Steris Amsco Eagle 3000 Series Sterilizers Operating instructions

Type
Operating instructions
OPERATOR MANUAL
Amsco® Eagle® 3000
Small and Medium Steam Sterilizers
(04/15/09) P150829-001
This Page Blank
i
Introduction and Table of Contents Operator Manual 150829-001
A WORD FROM STERIS CORPORATION
This manual contains important information on proper use of this sterilizer. All
personnel involved in the use of this equipment must carefully review and
comply with the warnings, cautions and instructions contained in this
manual.
STERIS carries a complete line of accessories for this unit to simplify, organize
and assure sterility of the sterilization process. Instrument trays, pouches,
and biological/chemical monitoring systems are all available to fulfill your
facility's processing needs. A STERIS representative will gladly review these
with you.
A summary of the
Safety Precautions
to be observed when operating this
equipment can be found in
SECTION 1
of this manual. Do not operate the sterilizer
until you have become familiar with this information.
Any alteration of the sterilizer not authorized or performed by STERIS which
could affect its operation will void the warranty, could adversely affect
sterilization efficacy, and could violate federal, state, and local regulations and
jeopardize your insurance coverage.
©2009, STERIS Corporation. All rights reserved. Printed in U.S.A.
Advisory
A thorough preventive maintenance program is essential to safe and proper
sterilizer operation. You are encouraged to contact STERIS concerning our
Annual Maintenance Agreements. Under the terms of this agreement, preven-
tive maintenance, adjustments and replacement of worn parts are done on a
scheduled basis to assure equipment performance at peak capability and to
help avoid untimely or costly interruptions. STERIS maintains a global staff of
well equipped, factory-trained technicians to provide this service, as well as
expert repair services. Contact STERIS and refer to the
SERVICE AND M AINTENANCE
PROCEDURES
manual for details.
Service Information
Indications for Use
Amsco® Eagle® 3000 Small and Medium Sterilizers are designed for steriliza-
tion of certain materials used in healthcare facilities, and are available in two
configurations:
Gravity - designed for sterilization of non-porous, heat- and moisture-
stabile goods, and sterilization of items intended for immediate use.
Gravity sterilizer is equipped with a Gravity cycle.
Prevacuum - designed for fast, efficient sterilization of porous, heat- and
moisture-stabile materials, in addition to the same sterilization capabilities
as a Gravity sterilizer. Prevacuum unit is equipped with Prevacuum,
Gravity, Vacuum Leak Test, and Dart® (Daily Air Removal Test) cycles.
ii
150829-001 Operator Manual Introduction and Table of Contents
If there is any doubt about a specific material or product, contact the
manufacturer of the product for the recommended sterilization technique.
This sterilizer is not designed to process flammable liquids or compounds.
The Eagle 3000 Series Steam Sterilizer is available in the following configu-
rations:
Gravity Prevacuum
Series 3011 Series 3013
16x16x26" 16x16x26"
(406x406x660 mm) (406x406x660 mm)
Series 3021 Series 3023
20x20x38" 20x20x38"
(508x508x965 mm) (508x508x965 mm)
Series 3031 Series 3033
24x36x36" 24x36x36"
(610x914x914 mm) (610x914x914 mm)
Series 3041 Series 3043
24x36x48" 24x36x48"
(610x914x1219 mm) (610x914x1219 mm)
Series 3051 Series 3053
24x36x60" 24x36x60"
(610x914x1524 mm) (610x914x1524 mm)
STERIS offers the Amsco Eagle 3000 Small and Medium sterilizers with
multiple default cycle parameters (see next page). These represent the most
common cycles used in healthcare applications. STERIS provides these
cycles for ease of configuration. It is the user's responsibility to confirm
configuration prior to use and to validate efficacy by use of biological
indicators. Reference AAMI Guidelines/Standards for a guide to qualifying
sterilization cycles and to ensure that proper sterility assurance level (SAL) as
well as moisture retention acceptance criteria are met.
Manufactured by:
STERIS Mexico, S. de R.L. de C.V.
Avenida Avante 790
Parque Industrial Guadalupe
Guadalupe, Nuevo Leon, Mexico C.P. 67190
Sales and Service:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
440-354-2600 • 800-444-9009
www.steris.com
The base language of this document is
ENGLISH. Any translations must be
made from the base language
document.
iii
Introduction and Table of Contents Operator Manual 150829-001
SMALL GRAVITY STERILIZERS (Series 3011 and 3021):
Default Recommended Sterilize Sterilize Dry Time
Cycles Load Temp Time
FLASH Unwrapped instrument tray 270°F 3 minutes 1 minute
with a single instrument (132°C)
FLASH Unwrapped instrument tray 270°F 10 minutes 1 minute
with a single instrument (132°C)
GRAVITY Up to two double-wrapped 270°F 15 minutes 30 minutes
instrument trays, max (132°C)
weight 17 lbs (7.7 kg)
GRAVITY Fabric Packs 250°F 30 minutes 15 minutes
See Table 2
(121°C)
MEDIUM GRAVITY STERILIZERS (Series 3031, 3041, and 3051):
Default Recommended Sterilize Sterilize Dry Time
Cycles Load Temp Time
GRAVITY Full Load, Instrument Trays 270°F 15 minutes 30 minutes
See Table 2
(132°C)
GRAVITY Full Load, Fabric Packs 270°F 30 minutes 15 minutes
See Table 2
(132°C)
GRAVITY Full Load, Instrument Trays 250°F 30 minutes 30 minutes
See Table 2
(121°C)
GRAVITY Full Load, Fabric Packs 250°F 30 minutes 15 minutes
See Table 2
(121°C)
SMALL PREVACUUM STERILIZERS (Series 3013 and 3023):
Cycles Recommended Sterilize Sterilize Dry Time
Default Load Temp Time
FLASH Unwrapped instrument tray 270°F 3 minutes 1 minute
with a single instrument (132°C)
FLASH Unwrapped instrument tray 270°F 10 minutes 1 minute
with a single instrument (132°C)
EXPRESS Single-wrapped instrument 270°F 4 minutes 3 minutes
tray with a single (132°C)
instrument. Non-porous
goods only.
PREVAC Up to two double-wrapped 270°F 4 minutes 20 minutes1
instrument trays, max (132°C)
weight 17 lbs (7.7 kg)
Continued. . .
1
Five minute Dry Time can be used for processing a single fabric pack.
Table 1. Sterilizer Cycle Values
iv
150829-001 Operator Manual Introduction and Table of Contents
Table 2. Maximum Chamber Load by Sterilizer Chamber Size
Chamber Size Instrument Trays Fabric Packs
16 dia. x 26" 22
(406 dia. x 406 mm)
16 x 16 x 26" 24
(406 x 406 x 660 mm)
20 x 20 x 38" 26
(406 x 406 x 965 mm)
24 x 36 x 36" 918
(610 x 914 x 914 mm)
24 x 36 x 48" 12 30
(610 x 914 x 1219 mm)
24 x 36 x 60" 15 36
(610 x 914 x 1524 mm)
MEDIUM PREVACUUM STERILIZERS (Series 3033, 3043, and 3053):
Default Recommended Sterilize Sterilize Dry Time
Cycles Load Temp Time
PREVAC One Fabric Pack 270°F 4 minutes 5 minutes
(132°C)
PREVAC Double-wrapped instrument 270°F 4 minutes 20 minutes
trays max weight 17 lbs (132°C)
(7.7 kg). Fabric Packs
See Table 2
GRAVITY Full Load, Fabric Packs 250°F 30 minutes 15 minutes
See Table 2
(121°C)
GRAVITY Full Load, Fabric Packs 250°F 30 minutes 15 minutes
See Table 2
(121°C)
v
Introduction and Table of Contents Operator Manual 150829-001
TABLE OF CONTENTS
Section Title Page
1 Listing of Safety Precautions ....................................................... 1-1
2 Component Identification .............................................................. 2-1
2.1 Printer ............................................................................................................................ 2-2
2.2 Optional Integral Steam Generator .............................................................................. 2-3
2.3 Operating End Control Panel........................................................................................ 2-4
2.4 Cycle Selector Touch Pads .......................................................................................... 2-5
2.5 Reset Function ............................................................................................................. 2-6
2.6 VALUES Touch Pads .................................................................................................... 2-6
2.7 Non-Operating End Control Panel (Double-Door Sterilizers) ....................................... 2-6
2.8 Manual Door Operation ................................................................................................. 2-6
2.9 Power Door Control Touch Pads (Optional) ................................................................. 2-7
2.10 Optional Power Door Operation .................................................................................. 2-7
2.11 Manual Operation of Optional Power Door ................................................................ 2-7
2.12 Information Regarding Fuses ..................................................................................... 2-8
3 Techniques of Sterilization ........................................................... 3-1
3.1 Special Information Regarding the Express Cycle ..................................................... 3-2
3.2 Special Information Regarding the Flash Cycle .......................................................... 3-3
3.3 Preparing Instruments for Flash Cycle Sterilization .................................................... 3-3
3.4 Control Measures For Verifying Sterilization Process ................................................. 3-4
3.4.1 Biological Monitors .............................................................................................. 3-4
3.4.2 Testing for Prevacuum Efficiency ......................................................................... 3-4
3.5 Dart (Bowie-Dick) Test .................................................................................................. 3-4
3.6 Vacuum Leak Test ........................................................................................................ 3-5
3.7 Recommendations for the Sterilization Process.......................................................... 3-5
4 Cycles and Cycle Values ............................................................... 4-1
5 Sterilizer Operation ........................................................................ 5-1
5.1 Before Operating the Sterilizer.................................................................................... 5-2
5.2 Preparing Loads for Sterilization Cycles ..................................................................... 5-4
5.3 Guidelines for Placement of Various Loads ................................................................ 5-4
5.4 Unloading the Sterilizer ................................................................................................ 5-5
5.5 Loading Car Instructions: Loading .............................................................................. 5-6
5.6 Loading Car Instructions: Unloading ........................................................................... 5-6
5.7 Loading/Unloading Sterilizer: Rack and Shelves ........................................................ 5-7
5.8 Automatic Operation: Gravity Cycle ........................................................................... 5-8
5.9 Automatic Operation: Prevacuum Cycle ................................................................... 5-10
5.10 Automatic Operation: Flash Cycle .......................................................................... 5-12
5.11 Automatic Operation: Express Cycle ...................................................................... 5-14
5.12 Vacuum Leak Test (Prevacuum Units Only) ............................................................ 5-16
5.13 DART (Bowie-Dick) Test (Prevacuum Units Only) ................................................... 5-18
6 Appendix Liquids Cycle ................................................................ 6-1
6.1 Automatic Operation: Liquids Cycle ........................................................................... 6-1
vi
150829-001 Operator Manual Introduction and Table of Contents
LIST OF FIGURES
Figure 2-1. Amsco® Eagle® 3000 Smal and Medium Sterilizers ................................................ 2-1
Figure 2-2. Optional Integral Steam Generator ........................................................................... 2-3
Figure 2-3. Control Panel ............................................................................................................. 2-4
Figure 2-4. Non-Operating End Control Panel............................................................................. 2-5
Figure 2-5. Power Door Operation ............................................................................................... 2-7
Figure 5-1. Control Locations ...................................................................................................... 5-1
Figure 5-2. Chamber Rack and Shelves ..................................................................................... 5-7
Figure 5-3. Typical Printout – Gravity Cycle ............................................................................... 5-8
Figure 5-4. Typical Printout – Prevacuum Cycle ........................................................................ 5-10
Figure 5-5. Typical Printout Flash Cycle ................................................................................. 5-12
Figure 5-6. Typical Printout Express Cycle............................................................................. 5-14
Figure 5-7. Typical Printout – Leak Test .................................................................................... 5-17
Figure 6-1. Open Door One Inch (25.4 mm) ............................................................................... 6-2
Figure 6-2. Typical Printout – Liquid Cycle.................................................................................. 6-3
Description Page
LIST OF TABLES
Table 1. Sterilizer Cycle Values ...................................................................................................... v
Table 2. Maximum Chamber Load by Sterilizer Chamber Size ..................................................... iv
Table 4-1. Sterilizer Cycle Values .............................................................................................. 4-1
Table 4- 2. Maximum Chamber Load by Sterilizer Chamber Size ............................................. 4-2
Table 6-1. Minimum Recommended Exposure Times for
Full Load of Autopour® Flasked Solutions Liquid Cycles Only ...................................... 6-1
Description Page
1-1
Listing of Safety Precautions Operator Manual 150829-001
The following
Safety Precautions
must be observed when operating or servicing this sterilizer. WARNING indicates the
potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain
Safety Precautions are repeated throughout the manual. It is important to review ALL
Safety Precautions
before
operating or servicing the unit.
Strictly following these
Safety Precautions
enhances your ability to safely and effectively utilize the unit and helps the
customer avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these
Safety Precautions
are not exhaustive; customers are encouraged to develop their own safety
policies and procedures to enhance and complement these
Safety Precautions
.
WARNING ELECTRIC SHOCK AND BURN HAZARD:
Disconnect all utilities to sterilizer before servicing. Do not service the sterilizer unless all utilities have been
properly locked out. Always follow OSHA Lockout-Tagout and electrical safety-related work practice standards.
(See CFR 1910.147 and .331 through .335.)
WARNING BURN HAZARD:
Sterilizer, rack/shelves and loading car will be HOT after cycle is run. Always wear protective gloves and apron
(also face shield if processing liquids) when removing a processed load. Protective gloves and apron should
also be worn when reloading sterilizer following previous operation.
Steam may be released from the chamber when door is opened. Step back from the sterilizer each time the door
is opened to minimize contact with steam vapor.
A steam supply malfunction, identified by an audible (warble tone) and visual alarm, can cause the sterilizer
chamber to fill with scalding water.
• Do not open the door.
Call for service immediately.
The water sensing system will not function if electrical power is not available (display is blank). During this time
a steam supply malfunction can cause the chamber to fill with scalding water. If the unit is operated manually:
• Unlock the door slowly.
• Re-tighten door if water leaks past the door gasket and call for service immediately. Do not attempt to operate
this sterilizer until repairs have been completed by a skilled, trained service technician.
It is inappropriate for a healthcare facility to sterilize liquids for direct patient contact.
After manual exhaust, steam may remain inside the chamber. Always wear protective gloves, apron and a face
shield when following emergency procedure to unload the sterilizer. Stay as far back from the chamber as
possible when opening the door.
When sterilizing liquids, to prevent personal injury or property damage resulting from bursting bottles and hot fluid,
you must observe the following procedure:
Use Liquid cycle only. No other cycle is safe for processing liquids.
Use only vented closures. Do not use screw caps or rubber stoppers with crimped seal.
Use only Type I borosilicate glass bottles. Do not use ordinary glass bottles or any container not designed
for sterilization.
Avoid sudden full opening of door at end of cycle. Open sterilizer door no more than one inch and wait at
least 10 minutes before unloading sterilizer.
Do not allow hot bottles to be jolted. This can cause hot-bottle explosions! Do not move bottles if any
boiling or bubbling is present.
Allow bottles to cool to touch before attempting to move them from sterilizer shelf or tray(s) to the storage
area..
..
.
1
LISTING OF SAFETY PRECAUTIONS
1-2
150829-001 Operator Manual Listing of Safety Precautions
WARNING BURN HAZARD (Continued):
It is inappropriate for a healthcare facility to sterilize liquids for patient contact.
Do not process flammable liquids.
WARNING - EXPLOSION HAZARD:
Do not process flammable liquids.
This sterilizer is not designed to process flammable liquids or compounds.
WARNING SLIPPING HAZARD:
To avoid slippery floor conditions, immediately wipe up any spilled liquids or condensation in sterilizer loading
or unloading area(s).
WARNING STERILITY ASSURANCE HAZARD:
Load sterility may be compromised if the biological indicator or air leak test indicates a potential problem. If these
indicators show a potential problem, refer the situation to a qualified service technician before using the sterilizer
further.
According to AAMI standards, a measured leak rate greater than one mm Hg/minute indicates a problem with
the sterilizer. Refer the situation to a qualified service technician before using the sterilizer further.
It is inappropriate to sterilize liquids for direct patient contact.
The Flash Cycle is intended to process unwrapped items intended for immediate use in the Eagle 3000 Series
Small Sterilizers. The Flash Cycle is not intended for use in the Eagle 3000 Series Medium Sterilizers.
The Flash Cycle is not intended for processing porous items.
The Express Cycle is intended to process unwrapped items intended for immediate use in the Eagle 3000 Series
Small Sterilizers. The Express Cycle is not intended for use in the Eagle 3000 Series Medium Sterilizers.
The Express Cycle is only intended for use with single-wrapped instrument tray with a single instrument in the
Eagle 3000 Series Small Sterilizers. The Express Cycle is not intended for use in the Eagle 3000 Series Medium
Sterilizers.
The Express Cycle is not intended for processing porous items (except tray wrappers).
WARNING PERSONAL INJURY HAZARD:
When closing the chamber door, keep hands and arms out of the door and ensure opening is clear of
obstructions.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment.
Contact your STERIS service representative to schedule preventive maintenance.
Repairs and adjustments to this equipment must be made only by fully qualified service personnel. Maintenance
performed by inexperienced, unqualified persons or installation of unauthorized parts could cause personal
injury or result in costly equipment damage.
1-3
Listing of Safety Precautions Operator Manual 150829-001
CAUTION POSSIBLE EQUIPMENT DAMAGE:
On medium sterilizers equipped with power door option, ensure shift handle is in either the extreme right (POWER)
or left (MANUAL) position at all times. Placing handle in an intermediate position could damage drive system of
door.
Never use a wire brush, abrasives or steel wool on door and chamber assembly. Do not use cleaners containing
chloride on stainless-steel surfaces. Chloride-based cleaners will deteriorate stainless steel, eventually leading
to failure of the vessel.
Do not use cleaners containing chlorides on loading cars. Chloride-based cleaners will deteriorate the loading
car metal.
Sterilization of chloride-containing solutions (e.g., saline) can cause chamber corrosion and is not recommended
by the manufacturer. If, however, chloride-containing solutions must be processed, clean the chamber after each
use.
1-4
150829-001 Operator Manual Listing of Safety Precautions
Symbol Definition
Transfer of Heat, Hot Surface
Steam Energy
Sterilizer Door Open/Close
Protective Earth (Ground)
Electrostatic Sensitive Device
Locked Cycle or Setup Options
Unlocked Cycle or Setup Options
Attention, Consult Manual for Further Instructions
A Amperage Rating of the unit
V Voltage Rating of the unit
~ Alternating Current
kW Power Rating of the unit
Hz Frequency of the unit
fPhase of the unit
Definition of
Symbols
2-1
Component Identification Operator Manual 150829-001
Figure 2-1. Amsco® Eagle® 3000 Small and Medium Steam Sterilizers
Medium
Sterilizer
Small
Sterilizer
STATUS
1 2 3 4 RESET
CYCLE
CYCLE
Display Window
Control Panel
Cycle Selector
Touch Pads
Lower
Access Door
(Storage Rack
for Printed
Materials on
Inside)
Water Supply
Steam Generator
(Electric Models
Only)
Chamber
Door
HI-LO
Steam
Control
Main
Power
Switch
Chamber
Pressure/
Vacuum
Gauge
Jacket
Pressure
Gauge
Water
Supply
Valve
Lower
Access Door
(Storage Rack
for Printed
Materials on
Inside)
Main
Power
Switch
Chamber
Door
Jacket
Pressure
Gauge
Chamber
Pressure/
Vacuum
Gauge
Upper
Access
Door
HI-LO
Steam
Control
Steam
Supply
Valve
Water
Supply
Valve
Steam
Supply
Valve
Printer
Door (Lowered
Position)
Cycle Printout
Stage 3
Control Column
Upper
Access
Door
Power
Switch
Printer
Function
Switch
2
COMPONENT IDENTIFICATION
2-2
150829-001 Operator Manual Component Identification
Become familiar with all control locations and functions before operating the
sterilizer (see Fig. 2-1).
Main Power Switch - -
- -
- behind upper access panel on small units and
behind lower access panel for medium units. Provides power to control
column. This switch should be left on at all times. (This switch does not
control power to optional electric steam generator.)
Sterilizer Control Power Switch - located behind the printer door.
This switch is used to switch the sterilizer control between Standby
and Ready conditions (Standby mode is usually used at night when the
sterilizer is not being operated). It also controls the optional integral
steam generator.
Located behind the upper access panel are the following:
Water Supply Valve - provides cold water supply to sterilizer; must
be open for sterilizer operation.
Steam Supply Valve - provides steam supply to sterilizer; must be
open for sterilizer operation.
HI-LO Steam Control Valve - regulates steam pressure; factory set
at: HI - for 270°F (132°C) cycles; LO - for 250°F (121°C) cycles.
Chamber Pressure/Vacuum And Jacket Pressure Gauges - show
respective steam pressure and chamber vacuum levels.
Power Door Control Touch Pads - will only be found on units
equipped with a power door.
Refer to Figure 2-3
. Printer records all cycle data on 2-1/4 inch wide, single-ply
thermal paper. See Service and Maintenance Procedures (P129362-976) for
paper ordering information and for paper changing procedure.
Printer Function Switch
(located behind printer door) controls two printer
functions - print and print values:
»
Print Press top of Printer Function switch to obtain a complete
duplicate printout of the last previous cycle (when unit is not in cycle)
or to obtain a printout of current cycle phase and conditions (when unit
is in cycle).
»
Print Values Press bottom of Printer Function switch to obtain a
printout of all currently set cycles and cycle values. (Press only when
unit is not in cycle.)
2.1 Printer
2-3
Component Identification Operator Manual 150829-001
Figure 2-2. Optional Integral Steam Generator
Generator Water
Supply Valve
The optional integral steam generator (if present) is located in the space below
the sterilizer chamber (on Small Series Sterilizers only). See
SERVICE AND
MAINTENANCE P ROCEDURES
(P129362-976).
Generator Water Supply Valve this valve should always be open. Ensure
that it is open before attempting to operate the sterilizer.
Generator Drain Valve this should be opened once a day, while the
generator is at room temperature, to flush the generator of residual solids
(present in most water) that may have accumulated in the generator's boiler.
After the boiler has been flushed ensure the drain valve is closed. Refer to
instructions in
SERVICE AND M AINTENANCE P ROCEDURES
(P129362-976).
2.2 Optional Integral
Steam Generator
2-4
150829-001 Operator Manual Component Identification
Refer to Figure 2-3.
The Operating (load) End Control panel is used to start
cycles, set cycles and cycle values, and obtain status printouts. The operating
end has:
Display Window - displays cycle status and control messages on a 2-line
x 20-character display.
Cycle Selector Touch Pads 1, 2, 3, and 4 - for selecting and starting
preset cycles; accessible when sliding door is in lowered position.
RESET Touch Pad - for resetting control, and terminating cycles quickly;
accessible when sliding door is in lowered position.
VALUES Touch Pads - for changing cycles and cycle values; accessible
when sliding door is in raised position.
Display Window
Control Panel
Door (Lowered
Position)
Cycle Printout
RESET
Touch Pad
Cycle Selector
Touch Pads
Printer
Printer
Function
Switch
STATUS
1 2 3 4 RESET
CYCLE
Power
Switch
2.3 Operating End
Control Panel
When not in cycle and sterilizer is not in the Ready mode, the display shows a
brief status of the sterilizer:
… DOOR UNLOCKED, or
… NOE (Non-Operating End) DOOR UNLOCKED, or
… BOTH DOORS UNLOCKED
Figure 2-3. Control Panel
2-5
Component Identification Operator Manual 150829-001
STATUS
STATUS
When not in cycle and sterilizer is in the Ready mode, the display alternately
shows available cycle selections … (example):
1. PREVAC 3. GRAVITY
2. PREVAC 4. GRAVITY
and
TIME T= XXX
(TIME OF DAY) P= XXX
When cycle is in progress, display shows cycle status and information. In the
event of an abnormal condition the display and printer will alert the operator. See
Section 5 of the Service and Maintenance Procedures (P129362-976) for a full
list of alarm conditions.
A sterilizer equipped with two doors will also be equipped with two control
panels. The control panel at the loading door of the sterilizer is referred to as
the “Operating End Control”. The control panel located at the unloading door
is referred to as the “Non-Operating End Control”. A single door sterilizer is
equipped with an “Operating End Control” only.
NOTE: Except for the presence of the printer (which is only present at the
Operating End Control of the unit), control panels at both ends of the unit are
similar and each can be used to start or abort the cycle.
Non-Operating
End Control
Panel
Chamber
Pressure
Gauge
Display Window
Figure 2-4. Non-Operating End Control Panel
Cycle
Selector
Touch
Pads
Four touch pads, numbered 1 through 4, are located below the display window.
These touch pads are used to select the cycle to be run. The type of cycle that
can be selected by the touch pad is shown on the display, next to the number
on the pad. For example, 1=PREVAC (on the display) means that by pressing
the #1 touch pad a Prevac cycle can be selected.
Pressing a touch pad once selects the cycle, pressing the same touch pad
twice, within five seconds, confirms selection and starts the cycle.
2.4 Cycle Selector
Touch Pads
2-6
150829-001 Operator Manual Component Identification
The RESET touch pad is used to abort a cycle and put the control into a known
condition. See Figure 2-3 for location of this touch pad.
If you push the wrong cycle selector touch pad when starting a cycle, proceed
as follows:
1. Press RESET touch pad to abort cycle. Display will show … ABORT ... and
the printer will record time RESET touch pad was pressed and will print
ABORT ... (time of day).
2. Wait until display shows … time of day.
3. Press touch pad for correct cycle and printer will record cycle selected and
proceed through correct cycle.
These touch pads are accessed by raising the sliding door located beneath the
display window and cycle selector touch pads. The VALUES touch pads are
used for changing the operating values used by cycles, changing the cycles
displayed on the main cycle selection menu, and for changing the operating
settings of the sterilizer. For more details on using these keys see Section 4 of
the Service and Maintenance Procedures (P129362-976).
The non-operating (unload) end control panel features four cycle selector touch
pads and RESET touch pad. Display window concurrently shows the same
message as shown in display window on the operating end of the sterilizer.
Cycle values cannot be changed from the non-operating end.
Cycles can be started or reset using the four cycle selector and
RESET touch pads.
Cycle status and control messages are concurrently shown on a 2-line
x 20-character display.
Medium sterilizers also have control of optional power door.
A handwheel operates the manual door for locking or unlocking. Door can be
swung open or closed from the unlocked position.
To open door from locked position rotate handwheel counterclock-
wise as far as it will go using normal hand pressure and swing door
open by hand.
To close and lock door from open position swing door closed by
hand and rotate handwheel clockwise as far as it will go using normal
hand pressure.
NOTE: Should the door not at first unlock, turn door wheel slightly clockwise,
press the center black button, then again turn door wheel counterclockwise
and open. Turn handwheel to extreme left to bring ends of holding arms inward
so that they will not strike door frame when door is opened or closed.
2.6 VALUES Touch
Pads
2.7 Non-Operating
End Control Panel
(Double-Door
Sterilizers)
2.5 Reset Function
2.8 Manual Door
Operation
2-7
Component Identification Operator Manual 150829-001
These touch pads are only visible on the control when sterilizer is equipped with
a power door. Note also that when the sliding door is in the up position, when
accessing the VALUES touch pads, the power door control pads will be
covered by the sliding door. For more details about using these pads, refer to
SECTION 2.10, OPTIONAL P OWER D OOR O PERATION
.
NOTE: Door motion stops whenever a resistance is encountered and motion
resumes when resistance is removed.
Door movement on sterilizers equipped with power door(s) is controlled by four
touch pads located on the control panel (four touch pads are also located on
the non-operating end control panel of double door units). Touch pads are
labeled CLOSE, LOCK, UNLOCK/OPEN and STOP (refer to Fig. 2-5) and
perform the following:
• CLOSE - press to swing the door closed.
• LOCK
- press to lock door (from closed position) or to close and lock
door (from open position).
• UNLOCK/OPEN - to open door from the locked position, press once
to unlock the door and press again to open the door.
•STOP
- press to stop movement of door at any time. Press the desired
movement control to resume operation of the door.
Power doors may be manually operated as follows:
1. Pull shift handle to left to engage handwheel (see Fig. 2-5). If handwheel
does not engage, turn it slowly while maintaining a slight pressure to the left
on the shift handle.
2. Operate door as described in
SECTION 2.8, MANUAL D OOR OPERATION.
NOTE: To return to power operation, push shift handle back to the right while
pressing the LOCK or UNLOCK/OPEN touch pad to ease engagement of gears.
2.9 Power Door
Control Touch Pads
(Optional)
2.10 Optional Power
Door Operation
2.11 Manual
Operation of
Optional Power Door
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: On
medium sterilizers equipped
with power door option, en-
sure shift handle is in either
the extreme right (POWER)
or left (MANUAL) position at
all times. Placing handle in
an intermediate position
could damage drive system
of door.
Figure 2-5. Power Door Operation
2-8
150829-001 Operator Manual Component Identification
2.12 Information
Regarding Fuses
This unit contains no operator replaceable fuses. Refer to
ILLUSTRATED P ARTS
BREAKDOWN
(P129362-978) and
SERVICE AND MAINTENANCE PROCUEDURES
(P129362-976) for further information.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52

Steris Amsco Eagle 3000 Series Sterilizers Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI