Wiring Diagram - Power Supply Assembly (End View)
Line Voltage at J2 Supply Power In
•TerminalsW4&W5
Line Voltage on PCB
•TerminalsW3&W5
•TerminalsW2&W5
•TerminalW5&onJ1plug
terminal 1 (Brown lead)
Line Voltage for 115 or 230 VAC
to Transformer
• LineVoltageatJ1plug
Terminals1&5
• 115VACatTerminals2&3
• 115VACatTerminals4&5
24 VAC Supply to Unit
• J4plugTerminals1&2
• J6plugTerminals1&3,2&4
(D2LEDisON).
20 VAC Supply to Light
• J5plugTerminals1&2
• J7plugTerminals1&2
(D3LEDisON)
12 VDC Supply to LIN Connections
(Chair & Assistant’s Unit)
• J3plugTerminals1&2
• J8plugTerminals1&2
(D1LEDisON)
Continuity Check of Transformer’s
Thermal Overload (O/L)
Continuity should be present at
TransformerBrown&Blackleads
• J1plugTerminals1&2
POWER SUPPLY PCB TEST POINTS
(RefertoWiringDiagram-PowerSupplyAssembly)
Note: Air Pressure Switch must be operated.(Contact COM & NO closed)
All fuses, F1, F2, F3, F4, & F5 must be good.
Transformer Thermal Overload (O/L) must be closed.
Note:
PE Gnd... ..Protective Earth Ground.
LIN Gnd.. ..Chair Control DCV Ground reference.
PSI kpa
30 206.8
50 344.7
80 551.6
100 689.5
Unit Power
(From J6)
Light
Power
(From J7)
12
VDC
LED
24
VAC
LED
20
VAC
LED
Note
This drawing is for clarity.
Supply Power
In
Line Power Out
to Chair
LIN Connections
(FromChair&Assistant’sUnit)
UnitLight
Air Pressure Switch
D3 D2 D1
Power to Unit
J6 Plug
1 PEGnd
2 LINGnd
3 24 VAC
Com
4 24 VAC
5 PEGnd
6 LIN
7 24 VAC
Com
8 24 VAC
Power to Light
J7 Plug
1 PEGnd
2 LINGnd
3 20 VAC Com
4 N/C
5 LIN
6 20 VAC
Chair - Power
1 Line
2 Neutral
3 PEGnd
Supply Power
In J2 Plug
1 Line
2 Neutral
3 PEGnd
LED Power
Indicators
D1 ONwhen12VDCis
suppliedtoJ3&J8
LINConnections.
D2 ONwhen24VACis
suppliedtoJ6Unit
Connector.
D3 ONwhen20VACis
suppliedtoJ7Light
Connector.
Light Harness - Reference
Delivery Unit Harnesses - Reference