticallyswitchtothe normaloperating
mode).
SDCardFORMATTING
1. AccesstheSdFparameter.
2. UsethearrowbuttonstoselectYEStoformatthecard(selectNOtoabort).Notethatall
dataonthecardwillbelostifformattingisattempted.
3. PressENTERtoconfirmselection.
4. PressENTERagaintoreconfirm.
5. The
meterwillautomaticallyreturntothenormaloperatingmodewhenformattingis
complete.Ifnot,pressandholdtheSETbuttonforatleast1.5secondstoexittothenormal
operationmode.
6 SDL730-en-GB_V1.0 12/16
SystemReset
Ifthemeter’skeysbecomeinoperableorifthedisplayfreezestheResetbuttoncanbeusedto
resettheinstrument.
Useapapercliporsimilaritemtomomentarilypresstheresetbuttonlocatedonthelower
rightsideoftheinstrumentunderthesnapoffcompartmentcover.
AfterpressingtheResetbutton,switchtheinstrumentONbypressingandholdingthe
POWERkeyforatleast1.5seconds.Ifusingthepoweradaptorunplugtheadaptorandthen
plugitbackinagaintopowerthemeter.
MaxMinReadingRecord
Foragivenmeasurementsession,thismetercanrecordthehighest(MAX)andthelowest(MIN)
readingsforlaterrecall.
1. PresstheMAXMINbuttonmomentarilytoaccessthismodeofoperation(RECicon
appears)
2. ThemeterisnowrecordingtheMAXandMINreadings.
3. PresstheMAX
MINbuttonagaintoviewthecurrentMAXreadings(MAXiconappears).
ThereadingsonthedisplayarenowthehighestreadingsencounteredsincetheRECicon
wasswitchedon(whentheMAXMINbuttonwasfirstpressed).
4. PresstheMAXMINbuttonagaintoviewthecurrentMINreadings(MINiconappears).The
readingsonthedisplayarenowthelowestreadingsencounteredsincetheRECiconwas
switchedon(whentheMAXMINbuttonwasfirstpressed).
5. ToexittheMAXMINmode,pressandholdtheMAXMINbuttonforatleast1.5seconds.
Themeter
willbeep,theRECMAXMINiconswillswitchoff,theMAXMINmemorywill
clear,andthemeterwillreturntothenormaloperatingmode.
FittingsConnection
ConnecttubingtotheP1(+),P2(),orbothinputport(s).Ifbothinputsareused(differential
mode),themeterdisplaysapositivepressurereadingiftheP1(+)pressureisgreaterthanP2()
andanegativereadingifP2()isgreaterthanP1(+).
Datalogging
TypesofDataRecording
ManualDatalogging:Manuallylogupto99readingsontoanSDcardviapushbuttonpress.
AutomaticDatalogging:AutomaticallylogdataontoanSDmemorycardwherethenumber
ofdatapointsisvirtuallylimitedonlybythecardsize.Readingsareloggedataratespecified
bytheuser.
SDCardInformation
InsertanSDcard(from1Gsizeupto16G)intotheSDcardslotatthebottomofthemeter.
Thecardmustbeinsertedwiththefrontofthecard(labelside)facingtowardtherearofthe
meter.
IftheSDcardisbeingusedforthefirsttimeitisrecommendedthatthecardbeformatted
andthelogger’sclocksettoallowforaccuratedate/timestampingduringdatalogging
sessions.RefertotheSetupModesectionforSDcardformattingandtime/datesetting
instructions.
7 SDL730-en-GB_V1.0 12/16
EuropeanandUSAnumericalformatsdiffer.ThedataontheSDcardcanbeformattedfor
eitherformat.ThemeterdefaultstoUSAmodewhereadecimalpointisusedtoseparate
unitsfromtenths,i.e.20.00.TheEuropeanformatusesacomma,i.e.20,00.Tochangethis
setting,
refertotheSetupModesection.
ManualDatalogging
InthemanualmodetheuserpressestheLOGbuttontomanuallylogareadingontotheSDcard.
1. Setthesamplingrateto‘0’secondsasdescribedintheSetupModesection.
2. PressandholdtheLOGbuttonforatleast1.5secondsandtheDATALOGGERiconwill
appearontheLCD;thelowerportionofthedisplaywillshowpn(n=memoryposition
number199).NotethatifPSIissetastheunitofmeasureitappearsasP51(wherea‘5’is
usedasan‘S’)inthesameareaoftheLCDwherememorylocationsareshown.Thiscanbe
disorientingatfirst.
3. MomentarilypresstheLOGbuttontostoreareading.TheDATALOGGERiconwillflashand
thebeeperwillsound(ifsettoON)eachtimeadatapointisstored.
4. Usetheandbuttons
toselectoneofthe99datamemorypositionsinwhichtorecord.
5. Toexitthemanualdataloggingmode,pressandholdtheLOGbuttonforatleast1.5
seconds.TheDATALOGGERiconwillswitchoff.
AutomaticDatalogging
Inautomaticdataloggingmodethemetertakesandstoresareadingatauserspecifiedsampling
rateontoanSDmemorycard.Themeterdefaultstoasamplingrateoftwoseconds.Tochange
thesamplingrate,refertotheSetupModesection(thesamplingratecannotbe‘0’for
automaticdatalogging):
1. SelectthesamplingrateintheSetupModetoavalueotherthanzero.
2. PressandholdtheLOGbuttonforatleast 1.5seconds.Themeterwillflashthe
DATALOGGERiconattheselectedsamplingrateindicatingthatreadingsarenowbeing
automaticallyrecordedtothe
SDcard.
3. Ifacardisnotinsertedorifthecardisdefective,themeterwilldisplayEMPTYandexitthe
DATALOGGERmode.Inthiscase,switchthemeterOFFandtryagainwithavalidSDcard.
4. PausethedataloggerbypressingtheLOGbuttonmomentarily.
TheDATALOGGERiconwill
stopflashingandthesampleratewilldisplayforashorttime.Toresumeloggingsimply
presstheLOGbuttonagainmomentarily.
5. ToterminatethedataloggingsessionpressandholdtheLOGbuttonforatleast1.5
seconds.
6. WhenanSDcardisusedfor
thefirsttimeafolderiscreatedonthecardandnamed
PMA01.Upto99spreadsheetdocuments(eachwith30,000readings)canbestoredinthis
folder.
7. WhendataloggingbeginsanewspreadsheetdocumentnamedPMA01001.xlsiscreatedon
theSDcardinthePMA01folder.Thedata
recordedwillbeplacedinthePMA01001.xls
documentuntil30,000readingsarereached.
8. Ifthemeasurementsessionexceeds30,000readings,anewdocumentwillbecreated
(PMA01002.xls)whereanother30,000readingscanbestored.Thismethodcontinuesfor
upto99documents,afterwhichanotherfolderiscreated(PMA02)
whereanother99
spreadsheetdocumentscanbestored.Thisprocesscontinuesinthissamefashionwith
foldersPMA03throughPMA10(lastallowablefolder).
8 SDL730-en-GB_V1.0 12/16
SDDataCardtoPCDataTransfer
1. Completeadataloggingsessionasdetailedaboveintheprevioussections.Hint:Forthe
firstfewtests,simplyrecordasmallamountoftestdata.Thisistoensurethatthe
dataloggingprocessiswellunderstoodbeforecommittingtocritical,largescale
datalogging.
2. WiththemeterswitchedOFF,remove
theSDCard.
3. PlugtheSDCarddirectlyintoaPCSDcardreader.IfthePCdoesnothaveanSDcardslot,
useanSDcardadaptor(availableatmostoutletswherecomputeraccessoriesaresold).
4. PowerthePCandrunaspreadsheetsoftwareprogram.Open
thesaveddocumentsinthe
spreadsheetsoftwareprogram(seeexamplespreadsheetdatascreenbelow).
Spreadsheetdataexample
ACPowerAdaptor
Thismeterisnormallypoweredbysix(6)1.5V‘AA’batteries.Anoptional9Vpoweradaptoris
available.Whentheadaptorisused,themeterispermanentlypoweredandthepowerbutton
willbedisabled.
9 SDL730-en-GB_V1.0 12/16
HoseCouplings
Thismeterissuppliedwithapairofhosecouplings.Theyfit4.0mm(0.157”)tubing.
Seediagramonhowtoconnectthesecouplingstothehoseandthemeter.
10 SDL730-en-GB_V1.0 12/16
BatteryReplacementandDisposal
Whenthelowbatteryicon appearsontheLCD,thebatteriesmustbereplaced.Several
hoursofaccuratereadingsarestillpossibleinthiscondition;however,batteriesshouldbe
replacedassoonaspossible:
Removethetwo(2)Phillipsscrewsfromtherearofthemeter(directlyabovethetopofthe
tiltstand).
Removeandsafelyplacethebatterycompartmentandscrewswheretheywillnotbe
damagedorlost.
Replacethesix(6)1.5V‘AA’batteriesobservingcorrectpolarity.
Replacethebatterycompartmentcoverwiththetwo(2)Phillipsscrews.

Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatterieswerepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeend
oflifedevicestoadesignatedcollectionpointforthedisposalof electricalandelectronic
equipment.
OtherBatterySafetyReminders
o Neverdisposeofbatteriesinafire.Batteriesmayexplodeorleak.
o Nevermixbatterytypes.Alwaysinstallnewbatteriesofthesametype.

11 SDL730-en-GB_V1.0 12/16
Specifications
GeneralSpecifications
Display BacklitLCD;LCDsize:51x37mm(2x1.5”)
Statusindicators Overra nge audiblebeepandlowbatte rydisplayicon
MeasurementUnits m
Bar,psi,Kg/cm
2
,hPA,mmHg,inchHg,metersH
2
O,kPA,inches
H
2
O,ATP
Accuracy (Meter) ±(2%FS)atambienttempera ture23°C5°C)
Inputcircuit Differentialinputs(P1+andP2‐)
InputPorts Twometal5.0mm(0.197”),barbedfor4.0mm(0.157”)IDtubing
Hosecouplings Replacementpairofhosecouplings(407915)
Sensor Builtinpiezoelectricsensors
Datalogger SamplingRate AUTOLOGGING:From1secondupto8hour s59min59
sec.(Datalosemayoccurwhenloggingata1second rate)
MANUALLOGGING:Setthesamplingrateto‘0’seconds.
Select1to99locati ons .
Dataerrornumber≤0.1%numberoftotalsaveddata
MemoryCa rd SDmemorycard;1Gto16GBsize
DataHold Freezethe displayedreading
MemoryReca ll RecordandRecalltheMaximumandMinimumreadings
Displayupdate rate Approx.1second
OperatingTempera tur e 0to50
°
C(32to122
°
F)
OperatingHumidit y 85%R.H.max.
AutoPowerOFFAfter10minutesofinactivity(canbed isa b le d)
PowerSupply Six(6)1.5VDCbatter ies(optional9VACadaptor)
PowerConsumptio n Normaloperation(backlightanddataloggerOFF):approx.
7mAdc
WithbacklightOFFanddatalogg ing ON:approx.25mAdc
WithbacklightONadd
approx.10mAdc
Weight 265g(0.59lbs.)
Dimensio ns 190x68x45mm(7.5x2.7x1.8”)
12 SDL730-en-GB_V1.0 12/16
Range/ResolutionSpecifications
Range/Units(Max.) Resolution
±7000mbar 5
±101.5psi 0.05/0.1
±7.135Kg/cm
2
0.005
±5250mmHg 5
±206.7inchesHg 0.1
±700.0kPa 0.5
±7000hPA 5
±2810inchesH
2
O 2
±71.35mH
2
O 0.05
±6.905ATP 0.005
Conversions
UsefulconversionFactors
From To Multiplier
inofH
2
O inofHg 0.07355
inofH
2
O cmofH
2
O 2.54
mmofHg inofH
2
O 0.53524
Copyright©2016FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
ISO9001Certified
www.extech.com
/