Invacare Etude Plus Low User manual

Type
User manual

This manual is also suitable for

Invacare®EtudePlus
EtudePlus,EtudePlusLow
ENBed
UserManual
ThismanualMUSTbegiventotheuseroftheproduct.
BEFOREusingthisproduct,readthismanualandsaveforfuture
reference.
©2013Invacare®Corporation
Allrightsreserved.Republication,duplicationormodificationinwholeorinpartisprohibitedwithoutpriorwritten
permissionfromInvacare.Trademarksareidentifiedby™and®.AlltrademarksareownedbyorlicensedtoInvacare
Corporationoritssubsidiariesunlessotherwisenoted.
ThisbedhasbeentestedandapprovedaccordingtoIEC60601–2–52(Bedswithtiltfunctioncan’tbeapprovedaccording
toIEC60601–2–52)
ThisbedhasundergoneariskanalysisaccordingtoENISO14971:2000
InvacareiscertifiedaccordingtoISO9001andISO13485
Contents
1General.......................................4
1.1Aboutthismanual...........................4
1.2IntendedUse...............................4
1.3Servicelife.................................4
1.4Warranty.................................4
2Safety.........................................5
2.1Generalwarning............................5
2.2Labelsandsymbolsontheproduct...............5
3Setup.........................................6
3.1Receivingthebed............................6
3.2Assemblyofthebed..........................6
3.3Wiring...................................6
3.4Dismantlethebed...........................7
4Operatingthebed..............................8
4.1Operatingthehandcontrolunit.................8
4.2Operatingthecastorbrakes....................8
4.3Adjustingthelegsection.......................8
4.4Emergencyreleaseofthebackrestorthigh/leg
section...................................8
5Accessories....................................9
5.1Listofavailableaccessories.....................9
5.2Siderails..................................9
Steelsiderails............................9
LineandBrittVsiderails....................9
5.3Liftingpole................................10
5.4Mattresssupportextension....................10
5.5Mattress..................................10
5.6Shippingbrackets/Transportbrackets............11
6Maintenance...................................12
6.1Maintenanceschedule.........................12
6.2Cleaning..................................12
7AfterUse......................................13
7.1Wastedisposal.............................13
8Troubleshooting................................14
8.1Troubleshootingelectricalsystem................14
9Technicaldata..................................15
9.1Dimensions................................15
9.2Weights..................................15
9.3Environmentalconditions......................15
9.4Electricalsystem............................15
9.5Electromagneticcompliance(EMC)...............16
EtudePlus
1General
1.1Aboutthismanual
ThankyouforchoosinganInvacareproduct.
Thisusermanualcontainsimportantinformationaboutthehandlingof
theproduct.Inordertoensuresafetywhenusingtheproduct,read
theusermanualcarefullyandfollowthesafetyinstructions.
Allindicationsofrightandleftarebasedonapersonlyingonhis
backinthebed.
Pleasenotethattheremaybesectionsinthisusermanual,whichare
notrelevanttoyourproduct,sincethismanualappliestoallexisting
modules(onthedateofprinting).
Symbolsinthismanual
InthisUserManualwarningsareindicatedbysymbols.Thewarning
symbolsareaccompaniedbyaheadingthatindicatestheseverityof
thedanger.
WARNING
Indicatesahazardoussituationthatcouldresultin
seriousinjuryordeathifitisnotavoided.
CAUTION
Indicatesahazardoussituationthatcouldresultin
minororslightinjuryifitisnotavoided.
IMPORTANT
Indicatesahazardoussituationthatcouldresultin
damagetopropertyifitisnotavoided.
TipsandRecommendations
Givesusefultips,recommendationsandinformation
forefficient,trouble-freeuse.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.Thelaunchdateofthis
productisstatedintheCEdeclarationofconformity.
1.2IntendedUse
Thebedhasbeendevelopedfordomesticcareandfeaturesa
comfortablesittingandlyingpositionforthepatient.Furthermore,
ergonomicaloperationforthecarerisensured.
Thebedisdefinedtobeusedaccordingto:
Applicationenvironment3;Long-termcareinamedicalareawhere
medicalsupervisionisrequiredandmonitoringisprovidedif
necessaryandmedicalelectricalequipmentusedinmedical
proceduresmaybeprovidedtohelpmaintainorimprovethe
conditionofthepatient.
Applicationenvironment4;Careprovidedinadomesticareawhere
medicalelectricalequipmentisusedtoalleviateorcompensate
foraninjury,disabilityordisease.
Thebedisnotintendedforchildrenunder12yearsandforpsychiatric
patients.
Thebedisnotintendedfortransportingthepatient.
WARNING!
Anyotherorincorrectusecouldleadtohazardous
situations.
1.3Servicelife
Theexpectedservicelifeofthisproductissevenyearswhenused
dailyandinaccordancewiththesafetyinstructions,maintenance
intervalsandcorrectuse,statedinthismanual.Theeffectiveservice
lifecanvaryaccordingtofrequencyandintensityofuse.
1.4Warranty
Thewarrantycoversallmaterialandproductiondefectsfortwoyears
fromthedateofdelivery,provideditcanbedemonstratedthatsuch
defectswerepresentbeforedelivery.Allmanufacturingfaultsor
defectsmustbepromptlyreported.
Invacare®mayrepairthefaultorreplacethecomponent.The
warrantyprovidedbyInvacare®doesnotcoveradditionalcosts
(transport,packaging,labour,sundryexpenses,etc.areforthe
customer’saccount).
Thewarrantydoesnotcover:
Damagecausedduringtransportthatisnotdirectlyreportedto
theforwarderatthemomentofdelivery.
Repairsperformedbyunauthorizedcentersandpersonnel.
Partssubjecttonormalwear.
Maliciousdamagesordamagedcausedbyimproperuseofthe
bed.
4
1567454-A
Safety
2Safety
2.1Generalwarning
WARNING!
Riskofentrapment/suffocation
There’sariskofentrapment/suffocationbetween
mattresssupport,siderailandbedend
Thebedmustnotbeusedbypatientsunder12years
ofage,orbypatientswithbodysizeequivalenttoan
average12yearsoldorsmaller.
Thebed,incombinationwithsiderailsmustnot
beusedforpersonsunder45kgorunder150cm
height.Eitherforpersonsthatarerestless(spasms)
orconfused,unlessaprofessionalriskassessmenthas
takenplaceandbeenaccepted.
WARNING!
Riskduetoelectromagneticinterference
Electromagneticinterferencebetweenthebedandother
electricalproductscanoccur.
Toreduceoreliminatesuchelectromagnetic
interference,increasethedistancebetweenthebed
andtheproductsorswitchthemoff.
Thismedicalbedcanbeusedtogetherwith
medicalelectricalequipmentconnectedtotheheart
(intracardially)orbloodvessels(intravascularly)provided
thatfollowingpointsarerespected:
Thebedshouldbeequippedwithmeansforpotential
equalizationconnectionmarkedoutbyasymbolshown
inthebackofthismanual.
Medicalelectricalequipmentshouldnotbefixedon
thebed’smetallicaccessoriessuchassiderails,lifting
pole,driprod,bedends,etc.
Themedicalelectricalequipmentpowersupplycord
shouldbekeptclearoftheaccessoriesoranyother
movingpartsofthebed.
WARNING!
Riskofpatientsslippingthroughtheopenings
Thebedfulfilsallrequirementsregardingmaximum
distances.However,itispossiblethatpatientswithsmall
bodydimensionsslipthroughtheopeningsbetweenthe
siderailsorthroughtheopeningbetweenthesiderail
andthemattresssupport.
Payspecialattention,ifthebedisusedforthecareof
patientswithsmallbodydimensions.
CAUTION!
Riskofinjury
Alwayslowerthebedtothelowestpositionbefore
leavingthepatientinthebedunattended.
Makesurethatthereisnothingunder,over,ornear
thebedthatcanobstructtheheightadjustment,like
forexamplefurniture,liftsorwindowframes.
Whenenteringorexitingthebed,usetheheight
adjustment.Usethebackrestassupport(yellow
buttononthehandcontrol).Thethigh-andleg
sectionmustbehorizontal,otherwisethereisariskof
overloadingthemattresssupport.
Thereisariskofentrapmentoffingersinthebed
movingparts.Payattentiontoyourfingers.
IMPORTANT!
Donotrolloverthemainscable.Donotbringmains
cableintomovingparts.
Disconnecttheplugfromthemainsbeforemoving
thebed.
Makesurethatnocables(mainsorfromother
equipment)arejammedordamaged,whenthebedis
used.
Donotplacethebedclosetoadirectheatsource(fire
place,radiatoretc.)andnotindirectsunlight.
2.2Labelsandsymbolsontheproduct
Theproduct-labelisplacedontheframeofthebed:
A
B
C
D
E
A
Typenumberofthebed
B
Serialnumber
C
Dateofproduction
D
Electricalinformationsee9.4Electricalsystem,page15
E
Manufacturer
Symbols
Max.patientweight
Max.safeworkingload
InsulationclassII
TypeBAppliedPart.Seealso9.4Electricalsystem,page
15
Refertotheusermanual.
Wastedisposalandrecyclinginformation.See7.1Waste
disposal,page13.
ThisproductcomplieswithDirective93/42/EEC
concerningmedicaldevices.Thelaunchdateofthis
productisstatedintheCEdeclarationofconformity.
1567454-A5
EtudePlus
3Setup
3.1Receivingthebed
Whenyoureceivethebed,checkthepackaging.Ifthebedshowsany
signsofdamageupondelivery,pleasefollowtermsofdelivery
0
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
00000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
0000000000000000
000000
000000
000000
000000
000000
000000
000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
0000000000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
00000000
000000000000
000000000000
00000000
A
B
C
D
E
F
G
H
I
J
Scopeofdelivery:
A
Mattresssupport,headend
B
4pcs.ofmattresssupportbracket
C
Pairofsiderails(woodorsteel)
D
Mattresssupport,footend
E
Motorforbedend
F
Castors
G
Motorforheadend
I
Motorforlegend
J
Pairofbedends
Handcontrol
Optional
H
Mattresssupportextension(10or20cm)
Liftingpole
Transportfitting
3.2Assemblyofthebed
WARNING!
Aftereachassembly,checkthatallfittingsaretightened
correctlyandthatthefunctionofallpartsisgiven.
Mattresssupport
1.Placethetwoinsertsintheheadendofthemattresssupport.
Theinsertsmustbemountedinsuchawaythatoneextends
furtheroutofthesidetubethantheother.
2.Looselyscrewinthetwothumbscrews.
3.Pushthefootendofthemattresssupportontothetwoinserts.
4.Tightenwithtwothumbscrews.
5.Retightenthetwothumbscrewsattheheadendofthetopframe.
6.Turnupthefourmattresssupportbrackets.
7.Turnthehandleofthelegsectionup.
8.PressthebracketAonthelockingring*andturnthelockingring
onthebedendstotheposition"open".
1
2
A
9.Latchthemattresssupporttothebedendsandpressfirmly
intoposition.
10.Makesurethatthebedendsareinthesamelevel.Itispossible
tolevelthebedinhighposition(1)andlowposition(2).
11.Turnthelockingringtotheposition"lock".
*Thelockingringsensurethatthemattresssupportisnotaccidentally
liftclearoffthebedends.
ControlBox
Thecontrolboxislatchedtothebackrestmotor.Thecontrolbox
isprovidedwithalabelwithsymbolsshowingwheretoconnectthe
motorplugs:
Backrestmotor
Thighsection’smotor
Bedendmotor
Handcontrol
3.3Wiring
IMPORTANT!
Thecablesmustbemountedinsuchawaythatthey
arekeptclearoftheflooranddonotblockthecastors.
Mountthemainscableonthehookforthispurpose,
seepicture.
61567454-A
Setup
Inordertopreventthecablesfrombeingtornapartwhenactivating
themotors,followtheinstructionsbelow.
1.Connectthecablesofthebedendmotors(headandfootend)
directlytothecontrolbox.
2.Theplugofthethighmotormustbedirectedthroughtheopening
ofthesupporterforthebackrestmotorandthenbeconnected
tothecontrolbox.
3.Connectthemaincableplugtothe230Vsocket.
4.Runthemotorsofthebedendstotheirtopposition.
5.Placethemotorcableofthefootendonthehookatthefootend.
6.Runthebackresttoitstopposition.
7.Placetheheadendmotorcablesonthefourhooksatthehead
end.
8.Fastenthesafetycliptothecontrolbox.
3.4Dismantlethebed
1.Dismantlesiderailsandliftingpole.
2.Bringthebedtoitslowestpositionandadjustallmattress
supportsectionstohorizontalposition.
3.Disconnectthe230Vsupply.Rollthecableontothehookof
theheadendofthebed.
4.Disconnectallthemotorplugsfromthemotors(donot
disconnectthebackrestmotorcable).
5.Dismantlethesafetycombfromthecontrolbymeansofatool,
e.g.aflatheadedscrewdriver.
6.Pulloutthecablesofthebedendmotorsandlegsectionmotor.
7.Separatethemattresssupportfromthebedends.
8.Dividetheheadandfootendofthemattresssupport.
9.Forbedswith24Vexternalpowersupply,verifythatmainscable
isattachedtocontrolunitpowersupplycable.
1567454-A7
EtudePlus
4Operatingthebed
4.1Operatingthehandcontrolunit
HB80HL80HL84
A
B
C
A
B
C
E
A
B
C
D
E
p
Up
q
Down
A
Sittingposition
B
Thightsection
C
Heightadjustmentofthebed
D
Tiltfunction
E
Keyforlockingandunlockingtheelectricalfunctions
Insertthekeybeloweachbuttontorespectivelylockand
unlockthehandcontrolfunctions.
1.UsethebuttonsAtoDonthehandcontroltooperatethe
electricalfunctionsofthebed.
2.InsertthekeyEbeloweachbuttonandturnittorespectively
lockandunlockthehandcontrolfunctions(handcontrolunit
withlockingfunctiononly).
Ifthebedisusedbyrestlessorconfusedpersonsorpersons
withspasms,thebedshouldbeequippedwithalockable
handcontrolunittoblockthehandcontrolfunctions.
Tiltfunction
Thebedcanbeequippedwithahandcontrol(HL84)thatactivates
thetiltofthemattresssupport.Thetiltfunctionisforuser’scomfort
whensittinginthebed.
CAUTION!
Riskofinjury
Onlymedicallytrainedpersonsmustoperatethetilt
function.
ThetiltfunctionisNOTaTrendelenburgerfunction
andmustnotbeusedformedicaltreatment.
Whenthetiltfunctionisnotneeded,lockthetilt
functionwiththekeyEonthehandcontrolHL84.
4.2Operatingthecastorbrakes
Eachofthecastorsisprovidedwithbrakesforbothlengthwiseand
crosswiselocking.Thebrakesarefoot-operated.
Whenthebedispositionedcorrectly,atleastonecastoratthehead
endandonecastoratthefootendmustbebraked.
WARNING!
Riskofinjury
Thebed’scastorsmustbelockedwhilstnursinga
patientinbedorwhenusinganypositioningfunction.
Thebed’scastorsmustbelockedwhenthepatient
movinginoroutofthebed.
CAUTION!
Riskoftrapping
Donotreleasethebrakewiththefingers.
1. 2.
1.Forbraking,steponthepedal
2.Forreleasingthebrake,steponthereleasepedal
Set-offfromthecastorsmightunderspecialconditions
appearatdifferenttypesofabsorbingfloorcoveringincluding
untreatedorbadlytreatedfloors.Inmattersofdoubt,it’s
recommendedtoplaceasuitablekindofprotectionbetween
thecastorsandthefloor.
4.3Adjustingthelegsection
Operatethelegsectionbyliftingthemattresshandle:
1.Up:Liftthemattresshandleonthelegsection.
2.Down:Liftthemattresshandleonthelegsectionrightupand
thenlowerthelegsection.
4.4Emergencyreleaseofthebackrestorthigh/leg
section
Anemergencyreleaseofthebackrestorthigh/legsectioncouldbe
necessaryinthecaseofpower-ormotorfailure.
CAUTION!
Riskofinjury
Aminimumoftwopersonsarerequiredtoreleasea
mattresssection.
AnemergencyreleaseoftheheightadjustmentisNOT
possible.
Removetheplugfromthemainssocketbefore
emergencyreleaseofthemattresssupport.
1.Bothpersonsholdthemattresssection.
2.Oneofthempullsoutthecotterpin(fromthemotorinquestion).
3.Bothpersonsslowlylowerthemattresssectionuntilitis
completelydown.
81567454-A
Accessories
5Accessories
5.1Listofavailableaccessories
Siderails
ScalaBasic2
3/4lengthcollapsiblesteelsiderail(168cmx40cm),mounted
onthesidetube.Releasebuttonintheheadsection
ScalaBasicPlus2
3/4lengthcollapsiblesteelsiderail(168cmx40cm),mounted
onthesidetube.Releasebuttonintheheadsection
ScalaMedium2
3/4lengthcollapsiblesteelsiderail(165cmx46cm),mounted
onthesidetube.Releasebuttonintheheadsection
ScalaDecubi2
3/4lengthcollapsiblesteelsiderail(168cmx54cm),mounted
onthesidetube.Releasebuttonintheheadsection
BrittV
Fulllengthsiderail(205cmx40cm)inwood,mountedon
thebedendrail.Twocenteredreleasebuttonsforfolding
ofthegate
Linestandard
Fulllengthsiderail(206cmx40cm)inaluminumorwood
finish,mountedonthebedendrail.Twocenteredrelease
buttonsforfoldingofthegate
Lineextendable
Fulllengthsiderailwithtelescopicextension(206-226cmx
40cm)inaluminumorwoodfinish,mountedonthebedend
rail.Twocenteredreleasebuttonsforfoldingofthegate
Mattresssupportextension10cm
Mattresssupportextension20cm
Liftingpole
Transportfitting
Handcontrolwithtilt,HB84
Handcontrolwithtilt,HL84,withintegratedACPfunction
Backupbatteryandbatterybox
Important
Useonlyoriginalaccessoriesandspareparts.Spare
partslistsandextrausermanualscanbeorderedfrom
Invacareorviahomepage.
5.2Siderails
WARNING!
Riskofentrapmentorsuffocation
There’sariskofentrapmentofsuffocationbetween
mattresssupport,siderailandbedend.
Alwaysensurecorrectfittingofthesiderails.
Steelsiderails
WARNING!
Riskofentrapment
There’sariskofentrapmentofsuffocationbetween
mattresssupport,siderailandbedend.
Whenusingsteelsiderails,alwaysmakesurethatthe
distancebetweenthebedendandthesiderail’shandle
upperedgeislessthan6cmintheheadendandmore
than32cminthefootendofthebed.
Astickerisplacedonthesiderailtoindicatetheminimumspace
betweensiderailandtheheadendofthebed:
CAUTION!
There’sariskofentrapmentoffingersonassemblyand
operationofthesiderail.
Mountinganddismounting
Thesiderailmustbemountedwiththereleasesystematthehead
endofthebed.
1.Mounting
Theforklinksonthesiderailmustbemountedaccordingto
theinstructiononthesiderail.
Tightenthesiderailwiththetwofingerscrews.
2.Dismounting
Loosenthetwofingerscrewsandremovethesteelsiderail.
Operating
1. 2.
1.Up:Liftandpullthetopbarofthesiderailtowardstheendwith
thelockingmechanism.
Ensure,itislockedinplace.
2.Down:Pressthereleasebuttonandpushthetopbaroftheside
railawayfromthelockingmechanism
CAUTION!
Ifthesiderailisnotlockedproperlyitcanfalldown.
Pull/pushthetopbarofthesiderailtoensurethatthe
lockingsystemisproperlyengaged.
LineandBrittVsiderails
WARNING!
Riskofentrapment/suffocation
Toavoidthepatientfromentrappingoraccidentally
fallingfromthemattresssupport,itisessentialthat
theglidingshoesarecorrectlyguidedintothebedend
rails.Adjustthesiderailbarsbyhand,toensurethey
aremountedcorrectly.
Mounting
CAUTION!
Makesurethatthestrapsbetweentherailsarenot
damagedorloose.
1.Mountoneendatatime.Lifttheuppersiderailbar(therelease
buttonsmustfaceup/outwards)
2.Pressthelockingpinattheendofthesiderailwithafinger.
3.Allthreeglidingshoesattheendofthebarsareguidedinto
thebedendrail,untilthelowerbarhasbeenlocked,thereby
preventingthesiderailfromfallingdown.
4.Repeatpoints2and3whenmountingtheotherendoftheside
rail.
1567454-A9
EtudePlus
Dismounting
1.Lowerthesiderail.
2.Dismountoneendatatime.Liftthelowerbar,sothatthespring
pawlisvisible.
3.Pressthespringpawlwithakeyorascrewdriver.
4.Holdthelowerbar.Lowerthesiderail,untilallthreebarsare
freeofthebedendrail.
5.Repeatpoints2to4whendismountingtheotherendoftheside
rail.
Operating
Asastandard,thesiderailscanonlybelockedintopposition.
1.Tolock/raisetoupperposition:Pullupthetopsiderailbar,
untilthelockingpinlockswithanaudibleclick
2.Toreleasetolowerposition:Liftthetopsiderailbarand
pressthetwolockingringstogetherandlowerthesiderail.
5.3Liftingpole
WARNING!
Riskofinjury
Iftheliftingpoleisusedwhilethehandlehasbeenturned
awayfromthebed,thebedcantipwhenthehandleis
used.
Positiontheliftingpoleinsuchawaythatthehandle
extendsinwardsacrossthebed.
Mounting
1.Removetheplugfromtheliftingpoletubeattheheadendofthe
bedwheretheliftingpoleistobeplaced.
2.Inserttheliftingpoleintotheliftingpoletubeandfixwiththe
thumbscrew.
Adjustingtheheightoftheliftingpolehandle
1.Loosenthecord.
2.Theliftinghandlecannowbeadjustedtothedesiredheight.
3.Pressthecordtogetherandcheckthatthecordislockedinthe
cordlockbypullingthehandle.
Thedistancebetweentheliftingpoleandthemattress
Minimum(at16cmmattress)36-59cm
Maximum(at10cmmattress)42-65cm
5.4Mattresssupportextension
Forpatients,tallerthantwometersitisrecommendedto
useamattresssupportextension.
Fittingthemattresssupportextension
1.Disassemblethebedandremovetheinserts.
2.Mountthemattresssupportextensionbetweenbothpartsofthe
mattresssupportandreassemblethebed.
5.5Mattress
WARNING!
Safetyaspectsregardingcombinationofsiderails
andmattresses:
Inordertoobtainthehighestpossiblesafetylevel,
whenusingthebedincombinationwithsiderails,it
isimportanttorespecttheminimumandmaximum
measuresformattressesstatedinthetable.
Tomitigatethepossibilityoftheusercrawlingoverthe
topofthesiderail,withtheriskofsevereinjurydue
toe.g.falling,aminimumverticaldistanceof22cm
betweenthetopofthemattressandthetopofthe
siderailaretoberespected.Themaximummattress
heightincombinationwithagivensiderailisstated
inthetable.
Topreventheadentrapmentbetweentheinside
surfaceofthesiderailandthemattress,withtherisk
ofe.g.suffocation,asafetymaximumspacehastobe
respected.Whenespeciallyverythickmattressesor
softmattresses(lowdensity)oracombinationofthese
itisimportanttofocusonthisrisk.Theminimum
width(andlength)ofmattressesincombinationwitha
givensiderailisstatedinthetable.
Mattressmeasures(incm):
Siderail:
Min.
height
Max.
height
Min.
width
Min.
length
ScalaBasic210178590200
ScalaBasicPlus210178590200
ScalaMedium210228590200
ScalaDecubi223318590200
Line*,mattresssupport
platforminlowposition
11228890200
Line*,mattresssupport
platforminhighposition
5
158890200
BrittV,mattresssupport
platforminlowposition
11228590200
BrittV,mattresssupport
platforminhighposition
5
158590200
101567454-A
Accessories
*bothLinesiderailsstandardandextendable.
Min.density:38kg/m
3
5.6Shippingbrackets/Transportbrackets
1.
2.
3.
4.
1567454-A
11
EtudePlus
6Maintenance
6.1Maintenanceschedule
Important
Serviceandmaintenanceofthebedmustonlybeperformedbypersonnelwhohavereceivedthenecessaryinstructionortraining.
Withnormaldailyuse,servicemustbecarriedoutaccordingtotheservicescheduleaftertwoyearsofuseandthereaftereverysecondyear.
Whenrelocatingthebed,servicemustbecarriedoutaccordingtotheserviceschedule.
BeforeUseEnsurethatallmanualandelectricalpartsarefunctioningcorrectlyandareinasecurestate.
AfterthreemonthEnsurethatallmanualandelectricalpartsarefunctioning,andtightenbolts,screws,nuts,etc.
Everytwoyears
Serviceshouldbeperformedaccordingtothemaintenancechart.Seebelow.
Bed/Id.no.:______________________________________
Date/Initials
Visualinspectionofallpartsofthebed(plasticdeformationand/orwearandtearof
weldedjoints)
Controlallcentresofrotation(motors+mattresssupportparts)
Checkthatallmotorsarerunningwithoutfailures(withregularspeedandatlownoise)
ChecktheRastofixfittingsandfunction
Visualinspectionofallcabinetsfordamages.
Checkthatmainscableandplugareintact.Checktherunningofthecable.
Checkallremainingcablesfordamages.Alsochecktherunningofthecables.
Checkofsiderails´fixingandlocking/movement.
Checkthewheels(security,brakingandfreerolling)
Lubricationplan
Werecommendlubricatingthebedaccordingtothefollowinginstructions:
Lubricatewithmedicallycleanoil,e.g.KEN-WO50:
Pointsofrotationinmattresssupportandbaseframe-lubricatewithoil.
Allofthemotors’tensionrod-lubricatewithoil.
AllmotorbearingsandRastofixfitting-lubricatewithoil.
Thewoodensiderailsglidingsystemmustnotbelubricatedwithoil-otherwisethewoodenbarswillmovesluggishly.
6.2Cleaning
Important
Makesurethepowerplugisremovedfromthesocketoutletbeforecleaning.
Allplugsmustbeinstalledduringwashingofthebed.
Thebedshouldbewasheddownusingawetsponge,clothorbrush.
Thebeddoesnottoleratecleaninginautomaticwashingplantsortheuseofwater-jetbasedcleaningequipment.
Useordinaryhouseholdcleaningagents.Neveruseacids,alkalinesorsolventssuchasacetoneorcellulosethinner.
Thehandcontrol,motorsandcontrolunitmaybewashedwithbrushandwater,butnotwithpressurizedwater.
Drythebedaftercleaning.
12
1567454-A
AfterUse
7AfterUse
7.1Wastedisposal
Thisproducthasbeensuppliedfromanenvironmentallyaware
manufacturerthatcomplieswiththeWasteElectricalandElectronic
Equipment(WEEE)Directive2012/19/EU.Thisproductmaycontain
substancesthatcouldbeharmfultotheenvironmentifdisposedofin
places(landfills)thatarenotappropriateaccordingtolegislation.
The»crossedoutwheeliebin«symbol
isplacedonthisproducttoencourage
youtorecyclewhereverpossible.
Pleasebeenvironmentallyresponsibleandrecyclethisproduct
throughyourrecyclingfacilityatitsendoflife.
Allwoodenpartsmustbedismantledandsentforincineration.
Allelectricpartsmustbedismantledandbedisposedofaselectric
components.
Plasticpartsmustbesentforincinerationorrecycling.
Steelpartsandcastorsmustbedisposedofaswastemetals.
Wastedisposalmustcomplywiththelawsandregulationspertaining
wastehandlingineachcountry.
1567454-A13
EtudePlus
8Troubleshooting
8.1Troubleshootingelectricalsystem
Symptom
Possiblecause
Remedy
Mainsisnotconnected
Connectmains
FuseinthecontrolunitisblownReplacethecontrolunit Mainsindicatordoesnotlightup
Controlunitisdefective
Replacethecontrolunit
Motorplugisnotfullyinsertedintothe
controlunit.
Insertthemotorplugproperlyintothe
controlunit
Themotorisdefective.
Replacethemotor
Motorcableisdamaged.Replacethecable
Mainsindicatorlightsup,butthemotorisnot
running.Therelayinthecontrolunitmakesa
clickingnoise.
Controlunitisdefective
Replacethecontrolunit
Controlunitisdefective
Replacethecontrolunit
Mainsindicatorlightsup,butthemotorisnot
running.Norelaysoundisheardfromthe
controlunit.
Handcontrolisdefective
Replacethehandcontrol
Controlunitisdefective
Replacethecontrolunit
Controlunitisinorderexceptforone
directionononechannel.
Handcontrolisdefective
Replacethehandcontrol
Motorisrunning,butthepistonroddoesnot
move.
Themotorcannotliftfullload.
Motornoise,butnomovementofpistonrod.
Pistonrodoperatesinwardsandnotoutwards.
MotorisdamagedReplacethemotor
14
1567454-A
Technicaldata
9Technicaldata
9.1Dimensions
Allmeasurementsarestatedincm.Allanglesarestatedindegrees.
Allmeasurementsandanglesarestatedwithouttolerances.
Invacare®reservestherighttochangethestatedmeasurementsand
angles.
A
B
D
C
EtudePluslowEtudePlus
Low
High
Low
High
A
219219219219
B
102102102102
C
20–6027–6733–7340–80
D
138–178145–185158–198165–205
Theanglebetweenthelowerlegsectionandhorizontalisadjustable
fromto15º.
Bedwithtiltfunction
9.2Weights
Max.patientweight
(providedthattheweightofthemattress
andtheaccessoriesdonotexceed35kg)
165kg
Max.safeworkingload
(patient+accessories)
200kg
EtudePlusstandardbed,complete,excluding
accessories
77.3kg
EtudePlusauto-regression,complete,excluding
accessories
82.7kg
EtudePlusbedendwithextraplate1piece20.6kg
EtudePlusbedendwithoutextraplate1piece17.9kg
Mattresssupport,standardheadend
24.1kg
Mattresssupportwithauto-regression,headend
27.9kg
Mattresssupportwithauto-regression,headend,
withoutactuatorandcontrolbox
24.1kg
Mattresssupport,footend17.4kg
ScalaBasic2siderail1piece
7.4kg
Liftingpole4.2kg
Mattresssupportextension(10cm)3.2kg
Mattresssupportextension(20cm)
5.5kg
9.3Environmentalconditions
Storage
Operationand
Transportation
Temperature
-10°Cto+50°C+5°Cto40°C
Relativehumidity
20%to75%
Atmospheric
pressure
800hPato1060hPa
Beawarethatwhenabedhasbeenstoredunderlow
temperatures,itmustbeadjustedtooperatingconditions
beforeuse.
9.4Electricalsystem
Voltagesupply:Uin=230Volt,AC,50Hz(AC=Alternating
current)
Maximumcurrentinput:Iinmax.=2A
Intermittent(periodicmotoroperation):Int=max.10%,2min/
18min
Degreeofprotection:IPx6*orIPx4**(dependingonconfiguration)
Thecontrolunit,externalpowersupplyandmotorsareprotected
accordingtoIPx6.WhenthebedisequippedwithaHL80hand
control,itisprotectedaccordingtoIPx4
Insulationclass:ClassII
TypeBAppliedPart
AppliedPartcomplyingwiththespecifiedrequirementsfor
protectionagainstelectricalshockaccordingtoIEC60601-1.
(Anappliedpartsisapartofthemedicalequipmentwhichis
designedtocomeintophysicalcontactwiththepatientorpartsthat
arelikelytobebroughtintocontactwiththepatient.)
Soundlevel:45to50dB(A)
*IPX6classificationmeansthattheelectricalsystemisprotected
againstwaterprojectedinpowerfuljetsfromanydirection.
**IPX4classificationmeansthattheelectricalsystemisprotected
againstwatersplashedagainstthecomponentfromanydirection.
Thebedisnotprovidedwithamainsswitch,sothemains
plugistheonlyseparationfromthemains.
1567454-A15
EtudePlus
9.5Electromagneticcompliance(EMC)
Guidanceandmanufacturer´sdeclarationelectromagneticemission
Themedicalbedisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuserofthebedshouldassure
thatitisusedinsuchanenvironment.
Emissionstest
Compliance
Electromagneticenvironment-guidance
RFemissions
CISPR11(partly)
GroupI
ThemedicalbedusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsare
verylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.
RFemissions
CISPR11(partly)
ClassB
Themedicalbedissuitableforuseinallestablishmentsincludingdomesticestablishments
andthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsupplies
buildingsusedfordomesticpurposes.
Harmonicemissions
IEC61000-3-2
ClassA
Voltagefluctuations/
flickeremissions
IEC61000-3-3
Complies
Guidanceandmanufacturer´sdeclarationelectromagneticimmunity
Themedicalbedisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuserofthebedshouldassure
thatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevel
Electromagneticenvironmentguidance
Electrostatic
discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.Iffloors
arecoveredwithsyntheticmaterial,therelativehumidity
shouldbeatleast30%.
Electrostatic
transient/burst
IEC61000-4-4
±2kVforpowersupply
lines
±1kVforinput/outputlines
±2kVforpowersupply
lines
±1kVforinput/outputlines
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.
Surge
IEC61000-4-5
±1kVdifferentialmode
±2kVcommonmode
±1kVdifferentialmode
±2kVcommonmode
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC61000-4-11
<5%U
T
(>95%dipinU
T
)
for0,5cycle
40%U
T
(60%dipinU
T
)for
5cycles
70%U
T
(30%dipinU
T
)for
25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
<5%U
T
(>95%dipinU
T
)
for0,5cycle
40%U
T
(60%dipinU
T
)for
5cycles
70%U
T
(30%dipinU
T
)for
25cycles
<5%U
T
(>95%dipinU
T
)
for5sec
Mainspowerqualityshouldbethatofatypicalcommercial
orhospitalenvironment.Iftheuserofthemedical
bedrequirescontinuedoperationduringpowermains
interruptions,itisrecommendedthatthemedicalbed
bepoweredfromanun-interruptiblepowersupplyor
abattery.
U
T
isthea.c.mainsvoltagepriortoapplicationofthe
testlevel.
Powerfrequency
(50/60Hz)magnetic
field
IEC61000-4-8
3A/m3A/m
Powerfrequencymagneticfieldsshouldbeatlevels
characteristicofatypicallocationinatypicalcommercial
orhospitalenvironment.
161567454-A
Technicaldata
PortableandmobileRFcommunicationsequipment
shouldbeusednoclosertoanypartofthemedical
bedincludingcables,thantherecommendedseparation
distancecalculatedfromtheequationapplicabletothe
frequencyofthetransmitter.
Recommendedseparationdistance:
ConductedRF
IEC61000-4-6
3V3V
RadiatedRF
IEC61000-4-3
3V/m3V/m
80MHzto800MHz
800MHzto2,5GHz
wherePisthemaximumoutputpowerratingofthe
transmitterinwatts(W)accordingtothetransmitter
manufactureranddistherecommendedseparation
distanceinmetres(m).
b
FieldstrengthsfromfixedRFtransmitters,asdetermined
byanelectromagneticsitesurvey,
a
shouldbelessthanthe
compliancelevelineachfrequencyrange.
b
Interferencemayoccurinthevicinityofequipment
markedwiththefollowingsymbol:
a
Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephonesandlandmobileradios,amateurradio,AM
andFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywithaccuracy.Toassesstheelectromagneticenvironmentduetofixed
RFtransmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthemeasuredfieldstrengthinthelocationinwhichthemedicalbedisused
exceedstheapplicableRFcompliancelevelabove,themedicalbedshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceis
observed,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingthemedicalbed.
b
Overthefrequencyrange150kHzto80MHz,fieldstrengthsshouldbelessthan[V1]V/m.
At80MHzand800MHz,thehigherfrequencyrangeapplies.
RecommendedseparationdistancesbetweenportableandmobileRFcommunicationsequipmentandthemedical
bed
ThemedicalbedisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.Thecustomeror
theuserofthemedicalbedcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobile
RFcommunicationsequipment(transmitters)andthemedicalbedasrecommendedbelow,accordingtothemaximumoutputpowerof
thecommunicationsequipment
Separationdistanceaccordingtofrequencyoftransmitter
[m]
Ratedmaximumoutputof
transmitter
[W]
150kHzto80MHz
80MHzto800MHz
800MHzto2,5GHz
0.010.120.120.23
0.10.370.370.74
11.171.172.33
103.693.697.38
10011.6711.6723.33
Fortransmittersratedatamaximumoutputpowernotlistedabovetherecommendedseparation,distancedinmetres(m)canbeestimated
usingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)
accordingtothetransmittermanufacturer.
At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreflectionfromstructures,
objectsandpeople.
1567454-A17
Notes
Notes
InvacareSalesCompanies
Ireland:
InvacareIrelandLtd,
Unit5SeatownBusinessCampus
SeatownRoad,Swords,County
Dublin
Tel:(353)18107084
Fax:(353)18107085
www.invacare.ie
UnitedKingdom:
InvacareLimited
PencoedTechnologyPark,Pencoed
BridgendCF355AQ
Tel:(44)(0)1656776222
Fax:(44)(0)1656776220
www.invacare.co.uk
Manufacturer:
InvacareLda
RuaEstradaVelha949
4465-784LeçadoBalio,Portugal
1567454-A2013-04-01
*1567454A*
MakingLife’sExperiencesPossible™
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12