11
IM 700938-01E
Specifications
Product Specifications
Compatible sensors 700937 Current Probe 700939FETProbe
701920 Differencial Probe 701930 Current Probe
701931 Current Probe 701932 Current Probe
701933 Current Probe
Number of power supply connectors
2
Output voltage ±12V±0.5V
Operating temperature and humidity range
0to40°C(32to104°F),80%RHorless(no
condensation)
Storage temperature and humidity range
-10to50°C(14to122°F),80%RHorless(no
condensation)
Location for use Indoor, altitude up to 2000 m (6562 feet)
Rated output current 600 mA (the total value of both outputs)
Rated supply voltage 100VAC(120,220,and240Vrequiresspecification)
(Voltagefluctuationof10%fromtheratedsupply
voltage are taken into account.)
Ripple supply voltage 3mVp-p(attheratedoutputcurrent)
Rated supply frequency 50/60Hz
Maximum rated power 60VA(attheratedoutputcurrent)
External dimensions Approx.73(W)×110(H)×186(D)mm
Approx.2.87”(W)×4.33”(H)×7.32”(D)
Mass Approx. 1.1 kg
Approx. 38.8 oz.
Accessories Power cord, Instruction manual,
Spare fuse (in the fuse holder)
100V,120V:F1.0AL/250V,20mm×5mmdia.
220V,240V:F0.5AL/250V,20mm×5mmdia.
Standards Applying
Safety EN61010-1
Pollution Degree 2
EMC Emission EN61326-1 Class B
EN55011 Class B, Group 1
EN6100-3-2 Class B
EN6100-3-3
EMC standards of Australia and New Zealand
EN55011 Class B, Group 1
Immunity EN61326-1 Table 1 (Basic immunity requirement)
Corrective Action to be Taken in the case of an Abnormality
• Ifanmessageappearsonthescreen,refertothefollowingpages.
• Ifmaintenanceserviceisrequired,oriftheinstrumentstilldoesnotoperate
properly even if the proper corrective action has been taken, contact your
nearest YOKOGAWA representatives, listed on the back cover of this manual.
Symptom Possible Cause Corrective Action
The power cannot be turned ON. The fuse is blown. Replace the fuse.