Steris Amsco Century Small Steam Sterilizer Operating instructions

Type
Operating instructions
OPERATOR MANUAL
International Hospital
Amsco® Century®
Prevacuum Sterilizers
(REV AC)P150829-026
Table of Contents Operator Manual 150829-026
A WORD FROM STERIS CORPORATION
This manual contains important information on proper use of this
sterilizer. All operators and department heads are urged to carefully
review and become familiar with the warnings, cautions, and
instructions contained herein.
This sterilizer is specifically designed to process goods using only
the cycles as specified in this manual. If there is any doubt about
a specific material or product, contact the manufacturer of the
product for the recommended sterilization technique.
STERIS carries a complete line of accessories for this unit to
simplify, organize, and assure sterility of the sterilization process.
Instrument trays, pouches, and biological/chemical monitoring
systems are all available to fulfill your facility’s processing needs.
A STERIS representative will gladly review these with you.
A summary of the
Safety Precautions
to be observed when operating
and servicing this equipment can be found in Section 1 of this
manual. Do not operate or service the equipment until you have
become familiar with this information.
Any alteration of the sterilizer not authorized or performed by STERIS
which could affect its operation will void the warranty, could adversely
affect sterilization efficacy, could violate national, state and local
regulations, and could jeopardize your insurance coverage.
The Amsco® Century® Sterilizer is designed for sterilization of certain
materials used in healthcare facilities and available as:
Prevacuum - designed for sterilization of heat and moisture-
stabile materials. Unit provides express processing for a
single instrument in a single wrapped tray (not for porous
items, except for tray wrapper). Prevacuum sterilizer is
equipped with Prevacuum, Flash, Express, Leak Test, and
Bowie-Dick cycles.
This sterilizer is specifically designed to only process goods using
the cycles as specified in this manual. If there is any doubt about
a specific material or product, contact the manufacturer of that
product for the recommended sterilization technique.
The Amsco Century sterilizer is offered in the following
configurations:
406mm x 406 mm x 660mm (16"x 16"x 26") Single Door Prevacuum
" " Double Door Prevacuum
508 mm x 508 mm x 965 mm (20"x 20"x 38" ) Single Door Prevacuum
" " Double Door Prevacuum
Amsco Century sterilizers are offered for sale with the factory set
sterilization cycles and cycle values shown in Table 6-1 of Section 6.
i
Indications for Use
Advisory
© 2019 - STERIS Corporation. All rights reserved.
150829-026 Operator Manual Table of Contents
ii
A thorough preventive maintenance program is essential to safe
and proper sterilizer operation. This manual contains maintenance
schedules and procedures which should be followed for satisfac-
tory equipment performance.
You are encouraged to contact STERIS concerning our comprehen-
sive annual maintenance agreements. Under terms of the agree-
ments, preventive maintenance, adjustments, and replacement of
worn parts are done on a scheduled basis to help assure optimal
equipment performance and to help avoid untimely or costly
interruptions. STERIS maintains a worldwide staff of well-equipped,
factory-trained technicians to provide this service, as well as expert
repair services. Contact STERIS for details.
Service Information
Manufactured by:
STERIS Mexico S. de R.L. de C.V.
Avenida Avante 790
Parque Industrial Guadalupe
Guadalupe, Nuevo Leon, Mexico C.P. 67190
Sales and Service:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
440-354-2600800-444-9009
www.steris.com
STERIS Limited
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865 Ireland
The base language of this document is
ENGLISH. Any translations must be
made from the base language
document.
2797
Table of Contents Operator Manual 150829-026
iii
A WORD FROM STERIS CORPORATION ............................... i
1 LISTING OF SAFETY PRECAUTIONS ................................ 1-1
2 INSTALLATION VERIFICATION......................................... 2-1
2.1 Installation Checklist .......................................................................................... 2-1
2.1.1 Service Checklist ......................................................................................... 2-1
2.1.2 Plumbing Services ....................................................................................... 2-1
2.1.3 Electrical Service ......................................................................................... 2-2
2.1.4 Sterilizer Final Check ................................................................................... 2-2
2.1.5 Cycle Operation ........................................................................................... 2-2
2.2 Technical Specifications .................................................................................... 2-2
2.2.1 Overall Size ................................................................................................. 2-2
2.2.2 Weight .......................................................................................................... 2-2
2.2.3 Utility Requirements ..................................................................................... 2-3
2.2.4 Environment Conditions ............................................................................... 2-3
3 STERILIZATION TECHNIQUES ......................................... 3-1
3.1 General .............................................................................................................. 3-1
3.2 Special Information Regarding the Express Cycle .............................................. 3-2
3.2.1 Preparing Instruments for Express Cycle Sterilization.................................. 3-2
3.2.2 Unloading ..................................................................................................... 3-3
3.3 Control Measures For Verifying Sterilization Process ......................................... 3-3
3.3.1 Biological Monitors ....................................................................................... 3-3
3.3.2 Testing for Prevacuum Efficiency ................................................................ 3-4
3.4 Bowie-Dick Test ................................................................................................. 3-4
3.5 Vacuum Leak Test ............................................................................................. 3-4
3.6 Recommendations for the Sterilization Process ................................................. 3-5
4 COMPONENT IDENTIFICATION ........................................ 4-1
4.1 General .............................................................................................................. 4-2
4.2 Main Sterilizer and Cycle Controls...................................................................... 4-2
4.3 Control Displays ................................................................................................. 4-3
4.4 Alarm Displays................................................................................................... 4-4
4.5 Operating End Control Panel .............................................................................. 4-4
4.6 Cycle Selection Touch-Screen Pads .................................................................. 4-6
4.6.1 Values Touch-Screen Pads.......................................................................... 4-6
4.6.2 Abort Touch -Screen Pad ............................................................................. 4-7
4.7 Emergency Stop Switch ..................................................................................... 4-8
4.7.1 Emergency Stop Key ................................................................................... 4-9
4.7.2 Emergency Stop Guard................................................................................ 4-9
4.7.3 Emergency Stop Label ................................................................................. 4-9
4.8 Front Access Panel Key Latch........................................................................... 4-9
4.9 Operator Troubleshooting (Single Door Sterilizer) ............................................. 4-10
4.10 Operator Troubleshooting (Double Door Sterilizer) ........................................... 4-11
4.11 Printer ............................................................................................................ 4-12
4.12 Printouts......................................................................................................... 4-13
4.13 Power Door Operation ..................................................................................... 4-14
TABLE OF CONTENTS
Page
Title
Section
150829-026 Operator Manual Table of Contents
iv
4.14 Emergency Door Opening Procedure.............................................................. 4-15
4.15 Optional Electric Steam Generator ................................................................. 4-16
4.16 Accessories ................................................................................................... 4-16
5 STERILIZER OPERATION ................................................. 5-1
5.1 Before Operating the Sterilizer............................................................................ 5-1
5.2 Preparing Loads for Sterilization Cycles ............................................................. 5-3
5.3 Guidelines for Placement of Various Loads ........................................................ 5-3
5.4 Unloading the Sterilizer....................................................................................... 5-4
5.5 Loading Car Instructions: Loading ...................................................................... 5-5
5.6 Loading Car Instructions: Unloading ................................................................... 5-6
5.7 Loading/Unloading Sterilizer Equipped With Rack and Shelves .......................... 5-6
5.8 Sterilizer (Factory) Cycle Settings ...................................................................... 5-6
5.9 134°C Flash Cycle.............................................................................................. 5-7
5.10 134°C Express Cycle ..................................................................................... 5-10
5.11 134° & 121° C Prevac Cycle........................................................................... 5-13
5.12 Bowie-Dick Test ............................................................................................. 5-16
5.13 Vacuum Leak Test ......................................................................................... 5-19
5.14 Aborting Cycles .............................................................................................. 5-21
5.15 Cycle Graphs ................................................................................................. 5-22
6 CYCLE AND CONTROL VALUE PROGRAMMING ................ 6-1
6.1 Change Values................................................................................................... 6-1
6.2 Cycle Values...................................................................................................... 6-2
6.3 Change Values................................................................................................... 6-3
6.4 Change Cycle Values ......................................................................................... 6-4
6.4.1 Overview...................................................................................................... 6-4
6.4.2 Step by Step ................................................................................................ 6-4
6.5 Changing Cycle Name, Sterilize Time, and Dry Time ......................................... 6-7
6.6 Change Time and Date....................................................................................... 6-8
6.7 Change Machine Setup .................................................................................... 6-10
6.8 Access Code.................................................................................................... 6-11
6.9 Lockout ............................................................................................................ 6-13
6.10 Utilities Control ............................................................................................... 6-15
6.11 Language ........................................................................................................ 6-18
6.12 Machine Number ............................................................................................ 6-19
6.13 Time Format................................................................................................... 6-20
6.14 Print Format ................................................................................................... 6-21
6.15 Audible Signals .............................................................................................. 6-22
6.16 Units............................................................................................................... 6-23
6.17 Date Format ................................................................................................... 6-23
6.18 Duplicate Print................................................................................................ 6-24
6.19 Leaving Change Values.................................................................................. 6-24
TABLE OF CONTENTS (Cont'd)
Title
Section Page
Table of Contents Operator Manual 150829-026
3-1 Cycle Validation Standards ....................................................................................................... 3-1
3-2 Item Processing Guidelines For Express Cycle ........................................................................ 3-2
4-1 Operator Troubleshooting – Emergency Stop Switch .............................................................. 4-11
4-2 Required Feed Water Quality for Carbon Steel Steam Generators .......................................... 4-14
4-3 Loading Equipment and Accessories for Century Sterilizers .................................................. 4-14
6-1 Cycle Default Values ................................................................................................................. 6-1
6-2 Factory-Set Cycles and Cycle Values ....................................................................................... 6-2
6-3 Change Machine Setup .......................................................................................................... 6-10
7-1 Preventive Maintenance Schedule for Century Sterilizer ........................................................... 7-1
7-2 Preventive Maintenance Schedule for Optional Electric Steam Generator ................................. 7-3
9-1 Replacement Parts ................................................................................................................... 9-5
7 ROUTINE MAINTENANCE ................................................. 7-1
7.1 Preventive Maintenance Guide........................................................................... 7-1
7.2 Daily Maintenance.............................................................................................. 7-4
7.2.1 Flush Optional Electric Steam Generator ..................................................... 7-4
7.2.2 Clean Chamber Drain Strainer ...................................................................... 7-6
7.2.3 Clean Chamber ............................................................................................ 7-6
7.3 Weekly Maintenance .......................................................................................... 7-7
7.4 As Required ....................................................................................................... 7-8
7.4.1 Change Printer Paper Roll ............................................................................ 7-8
7.4.2 Change Printer Ink Cartridge ...................................................................... 7-10
7.4.3 Replace Door Seal ..................................................................................... 7-12
8 TROUBLESHOOTING ........................................................ 8-1
8.1 General .............................................................................................................. 8-1
8.1.1 Typical Alarm Screen................................................................................... 8-1
8.1.2 Typical Alarm Printout.................................................................................. 8-2
8.2 In-Cycle Alarms.................................................................................................. 8-2
8.3 Out-of-Cycle Alarms ........................................................................................... 8-9
8.4 Sensor Alarms ................................................................................................. 8-12
9 SERVICE PROCEDURES .................................................... 9-1
9.1 Air Filter Replacement ........................................................................................ 9-1
9.2 Clean Strainers ................................................................................................... 9-1
9.3 Steam Trap Replacement................................................................................... 9-3
9.4 Clean or Replace Check Valves ......................................................................... 9-3
9.5 Clean or Replace Manifold Check Valves ........................................................... 9-3
9.6 Rebuild Manifold Solenoid Valves ...................................................................... 9-4
9.7 Safety Valve Test .............................................................................................. 9-4
9.8 Pressure Regulating Valve Adjustment .............................................................. 9-4
9.9 Recommended Spare Parts................................................................................ 9-4
9.10 Waste Products Disposal ................................................................................. 9-6
v
LIST OF TABLES
Page
Title
Table
TABLE OF CONTENTS (Cont'd)
Title
Section Page
150829-026 Operator Manual Table of Contents
vi
Healthcare
Capital Equipment
1-1
Listing of Safety Precautions Operator Manual 150829-026
Following is a list of the
Safety Precautions
which must be observed when operating this equipment. WARNINGS
indicate the potential for danger to personnel, and CAUTIONS indicate the potential for damage to equipment. These
precautions are repeated (in whole or in part), where applicable, throughout the manual. This is a listing of all
Safety
Precautions
appearing in the manual. Carefully read them before proceeding to use or service the unit.
WARNING - SHOCK AND BURN HAZARD:
Disconnect all utilities to sterilizer before servicing. Do not service the sterilizer unless all utilities have been
properly locked out. Always follow appropriate Lockout-Tagout and electrical safety-related work practice
standards.
Sterilizer, rack/shelves, and loading car will be hot after cycle is run. Always wear protective gloves and
apron when removing a processed load. Protective gloves and apron must be worn when reloading sterilizer
following the previous operation.
Before daily flushing of the generator, generator must be at 0.0 psig and cooled to room temperature.
Do not attempt to open the sterilizer door if a WATER IN CHAMBER ALARM condition exists. Call a qualified
service technician before attempting to use sterilizer further.
Sterilizer operator may be severely burned by scalding water if the water level control malfunctions. The
steam generator level control may malfunction if the supply water exceeds 26,000 ohms/cm (38.5 microhms
conductivity min.). Do not connect to treated water (e.g., distilled, reverse osmosis, deionized) unless water
resistivity is determined to be acceptable. If water exceeds 26,000 ohms/cm, contact STERIS for information
concerning modifications required to the generator control system.
After manual exhaust, steam may remain inside the chamber. Always wear protective gloves, apron and a
face shield when following emergency procedure to unload sterilizer. Stay as far back from the chamber opening
as possible when opening the door.
Sterilizer and shelves will be hot after cycle is run. Protective gloves and apron must be worn when reloading
sterilizer following the previous operation.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment.
Contact STERIS to schedule preventive maintenance.
Allow sterilizer and generator (if applicable) to cool to room temperature before performing any cleaning
or maintenance procedures.
Failure to shut off the steam supply when cleaning or replacing strainers can result in serious injury.
Jacket pressure must be 0.0 psig before beginning work on the steam trap.
Proper testing of the safety valve requires the valve to be operated under pressure. Exhaust from the safety
valve is hot and can cause burns. Proper safety attire (gloves, eye protection, insulated overall) as designated
by OSHA, is required. Testing is to be performed by qualified service personnel only.
WARNING - EXPLOSION HAZARD:
This sterilizer is not designed to process flammable compounds.
LISTING OF SAFETY PRECAUTIONS AND SYMBOLS 1
1-2
150829-026 Operator Manual Listing of Safety Precautions
WARNING - SLIPPING HAZARD:
To avoid slippery floor conditions, immediately wipe up any spilled liquids or condensation in sterilizer loading
or unloading area.
WARNING - PERSONAL INJURY HAZARD:
When closing the chamber door, keep hands and arms out of the door opening and ensure opening is clear
of obstructions.
In case of any emergency situation, press the Emergency Stop Button to halt sterilizer operation and contact
a qualified service technician to restart sterilizer to normal operation.
The door safety mechanism is not functional during manual operation. Always confirm that door openings are
unobstructed before closing.
Repairs and adjustments to this equipment must be made only by fully qualified service personnel.
Maintenance performed by inexperienced, unqualified persons or installation of unauthorized parts could cause
personal injury or result in costly equipment damage.
WARNING - STERILITY ASSURANCE HAZARD:
Load sterility may be compromised if the biological indicator or air leak test indicates a potential problem.
If these indicators show a potential problem, refer the situation to a qualified service technician before using the
sterilizer further.
According to EN 285 and AAMI standards, a measured leak rate greater than 1.0 mm Hg/minute (1.3 mbar/
min) indicates a problem with the sterilizer. Refer the situation to a qualified service technician before using
the sterilizer further.
The Express cycle is only intended for use with a single instrument in a single wrapped instrument tray.
The Express cycle is not intended for processing porous items (except the tray wrapper).
1-3
Listing of Safety Precautions Operator Manual 150829-026
CAUTION - POSSIBLE EQUIPMENT DAMAGE:
Gasket must be fully retracted prior to operating sterilizer door.
If zero dry time is selected, sterilizer automatically initiates a vapor removal phase in place of drying. This phase
can still draw a vacuum to 5.0 inHg. Consult device manufacturer’s recommendations to ensure devices being
processed can withstand this depth of vacuum.
Never use a wire brush, abrasives or steel wool on door and chamber assembly. Do not use cleaners
containing chloride on stainless-steel surfaces. Chloride-based cleaners will deteriorate stainless steel,
eventually leading to failure of the vessel.
Allow thermostatic traps to cool down to room temperature before removing cover. Since there is nothing to
limit expansion, the bellows may rupture or fatigue if trap is opened while hot.
Actuation at less than 75 percent of rated pressure can allow debris to contaminate the seat and cause the safety
valve to leak. A leaking safety valve must be replaced.
Insufficient service clearance will make repairs more difficult and time consuming.
Piping sized too small may cause water hammer, resulting in damage to the sterilizer.
After installation, it is mandatory to brace piping at the drain funnel so that it will not move vertically.
Ensure door opening is clear of any obstruction before closing the door(s).
Do not attempt to manually open sterilizer door unless chamber is at 0.0 psig.
Ensure gasket is fully retracted prior to manually operating sterilizer door.
Failure to flush generator daily could result in malfunction of the generator. Warranty on the generator will be
voided unless flushed daily.
Before flushing generator, ensure generator drain valve is fully open to prevent generator heaters from turning
on during flush phase.
IMPORTANT INSTALLATION NOTE:
Cycle Control Access: Sterilizer is shipped with access to cycle change control Locked Out. The default access
code number is 1234. See Access Code section of operator's manual for instructions for resetting the access
code number.
1-4
150829-026 Operator Manual Listing of Safety Precautions
Symbol Definition
Transfer of Heat, Hot Surface
Steam Energy
Sterilizer Door Open
Sterilizer Door Closed
Protective Earth (Ground)
Electrostatic Sensitive Device
Locked Cycle or Setup Options
Unlocked Cycle or Setup Options
Attention, Consult Manual for
Further Instructions
AAmperage Rating of the Unit
VVoltage Rating of the Unit
~Alternating Current
kW Power Rating of the Unit
Hz Frequency of the Unit
φφ
φφ
φPhase of the Unit
SN Serial Number of Unit
Definition of Symbols
2-1
Installation Verification Operator Manual 150829-026
INSTALLATION VERIFICATION 2
An Equipment Drawing showing all utility and space requirements
was supplied with the sterilizer. Clearance space shown on the
drawing is necessary for ease of installation and to assure proper
operation and maintenance of equipment. Uncrating and Installation
Instructions were also furnished with the sterilizer. If any of these
documents are missing or misplaced, contact STERIS giving the
serial and model numbers of the equipment. Replacement copies
will be sent to you promptly.
After installing this unit according to the instructions provided,
complete the following checklist to assure that your installation is
complete and correct. Or, if you desire, contact STERIS for a
technician to be scheduled to test your installation and demonstrate
proper equipment operation.
Clearance as specified on the Equipment Drawing must be
available.
Feed Water:
All supply line shutoffs must be provided with lockout
capability.
Backflow prevention is by others.
Water Pressure measured (specification is 2.1 to 3.5 bar [30 to
50 psig], dynamic).
Water pressure supplied must be within
specifications as shown on the Equipment Drawing. If pressure is
too high, a regulator must be installed. If water pressure is too low,
equipment performance will be affected.
Water Quality supplied must be within specifications. Im-
proper water quality adversely affects equipment operation,
especially units equipped with an electric steam generator.
Damage to the equipment from improper water quality is not
covered under warranty.
Steam Supply:
Shut-offs (with provisions for lockout and tagout) located nearby.
Supply piping adequately sized.
Supply pressure measured (specification is 3.5 to 5.2 bar [50
to 80 psig], dynamic).
Steam Piping to the sterilizer is to be trapped at a point before
connection to the sterilizer. Trap is required to remove conden-
sate from the steam line.
Drain Piping must be sloped properly, and sized to handle the
maximum waste flow from the sterilizer.
CAUTION - POSSIBLE EQUIPMENT
DAMAGE: Insufficient service
clearance will make repairs more
difficult and time consuming.
CAUTION - POSSIBLE EQUIPMENT
DAMAGE: Piping sized too small
may cause water hammer, resulting
in damage to the sterilizer.
2.1 Installation
Checklist
CAUTION - POSSIBLE EQUIPMENT
DAMAGE: After installation, it is
mandatory to brace piping at the
drain funnel so that it will not
move vertically.
2.1.1 Service Clearance
2.1.2 Plumbing Services
2-2
150829-026 Operator Manual Installation Verification
2.1.3 Electrical Service
WARNING - EXPLOSION HAZARD:
This sterilizer is not designed to
process flammable compounds.
2.2.1 Overall Size 16" Sterilizer: 660 mm wide x 1892 mm high x 908 mm deep
(26" wide x 74-1/2" high x 35-3/4" deep)
20" Sterilizer: 762 mm wide x 1892 mm high x 1146 mm
( 30" wide x 74-1/2" high x 45-1/8" deep)
16" Sterilizer: Single Door, Steam 340 kg (750 lbs)
Double Door, Steam 449 kg (989 lbs)
Singel Door, Electric 404 kg (890 lbs)
20" Sterilizer: Single Door, Steam 558 kg (1231 lbs)
Double Door, Steam 728 kg (1606 lbs)
Single Door, Electric 622 kg (1371 lbs)
Double Door, Electric 782 kg (1726 lbs)
2.2 Technical
Specifications
2.1.4 Sterilizer Final Check
2.1.5 Cycle Operation
All disconnects must be provided with lockout capability.
Electric one phase service to the unit must be as specified on the
Equipment Drawing and on the Machine Data Plate.
Electric one phase service requires a clearly marked disconnect
located near the sterilizer.
Electric one phase service should be on a separate circuit, and
not tied into circuits containing large reactive loads (e.g., motors).
Protective earthing to the unit must be connected to terminal
block TB-1 in the unit power box.
Three-phase power for optional electric steam generators must
meet specifications on the equipment drawing.
Three-phase service requires a clearly marked disconnect lo-
cated near the sterilizer.
Door runs up and down smoothly.
Door open/close time is seven to nine seconds.
Door up switches adjusted correctly.
Chamber strainer in place.
Rack and shelves and/or loading car operates correctly.
Paper loaded in printer.
Unit powers up correctly.
If prevacuum sterilizer, run Leak Test cycle — leak rate is to be
less than 1.0 mm Hg/minute.
Verify operation of a typical cycle (134° C prevacuum).
2.2.2 Weight
2-3
Installation Verification Operator Manual 150829-026
2.2.3 Utility Requirements
2.2.4 Environmental
Conditions
Electric:
Direct Steam: 240 V, 1.0 A, One Phase
Steam Generator: 240 V, 5.0 A, One Phase (control)
(if equipped) 400 V, 65 A, Three Phase (heaters)
Water:
Pressure:
Sterilizer 2.1 to 3.5 bar (30 to 50 psig)
Steam Generator 1.4 to 3.5 bar (20 to 50 psig)
(if equipped)
Temperature:
Sterilizer 21°C (70°F) maximum
Steam Generator 60°C (140°F) maximum
(if equipped)
Consumption:
Sterilizer 23 lpm (6.0 gpm)
Steam Generator 23 Ipm (6.0 gpm)
(if equipped)
Steam:
Pressure: 3.5 to 5.2 bar (50 to 80 psig)
Consumption:
16" Direct Steam 38 kg/h (83 lb/hr)
20" Direct Steam 53 kg/hr (116 lb/hr)
Temperature: 10° to 32°C (50° to 90°F)
Humidity: 10 to 90% noncondensing
Pollution Degree: 2
Installation Category (Overvoltage Category): II
A-Weighted Sound Power Level: 68 dBA (mean) -78 dBA (maximum)
2-4
150829-026 Operator Manual Installation Verification
Healthcare
Capital Equipment
3-1
Sterilization Techniques Operator Manual 150829-026
STERILIZATION TECHNIQUES 3
The information in this section is intended as a general guide to
steam sterilization techniques. Also recommended is reference to
EN554 sterilization of medical devices: Validation and Routine
Control of Sterilization by Moist Heat.
Prior to sterilization, all materials and articles must be thor-
oughly cleaned.
After sterilization, goods should be stored in conditions that will
not compromise the barrier quality of their wrapping materials.
Always carefully segregate items which have been sterilized
using conventional cycles from those sterilized using Flash or
Express cycles.
IMPORTANT: Applicable cycles have been validated to satisfy the
requirements outlined in Table 3-1. If cycle parameters (sterilize
time, dry time, temperature) other than those in Table 6-1 are
required, it is the responsibility of the healthcare facility to validate
the cycle. Reference EN285 or AAMI for a guide to validation
sterilization cycles and to ensure drying efficiencies are met.
NOTE: Contact STERIS for information on a wide range of educa-
tion/training programs designed to meet the educational needs of
healthcare industries.
3.1 General
121°C Prevac
134°C Prevac
Leak Test
Bowie-Dick Test
Validated Per EN285
Table 3-1. Cycle Validation Standards
Validated Per AAMI ST-37
Validated Per AAMI ST-8
134°C Flash
134°C Express
121°C Gravity
3-2
150829-026 Operator Manual SterilizationTechniques
The Express cycle is an abbreviated prevacuum cycle. Appro-
priate parameters for sterilization are factory-preset. The Express
cycle is designed to permit flash sterilization of a single instrument,
using a single
wrapper (non-woven or textile, but not a peel pouch)
on the instrument tray.
Rationale: The single wrapper serves to confine and contain the
sterilized item from environmental contaminants that may be
encountered enroute from the sterilizer to the point of use.
Single- wrapped items sterilized with the Express cycle do not
have a shelf life.
The Express cycle is useful in providing quick turnaround of an
instrument using the wrapped technique for transport from the
sterilizer to the point of use. Up to two instrument trays in single
wrappers may be processed at one time in the sterilizer.
IMPORTANT: Instrument trays processed using the Express cycle
are intended for immediate use.
Decontaminate and then prepare the surgical instrument (open and/
or disassembled), and place in a perforated or mesh bottom
instrument tray. Only non-porous items (except for the single
wrapper) should be processed using this cycle.
Items with lumens (e.g., needles for injection and diagnostics or
metal suction cannulae) are considered porous items, and must not
be processed using this cycle.
The Express cycle has fewer prevacuum pulses so air removal and
subsequent steam contact within lumens may be difficult to achieve.
3.2.1 Preparing Instruments
for Express Cycle Sterilization
Examples of non-porous items that can be processed using
the Express cycle:
Forceps,
Needle Holders,
Scissors and other routine metal instruments.
Examples of porous items that CANNOT be processed using
the Express cycle:
Towels,
Rubber or Plastic Items,
Items with Lumens,
Items with sliding parts that prevent sterilant contact with
surfaces.
WARNING STERILITY ASSUR-
ANCE HAZARD: The Express cycle
is only intended for use with a
single instrument in a single
wrapped instrument tray.
Table 3-2. Item Processing Guidelines for
Express Cycle
WARNING STERILITY ASSURANCE
HAZARD: The Express cycle is not
intended for processing porous items
(except the tray wrapper).
3.2 Special Information
Regarding the Express
Cycle
3-3
Sterilization Techniques Operator Manual 150829-026
3.2.2 Unloading Use instruments processed in an Express cycle immediately. Do
not store for later use.
When opening and unloading the sterilizer at the conclusion of the
Express cycle, always use aseptic techniques. Sterile towels and
sterile gloves are used to remove the tray from the sterilizer. The still-
wrapped tray is transported to the point of use, being careful not to
contact any unsterile surfaces. The tray is placed on a previously
draped sterile surface field and the wrapper opened by the circulat-
ing nurse. The draping material used must be impervious to liquids
and able to withstand contact with hot surfaces.
To shorten the sterilization process, drying time has been reduced
in the Express cycle. Thus, the single wrapper will appear dry at the
conclusion of the cycle, but the contents of the package will, in all
probability, still be wet with condensation formed during steriliza-
tion. Water will penetrate the wrapper, especially on the under side
of the tray because of gravity. The wrapper only serves the function
of protecting the tray and its contents from particulate matter
encountered on the way from the sterilizer to the point of use.
Any moisture present can strike through the wrapper, so procedures
should be developed to avoid contamination by contact with
unsterile surfaces. For example:
Wear sterile gloves and use sterile towels as “pot holders” when
removing the goods from the sterilizer.
Never place the wrapped tray on an unsterile surface.
As part of the operator’s verification of the sterilization process,
biological indicators may be used to demonstrate that sterilization
conditions have been met.
NOTE: Contact STERIS for information on specific biological
indicators recommended for use with this sterilizer.
A live spore test utilizing
G. stearothermophilus
is the most reliable
form of biological monitoring. This type of product utilizes controlled
populations of a controlled resistance, so that survival time and kill
time can be demonstrated.
To verify the process, insert the biological indicator in a test pack
and place pack on the bottom shelf. Run test pack through a typical
cycle. On completion, forward test pack and monitor to appropriate
personnel for evaluation. Refer to AAMI guidelines to conduct
routine biological monitoring.
3.3 Control Measures
For Verifying
Sterilization Process
3.3.1 Biological
Monitors
3-4
150829-026 Operator Manual SterilizationTechniques
WARNING STERILITY ASSURANCE
HAZARD: Load sterility may be
compromised if the biological indicator
or air leak test indicates a potential
problem. If these indicators show a
potential problem, refer the situation
to a qualified service technician before
using the sterilizer further.
3.4 Bowie-Dick Test
3.5 Vacuum Leak Test
3.3.2 Testing for
Prevacuum Efficiency
WARNING STERILITY ASSURANCE
HAZARD: According to EN285 and
AAMI standards, a measured leak
rate greater than 1.0 mm Hg/minute
indicates a problem with the sterilizer.
Refer the situation to a qualified
service technician before using the
sterilizer further.
The first prevacuum cycle of each day should be used to test the
adequacy of air removal from the chamber and load, so that steam
can penetrate the load. It is not a test for adequate exposure to heat
in terms of time-at-temperature.
The Bowie-Dick Test is designed to document the removal of
residual air from a sample challenge load.
In the case of this test, following exposure in a prevacuum
sterilizing cycle, the pack is opened, the indicator examined
and conclusions are drawn as to the pattern of residual air, if any,
that remained in the pack during the sterilizing cycle. Any
indication of a malfunction must be reported to the supervisor,
who will determine the disposition of the sterilizer, i.e., whether
it should be retested, serviced, or remain in use.
According to EN554 and AAMI-46, a steam penetration test shall
be carried out at the beginning of each day the sterilizer is to be
used. Refer to instructions for running the
Bowie-Dick Test
cycle
in Section 5. Bowie-Dick test packs are designed to expose the
pattern and document the removal of residual air from the
sample load. Any test package must be constructed in accor-
dance with instructions given in the EN285 standard.
NOTE: The Bowie-Dick test cycle is not a test for adequate
exposure to heat in terms of time-at-temperature.
The Leak Test (see appropriate cycle description in Section 5)
measures the integrity of the sealed pressure vessel and
associated piping to assure air is not being admitted to the
sterilizer during the vacuum draw downs. Refer to EN285,
section 20, for leak test instructions.
After running a Leak Test cycle, a value or leak rate will be printed
on the printer tape. This value will help define a trend over a period
of time if the integrity of the system begins to deteriorate (i.e.,
allowing air to enter the system). By running a Leak Test cycle
weekly, the operator or maintenance personnel can always monitor
the air tightness of the system and make repairs or adjustments
when necessary.
NOTE: A leak rate of greater than 1.0 mm Hg/minute indicates a
problem with the sterilizer that must be addressed.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126