  • Hello! I am an AI chatbot trained to assist you with the BIO RAD 760-2030 Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the BioLogic DuoFlow System USB Bitbus Communicator?
    How do I power the USB Bitbus Communicator?
    How do I install the USB Bitbus Communicator?
    How do I install the USB Bitbus Communicator with SIM devices?
    What is the PWR Select switch for?
    How do I connect a single SIM device?
    How do I connect two SIM devices?
Dear Customer,
The Biologic DuoFlow System USB Bitbus Communicator is the communication link between the DuoFlow
Controller (computer), the DuoFlow Workstation and associated peripherals such as the DuoFlow Maximizer,
Signal Import Module (SIM), Econo Gradient Pump (EGP), and fraction collectors.
The USB Bitbus Communicator is powered directly from the Controller/
computer via the USB cable. In this case, the PWR Select switch should be
set to INT.
If one or two SIM devices are used, the USB Bitbus Communicator must be
powered by connecting a universal AC/DC in-line adaptor (Catalog number
760-2034). The PWR SELECT switch should then be set to EXT.
Installation Instructions
1. With the power switched off, connect the USB cable between an available USB port on the Controller/
computer (front or rear) and the USB port on the USB Bitbus Communicator. (front view)
2. With the power switched off, connect the instrument bus cable between the Workstation and the INSTR BUS
port on the USB Bitbus Communicator. (rear view)
3. The PWR Select switch should be set to INT.
4. Connect peripheral devices such as fraction collectors and auxiliary pumps to the Workstation via instrument
bus cables or the Aux connector.
FRONT VIEW
USB CABLE
INSTRUMENT BUS
REAR VIEW
REAR VIEW
REAR VIEW
WORKSTATION
USB BIT BUS COMMUNICATOR
BIOLOGIC CONTROLLER/COMPUTER
BioLogic DuoFlow System
USB Bitbus Communicator
Catalog Number 760-2030
FRONT VIEW
REAR VIEW
USB
PWR
SELECT
EXT INT
BUS
PWR
INSTR BUS
ON
POWER
+ 5 V
Installation with SIM Devices
1. With the power switched off, connect the USB cable between an available USB port on the Controller/computer
and the USB port on the USB Bitbus Communicator. (front view)
2. The PWR Select switch should be set to EXT.
3. Connect the universal AC/DC in-line adaptor (Catalog number 760-2034) to the USB Bitbus Communicator.
When the AC/DC in-line adaptor is plugged into a wall socket, power is supplied to the USB Bitbus
Communicator.
4. Switch on the Controller/computer and allow the Windows operating system to launch.
5. Verify that the BUS PWR light on the USB Bitbus Communicator is illuminated.
6. Connect one or two SIM devices to the USB Bitbus using the INSTR BUS port and cable. If two SIM devices
are connected, ensure that one is set to device #1 and the other to device #0 using the small selector switch.
7. With the power switched off, connect the Workstation instrument bus port to a SIM device.
8. When all devices are connected, turn on power to the Workstation.
2
REAR VIEW
TO WORKSTATION
TO WORKSTATION
SET TO 1
FOR SIM 1
FRONT VIEW
REAR VIEW
REAR VIEW
USB BITBUS COMMUNICATOR
USB BITBUS COMMUNICATOR
BIOLOGIC CONTROLLER/COMPUTER
SIGNAL IMPORT MODULE
CONNECTING A SINGLE SIM
CONNECTING TWO SIMs
REAR VIEW
SIM
UNIVERSAL
AC/DC INLINE ADAPTOR
WALL AC POWER
SET TO 0
FOR SIM 2
USB CABLE
INSTRUMENT BUS
REAR VIEW
USB BITBUS COMMUNICATOR
INSTRUMENT BUS
1