Adam Equipment GFC 165a, GBC 130a, GBC 60, GBC GFC, GFC Series User manual

  • Hello! I am an AI chatbot trained to assist you with the Adam Equipment GFC 165a User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
© Adam Equipment Company 2014
AdamEquipment
User Manual (EN)
GBC/GFC Series
(P.N. 3056610587, Revision B4, Apr 2014)
Notice Utilisation (FR)
Séries GBC/GFC
Bedienungsanleitung (DE)
GBC / GFC REIHEN
SoftwareRev2.18andup
© Adam Equipment Company 2014
EasyReference:

Modelnameoftheindicator/scale:
Serialnumberoftheunit:
Softwarerevisionnumber
(Displayedwhenpowerisfirstturnedon):
DateofPurchase:
Nameofthesupplierand place:

Page 3 © Adam Equipment Company 2014
2.0 INTRODUCTION
TheGBC/GFCseriesprovidesaccurate,fastandversatilepartscountingscales.
There are 2 types of scales within the GBC/GFCseries:GBC/GFCscalesare
kilogram/gramscalesandtheGBC/GFC‐ascalesarechangeablefrompoundsto
kilogramsiftheuserrequiresit.Thescaleshavethesamefunctionsexceptthat
GBC/GFC‐ascaleshavetheabilitytotogglebetweentheweighingunits.
AllscalesaresuppliedwithaRS‐232bi‐directionalinterfaceandrealtimeclock
(RTC).
The GBC/GFC has a sealed keypad with colour coded membrane switches and
thereisalarge,easytoreadliquidcrystaltypedisplay(LCD).TheLCDissupplied
withabacklight.
TheGBC/GFCincludeautomaticzerotracking,audiblealarmforpre‐setcounts,
automatictare,pre‐settare,anaccumulationfacilitythatallowsthecounttobe
storedandrecalledasanaccumulatedtotal.
Page 4 © Adam Equipment Company 2014
SPECIFICATIONS
GBCModels
Model#
GBC16 GBC32 GBC60 GBC35a GBC70a GBC130a
MaximumCapacity
16kg 32kg 60kg 16kg/35lb 32kg/70lb 60kg/130lb
Readability
0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb 2g/0.005lb
Repeatability(StdDev)
0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb 2g/0.005lb
Linearity+
1g 2g 4g 1g/0.002lb 2g/0.004lb 4g/0.01lb
UnitsofMeasure KilogramsandGrams.PoundsforGBCXXaunits
StabilizationTime 2‐3Secs
OperatingTemperature ‐10°Cto+40°C/+14°Fto+104°F
PowerSupply 230VAC50/60Hz.inEurope,AsiaandSouthAfrica.
12vDC800mAUL/CSAadapterforUSA
Calibration External
CalibrationMass UserSelectable
Display BacklitGreendisplay25mmwithcapacitytracker
BalanceHousing Castaluminiumbase,Pantonecoolgreypaintedbase,stainlesssteelgrade304Toppan,ABSCoolgreyindicatorhousing
PanSize
300mmx400mmx50mm/12”x15.7”x2”
Overall Dimensions
(wxdxh)
300mmx520mmx660mm/11.8”x20.5”x26”
NetWeight 7.6kg/16.8Lb
Features Weighing/Counting/CheckweighingwithLEDlights/Percentage/Holdfunction/RS232
Page 5 © Adam Equipment Company 2014
GFCModels
Model#
GFC75 GFC150 GFC300 GFC165a GFC330a GFC660a
MaximumCapacity 75kg 150kg 300kg 75kg/165lb 150kg/330lb 300kg/660lb
Readability 5g 10g 20g 5g/0.01lb 10g/0.02lb 20g/0.05lb
Repeatability(StdDev) 5g 10g 20g 5g/0.01lb 10g/0.02lb 20g/0.05lb
Linearity+ 10g 20g 40g 10g/0.02lb 20g/0.04lb 40g/0.1lb
UnitsofMeasure KilogramsandGrams.PoundsforGFCXXaunits
StabilizationTime 2‐3Secs
OperatingTemperature ‐10°Cto+40°C/+14°Fto+104°F
PowerSupply 230VAC50/60Hz.inEurope,AsiaandSouthAfrica.
12vDC800mAUL/CSAadapterforUSA
Calibration External
CalibrationMass UserSelectable
Display BacklitGreendisplay25mmwithcapacitytracker
BalanceHousing Castaluminiumbase,Pantonecoolgreypaintedbase,stainlesssteelgrade304Toppan,ABSCoolgreyindicatorhousing
PanSize
400mmx500mmx65mm/15.7”x19.7”x2.6”
OverallDimensions
(wxdxh)
400mmx620mmx860mm/15.7”x24.4”x33.9”
NetWeight 12.5kg/27.6 Lb
Features Weighing/Counting/CheckweighingwithLEDlights/Percentage/Holdfunction/RS232
© Adam Equipment Company 2011
3.0 INSTALLATION
3.1 UNPACKING
Theuser’sapplicationandthetechnicalspecificationsoftheplatformorloadcellwill
determinethenecessaryconfiguration.
3.2 LOCATING
Thescalesshouldnotbeplacedinalocationthat
willreducetheaccuracy.
Avoidextremesoftemperature.Donotplacein
directsunlightornearairconditioningvents.
Avoidunsuitabletables.Thetableorfloormust
berigidandnotvibrate.
Avoidunstablepowersources.Donotusenear
large users of electricity such as welding
equipmentorlargemotors.
Donotplacenearvibratingmachinery.
Avoid high humidity that might cause
condensation. Avoid direct contact with water.
Donotsprayorimmersethescalesinwater.
Avoidairmovementsuchasfromfansoropening
doors. Do not place near open windows or air
conditioningvents.
Keepthescalesclean.Donotstackmaterialon
thescaleswhentheyarenotinuse.

© Adam Equipment Company 2011
3.3 SETTINGUPTHESCALES
Thepillarisattachedtothebaseusingabracketthatmustbeattached
tothebaseframefirstusingthe4boltssupplied.Thepillarissecuredto
thebracketusing2setsofscrews.Thecablefromthebasetothe
indicator module is run through the tube and taken out through the
plasticsupportatthetop.Excesscablecanbestoredwithinthetube.
The GBC/GFC Series comes with a stainless steel platform packed
separately.Placetheplatforminthebase.
Levelthescalebyadjustingthefourfeet.Ifthescalerocksre‐adjustthe
feet.
Attachtheindicatormoduletothepillarbyslidingitoverthebracket
withtheflangesengagedinthegrovesonthebase.Attachthecable
fromthebasetotheconnectorontherearoftheindicator.
Attachthepowertotheindicator.Pressthe[On/Off]keywhichisatthe
rear of the indicator. The software revision number will be displayed
followedbyaselftestshowingalldigitsbeforethezeroisdisplayed
alongwiththeunitofweightthatwasselectedlast.
© Adam Equipment Company 2011
4.0 KEY
DESCRIPTIONS
Keys
Functions
[0‐9,.] Numericentrykeys,usedtomanuallyenteravaluefortareweights,unit
weight,andsamplesize.
[CE] Usedtocleartheunitweightoranerroneousentry.
[M+] Addthecurrentcounttotheaccumulator.Upto99valuesorfullcapacityof
theweightdisplaycanbeadded.AlsoprintsthedisplayedvalueswhenAuto
printisswitchedoff.
[MR] Torecalltheaccumulatedmemory.
[Pst]
Tosettheupperlimitforthenumberofitemscounted.Whenthisupperlimit
isexceededthescalewillsoundthebeeper.
[Print] Toprinttheweight/unitweightandcounttoaPCorprinterusingtheRS‐232
interface.Printthetotalsfrommemorywhen[MR]ispressedfirst.
[Smpl] Usedtoinputthenumberofitemsinasample.
[U.Wt./Units] Usedtoentertheweightofasamplemanually.
Selecttheweighingunitwhenthe“UnitWeight”displayisatzeroifenabled.
[Tare] Taresthescale.Storesthecurrentweightinmemoryasatarevalue,subtracts
thetarevaluefromtheweightandshowstheresults.Thisisthenetweight.
Enteringavalueusingthekeypadwillstorethatasthetarevalue.
[Zero] Setsthezeropointforallsubsequentweighingtoshowzero.
© Adam Equipment Company 2011
5.0 DISPLAYS
Thescaleshavethreesections.Theseare“Weight”,“UnitWeight”and“Count”.
TheLCDdesignis:
WEIGHT UNIT WEIGHT
COUNT
5.1 WEIGHT
DISPLAY
Ithas6‐digitdisplaytoindicatetheweightonthescale.
NetWeightDisplay,"Net"
Stabilityindicator, symbol
Zeroindicator,0symbol
Theweighingunit: kg, lb
© Adam Equipment Company 2011
5.2 UNITWEIGHTDISPLAY
Thisdisplaywillshowtheunitweightofasample.Thisvalueiseitherinputbythe
user or computed by the scale. The unit of measurement is gramsifgramsif
kilogramsisselectedfortheweighingunit,orpoundsifpoundsisselected.
5.3 COUNT
DISPLAY
This display will show the number of items on the scale or the value of the
accumulatedcount.SeethenextsectiononOPERATION.
5.4 OTHER
SYMBOLS
The symbol, when there is insufficient number of samples to
accuratelydeterminethecount.
Whentheunitweightisnotlargeenoughtodetermineanaccuratecount,
the symbolwillbeon.
Inbothcasesthescalecontinuestooperateandtheindicationsaretoalert
theuserforapotentialproblem.
Ifapresetcounthasbeenstoredthe symbolwillbeon.
Alowbatterysymbolwillturnonifthebatteryshouldberecharged.
The Lo, OK, Hisymbolswillbeusedwhenthepresetcounthasbeenset.
Thesymbol“Mem”willbeseenwhentheinternalmemoryhasarecordof
previousweighingsstoredinit.
© Adam Equipment Company 2011
6.0 OPERATION
NOTE:GBC/GFC‐aScalesonly
SETTINGTHEWEIGHINGUNIT,lborkg
Thescalewillturnondisplayingthelastweighingunitselected,eitherkilogramsor
pounds.Tochangetheweighingunitpressthe[U.Wt./Units]keywhenthe“Unit
Weight”displayshowszero.Ifnecessarypressthe[CE]keytocleartheunitweight
beforechanging.
6.1 ZEROINGTHEDISPLAY
Youcanpressthe[Zero]keyatanytimetosetthezeropointfromwhich
allotherweighingandcountingismeasured.Thiswillusuallybenecessary
only when the platform is empty. When the zero point is obtained the
“Weight”displaywillshowtheindicatorforzero,“>Z<”.
Thescalehasanautomaticrezeroingfunctiontoaccountforminordrifting
oraccumulationof materialon the platform.However youmayneedto
press[Zero]tore‐zerothescaleifsmallamountsofweightarestillshown
whentheplatformisempty.
6.2 TARING
Zerothescalebypressingthe[Zero]keyifnecessary.Theindicator“>Z<”
willbeON.
Placeacontainerontheplatformanditsweightwillbedisplayed.
Press[Tare]totarethescale.Theweightthatwasdisplayedisstoredas
thetarevaluewhichissubtractedfromthedisplay,leavingzeroonthe
display.Theindicators“Netand“>Z<”willbeON.
Asaproductisaddedonlytheweightoftheproductwillbeshown.The
scalecouldbetaredasecondtimeifanothertypeofproductwastobe
addedtothefirstone.Againonlytheweightthatisaddedaftertaring
willbedisplayed.
Whenthecontainerisremovedanegativevaluewillbeshown.Ifthe
scalewastaredjustbeforeremovingthecontainer,thisvalueisthegross
weight of the container plus all products those were removed. The
indicator “>Z<”willalsobeONbecausetheplatformisbacktothesame
conditionasitwaswhenthe[Zero]keywaspressedlast.
6.2.1 PresetTare
Theusercanenterapresettarevalueifthedisplayisatzeroorgross
weight.Enterthevalueforpresettareusingthekeypad,thenpressthe
© Adam Equipment Company 2011
[Tare]keytosetthepresetzerovalue.
Ifthedisplaywasatzerotheweightwillshowanegativevalueequalto
thepresettarevalueenteredandthe“Net”indicatorwillbeon.
6.3 PARTS
COUNTING
6.3.1 SettingUnitWeight
Inordertodopartscountingitisnecessarytoknowtheaverageweightoftheitems
tobecounted.Thiscanbedonebyweighingaknownnumberoftheitemsandletting
thescaledeterminetheaverageunitweightorbymanuallyinputtingaknownunit
weightusingthekeypad.
A.WeighingasampletodeterminetheUnitWeight
Todeterminetheaverageweightoftheitemstobecounted,youwillneedto
placeaknownquantityoftheitemsonthescaleandthentokeyinthenumber
of items being weighed. The scale will then divide the total weight by the
numberofitemsand display theaverageunitweight.Press[CE]anytimeto
cleartheunitweight.
Zerothescalebypressingthe[Zero]keyifnecessary.Ifacontaineristo
beused,placethecontaineronthescaleandtarebypressing[Tare]as
discussedearlier.
Placeaknownquantityofitemsonthescale.Aftertheweightdisplayis
stable,enterthequantityofitemsusingthenumerickeysandthenpress
the[Smpl]key.
Thenumberofunitswillbedisplayedonthe"Count"displayandthe
computedaverageweightwillbeshownonthe"UnitWeight"display.
Asmoreitemsareaddedtothescale,theweightandthequantitywill
increase.
Ifaquantitywhichissmallerthanthesampleisplacedonthescale,then
thescalewillautomaticallyenhancetheUnitWeightbyre‐calculatingit.
TolocktheUnitWeightandavoidre‐sampling,press[U.Wt./Units].
If the scale is not stable, the calculation will not be completed. If the
weightisbelowzero,the“Count”displaywillshownegativecount.
© Adam Equipment Company 2011
B.EnteringaknownUnitWeight
Iftheunitweightisalreadyknownthenitispossibletoenterthatvalue
usingthekeypad.
Enter the value of the unit weight in grams, using the numeric keys
followedbypressingthe[U.Wt./Units]key.The"UnitWeight"display
willshowthevalueasitwasentered.
Thesampleisthenaddedtothescaleandtheweightwillbedisplayedas
wellasthequantity,basedontheunitweight.
6.3.2 Countingmoreparts
Aftertheunitweighthasbeendeterminedorentered,itispossibleto
usethescaleforpartscounting.Thescalecanbetaredtoaccountforthe
containerweightasdiscussedintheearliersection.
Afterthescaleistaredtheitemstobecountedareaddedandthe
"Count"displaywillshowthenumberofitems,computedusingthetotal
weightandtheunitweight.
It is possible to increase the accuracy of the unit weight at any time
duringthecountingprocessbyenteringthecountdisplayedandthen
pressingthe[Smpl]key.Youmustbecertainthatthequantitydisplayed
matchesthequantityontheplatformbeforepressingthekey.Theunit
weightcanbeadjustedbaseduponalargersamplequantity.Thiswill
givegreateraccuracywhencountinglargersamplesizes.
6.3.3 Check‐counting
Check‐counting is a procedure to cause an alarm to sound when the
numberofitemscountedonthescalemeetsorexceedsanumberstored
inthememorybyusingthe[Pst]key.
Thevaluetobestoredisenteredfromthekeyboard.Enterthenumeric
valuetobestoredusingthenumerickeys.Thenpressthe[Pst]keyto
storethevalue.
Toclearthevaluefromthememoryandtherebyturnoffthecheck‐
countingfeature,enterthevalue"0"andpress[Pst].
Duringcountingifthecountisbelowthesetpointthe“Lo”indicatorwill
beon,ifitisequaltothesetpointthe “OK” indicatorwillbeonandifitis
abovethelimitthe“Hi”indicatorwillbeon.
6.3.4 ManuallyAccumulatedTotals
Thevalues(weightandcount)shownonthedisplaycanbeaddedtothe
© Adam Equipment Company 2011
valuesinthememorybypressingthe[M+]key.The"Weight"displaywill
showthetotalweight,the"Count" display will show the total
accumulatedcountandthe"UnitWeight"displayshowsthenumberof
times,theitemshavebeenaddedtothememoryforaccumulation.The
valueswillbedisplayedfor2secondsbeforereturningtonormal.Ifthe
accumulatednumberexceeds6digitsthedisplaywillshow“diSPEr”,the
countisstillbeingstoredcorrectlyandtheRS232printoutwillstillshow
thecorrectvalues.
The “Mem” symbol will be on when there are values stored in the
accumulationmemory.
The scale must return to zero or a negative number, before another
samplecanbeaddedtothememory.
Moreproductscanthenbeaddedandthe[M+]keytobepressedagain.
Thiscancontinueforupto99entriesoruntilthecapacityofthe
“Weight” display is exceeded. When the weight exceeds 6 digits the
displaywillshow“diSPEr”.
Toobservethetotalstoredvalue,pressthe[MR]key.Thetotalwillbe
displayedfor2seconds.
Toclearthememoryfirstpress[MR]torecallthetotalsfrommemory
andthenpressthe[CE]keytoclearallvaluesfromthememory.
6.3.6AutomaticAccumulatedTotals
Thescalecanbesettoautomaticallyaccumulatetotalswhenaweightis
placedonthescale.Thiseliminatestheneedtopressthe[M+]keyto
storevaluesintothememory.Howeverthe[M+]keyisstillactiveand
canbepressedtostorethevaluesimmediately.Inthiscasethevalues
willnotbestoredwhenthescalereturnstozero.
See the Section 9.0 on RS‐232 Interface for details on how to enable
AutomaticAccumulation.
© Adam Equipment Company 2011
7.0 USERPARAMETERS
Theparametersaresettocustomisethescaletosuittheweighingapplications.You
needtoenterasecuremenubyenteringapasswordwhenrequested.
Press[Tare] once, during the initial counting of the display after the
poweristurnedon.
The “Weight” display will show "PIn " requesting for the password
number.
Thedefaultpasswordis"0000"butothernumberscanbesetusingthe
parametermenus.Pressthe[0]keyfourtimes.
Pressthe[Tare]key.
Thereare3functionsthatcanbecycledthroughusingthe[U.Wt./Units]
key.TheWeightwindowwillshowtheparameternumberandtheUnit
Pricewindowwillshowtheworddescribingthefunction.
Thefirstparameteris “F1” “CAL”.(Seesection8.0fordetails)
ToenterintoCalibrationfunction,pressthe[Tare]key.
Toexitaparameter,pressthe[Zero]key.
Toselectthesecondparameterpressthe[U.Wt./Units]key.
© Adam Equipment Company 2011
SETTINGOFPIN
Display will show “F2” “Pin”.Thisparameterallowssettingofanew
passwordnumber.Thedefaultpasswordis “0000”.
Press[Tare].The“Weight”displaywillshow"Pin 1".
Enterthenewpasswordnumber.The“UnitWeight”displaywillshow
dashes.Press[Tare].
The“Weight”displaywillchangeto"Pin 2 ",Enterthepasswordagain
andpress[Tare].
The display will show "donE" to show the new password has been
acceptedand willreturn to the menu. Record the new passwordin a
securedplace.
Toexitaparameter,pressthe[Zero]key.
ENABLEWEIGHINGUNITS
Displaywillshow“F3” “Unit”.Thisparameterallowsweighingunitstobe
enabledordisabled.Ifenabledtheweighingunitswillchangewhenthe
[U.Wt./Unit]keyispressed.Seesection6.0.
Toenterthisfunctionpressthe[Tare]key.
ThedisplaywillshowthefirstweighingunitandifitiscurrentlyOnor
Off.Pressthe[U.Wt./Unit]keytochangethesettingandthenpressthe
[Tare]keytogotothenextweighingunit.Theweighingunitsselected
willdependuponthemodelandthecapacityofthescales.SeeSection
12.2,ServiceParameters.
Pressthe[Zero]keytoreturntotheweighingmode.
© Adam Equipment Company 2011
8.0 CALIBRATION
TheGBC/GFCscalesarecalibratedusingmetricweightsandGBC/GFC‐a
scalesarecalibratedusingmetricorpoundweightsdepending onthe
unitinuse before calibration.SeetheServiceparameters section for
moreinformation.
Thescalewilldisplayavalueoftheweighttobeusedforcalibration,this
valueisthelastweightusedforcalibration.Youcanenteradifferent
value,ifdesired.
ForenteringCalibration,seesection7.0Parameters
When“F1” “CAL”isdisplayed,press[Tare].
Thedisplaywillthenshow"unLoAd"torequestallweightberemoved
fromtheplatform.
Pressthe[Tare]keytosetthezeropoint.
Thedisplayswillthenshowthecalibrationweightsuggested,thevalue
isshownasanintegernumber,itisnotpossibletohavefractionsofa
kilogramorpound.Forexample:
LoAd 00004
KiLoS
Placethecalibrationweightontheplatform.Ifthecalibrationweightis
differentfromthevalueshown,Press[CE]toclearthecurrentvaluethen
enterthecorrectvalue.
Press[Tare]tocalibrate.
Whencalibrationisdonethedisplaywillshow“SPAn” “PASS”.
Removetheweight.Thescalewillreturntonormalweighing.
Ifanerrormessage“SPAn” “FAIL”isdisplayed,thenrepeattheprocess.
Aftercalibration,thescaleshouldbecheckedwhetherthecalibrationis
correct.Ifnecessary,repeatcalibration.
NOTE:GBC/GFC‐ascaleswillhavethelborkgindicatoron,toshowtheunitofthe
weightrequested.Ifthescalewasinpoundsbeforestartingthe calibration, the
weightsrequestedwillbeinpoundvaluesorifthescalewasweighinginkilograms
thenmetricweightswillberequested.
© Adam Equipment Company 2011
9.0 RS‐232INTERFACE
The GBC/GFC Series are supplied with a RS‐232 bi‐directional interface. The scale
whenconnectedtoaprinterorcomputerthroughtheRS‐232interface,outputsthe
weight,unitweightandcount.
Specifications:
Connector:
ThescalecanbesettoprinttextinEnglish,French,GermanorSpanish.SeetheRS‐
232parameterssectionfordetails.
Alllinesendwithlinefeed<lf>,andcarriagereturn<cr>characters.
RS‐232outputofweighingdata
ASCIIcode
AdjustableBaudrate,600,1200,2400,4800,9600and19200baud
8databits
NoParity
9pinD‐subminiaturesocket
Pin3Output
Pin2Input
Pin5SignalGround
© Adam Equipment Company 2011
DataFormat‐NormalOutput:
DataFormat‐MemoryRecallPrint:
<lf><cr>
<lf><cr>
DATE 12/09/2006
TIME 14:56
<lf><cr>
GROSS Wt 1.234 Kg NetWt.ifnetweightisdisplayed
Unit Wt. 123 g Kgorgformetricweightsandlbforpounds.
Pcs 10 pcs
<lf><cr> Includes2linefeedswithcarriagereturn
<lf><cr> atbeginningandendoftheform
<lf><cr>
DATE 12/09/2006
TIME 14:56
<lf><cr>
------------------
TOTAL
No. 5
Wt. 1.234 Kg
Pcs 10 pcs
<lf><cr>
Includes1linefeed
------------------
<lf><cr>
2linefeeds,carriagereturn
<lf><cr>
© Adam Equipment Company 2011
DataFormat‐ContinuousPrint:
Inotherlanguagestheformatisthesamebutthetextwillbeinthelanguageselected.
SeeSection9.2
Description ENGLISH FRENCH GERMAN SPANISH
Print gross
weight
Gross Wt Pds Brut Brut-Gew Pso Brut
Netweight
Net Wt. Pds Net Net-Gew Pso Net
Weight per
unitcounted
Unit Wt. Pds unit Gew/Einh Pso/Unid
Number of
items
counted
Pcs Pcs Stck. Piezas
Number of
weighings
added to
subtotals
No. Nb. Anzhl Num.
Total weight
and count
printed
Total Total Gesamt Total
Printdate
Date Date Datum Fecha
Printtime
Time Heure Zeit Hora
ST,GS, 0.9080 kg
U.W. 0.90798 g/pcs
PCS 1000 pcs
<lf><cr> Includes2linefeedswithcarriagereturn
<lf><cr> betweensetsofdata
/