A&D AD-8922A User manual

Type
User manual
AD-8922A
Remote Controller
INSTRUCTION MANUAL
1WMPD4002022
© 2009 A&D Company Ltd. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or translated
into any language in any form by any means without the written permission of A&D
Company Ltd.
The contents of this manual and the specifications of the instrument covered by this
manual are subject to change for improvement without notice.
1
CONTENTS
INTRODUCTION .................................................................................................................................... 2
- Outline .......................................................................................................................................................2
- Applicable Instruments ..............................................................................................................................2
- Compliance................................................................................................................................................3
DESCRIPTION OF EACH PART ............................................................................................................ 4
- Display .......................................................................................................................................................5
- Key Operation............................................................................................................................................5
- Connectors ................................................................................................................................................5
GETTING READY................................................................................................................................... 6
- Setting the Weighing Instrument and the AD-8922A.................................................................................6
- Connecting the AD-8922A .........................................................................................................................6
- Turning the Power on ................................................................................................................................7
- Operation ...................................................................................................................................................7
- Example of Use .........................................................................................................................................7
FUNCTION SETTINGS........................................................................................................................... 9
- Display and Keys.......................................................................................................................................9
- Function Table .........................................................................................................................................10
- Initialization ..............................................................................................................................................12
TROUBLESHOOTING.......................................................................................................................... 13
SPECIFICATIONS ................................................................................................................................ 14
RS-232C SERIAL INTERFACE ............................................................................................................ 15
EXTERNAL DIMENSIONS ................................................................................................................... 16
OPTIONS.............................................................................................................................................. 17
- Confirming the AD-8922A Software Version............................................................................................17
- Installing the Option.................................................................................................................................18
10 BCD OUTPUT (AD-8922A-01) ......................................................................................................... 19
10- BCD Output Specifications...................................................................................................................19
10- Setting of the Decimal Point Position...................................................................................................24
11 COMPARATOR OUTPUT (AD-8922A-04)........................................................................................ 25
11- Comparator Output Specifications .......................................................................................................25
11- Using the Comparator Output ..............................................................................................................27
11- Setting the Upper and Lower Limit Values...........................................................................................28
12 CURRENT LOOP INPUT (AD-8922A-05)......................................................................................... 30
12- Current Loop Input Specifications ........................................................................................................30
13 ANALOG OUTPUT (AD-8922A-06).................................................................................................. 32
13- Analog Output Specifications...............................................................................................................32
13- Function setting....................................................................................................................................34
13- Switching Output Voltage.....................................................................................................................35
13- Output Voltage Fine Adjustment ..........................................................................................................35
13- Fixed Output Voltage ...........................................................................................................................36
14 ACCESSORIES (CABLE LIST)........................................................................................................ 37
2
INTRODUCTION
The AD-8922A is a remote controller. Read this manual completely before using the AD-8922A in order to
ensure a sufficient understanding for proper use.
- Outline
The AD-8922A is connected to an A&D manufactured weighing instrument, using the RS-232C serial interface.
Displays the weighing data transmitted by the weighing instrument.
Key operations remotely control the weighing instrument.
(Entering the function setting mode of the weighing instrument is not available. Available operations
depend on the weighing instrument used. See Table 2 in "1-2 Applicable Instruments".)
The data the AD-8922A receives can be output, using the RS-232C serial interface. So, external
devices such as a personal computer or a printer can be connected to the AD-8922A.
Can be panel-mounted.
Various options such as BCD output, comparator output, current loop input and analog output are
available. For details, refer to "9 OPTIONS" and the following relevant chapters.
When connected to the AD-4212C, the AD-8922A can change the response characteristic, calibrate
the AD-4212C using the external weight, and both instruments can share power.
(Power can be supplied to both instruments by plugging the AC adapter into either the AD-4212C or
the AD-8922A. For details, refer to "3-3 Turning the Power on ".)
- Applicable Instruments
Applicable weighing instruments and what is required are shown in the table below.
Table 1 Applicable weighing instruments and what is required
What is required to connect to
a weighing instrument
Cable required to connect to
an external device such as
Weighing
instrument Option for the
instrument
Communications cable*
(Length 2 m)
Personal
computer
AD-8121B
compact printer
AD-4212C None
(D-Sub 9-pin)
None
(Use the cable provided
for AD-4212C)
GX, GF, GX-K,
GF-K, GP, FP,
AD-4212A/B, GR, HR
None
(D-Sub 25-pin) AX-KO1710-200
EK-i, EW-i, FC-i,
FC-Si, GH, HR-i,
FZ-i, FX-i
None
(D-Sub 9-pin) AX-KO2466-200
EK-G, EK-H,
ET-W, EW-G
OP-03
(D-Sub 25-pin) AX-KO1710-200
HV-G, HV-WP,
HW-G, HW-WP
None
(DIN 7-pin) AX-KO1786-200
FG OP-03
(DIN 7-pin) AX-KO1786-200
FS, FS-KL OP-03
(DIN 8-pin) AX-KO1786-200
FG-L, FG-M OP-23
(DIN 8-pin) AX-KO1786-200
AX-KO1786-200 AX-KO462-200
Note (*) - The standard communications cable is AX-KO1710-200. A substitute communications
cable may be provided as specified when ordering the AD-8922A.
- When connecting to the AD-4212C, use the AX-KO3590-500 (5m) cable, provided as standard
for the AD-4212C.
3
The AD-8922A functions in two ways as follows, depending on the weighing instrument used:
A remote controller that displays the weighing data and remotely controls the weighing instrument.
A remote display that displays the weighing data.
Available key operations depend on the weighing instrument used, as shown below. Set the command set
"C5et" of the function setting, appropriate to the weighing instrument.
Table 2 Applicable weighing instruments and key operations
AD-8922A key Weighing
instrument ON:OFF CAL SAMPLE PRINT MODE RE-ZERO
Command
set *5
AD-4212C
Calibrates
using the
external
mass.
Switches
the
response
characteristic.
C5et 1
GX, GX-K, GP,
GH, ET-W, FZ-i C5et 1
GR
Calibrates
using the
internal
mass. C5et 2
GF, GF-K,
EK-H,
AD-4212A/B, HR,
HR-i, FX-i
Turns the
weighing
instrument
display
on or off.
*1
Switches
the
minimum
display.
*2
C5et 3
EK-i, EW-i
Switches
the unit
displayed.
*4
Sets the
display to
zero.
C5et 4
EK-G, EW-G, FC-i,
FC-Si, FG, FG-L,
FG-M, FP, FS, FS-KL,
HV-G, HV-WP, HW-G,
HW-WP
Outputs
the
received
data to
an
external
device.
*3
C5et 0
Note: "" in the table indicates that the key operation is not available.
*1: Switching the standby or weighing mode is available for the AD-4212C.
*2: Not applicable to the counting mode and the percent mode. Switching the minimum display
is not available for the ET-W.
*3: Available when the AD-8922A is in key mode ("out 1" or "out 2" of the function setting).
*4: Not available for the ET-W and AD-4212
*5: AD-8922A function settings
-Compliance
-- Compliance with FCC Rules
Please note that this equipment generates, uses and can radiate radio frequency energy. This
equipment has been tested and has been found to comply with the limits of a Class A computing
device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide
reasonable protection against interference when equipment is operated in a commercial
environment. If this unit is operated in a residential area, it may cause some interference and under
these circumstances the user would be required to take, at his own expense, whatever measures
are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
-- Compliance with EMC Directives
This device features radio interference suppression in compliance with valid EC
Regulation 89/336/EEC.
4
DESCRIPTION OF EACH PART
Note: Please confirm that the AC adapter type is correct for your local voltage and receptacle
type.
A substitute communications cable, listed below, may be provided as specified when
ordering the AD-8922A.
D-Sub 9-pin to D-Sub 9-pin (AX-KO2466-200)
DIN 7-pin to D-Sub 9-pin (AX-KO1786-200)
When connecting to the AD-4212C, use the cable provided for the AD-4212C.
Do not use the cable provided for the AD-8922A.
Rear
Connector
to the weighing instrument
(D-Sub 9-pin )
Connector
to the external device
Display
Keys
Angle adjustment knob
AC adapter
jack Grounding terminal
Accessories
AC adapter AC adapter identification label
See Note below.
Communications cable
to connect to a weighing
instrument (Length: 2 m, 1 pc.)
D-Sub 25-pin to D-Sub 9-pin
(AX-KO1710-200)
(DIN 8-pin )
Stand
Main unit
5
- Display
Displays the weighing data and the unit (or mode) received. The unit (or mode) may be different from
that of the weighing instrument.
Turns on the stabilization indicator when the header of the weighing data received is "ST", "QT", or "WT".
Turns on the comparator indicator when the comparison results are added to the data received.
Applicable to GX-K, GF-K, GP and AD-4212A/B. (Function setting "Cp-r")
When nothing has been received for two seconds or more, the bar display [- - - - -] appears. By
changing the function setting, the previous data received is displayed until the next data is received
(Hold display). During the hold display, the hold indicator is turned on.
Note: When the data is in eight digits, the highest-order digit is displayed in the upper left
corner as shown below.
For example, if the weighing data is 101.00000 g, the display is like .
- Key Operation
Available key operations to control the weighing instrument depend on the weighing instrument used. For
details, see Table 2 in " 1-2 Applicable Instruments".
Set the command set "C5et" of the function setting, appropriate to the weighing instrument.
To enter the function setting mode of the AD-8922A, press and hold the ON:OFF key and press the CAL
key.
The weighing instrument has its own function settings and the AD-8922A can not change those settings.
- Connectors
--Connector to the weighing instrument (BALANCE / SCALE) ... D-Sub 9-pin male
Used to connect to the weighing instrument. The cable used for connection depends on the instrument.
For details, refer to the instruction manual of the weighing instrument.
--Connector to the external device (PC / PRINTER) ... DIN 8-pin female
Used to connect to an external device such as a personal computer or a printer (AD-8121B). The cable
used for connection depends on the device. For details, refer to the instruction manual of the device.
6
GETTING READY
- Setting the Weighing Instrument and the AD-8922A
1. Set each item as shown below. Set the same value for the weighing instrument and the AD-8922A.
Table 3
Item Weighing instrument AD-8922A
Baud rate 600, 1200, 2400, 4800, 9600. 19200 bps
Data bits, Parity bit 7 bits - EVEN, 7 bits - ODD, 8 bits - NONE
Stop bits 1 bit or 2 bits
Terminator <CR> or <CR><LF>
Data format A&D standard format
Communication control No RTS/CTS control
Data output mode Stream mode*
Items in bold face type: Factory settings for both the AD-8922A and the weighing instrument.
* When connected to an external device, the settings can be changed to suit the use.
2. Set the command set "C5et" of the function setting, appropriate to the weighing instrument. See Table 2
in "1-2 Applicable Instruments".
Note: Some items may not be available for the instrument used. For details on setting, refer to
the instruction manual of each weighing instrument.
When connected to an external device such as a personal computer or a printer, set the output
mode "out" and time out "Hold" of the function setting, and set the data output mode of the
weighing instrument, appropriate to the use. (See Table 3 in "3-5 Example of Use".)
- Connecting the AD-8922A
For information on cables required for connection, see Table 1 in " 1-2 Applicable Instruments".
Connect the weighing instrument to the AD-8922A D-Sub 9-pin connector.
Connect an external device such as a personal computer or a printer to the AD-8922A DIN 8-pin
connector.
7
- Turning the Power on
Insert the AC adapter plug into the AC adapter jack located on
the rear of the AD-8922A. Plug the AC adapter into an
appropriate electrical outlet.
When connected to the AD-4212C, power can be supplied to
both instruments by plugging the AC adapter into either the
AD-4212C or the AD-8922A.
(Both instruments can have their AC adapter connected at the
same time.)
- Operation
The AD-8922A displays the weighing data transmitted by the weighing instrument used.
The AD-8922A key operations remotely control the weighing instrument.
Available operations depend on the weighing instrument. See Table 2 in "1-2 Applicable Instruments".
- Example of Use
A personal computer is connected to the AD-8922A and the weighing data can be transmitted to the
personal computer, using Windows communication tools (WinCT).
To protect against inadvertent operations, the AD-8922A keys can be disabled. (Function setting
"C5et 0").
When nothing has been received for two seconds or more, the bar display [- - - - -] appears. By
changing the function setting, the previous data received is displayed until the next data is received
(Hold display). (Function setting "Hold 1 ")
About details on the settings of the weighing instrument or the external device, see the relevant
instruction manual.
External device
AD-8121B
Personal computer
PLC
Weighing
instrument
BALANCE/SCALE
Used to connect to
the weighing instrument
PC/PRINTER
Used to connect to
the external device
AD-8922A
8
Table 4 Use of the AD-8922A
Example
of use
Weighing
instrument AD-8922A setting External device
setting
Through mode "out 0"
Sends all of the
received weighing
data to the external
device.
[AD-8121B]
MODE 2
(Prints data in
conjunction with
theAD-8121B
DATA key or
interval setting.)
Key mode 1 "out 1"
When the AD-8922A
PRINT key is
pressed, sends the
latest weighing data
received to the
external device,
regardless of the data
status.
Stream mode
(Outputs the
weighing data
continuously.)
Key mode 2 "out 2"
When the AD-8922A
PRINT key is
pressed, sends the
latest weighing data
received to the external
device when the data is
stable.
[AD-8121B]
MODE 1
(Prints data when
the AD-8121B
DATA key is
pressed.)
Monitors the
weighing data
of the
weighing
instrument
remotely.
Key mode or
Auto print mode
(Outputs the
weighing data
when the key
is pressed or
outputs the data
automatically
when stable.)
Through mode "out 0"
Sends all of the
received weighing
data to the external
device.
[AD-8121B]
MODE 1
(Prints data
according to the
data output
mode of the
weighing
instrument.)
Monitors the
weighing data
of the
weighing
instrument that
is built into a
weighing system.
Command
mode *
(Outputs the
weighing data
by the data
request
command.)
Through mode "out 0"
Sends all of the
received weighing
data to the external
device.
[Personal computer
or PLC]
(The program to
control the
device is
required.)
* The command mode may not be available for weighing instruments of which command is always
valid.
9
FUNCTION SETTINGS
The function settings specify the parameters for the AD-8922A performance. These parameters stored,
even if the AC adapter is removed, are maintained in non-volatile memory.
The function menu consists of two layers. The first layer is the “Class” and the second layer is the “Item”.
Each item is selected by the SAMPLE key. The parameter of the selected item is changed by the
RE-ZERO key. Finally, the parameter is stored and is enabled by the PRINT key.
Example
This example sets “9600 bps” for “Baud rate” and “8 bits NONE” for “Data bits, parity bit”.
Note: Use much care when changing parameters. The AD-8922A may not function properly
when the settings and operational environment are not appropriate.
- Display and Keys
Table 5
Display/Key Description
The symbol “ ” indicates that the parameter displayed is in effect.
Selects the class or item in the function setting mode.
Changes the parameter.
When a class is displayed, moves to an item in the class.
When an item is displayed, stores the new parameter and displays the next class.
When an item is displayed, cancels the new parameter and displays the next class.
When a class is displayed, exits the function setting mode.
10
- Function Table
Table 6
Class Item
Parameter
Description
0
Set 0
When connected to EK-G, EW-G,
FC-i, FC-Si, FG, FP, FS, FS-KL
HV-G, HV-WP, HW-G or HW-WP.
Disables the AD-8922A keys for use
as a remote display.
1 Set 1 When connected to ET-W, GH, GP,
GX, GX-K, FZ-i, AD-4212C.
2 Set 2 When connected to a GR series
balance.
3 Set 3 When connected to GF, GF-K,
AD-4212A/B, HR, HR-i. EK-H, FX-i
C5et *
Command set
4
Set 4 When connected to an EK-i/EW-i
series balance.
0
Through mode
Always outputs the data received by
the D-Sub 9-pin connector, to the
DIN 8-pin connector.
1 Key mode 1
Outputs the latest data received by
the D-Sub 9-pin connector, to the
DIN 8-pin connector, when the
AD-8922A PRINT key is pressed.
out
Output mode
2 Key mode 2
Outputs the latest stable data
received by the D-Sub 9-pin
connector, to the DIN 8-pin
connector, when the AD-8922A
PRINT key is pressed.
0
Bar display Bar display if nothing has been
received for two seconds or more.
Hold
Timeout 1 Hold display
Displays the previous data received
if nothing has been received for two
seconds or more, and turns on the
hold indicator.
0
Not used No buzzer upon data receipt.
1 At intervals
Sounds buzzer when the data is
received with an interval of two
seconds or more.
bep
Data receipt
buzzer
2 Used Sounds buzzer upon each data
receipt.
-
Not set
0
dpp
Position of
decimal point 6
Set
Sets the decimal at a specific position. When
changing the minimum display digit, pressing
SAMPLE key does not change the decimal
point position. (For details, refer to “9-2”.)
0
No function
The SAMPLE key does not function.
fnc
Environment
Display
5apl
Function of the
SAMPLE key 1 Function
The SAMPLE key will function.
0
600 bps
1 1200 bps
2 2400 bps
3 4800 bps
4
9600 bps
5if
Serial
interface
bp5
Baud rate
5 19200 bps
Choose a parameter appropriate to
the weighing instrument.
When the AD-8121B compact
printer is to be connected, leave the
factory settings of the AD-8922A as
is and set the weighing instrument.
Factory setting
* See Table 2 in "1-2 Applicable Instruments" to set the command set "C5et" of the function setting,
appropriate to the weighing instrument.
11
Class Item
Parameter
Description Class
0
7 bits - EVEN
1 7 bits - ODD
btpr
Data bits
Parity bit 2 8 bits - NONE
0
1 bit
5top
Stop bits 1 2 bits
0
CR/LF
5if
Serial
interface
Crlf
Terminator 1 CR
Choose a parameter appropriate to
the weighing instrument.
When the AD-8121B compact
printer is to be connected, leave the
factory settings of the AD-8922A as
is and set the weighing instrument.
0
ON when 0
data
data 1 ON when 1
0
ON when positive or 0
polp
Polarity 1 ON when negative
0
ON when the stabilization indicator is turned off
5tbp
Stability 1 ON when the stabilization indicator is turned on
0
OFF when e or -e
0erp
Over 1 ON when e or -e
0
Data refresh is complete when ON OFF
5trp
Strobe 1 Data refresh is complete when OFF ON
0
Approx. 10 ms
1 Approx. 20 ms
5trt
Strobe pulse
width 2 Approx. 50 ms
Strobe pulse width after data refresh
0
BUSY input function
bcd
BCD
Available only
when the
BCD output
is installed.
5trp
(Input terminal
function) 1 RE-ZERO input function
0
No comparison
1 Comparison, excluding "near zero" when the value is
stable or overloaded
2 Comparison, including "near zero" when the value is
stable or overloaded
3 Continuous comparison, excluding "near zero"
4
Continuous comparison, including "near zero"
Cp
Comparator
mode
5 Contact-outputs the second header information of the data
received. (Applicable to AD-4212A, GP and GX-K)
0
OFF
bep
LO buzzer 1 ON
Selects whether or not to sound the buzzer
when LO.
0
OFF
bep-
OK buzzer 1 ON
Selects whether or not to sound the buzzer
when OK.
0
OFF
Cp fnc
Comparator
Available only
when the
comparator
output is
installed.
bep
HI buzzer 1 ON
Selects whether or not to sound the buzzer
when HI.
0
2-digt
output
Converts the consecutive 2 digits, with the digit
selected in 5el as the least, to voltage and outputs.
an
Analog output
mode 1 3-digt
output
Converts the consecutive 3 digits, with the digit
selected in 5el as the least, to voltage and outputs.
0
Selects the first digit as the least.
1 Selects the second digit as the least.
2 Selects the third digit as the least.
3
Selects the fourth digit as the least.
4
Selects the fifth digit as the least.
aout
Analog
output
Available only
when the
analog output
is installed.
5el
Analog output
digit selection
5 Selects the sixth digit as the least.
Factory setting
12
- Initialization
Initialization restores the AD-8922A function settings to factory settings.
1. Connect the AC adapter. The bar display or the weighing data
display appears.
2. While pressing and holding the ON:OFF key, press the PRINT
key. "Clr" appears in the display.
3. Press the PRINT key. (To cancel the operation, press the CAL
key.)
4. Press the RE-ZERO key to select "go".
5. Press the PRINT key to perform initialization.
After initialization, the bar display or the weighing data display
appears.
13
TROUBLESHOOTING
Shown below is a list of various phenomena of the AD-8922A and the remedies.
Table 7
Phenomenon Cause and remedy
error10 appears.
The communications settings of the AD-8922A and the weighing
instrument are not the same.
Check the settings.
error11 appears.
The weighing instrument data format is incorrect.
Set the data format to A&D standard format. Also check whether or not
data other than weighing data is output.
The bar display
remains and no
weighing data
appears.
Is the weighing instrument in the stream mode?
Only the stream mode displays the weighing data continuously. Other
modes display the data only when received. When the hold display is
selected in the AD-8922A function setting, the previous data received is
displayed until the next data is received.
Are the communications settings correct?
Is the cable correct?
The display flickers.
Electrical noises may affect the display.
Using the grounding terminal located on the rear of the AD-8922A,
ground the AD-8922A.
14
SPECIFICATIONS
Power supply : AC adapter
(Please confirm that the AC adapter type is correct for your local voltage
and receptacle type.)
Power consumption : Approx.11 VA (Including the AC adapter, AD-8922A: approx. 1.5 VA)
Transmission system : RS-232C
Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Refresh rate : Approx. 10 times/second (When Baud rate is 2400 bps or greater).
Applicable connectors : D-Sub 9-pin (Male) to connect to the weighing instrument
DIN 8-pin (Female) to connect to an external device
Dimensions : 238 (W) x 132 (D) x 170 (H) mm
Mass : Approx. 1.0 kg
Standard accessories : AC adapter, Communications cable (Approx. 2 m)
(Please confirm that the AC adapter type is correct for your local voltage
and receptacle.)
15
RS-232C SERIAL INTERFACE
RS-232C
Transmission system : EIA RS-232C
Transmission form : Asynchronous, bi-directional, half duplex
Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits : 7 bits or 8 bits
Parity bit : EVEN, ODD (Data bits 7 bits)
NONE (Data bits 8 bits)
Stop bits : 1 bit or 2 bits
Code : ASCII
Terminator : <CR> or <CR><LF>
Circuit
Connection to the weighing instrument Connection to an external device
D-Sub 9-pin DIN 8-pin
Pin
No.
Signal
name Direction Description Pin
No.
Signal
name Direction Description
1 (Vs) Used internally 1 FG Frame ground
2 RXD Input Receive data 2 RXD Input Receive data
3 TXD Output Transmit data 3 TXD Output Transmit data
4 N.C. 4 RTS Input Request to send
5 SG Signal ground 5 SG Signal ground
6 DSR Input Data set ready 6 CTS Output Clear to send
7 RTS Output Request to send 7 DSR Output Data set ready
8 CTS Input Clear to send 8 N.C.
9 (Va) Used internally
(The signal names except TXD and RXD
apply to the DTE such as a personal
computer.)
(AD-8922A is a DTE. Connects to a DCE
such as a weighing instrument using a
straight through cable.)
When making the cable yourself, do no
t
connect to the internally used terminals.
9876
6
1
4
2
5
3
7
SG
CTS
RTS
DSR
FG
TXD
RXD
SG
AD-8922A interior
To a weighing instrument To an external device such as
a personal computer or a printer
54321
TXD
RXD
AD-8922A interior
8
Inch screw thread
(#4-40UNC)
(Va)
(Vs)
16
EXTERNAL DIMENSIONS
Unit: mm
238
130
132
57
170
192
185
96
91
108
72
100
17
OPTIONS
The AD-8922A has various options available as follows.
For details on each option, refer to the relevant chapter.
AD-8922A-01 BCD output
Outputs the weighing data received from the weighing instrument using the RS-232C serial interface,
in BCD format.
AD-8922A-04 Comparator output
Compares the weighing data received from the weighing instrument using the RS-232C serial
interface with the upper or lower limit value and contact-outputs the results.
When connected to the AD-4212C, the both instruments can share power.
AD-8922A-05 Current loop input
Receives the current loop output from the weighing instrument and displays the weighing data.
The weighing data received can be output using the RS-232C serial interface.
The AD-4212C does not have current loop output. Therefore, this option can not be used.
AD-8922A-06 Analog output
Converts the specified digits of the weighing data received from the weighing instrument, using the
RS-232C serial interface, into voltage and outputs the value.
- Confirming the AD-8922A Software Version
Before installing an option, confirm the AD-8922A software version.
The software version is displayed as "px.xx" for approx. 1 second before entering the function setting
mode.
18
- Installing the Option
Install the option as shown below. The installation procedure is the same for all the options.
1. Disconnect the AC adapter.
2. Remove the two screws and pull out the RS-232C board
provided as standard as shown in the illustration.
3. Insert the option board, along the guides on the left and
right sides.
4. Using the two screws removed in step 2, fasten the option
board.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42