Adam Equipment GFC Series User manual

Type
User manual

This manual is also suitable for

© Adam Equipment Company 2014
AdamEquipment
User Manual (EN)
GBC/GFC Series
(P.N. 3056610587, Revision B4, Apr 2014)
Notice Utilisation (FR)
Séries GBC/GFC
Bedienungsanleitung (DE)
GBC / GFC REIHEN
SoftwareRev2.18andup
© Adam Equipment Company 2014
EasyReference:

Modelnameoftheindicator/scale:
Serialnumberoftheunit:
Softwarerevisionnumber
(Displayedwhenpowerisfirstturnedon):
DateofPurchase:
Nameofthesupplierand place:

Page 3 © Adam Equipment Company 2014
2.0 INTRODUCTION
TheGBC/GFCseriesprovidesaccurate,fastandversatilepartscountingscales.
There are 2 types of scales within the GBC/GFCseries:GBC/GFCscalesare
kilogram/gramscalesandtheGBC/GFC‐ascalesarechangeablefrompoundsto
kilogramsiftheuserrequiresit.Thescaleshavethesamefunctionsexceptthat
GBC/GFC‐ascaleshavetheabilitytotogglebetweentheweighingunits.
AllscalesaresuppliedwithaRS‐232bi‐directionalinterfaceandrealtimeclock
(RTC).
The GBC/GFC has a sealed keypad with colour coded membrane switches and
thereisalarge,easytoreadliquidcrystaltypedisplay(LCD).TheLCDissupplied
withabacklight.
TheGBC/GFCincludeautomaticzerotracking,audiblealarmforpre‐setcounts,
automatictare,pre‐settare,anaccumulationfacilitythatallowsthecounttobe
storedandrecalledasanaccumulatedtotal.
Page 4 © Adam Equipment Company 2014
SPECIFICATIONS
GBCModels
Model#
GBC16 GBC32 GBC60 GBC35a GBC70a GBC130a
MaximumCapacity
16kg 32kg 60kg 16kg/35lb 32kg/70lb 60kg/130lb
Readability
0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb 2g/0.005lb
Repeatability(StdDev)
0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb 2g/0.005lb
Linearity+
1g 2g 4g 1g/0.002lb 2g/0.004lb 4g/0.01lb
UnitsofMeasure KilogramsandGrams.PoundsforGBCXXaunits
StabilizationTime 2‐3Secs
OperatingTemperature ‐10°Cto+40°C/+14°Fto+104°F
PowerSupply 230VAC50/60Hz.inEurope,AsiaandSouthAfrica.
12vDC800mAUL/CSAadapterforUSA
Calibration External
CalibrationMass UserSelectable
Display BacklitGreendisplay25mmwithcapacitytracker
BalanceHousing Castaluminiumbase,Pantonecoolgreypaintedbase,stainlesssteelgrade304Toppan,ABSCoolgreyindicatorhousing
PanSize
300mmx400mmx50mm/12”x15.7”x2”
Overall Dimensions
(wxdxh)
300mmx520mmx660mm/11.8”x20.5”x26”
NetWeight 7.6kg/16.8Lb
Features Weighing/Counting/CheckweighingwithLEDlights/Percentage/Holdfunction/RS232
Page 5 © Adam Equipment Company 2014
GFCModels
Model#
GFC75 GFC150 GFC300 GFC165a GFC330a GFC660a
MaximumCapacity 75kg 150kg 300kg 75kg/165lb 150kg/330lb 300kg/660lb
Readability 5g 10g 20g 5g/0.01lb 10g/0.02lb 20g/0.05lb
Repeatability(StdDev) 5g 10g 20g 5g/0.01lb 10g/0.02lb 20g/0.05lb
Linearity+ 10g 20g 40g 10g/0.02lb 20g/0.04lb 40g/0.1lb
UnitsofMeasure KilogramsandGrams.PoundsforGFCXXaunits
StabilizationTime 2‐3Secs
OperatingTemperature ‐10°Cto+40°C/+14°Fto+104°F
PowerSupply 230VAC50/60Hz.inEurope,AsiaandSouthAfrica.
12vDC800mAUL/CSAadapterforUSA
Calibration External
CalibrationMass UserSelectable
Display BacklitGreendisplay25mmwithcapacitytracker
BalanceHousing Castaluminiumbase,Pantonecoolgreypaintedbase,stainlesssteelgrade304Toppan,ABSCoolgreyindicatorhousing
PanSize
400mmx500mmx65mm/15.7”x19.7”x2.6”
OverallDimensions
(wxdxh)
400mmx620mmx860mm/15.7”x24.4”x33.9”
NetWeight 12.5kg/27.6 Lb
Features Weighing/Counting/CheckweighingwithLEDlights/Percentage/Holdfunction/RS232
© Adam Equipment Company 2011
3.0 INSTALLATION
3.1 UNPACKING
Theuser’sapplicationandthetechnicalspecificationsoftheplatformorloadcellwill
determinethenecessaryconfiguration.
3.2 LOCATING
Thescalesshouldnotbeplacedinalocationthat
willreducetheaccuracy.
Avoidextremesoftemperature.Donotplacein
directsunlightornearairconditioningvents.
Avoidunsuitabletables.Thetableorfloormust
berigidandnotvibrate.
Avoidunstablepowersources.Donotusenear
large users of electricity such as welding
equipmentorlargemotors.
Donotplacenearvibratingmachinery.
Avoid high humidity that might cause
condensation. Avoid direct contact with water.
Donotsprayorimmersethescalesinwater.
Avoidairmovementsuchasfromfansoropening
doors. Do not place near open windows or air
conditioningvents.
Keepthescalesclean.Donotstackmaterialon
thescaleswhentheyarenotinuse.

© Adam Equipment Company 2011
3.3 SETTINGUPTHESCALES
Thepillarisattachedtothebaseusingabracketthatmustbeattached
tothebaseframefirstusingthe4boltssupplied.Thepillarissecuredto
thebracketusing2setsofscrews.Thecablefromthebasetothe
indicator module is run through the tube and taken out through the
plasticsupportatthetop.Excesscablecanbestoredwithinthetube.
The GBC/GFC Series comes with a stainless steel platform packed
separately.Placetheplatforminthebase.
Levelthescalebyadjustingthefourfeet.Ifthescalerocksre‐adjustthe
feet.
Attachtheindicatormoduletothepillarbyslidingitoverthebracket
withtheflangesengagedinthegrovesonthebase.Attachthecable
fromthebasetotheconnectorontherearoftheindicator.
Attachthepowertotheindicator.Pressthe[On/Off]keywhichisatthe
rear of the indicator. The software revision number will be displayed
followedbyaselftestshowingalldigitsbeforethezeroisdisplayed
alongwiththeunitofweightthatwasselectedlast.
© Adam Equipment Company 2011
4.0 KEY
DESCRIPTIONS
Keys
Functions
[0‐9,.] Numericentrykeys,usedtomanuallyenteravaluefortareweights,unit
weight,andsamplesize.
[CE] Usedtocleartheunitweightoranerroneousentry.
[M+] Addthecurrentcounttotheaccumulator.Upto99valuesorfullcapacityof
theweightdisplaycanbeadded.AlsoprintsthedisplayedvalueswhenAuto
printisswitchedoff.
[MR] Torecalltheaccumulatedmemory.
[Pst]
Tosettheupperlimitforthenumberofitemscounted.Whenthisupperlimit
isexceededthescalewillsoundthebeeper.
[Print] Toprinttheweight/unitweightandcounttoaPCorprinterusingtheRS‐232
interface.Printthetotalsfrommemorywhen[MR]ispressedfirst.
[Smpl] Usedtoinputthenumberofitemsinasample.
[U.Wt./Units] Usedtoentertheweightofasamplemanually.
Selecttheweighingunitwhenthe“UnitWeight”displayisatzeroifenabled.
[Tare] Taresthescale.Storesthecurrentweightinmemoryasatarevalue,subtracts
thetarevaluefromtheweightandshowstheresults.Thisisthenetweight.
Enteringavalueusingthekeypadwillstorethatasthetarevalue.
[Zero] Setsthezeropointforallsubsequentweighingtoshowzero.
© Adam Equipment Company 2011
5.0 DISPLAYS
Thescaleshavethreesections.Theseare“Weight”,“UnitWeight”and“Count”.
TheLCDdesignis:
WEIGHT UNIT WEIGHT
COUNT
5.1 WEIGHT
DISPLAY
Ithas6‐digitdisplaytoindicatetheweightonthescale.
NetWeightDisplay,"Net"
Stabilityindicator, symbol
Zeroindicator,0symbol
Theweighingunit: kg, lb
© Adam Equipment Company 2011
5.2 UNITWEIGHTDISPLAY
Thisdisplaywillshowtheunitweightofasample.Thisvalueiseitherinputbythe
user or computed by the scale. The unit of measurement is gramsifgramsif
kilogramsisselectedfortheweighingunit,orpoundsifpoundsisselected.
5.3 COUNT
DISPLAY
This display will show the number of items on the scale or the value of the
accumulatedcount.SeethenextsectiononOPERATION.
5.4 OTHER
SYMBOLS
The symbol, when there is insufficient number of samples to
accuratelydeterminethecount.
Whentheunitweightisnotlargeenoughtodetermineanaccuratecount,
the symbolwillbeon.
Inbothcasesthescalecontinuestooperateandtheindicationsaretoalert
theuserforapotentialproblem.
Ifapresetcounthasbeenstoredthe symbolwillbeon.
Alowbatterysymbolwillturnonifthebatteryshouldberecharged.
The Lo, OK, Hisymbolswillbeusedwhenthepresetcounthasbeenset.
Thesymbol“Mem”willbeseenwhentheinternalmemoryhasarecordof
previousweighingsstoredinit.
© Adam Equipment Company 2011
6.0 OPERATION
NOTE:GBC/GFC‐aScalesonly
SETTINGTHEWEIGHINGUNIT,lborkg
Thescalewillturnondisplayingthelastweighingunitselected,eitherkilogramsor
pounds.Tochangetheweighingunitpressthe[U.Wt./Units]keywhenthe“Unit
Weight”displayshowszero.Ifnecessarypressthe[CE]keytocleartheunitweight
beforechanging.
6.1 ZEROINGTHEDISPLAY
Youcanpressthe[Zero]keyatanytimetosetthezeropointfromwhich
allotherweighingandcountingismeasured.Thiswillusuallybenecessary
only when the platform is empty. When the zero point is obtained the
“Weight”displaywillshowtheindicatorforzero,“>Z<”.
Thescalehasanautomaticrezeroingfunctiontoaccountforminordrifting
oraccumulationof materialon the platform.However youmayneedto
press[Zero]tore‐zerothescaleifsmallamountsofweightarestillshown
whentheplatformisempty.
6.2 TARING
Zerothescalebypressingthe[Zero]keyifnecessary.Theindicator“>Z<”
willbeON.
Placeacontainerontheplatformanditsweightwillbedisplayed.
Press[Tare]totarethescale.Theweightthatwasdisplayedisstoredas
thetarevaluewhichissubtractedfromthedisplay,leavingzeroonthe
display.Theindicators“Netand“>Z<”willbeON.
Asaproductisaddedonlytheweightoftheproductwillbeshown.The
scalecouldbetaredasecondtimeifanothertypeofproductwastobe
addedtothefirstone.Againonlytheweightthatisaddedaftertaring
willbedisplayed.
Whenthecontainerisremovedanegativevaluewillbeshown.Ifthe
scalewastaredjustbeforeremovingthecontainer,thisvalueisthegross
weight of the container plus all products those were removed. The
indicator “>Z<”willalsobeONbecausetheplatformisbacktothesame
conditionasitwaswhenthe[Zero]keywaspressedlast.
6.2.1 PresetTare
Theusercanenterapresettarevalueifthedisplayisatzeroorgross
weight.Enterthevalueforpresettareusingthekeypad,thenpressthe
© Adam Equipment Company 2011
[Tare]keytosetthepresetzerovalue.
Ifthedisplaywasatzerotheweightwillshowanegativevalueequalto
thepresettarevalueenteredandthe“Net”indicatorwillbeon.
6.3 PARTS
COUNTING
6.3.1 SettingUnitWeight
Inordertodopartscountingitisnecessarytoknowtheaverageweightoftheitems
tobecounted.Thiscanbedonebyweighingaknownnumberoftheitemsandletting
thescaledeterminetheaverageunitweightorbymanuallyinputtingaknownunit
weightusingthekeypad.
A.WeighingasampletodeterminetheUnitWeight
Todeterminetheaverageweightoftheitemstobecounted,youwillneedto
placeaknownquantityoftheitemsonthescaleandthentokeyinthenumber
of items being weighed. The scale will then divide the total weight by the
numberofitemsand display theaverageunitweight.Press[CE]anytimeto
cleartheunitweight.
Zerothescalebypressingthe[Zero]keyifnecessary.Ifacontaineristo
beused,placethecontaineronthescaleandtarebypressing[Tare]as
discussedearlier.
Placeaknownquantityofitemsonthescale.Aftertheweightdisplayis
stable,enterthequantityofitemsusingthenumerickeysandthenpress
the[Smpl]key.
Thenumberofunitswillbedisplayedonthe"Count"displayandthe
computedaverageweightwillbeshownonthe"UnitWeight"display.
Asmoreitemsareaddedtothescale,theweightandthequantitywill
increase.
Ifaquantitywhichissmallerthanthesampleisplacedonthescale,then
thescalewillautomaticallyenhancetheUnitWeightbyre‐calculatingit.
TolocktheUnitWeightandavoidre‐sampling,press[U.Wt./Units].
If the scale is not stable, the calculation will not be completed. If the
weightisbelowzero,the“Count”displaywillshownegativecount.
© Adam Equipment Company 2011
B.EnteringaknownUnitWeight
Iftheunitweightisalreadyknownthenitispossibletoenterthatvalue
usingthekeypad.
Enter the value of the unit weight in grams, using the numeric keys
followedbypressingthe[U.Wt./Units]key.The"UnitWeight"display
willshowthevalueasitwasentered.
Thesampleisthenaddedtothescaleandtheweightwillbedisplayedas
wellasthequantity,basedontheunitweight.
6.3.2 Countingmoreparts
Aftertheunitweighthasbeendeterminedorentered,itispossibleto
usethescaleforpartscounting.Thescalecanbetaredtoaccountforthe
containerweightasdiscussedintheearliersection.
Afterthescaleistaredtheitemstobecountedareaddedandthe
"Count"displaywillshowthenumberofitems,computedusingthetotal
weightandtheunitweight.
It is possible to increase the accuracy of the unit weight at any time
duringthecountingprocessbyenteringthecountdisplayedandthen
pressingthe[Smpl]key.Youmustbecertainthatthequantitydisplayed
matchesthequantityontheplatformbeforepressingthekey.Theunit
weightcanbeadjustedbaseduponalargersamplequantity.Thiswill
givegreateraccuracywhencountinglargersamplesizes.
6.3.3 Check‐counting
Check‐counting is a procedure to cause an alarm to sound when the
numberofitemscountedonthescalemeetsorexceedsanumberstored
inthememorybyusingthe[Pst]key.
Thevaluetobestoredisenteredfromthekeyboard.Enterthenumeric
valuetobestoredusingthenumerickeys.Thenpressthe[Pst]keyto
storethevalue.
Toclearthevaluefromthememoryandtherebyturnoffthecheck‐
countingfeature,enterthevalue"0"andpress[Pst].
Duringcountingifthecountisbelowthesetpointthe“Lo”indicatorwill
beon,ifitisequaltothesetpointthe “OK” indicatorwillbeonandifitis
abovethelimitthe“Hi”indicatorwillbeon.
6.3.4 ManuallyAccumulatedTotals
Thevalues(weightandcount)shownonthedisplaycanbeaddedtothe
© Adam Equipment Company 2011
valuesinthememorybypressingthe[M+]key.The"Weight"displaywill
showthetotalweight,the"Count" display will show the total
accumulatedcountandthe"UnitWeight"displayshowsthenumberof
times,theitemshavebeenaddedtothememoryforaccumulation.The
valueswillbedisplayedfor2secondsbeforereturningtonormal.Ifthe
accumulatednumberexceeds6digitsthedisplaywillshow“diSPEr”,the
countisstillbeingstoredcorrectlyandtheRS232printoutwillstillshow
thecorrectvalues.
The “Mem” symbol will be on when there are values stored in the
accumulationmemory.
The scale must return to zero or a negative number, before another
samplecanbeaddedtothememory.
Moreproductscanthenbeaddedandthe[M+]keytobepressedagain.
Thiscancontinueforupto99entriesoruntilthecapacityofthe
“Weight” display is exceeded. When the weight exceeds 6 digits the
displaywillshow“diSPEr”.
Toobservethetotalstoredvalue,pressthe[MR]key.Thetotalwillbe
displayedfor2seconds.
Toclearthememoryfirstpress[MR]torecallthetotalsfrommemory
andthenpressthe[CE]keytoclearallvaluesfromthememory.
6.3.6AutomaticAccumulatedTotals
Thescalecanbesettoautomaticallyaccumulatetotalswhenaweightis
placedonthescale.Thiseliminatestheneedtopressthe[M+]keyto
storevaluesintothememory.Howeverthe[M+]keyisstillactiveand
canbepressedtostorethevaluesimmediately.Inthiscasethevalues
willnotbestoredwhenthescalereturnstozero.
See the Section 9.0 on RS‐232 Interface for details on how to enable
AutomaticAccumulation.
© Adam Equipment Company 2011
7.0 USERPARAMETERS
Theparametersaresettocustomisethescaletosuittheweighingapplications.You
needtoenterasecuremenubyenteringapasswordwhenrequested.
Press[Tare] once, during the initial counting of the display after the
poweristurnedon.
The “Weight” display will show "PIn " requesting for the password
number.
Thedefaultpasswordis"0000"butothernumberscanbesetusingthe
parametermenus.Pressthe[0]keyfourtimes.
Pressthe[Tare]key.
Thereare3functionsthatcanbecycledthroughusingthe[U.Wt./Units]
key.TheWeightwindowwillshowtheparameternumberandtheUnit
Pricewindowwillshowtheworddescribingthefunction.
Thefirstparameteris “F1” “CAL”.(Seesection8.0fordetails)
ToenterintoCalibrationfunction,pressthe[Tare]key.
Toexitaparameter,pressthe[Zero]key.
Toselectthesecondparameterpressthe[U.Wt./Units]key.
© Adam Equipment Company 2011
SETTINGOFPIN
Display will show “F2” “Pin”.Thisparameterallowssettingofanew
passwordnumber.Thedefaultpasswordis “0000”.
Press[Tare].The“Weight”displaywillshow"Pin 1".
Enterthenewpasswordnumber.The“UnitWeight”displaywillshow
dashes.Press[Tare].
The“Weight”displaywillchangeto"Pin 2 ",Enterthepasswordagain
andpress[Tare].
The display will show "donE" to show the new password has been
acceptedand willreturn to the menu. Record the new passwordin a
securedplace.
Toexitaparameter,pressthe[Zero]key.
ENABLEWEIGHINGUNITS
Displaywillshow“F3” “Unit”.Thisparameterallowsweighingunitstobe
enabledordisabled.Ifenabledtheweighingunitswillchangewhenthe
[U.Wt./Unit]keyispressed.Seesection6.0.
Toenterthisfunctionpressthe[Tare]key.
ThedisplaywillshowthefirstweighingunitandifitiscurrentlyOnor
Off.Pressthe[U.Wt./Unit]keytochangethesettingandthenpressthe
[Tare]keytogotothenextweighingunit.Theweighingunitsselected
willdependuponthemodelandthecapacityofthescales.SeeSection
12.2,ServiceParameters.
Pressthe[Zero]keytoreturntotheweighingmode.
© Adam Equipment Company 2011
8.0 CALIBRATION
TheGBC/GFCscalesarecalibratedusingmetricweightsandGBC/GFC‐a
scalesarecalibratedusingmetricorpoundweightsdepending onthe
unitinuse before calibration.SeetheServiceparameters section for
moreinformation.
Thescalewilldisplayavalueoftheweighttobeusedforcalibration,this
valueisthelastweightusedforcalibration.Youcanenteradifferent
value,ifdesired.
ForenteringCalibration,seesection7.0Parameters
When“F1” “CAL”isdisplayed,press[Tare].
Thedisplaywillthenshow"unLoAd"torequestallweightberemoved
fromtheplatform.
Pressthe[Tare]keytosetthezeropoint.
Thedisplayswillthenshowthecalibrationweightsuggested,thevalue
isshownasanintegernumber,itisnotpossibletohavefractionsofa
kilogramorpound.Forexample:
LoAd 00004
KiLoS
Placethecalibrationweightontheplatform.Ifthecalibrationweightis
differentfromthevalueshown,Press[CE]toclearthecurrentvaluethen
enterthecorrectvalue.
Press[Tare]tocalibrate.
Whencalibrationisdonethedisplaywillshow“SPAn” “PASS”.
Removetheweight.Thescalewillreturntonormalweighing.
Ifanerrormessage“SPAn” “FAIL”isdisplayed,thenrepeattheprocess.
Aftercalibration,thescaleshouldbecheckedwhetherthecalibrationis
correct.Ifnecessary,repeatcalibration.
NOTE:GBC/GFC‐ascaleswillhavethelborkgindicatoron,toshowtheunitofthe
weightrequested.Ifthescalewasinpoundsbeforestartingthe calibration, the
weightsrequestedwillbeinpoundvaluesorifthescalewasweighinginkilograms
thenmetricweightswillberequested.
© Adam Equipment Company 2011
9.0 RS‐232INTERFACE
The GBC/GFC Series are supplied with a RS‐232 bi‐directional interface. The scale
whenconnectedtoaprinterorcomputerthroughtheRS‐232interface,outputsthe
weight,unitweightandcount.
Specifications:
Connector:
ThescalecanbesettoprinttextinEnglish,French,GermanorSpanish.SeetheRS‐
232parameterssectionfordetails.
Alllinesendwithlinefeed<lf>,andcarriagereturn<cr>characters.
RS‐232outputofweighingdata
ASCIIcode
AdjustableBaudrate,600,1200,2400,4800,9600and19200baud
8databits
NoParity
9pinD‐subminiaturesocket
Pin3Output
Pin2Input
Pin5SignalGround
© Adam Equipment Company 2011
DataFormat‐NormalOutput:
DataFormat‐MemoryRecallPrint:
<lf><cr>
<lf><cr>
DATE 12/09/2006
TIME 14:56
<lf><cr>
GROSS Wt 1.234 Kg NetWt.ifnetweightisdisplayed
Unit Wt. 123 g Kgorgformetricweightsandlbforpounds.
Pcs 10 pcs
<lf><cr> Includes2linefeedswithcarriagereturn
<lf><cr> atbeginningandendoftheform
<lf><cr>
DATE 12/09/2006
TIME 14:56
<lf><cr>
------------------
TOTAL
No. 5
Wt. 1.234 Kg
Pcs 10 pcs
<lf><cr>
Includes1linefeed
------------------
<lf><cr>
2linefeeds,carriagereturn
<lf><cr>
© Adam Equipment Company 2011
DataFormat‐ContinuousPrint:
Inotherlanguagestheformatisthesamebutthetextwillbeinthelanguageselected.
SeeSection9.2
Description ENGLISH FRENCH GERMAN SPANISH
Print gross
weight
Gross Wt Pds Brut Brut-Gew Pso Brut
Netweight
Net Wt. Pds Net Net-Gew Pso Net
Weight per
unitcounted
Unit Wt. Pds unit Gew/Einh Pso/Unid
Number of
items
counted
Pcs Pcs Stck. Piezas
Number of
weighings
added to
subtotals
No. Nb. Anzhl Num.
Total weight
and count
printed
Total Total Gesamt Total
Printdate
Date Date Datum Fecha
Printtime
Time Heure Zeit Hora
ST,GS, 0.9080 kg
U.W. 0.90798 g/pcs
PCS 1000 pcs
<lf><cr> Includes2linefeedswithcarriagereturn
<lf><cr> betweensetsofdata
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108

Adam Equipment GFC Series User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI