Medtronic CapnoLine User manual

Type
User manual
1
CapnoLine
®
H O
2
For Non-intubated Patients
Infant and Neonatal
Directions for Use
ETCO
2
SAMPLING
CapnoLine
®
H O
2
Length=3m/10ft
© 2015 Oridion Medical 1987 Ltd.
All rights reserved.
PN 012345E
US Patents: www.covidien.com/patents.
3
CapnoLine
®
H O
2
For Non-intubated Infant and
Neonatal Patients
Instructions for use:
Connect the connector of the CapnoLine
®
H O
2
sam-
pling line to the CO
2
port on the Microstream monitor.
. Screw the connector into the monitor clockwise until
it can no longer be turned.
Attach the O
2
tubing to an O
2
delivery system.
Connect the sampling line to the patient by gently in-
serting the nare sections of the patient face piece into
the nares of the patient. The tubing may be looped
around the patient’s ears (as seen in gure 3), if desired,
to hold the sampling line in place better.
Ensure that the tubing is not twisted or crimped.
1.
2.
3.
4.
2
3
1
1
3
2
≤ 3 L/M
O
2
5
en
CapnoLine
®
H O
2
for Non-intubated Patients
Infant and Neonatal
For use with all Microstream-enabled monitors
Single patient use, not to be re-processed
Non-sterile / non-pyrogenic
Does not contain natural rubber latex
Warning: Loose or damaged connections may compromise ventila-
tion or cause an inaccurate measurement of respiratory
gases. Securely connect all components and check con-
nections for leaks according to standard clinical proce-
dures.
Warning: Do not cut or remove any part of the sample line. Cutting
the sample line could lead to erroneous readings.
Warning: Do not cover the Naon
®
* ** on the CO
2
tubing. Covering
the Naon
®
* will limit its effectiveness in reducing moisture
and occlusions in the sampling line.
Warning: Check CO
2
and O
2
tubing regularly during use to ensure
that no kinks are present. Kinked tubing may cause inac-
curate CO
2
sampling or affect O
2
delivery to patient.
Caution: Ensure that tubing is not stretched during use.
Caution: Clear the patient’s nose of mucus and other obstructions
before use to optimize duration of use, CO
2
reading ac-
curacy and O
2
delivery. Duration of use, CO
2
readings as
well as O
2
inhalation may be affected if patient nares are
not patent.
Caution: Do not attempt to clean, disinfect, sterilize or ush any
part of the sampling line as this can cause damage to the
monitor.
Caution: Dispose of sampling lines according to standard operating
procedures or local regulations for the disposal of contami-
nated medical waste.
Caution: Use of a CO
2
sampling line with H in its name (indicating
that it is for use in humidied environments) during MRI
scanning may cause interference. These sampling lines
include CapnoLine H O
2
. The use of non H sampling lines
is advised.
Note: Replace the sampling line according to hospital protocol
or when a blockage is indicated by the device. Excessive
patient secretions or a build-up of liquids in the airway tub-
ing may occlude the sampling line, requiring more frequent
replacement.
Note: When connecting a sampling line to the monitor, insert the
sampling line connecter clockwise into the monitor CO
2
port and turn the connector until it can no longer be turned,
to ensure that it is connected securely to the monitor. This
will assure that there is no leak of gases during measure-
ment at the connection point and that measurement ac-
curacy is not compromised.
Note: Following connection of the CO
2
sampling line, check that
CO
2
values appear on the monitor display.
Note: Expect delay times for the following products to be longer
than the standard by the given times: Long (4m) Capno-
Lines - 3 seconds; 3m CapnoLines - 1.5 seconds.
Note: Note: Sampling lines with H in their names include a mois-
ture reduction component (Naon
®
* or its equivalent) for
use in higher humidity environments where long duration
use of CO
2
sampling is required.
Note: Note: Sampling lines are recommended for use with oxy-
gen provided at up to 3l/min. At higher levels of oxygen
provision, dilution of CO
2
readings may occur, leading to
lower CO
2
values.
Federal law (U.S.) restricts this device to sale by or on the order of a
physician.
** or alternate moisture reduction technology
31
Does not contain natural rubber latex
Storage temperature
Rx Only
Caution, consult accompanying documents
CE Mark
Caution, consult
accompanying
documents
-4°F
-20°C
131°F
55°C
Temperature
limitation
This product does
not contain natural
rubber latex
0482
Fire hazard
Fire Hazard
MR Unsafe
Not safe for use in MR environments
Single
patient use
For use with a single patient
Consult instructions for use
Consult
instructions
for use
Caution, consult
accompanying
documents
-4°F
-20°C
131°F
55°C
Temperature
limitation
This product does
not contain natural
rubber latex
0482
Fire hazard
Ne contient pas de latex naturel
Température de stockage
Sur prescription uniquement
Attention, consulter la documentation fournie
Marquage CE
Risque d'incendie
MR Unsafe
Dangereuse dans les environnements de RM
Utiliser pour un seul patient
Single
patient use
Consult
instructions
for use
Consulter le mode d’emploi
32
www.oridion.com
Oridion Medical 1987 Ltd.
7 Hamarpe St.
P.O. Box 45025
Jerusalem 9777407, Israel
US Toll Free: 1-888-674-3466
1-888-ORIDION
Obelis S.A.
Bd Général Wahis 53
1030 Brussels
Belgium
Oridion
®
, Microstream
®
, FilterLine
®
, Smart CapnoLine
®
, CapnoLine
®
, and Smart CapnoLine Guardian™ are trademarks
or registered trademarks of Oridion Medical 1987 Ltd.
TM
* and
®
* are trademarks of their respective owners.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Medtronic CapnoLine User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI