wtw inoLab pH/ION 735 Operating instructions

Category
Measuring, testing & control
Type
Operating instructions
inoLab pH/ION 735
ba75449e02 03/2005
Operating manual
pH/ION meter
p
H
/IO
N
735
Accuracy when
going to press
The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may
result in differences between this operating manual and your
instrument. Also, we cannot guarantee that there are absolutely no
errors in this manual. Therefore, we are sure you will understand that
we cannot accept any legal claims resulting from the data, figures or
descriptions.
Warranty We guarantee the instrument described for 3 years from the date of
purchase.
The instrument warranty covers manufacturing faults that are
discovered within the warranty period.
The warranty does not cover components that are replaced during
maintenance work, e. g. rechargeable batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper
handling or unauthorized opening of the instrument invalidates any
warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
Topicality of firmware The continuous development of our meter software is part of the
process of consequently improving our products. The current meter
software for the inoLab pH/ION 735 can be found on the Internet. It can
easily be transmitted to your meter with the aid of the supplied
AK 340/B cable and a PC.
More information is given in the appendix of this operating manual or
on the Internet under http://www.WTW.com.
Copyright
© Weilheim 2005, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the
express written permission of WTW GmbH, Weilheim.
Printed in Germany.
inoLab pH/ION 735 Contents
3
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 General features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 General operating principles . . . . . . . . . . . . . . . . . . . . . 14
4.2.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 14
4.2.2 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2.3 Navigation example 1: Setting the language . . . 17
4.2.4 Navigation example 2: Setting the date and time 19
4.3 System settings (system menu) . . . . . . . . . . . . . . . . . . . 21
4.3.1 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3.2 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.4 pH value / ORP voltage . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.4.1 General information . . . . . . . . . . . . . . . . . . . . . . 23
4.4.2 Measuring the pH value . . . . . . . . . . . . . . . . . . 24
4.4.3 Measuring the ORP voltage . . . . . . . . . . . . . . . 25
4.4.4 Settings for pH and ORP measurements . . . . . 25
4.4.5 pH calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.4.6 Carrying out the TEC and NIST/DIN calibration 30
4.4.7 Carrying out a ConCal calibration . . . . . . . . . . . 34
4.5 Ion concentration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.5.1 General information . . . . . . . . . . . . . . . . . . . . . . 37
4.5.2 Measuring the ion concentration . . . . . . . . . . . . 38
4.5.3 Settings for ISE measurements . . . . . . . . . . . . 38
4.5.4 Calibrating for ISE measurements . . . . . . . . . . 40
4.5.5 Selecting the measuring method . . . . . . . . . . . . 45
4.5.6 Standard addition (Std. add.) . . . . . . . . . . . . . . 46
4.5.7 Standard subtraction (Std. sub.) . . . . . . . . . . . . 48
4.5.8 Sample addition (Sample add.) . . . . . . . . . . . . . 51
4.5.9 Sample subtraction (Sample sub.) . . . . . . . . . . 53
4.5.10 Standard addition with blank value correction
(Blank add.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Contents inoLab pH/ION 735
4
4.6 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4.6.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . 59
4.6.2 Automatic storage at intervals . . . . . . . . . . . . . . 60
4.6.3 Editing the measured value storage . . . . . . . . . 62
4.6.4 Erasing the measured value storage . . . . . . . . . 65
4.6.5 Displaying and outputting calibration records . . 66
4.7 Transmitting data (RS 232 interface) . . . . . . . . . . . . . . . 68
4.7.1 Options for data transmission . . . . . . . . . . . . . . 68
4.7.2 Connecting a PC/external printer . . . . . . . . . . . . 69
4.7.3 Operation with MultiLab pilot . . . . . . . . . . . . . . . 69
4.8 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4.8.1 Resetting the sensor settings . . . . . . . . . . . . . . 70
4.8.2 Resetting the system settings . . . . . . . . . . . . . . 71
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 73
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
5.1.1 Exchanging the rechargeable batteries . . . . . . . 74
5.1.2 Exchanging the roll of paper . . . . . . . . . . . . . . . 75
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
6.1 pH and ORP measurement . . . . . . . . . . . . . . . . . . . . . . 77
6.2 ISE measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
6.3 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
7.2 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 84
7.2.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
7.2.2 ISE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
9 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Appendix: Updating the meter software . . . . . . . . . . . . . . . 91
inoLab pH/ION 735 Overview
5
1Overview
1.1 General features
The inoLab pH/ION 735 compact precision pH meter enables you to
carry out pH measurement, ORP measurements and concentration
measurements with ion sensitive electrodes quickly and reliably.
The inoLab pH/ION 735 provides the maximum degree of operating
comfort, reliability and measuring certainty for all applications.
The integrated printer* enables to document measurements according
to GLP requirements.
The proven MultiCal
®
calibration procedures and the special AutoRead
function support you in your work with the meter.
Note
If you need further information or application notes, you can obtain the
following material from WTW:
! Application reports
! Primers
! Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
* only on the inoLab pH/ION 735 with integrated printer
1 Keypad
2 Display
3 Socket field
4 Printer*
pH
/IO
N
735
3
2
1
4
Overview inoLab pH/ION 735
6
1.2 Keypad
1 Paper feed / line feed*
(switch printer off - Print LED lights up red)
* only on the inoLab pH/ION 735 with integrated printer
2 Switch printer on or off*
3 Print-LED*
red: printer off (off-line) or malfunction
green: printer ready for operation (on-line)
4 Call up calibration procedure
5 Switch between measuring mode pH/ORP and measuring
mode ISE concentration
6 Transmit / print displayed data
7 Switch the meter on/off
8 Cancel inputs
9 Store a measured value
10 Switch between pH measurement and ORP measurement, in-
crease values, scroll
11 Confirm entries, start AutoRead
12 Switch between pH measurement and ORP measurement,
decrease values, scroll
1
2
3
4
9
10
11
5
6
7
8
12
inoLab pH/ION 735 Overview
7
1.3 Display
The graphical display can indicate up to three measuring windows at
the same time. The illumination enables to read the display even in the
darkness. You can modify the display to meet your requirements in
multiple ways.
1.4 Socket field
Connections:
1 pH-Ion electrode
2 Temperature probe
3 Reference electrode
4 RS 232 interface/analog output
5 Power pack
pH
6.949
25.0
°C
05.08.04 09:25
Measured variable
Measured value (with unit)
Sensor symbol
Temperature display
Status line
1
2
3
4
5
Overview inoLab pH/ION 735
8
Warning
Only connect sensors to the meter that cannot return any voltages or
currents that are not allowed (> SELV and > current circuit with current
limiting).
Almost all sensors - in particular WTW sensors - fulfill these conditions.
inoLab pH/ION 735 Safety
9
2 Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating man-
ual before working with the measuring system. The operating manual
must always be available within the vicinity of the measuring system.
Target group The meter was developed for work in the laboratory.
We assume that, as a result of their professional training and experi-
ence, the operators will know the necessary safety precautions to take
when handling chemicals.
Safety instructions The individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manu-
als.
2.1 Authorized use
This meter is authorized exclusively for pH, ORP and ISE concentra-
tions measurements in the laboratory.
The technical specifications as given in chapter 7 T
ECHNICAL DATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized. Any other
use is considered to be unauthorized.
Safety inoLab pH/ION 735
10
2.2 General safety instructions
This instrument is built and inspected according to the relevant guide-
lines and norms for electronic meters (see chapter 7 T
ECHNICAL DATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during opera-
tion.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty func-
tioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
Safe operation If safe operation is no longer possible, the instrument must be taken out
of service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
! has been damaged in transport
! has been stored under adverse conditions for a lengthy period of
time
! is visibly damaged
! no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
! EEC directives for protective labor legislation
! National protective labor legislation
! Safety regulations
! Safety datasheets of the chemical manufacturers.
inoLab pH/ION 735 Commissioning
11
3 Commissioning
3.1 Scope of delivery
! inoLab pH/ION 735 laboratory meter
! Operating manual and short instructions in the printer cover
! 4 NiMH rechargeable batteries 1.2 V Mignon type AA
! Power pack
3.2 Power supply
Line-powered operation
and charging the batter-
ies
You can operate the meter either with the built-in rechargeable batter-
ies or with the power pack. The power pack supplies the meter with low
voltage (9 V DC). At the same time, the rechargeable batteries are
charged. The batteries are charged even when the instrument is
switched off.
Warning
Always use the specified NiMH rechargeable batteries (see chap-
ter 7 T
ECHNICAL DATA). Using normal batteries can damage the
meter.
Charging time of the
batteries
approx. 36 hours. The LoBat display indicator appears when the batter-
ies are nearly empty and have to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input volt-
age range of the original power pack (see chapter 7 T
ECHNICAL DA-
TA).
Warning
Use original power packs only (see chapter 7 TECHNICAL DATA).
Note
The batteries should not be completely discharged. If you do not oper-
ate the instrument for a longer period of time you should charge the bat-
teries every six months.
Automatic switchoff The instrument has an automatic switchoff function in order to save the
batteries (see section 4.3.2).
Commissioning inoLab pH/ION 735
12
Display illumination
with battery operation
To save the batteries, the meter has a switchoff function for the display
illumination that can be switched on and off (see section 4.3.2).
Note
You can carry out measurements without the power pack. The func-
tions of the integrated printer*, however, can only be used if the power
pack is connected.
3.3 Initial commissioning
Perform the following activities:
! For line-powered operation and charging the batteries: Connect the
power pack (see section 3.2 P
OWER SUPPLY).
! Switch on the meter (see section 4.1)
! Set the language (see section 4.2.3)
! Set the date and time (see section 4.2.4)
Note
When you set the language, date and time according to the mentioned
sections of this operating manual you will quickly be familiar with the
simple operation of the inoLab pH/ION 735.
1 Connect the plug (1) to the socket (2) of the meter.
2 Connect the original WTW power pack (3) to an easily
accessible power socket.
* only on the inoLab pH/ION 735 with integrated printer
1
2
3
inoLab pH/ION 735 Operation
13
4 Operation
4.1 Switching on the meter
Switching on Press the e key.
The measured value display appears.
Switching off Press the e key.
Automatic switchoff The instrument has an automatic switchoff function in order to save the
batteries (see section 4.3.2). The automatic switchoff switches off the
meter if no key is pressed for an adjustable period.
The automatic switchoff is not active
! if the power is supplied by the power pack,
! if the Autostore function is active, or with automatic data transmis-
sion
! if the communication cable and a PC with a running communication
program are connected,
! if the printer cable is connected (for external printers).
Display illumination
with battery operation
During battery operation, the meter automatically switches off the dis-
play illumination if no key has been pressed for 15 seconds. The illumi-
nation is switched on with the next keystroke again.
pH
6.949
25.0
°C
20.09.04 10:37
Operation inoLab pH/ION 735
14
4.2 General operating principles
This section contains basic information of the operation of the inoLab
pH/ION 735.
Operating elements,
display
An overview of the operating elements and the display is given in
section 1.2 and section 1.3.
Operating modes,
navigation
An overview of the operating modes and navigation of the inoLab pH/
ION 735 is given in section 4.2.1 and section 4.2.2.
4.2.1 Operating modes
The instrument has the following operating modes:
! Measuring
The measuring data of one to three sensors is displayed in the
measured value display
! Calibration
The course of a calibration with calibration information, functions
and settings is displayed
! Storage
The meter stores measuring data automatically or manually
! Transmitting data
The meter transmits measuring data and calibration records to the
serial interface automatically or manually.
! Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed
inoLab pH/ION 735 Operation
15
4.2.2 Navigation
Measured value display In the measured value display, you can
! select a measuring window with 82 and
open the relevant measuring menu by shortly
pressing g.
! open the system menu with the sensor-independent settings by
pressing g for a long
time (approx. 1 s).
! change the display in the selected measuring window (e. g. pH <−>
mV) by pressing m.
Menus and dialogs The menus for settings and dialogs in courses contain further sub-
menus. The selection is made with the 82 keys. The current selec-
tion is displayed in reverse video.
! Submenus
The name of the submenu is displayed at the upper edge of the
frame. Submenus are opened by confirming with g.
Example:
! Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. The setting mode is opened with g. Subse-
quently, the setting can be changed with 82 and g.
Example:
General
Interface
Clock function
Reset
System
Language: Deutsch
Beep: Off
Illumination: On
Contrast
: 48 %
Temperature unit: °C
Switchoff time: 30 min
System
Operation inoLab pH/ION 735
16
! Functions
Functions are designated by the name of the function. They are im-
mediately carried out by confirming with g.
Example: Display the Calibration record function.
Messages Information or operating instructions are designated by the i symbol.
They cannot be selected.
Example:
Note
The principles of navigation are explained in the two following sections
by reference of examples:
! Set the language (see section 4.2.3)
! Set the date and time (see section 4.2.4).
Cal. type: TEC
Calibration interval
: 7 d
Unit for slope: mV/pH
Calibration record
:
i 2.00 4.01 7.00 10.01
pH
i Buffer recognition TEC
i Immerse sensor in buffer 1
Set temperature: 25 °C
Continue
pH
inoLab pH/ION 735 Operation
17
4.2.3 Navigation example 1: Setting the language
1 Press the e key.
The measured value display appears.
The instrument is in the measuring mode.
2 Open the system menu by pressing g for a long time
(approx. 1 s).
The instrument is in the setting mode.
3 Select the System submenu with 82.
The current selection is displayed in reverse video.
4 Open the System
submenu with g.
5 Select the General submenu with 82.
The current selection is displayed in reverse video.
pH
7.00
25.0
°C
20.09.04 10:37
Data storage
System
Storage & config
General
Interface
Clock function
Service information
Reset
System
Operation inoLab pH/ION 735
18
6 Open the General submenu with g.
7 Open the setting mode for the Language with g.
8 Select the required language with 82.
9 Confirm the setting with g.
The setting becomes active the next time the system menu is
called up.
Language: Deutsch
Beep
: Off
Illumination: On
Contrast: 48 %
Temperature unit
: °C
Switchoff time: 30 min
System
Language: Deutsch
Beep
: Off
Illumination: On
Contrast: 48 %
Temperature unit
: °C
Switchoff time: 30 min
System
inoLab pH/ION 735 Operation
19
4.2.4 Navigation example 2: Setting the date and time
The meter has a clock with a date function. The date and time are indi-
cated in the status line of the measured value display. The indication
can be switched off. When storing measured values and calibrating, the
current date and time are automatically stored as well.
The correct setting of the date and time and date format is important for
the following functions and displays:
! Current date and time
! Calibration date
! Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries), the date and time are
reset to 01.01.2003, 00:00 hours.
Setting the date, time
and date format
The data format can be switched from the display of day, month, year
(dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1 In the measured value display:
Open the system menu by pressing g for a long
time
(approx. 1 s).
The instrument is in the setting mode.
2 Select and confirm the System / Clock function menu with
82 and g.
The setting menu for the date and time appears.
3 Select and confirm Time with 82 and g.
The seconds are highlighted.
4 Change and confirm the setting with 82 and g.
The minutes are highlighted.
5 Change and confirm the setting with 82 and g.
The hours are highlighted.
Time: 14:53:40
Date: 30.10.03
Date format
: dd.mm.yy
System
Operation inoLab pH/ION 735
20
6 Change and confirm the setting with 82 and g.
The time is set.
7 If necessary, set the Date and Date format. The setting is made
similarly to that of the time.
8 If necessary, select and set the Date with 82 and g.
9 To make further settings, switch to the next higher menu level
with #.
or
Switch to the measured value display with m.
The instrument is in the measuring mode.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93