Medtronic Surestream User manual

Type
User manual
1
© 2013 Oridion Medical 1987 Ltd.
All rights reserved.
PN 011045F
Products for Non-Intubated Patients
Directions for Use
ETCO
2
SAMPLING
CO
2
Oral/Nasal
Sampling with O
2
CO
2
Oral/
Nasal
Sampling
with O
2
CO
2
Oral/
Nasal
Sampling
CO
2
Nasal
Sampling
CO
2
Nasal
Sampling
with O
2
O
2
O
2
≤ 5 L/min
2
en
Surestream
TM
EtCO
2
Sampling
for Non-Intubated Patients
Single patient use, not to be re-processed.
Non-sterile / non-pyrogenic.
Does not contain natural rubber latex
Warning: Loose or damaged connections may compro-
mise ventilation or cause an inaccurate mea-
surement of respiratory gases. Securely connect
all components and check connections for leaks
according to standard clinical procedures.
Warning: Do not cut or remove any part of the sample
line. Cutting the sample line could lead to er-
roneous readings.
Warning: Check CO
2
and O
2
tubing regularly during use to
ensure that no kinks are present. Kinked tubing
may cause inaccurate CO
2
sampling or affect O
2
delivery to patient.
Caution: Ensure that tubing is not stretched during use.
Caution: Do not attempt to clean, disinfect, sterilize or
ush any part of the sampling line as this can
cause damage to the monitor.
Caution: Dispose of sampling lines according to standard
operating procedures or local regulations for the
disposal of contaminated medical waste.
Caution: Use of a CO
2
sampling line with H in its name
(indicating that it is for use in humidied envi-
ronments) during MRI scanning may cause in-
terference. These sampling lines include Smart
SureLine Plus H O
2
and SureLine H O
2
. The use
of non H sampling lines is advised.
Note: Replace the sampling line according to hospital
protocol or when a blockage is indicated by the
device. Excessive patient secretions or a build-
up of liquids in the airway tubing may occlude
the sampling line, requiring more frequent re-
placement.
Federal law (U.S.) restricts this device to sale by or on the
order of a physician.
Do not reuse
Does not contain natural rubber latex
Fire Hazard
Storage temperature
Rx Only
Caution, consult accompanying documents
CE Mark
ONLY
Does not
contain natural
rubber latex
28
Oridion Medical 1987 Ltd.
Har Hotzvim Science-
Based Industrial Park
P.O. Box 45025
Jerusalem 91450, Israel
Tel: 972-2-589-9111
Fax: 972-2-582-5873
Oridion Capnography Inc.
Needham MA 02494
USA
Tel: (781) 453-0500
Toll Free: 1-800-ORIDION
Obelis S.A.
Av. De Tervuren, 34 Bte. 44
B-1040 Brussels
Belgium
Tel: +32.2.34.33.87
This product is covered by one or more
of the following US patents: 6,422,240;
5,857,461; 6,437,316 6,926,005;
5,657,750; 7,383,839; 7,500,483;
6,783,573; and their foreign equivalents.
Additional patent applications pending.
www.oridion.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28

Medtronic Surestream User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI