Medtronic Surestream User manual

Type
User manual
1
Smart SureLine
Guardian
TM
Directions for Use
Sampling lines: 7 ft / 2 m or 14 ft / 4 m
Smart
SureLine
Guardian
TM
bite block
60 Fr.
Surestream
TM
oral-nasal
breath
sampling line
ETCO
2
SAMPLING
© 2015 Oridion Medical 1987 Ltd.
All rights reserved.
PN 012575F
US Patents: www.covidien.com/patents.
2
en
Surestream
TM
EtCO
2
Sampling
Smart SureLine Guardian
TM
Intended Use
The Smart SureLine Guardian™ set is intended to sample oral and
nasal CO
2
and administer supplemental oxygen via the nose and mouth,
for patients who can wear a 60 fr. bite block, during upper endoscopy
type procedures. The set is intended for single patient use only.
Contents:
25 Smart SureLine Guardian™ kits that contain:
• 1 Smart SureLine Guardian™ bite block
• 1 Surestream
TM
oral/nasal breath sampling line with O
2
delivery
Note: A hook and loop strap (PN 012542) is available for use instead
of the rubber strap, for enhanced securing of the Guardian
TM
on the
patient. See insert for usage instructions.
Instructions for Use
See the sketches below for instructions for use. Please note:
• Connect the sampling line to the monitor by inserting the sampling
line connecter clockwise into the monitor CO
2
port and turning
the connector until it can no longer be turned, to ensure that it is
connected securely to the monitor. This will assure that there is no
leak of gases during measurement at the connection point and that
measurement accuracy is not compromised.
• Place the oral/nasal sampling line on patient. Ensure that the
unattached end of the O
2
connector on the right side of the sampling
line is facing downward (see sketch 1). Adjust length of oral prong (A
in sketch 3) until fully extended (sketch 1).
• Place the Smart SureLine Guardian™ bite block on the patient
after the oral/nasal sampling line is in place. Ensure that the strap
is tightened and that it remains under the O
2
and CO
2
lines during
the procedure. Once both parts are in place, adjust the length of the
oral prong of the sampling line so that it rests in the channel of the
Smart SureLine Guardian™ bite block (B in sketch 4) and is visible
through the window. (This is to accommodate different patient face
sizes.) Please note that the oral prong may need to be repositioned to
keep it in the channel during the procedure.
• The oral/nasal sampling line includes an O
2
luer connector. Attach
this connector to the Smart SureLine Guardian™ bite block as
required (see sketch 5). If sampling line and Smart SureLine
Guardian™ bite block are not connected via the O
2
luer, 100% of
the O
2
owwillbedeliveredtothepatient’snares.Ifsamplingline
and Smart SureLine Guardian™ bite block are connected via the
O
2
luer (as recommended), O
2
will simultaneously be delivered to
thepatient’snaresandmouth,with80%oftheO
2
owgoingtothe
patient’smouthand20%dispersedthroughtheO
2
holes below the
patient’snares.
• After the Smart SureLine Guardian™ bite block is removed from
the patient at the end of the procedure, CO
2
measurement and O
2
delivery can continue during recovery with the sampling line alone.
O
2
delivery should be <5L/min when being delivered only through
the nasal cannula.
3
2
1
O
2
A
B
3
O
2
O
2
4
5
6
O
2
O
2
Warning
• Check CO
2
and O
2
tubing regularly during use to ensure that
no kinks are present. Kinked tubing may cause inaccurate
CO
2
sampling or affect O
2
delivery to patient.
Cautions
• Securely connect all components. Loose connections or failure to
correctly insert the oral prong into the Smart SureLine Guardian™
channel may cause an inaccurate measurement of respiratory gases.
The oral prong must remain in the channel throughout the procedure.
• Ensure that tubing is not stretched during use.
• Sedation may cause hypoventilation and CO
2
waveform distortion
or disappearance. Waveform attenuation or disappearance is an
indicatorthatthestatusofthepatient’sairwayshouldbeassessed.
• Partial blockage of the oral airway due to endoscope positioning
may cause periods of low readings and rounded waveforms. The
occurrence will be more pronounced with high oxygen delivery
levels.
• If O
2
isdeliveredatowratesgreaterthan10l/min,thepatient’s
exhaled CO
2
may be diluted, producing attenuated waveforms and
EtCO
2
values.
• If CO
2
insufationisperformedduringCO
2
monitoring, the EtCO
2
valueswillaccordinglyriseverysignicantlyandthismayresult
in device alarms and abnormally high waveforms until the CO
2
is
evacuated from the patient.
• Expect delay times for the following products to be longer than
the standard by the given times: Long (4m) Smart SureLine
Guardian™ - 3 seconds.
• The Guardian is recommended for use with oxygen provided at up
to 10l/min, or 5l/min with the sampling line alone. At higher levels
of oxygen provision, dilution of CO
2
readings may occur, leading to
lower CO
2
values.
55
Does not contain natural rubber latex
Storage temperature
Rx Only
Caution, consult accompanying documents
CE Mark
Caution, consult
accompanying
documents
-4°F
-20°C
131°F
55°C
Temperature
limitation
This product does
not contain natural
rubber latex
0482
Fire hazard
Fire Hazard
DEHP
Consult instructions for use
Consult
instructions
for use
Single
patient use
For use with a single patient
Caution, consult
accompanying
documents
-4°F
-20°C
131°F
55°C
Temperature
limitation
This product does
not contain natural
rubber latex
0482
Fire hazard
Ne contient pas de latex naturel
Température de stockage
Sur prescription uniquement
Attention, consulter la documentation fournie
Marquage CE
Risque d'incendie
DEHP
Consult
instructions
for use
Consulter le mode d’emploi
Utiliser pour un seul patient
Single
patient use
56
www.oridion.com
Oridion Medical 1987 Ltd.
7 Hamarpe St.
P.O. Box 45025
Jerusalem 9777407, Israel
US Toll Free: 1-888-674-3466
1-888-ORIDION
Obelis S.A.
Bd Général Wahis 53
1030 Brussels
Belgium
Oridion
®
, Microstream
®
, FilterLine
®
, Smart CapnoLine
®
, CapnoLine
®
, Smart CapnoLine Guardian™, Surestream
TM
, and
Smart SureLine Guardian
TM
are trademarks or registered trademarks of Oridion Medical 1987 Ltd.
TM
* and
®
* are trade-
marks of their respective owners.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56

Medtronic Surestream User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI