cobas c 513 analyzer
Operator's Manual
Version 5.0
Software Version 02-04
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
2
Publication information
Edition notice This publication is intended for operators of the
cobas c 513 analyzer.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version.
Where to find information The User Assistance contains all information about the
product.
The Safety Manual contains important safety
information. You must read the Safety Manual before
operating the system.
The Operator’s Manual focuses on operation and
maintenance. It contains the content below.
• System description
• Routine operation
• Maintenance
• Advanced operation
• Troubleshooting information
• Software description
Publication version Software version Revision date Change description
1.0 01-01 September 2015 Version 1.0
2.0 02-01 January 2016 Modified rinse station.
New air supply filter.
New tasks for cell detergent prime, downloading diluent
and detergent information, and masking of reagent
cassettes.
Minor revisions.
3.0 02-02 October 2016 New maintenance actions.
Added control for individual clot detection settings.
u What is new in publication version 3.0 (12)
4.0 02-03 July 2017 Added information about connecting to a CLAS.
Added task for programming calculated tests.
Added information about data alarm priority.
u What is new in publication version 4.0 (13)
5.0 02-04 January 2019 Updated task for cleaning the surfaces of the analyzer.
Minor revisions.
u What is new in publication version 5.0 (13)
y Revision history
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
3
!
General attention
To avoid serious or fatal injury, ensure that you are
familiar with the system and safety information before you
use the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the system in a way that is not described in
this publication.
r Store all publications in a safe and easily retrievable
place.
Training Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave tasks that are not described in the
user documentation to trained Roche Service
representatives.
Images The screenshots and hardware images in this publication
have been added exclusively for illustration purposes.
Configurable and variable data in screenshots, such as
tests, results, or path names visible therein must not be
used for laboratory purposes.
Warranty Any customer modification to the system renders the
warranty or service agreement null and void.
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Always leave software updates to a Roche Service
representative or perform such updates with their
assistance.
Copyright © 2015-2019, Roche Diagnostics GmbH. All rights
reserved.
Trademarks The following trademarks are acknowledged:
COBAS and COBAS C are trademarks of Roche.
All other trademarks are the property of their respective
owners.
Feedback Every effort has been made to ensure that this publication
fulfills the intended use. All feedback on any aspect of
this publication is welcome and is considered during
updates. Contact your Roche representative, should you
have any such feedback.
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
4
Approvals The cobas c 513 analyzer meets the requirements laid
down in:
Directive 98/79/EC of the European Parliament and of the
Council of 27 October 1998 on in-vitro diagnostic medical
devices.
Directive 2011/65/EU of the European Parliament and of
the Council of 8 June 2011 on the restriction of the use of
certain hazardous substances in electrical and electronic
equipment.
The following marks demonstrate compliance.
The cobas c 513 analyzer is manufactured and tested
according to the international safety standards below.
• IEC 61010-1
• IEC 61010-2-101
The RFID readers of the cobas c 513 analyzer meet the
requirements laid down in:
Directive 2014/53/EU of the European Parliament and of
the Council of 16 April 2014 on the harmonization of the
laws of the Member States relating to the making
available on the market of radio equipment and repealing
Directive 1999/5/EC.
The full text of the 2014/53/EU declaration of conformity
is available at the following internet address:
http://e-labdoc.roche.com
Complies with the IVD directive 98/79/EC and the
directive 2011/65/EU on RoHS.
Complies with the IVD directive 98/79/EC.
Complies with the directive 2011/65/EU on RoHS.
Issued by Intertek to prove compliance with safety
standards for Canada and the US.
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
5
Fluorinated greenhouse gas The product contains a fluorinated greenhouse gas in the
hermetically sealed refrigeration.
The insulation of the chamber includes foam blown with
fluorinated greenhouse gas.
License agreement for UltraVNC software on
the control unit
UltraVNC is a piece of free software for all commercial
uses. It is installed on the control unit of the
cobas c 513 analyzer.
You can redistribute the software and/or modify it under
the terms of the GNU General Public License (version 2 or
later), as published by the Free Software Foundation.
A copy of the GNU General Public License (version 2) is
stored on the control unit of the cobas c 513 analyzer.
The pathway for the license is C:\Program Files\uvnc
bvba\UltraVNC.
The software is distributed without warranty. There is no
implied warranty of merchantability or fitness for a
particular purpose. For more information, see the GNU
General Public License at http://www.gnu.org/licenses.
The source code for the software is stored on the control
unit PC of the cobas c 513 analyzer. The pathway for the
source code is C:\DriversAndTools\UltraVNC.
Type
Charge weight
(kg)
CO
2
equivalent
(tonne)
Global
warming
potential
R-134a 0.160 0.23 1430
y Fluorinated greenhouse gas detail
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
6
Contact addresses
Inside the European Union and
EFTA member states
Outside the European Union and
EFTA member states
Manufacturer of
cobas c 513 analyzer
Hitachi High-Technologies
Corporation
1-24-14 Nishi-Shimbashi
Minato-ku Tokyo 105-8717
Japan
Authorized
representative
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Manufactured by: Hitachi High-Technologies
Corporation
Manufactured for: Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Table of contents 7
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
Table of contents
Publication information 2
Contact addresses 6
Table of contents 7
Intended use 11
Symbols and abbreviations 11
What is new in publication version 3.0 12
What is new in publication version 4.0 13
What is new in publication version 5.0 13
System description
1 Description of the cobas c 513 analyzer
Overview of the cobas c 513 analyzer 19
About the control unit 21
About the analyzer 23
Sample types, containers, and racks 43
2 Specifications
List of system specifications 51
List of sampler unit specifications 54
List of analytical unit specifications 55
List of external media specifications 57
List of sample container specifications 58
List of barcode specifications 59
List of specifications for radio equipment 61
Operation and maintenance
3 Overview of operation
System modes 67
Overview of the software 71
Using the User Assistance 78
Using cobas
®
link 80
Overview of the daily workflow 86
Quick start guide 87
4 Before operation
Starting the system 95
Logging on to the software 97
Changing your password 98
Defining a new password when your password
expires 99
Preparing the system for operation 100
5 Reagents
About reagents 121
About reagent cassettes 123
Monitoring reagents 124
Viewing the reagent load/unload list 127
Loading reagent cassettes 129
Unloading reagent cassettes 131
Masking reagent cassettes 133
Unmasking reagent cassettes 134
Replacing system reagents 135
6 Calibration
About calibration 143
About recommended and manual
calibration 145
About lot calibration and cassette
calibration 147
About calibration methods 148
Configuring calibration 149
Performing calibration 161
Calibration results 168
7 Quality Control
About QC 179
About QC requests 180
About QC accumulation methods 182
Configuring QC 183
Performing QC 197
QC results 207
8 Orders and results
About orders and results 227
Ordering tests manually 228
Repeating test selections in non-barcode
mode 230
Loading samples on the analyzer 231
Testing samples 236
Test results 242
Rerunning samples 263
Unloading racks 266
9 After operation
After operation tasks 269
Logging off from the software 277
Shutting down the system 278
Extended shutdown 286
8 Table of contents
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
10 Maintenance
Overview of maintenance 291
Interlock override mode 294
Maintenance pipes and maintenance pipe
functions 298
Maintenance types 305
Maintenance status 310
Automatic maintenance actions and
checks 312
Periodic maintenance actions 317
As-needed maintenance actions 364
Advanced operation
11 Advanced operation
Operator IDs 397
Rack ranges and sample containers 403
Rack reception mode 412
Operating the system with a CLAS 416
Downloading application parameters 423
Programming calculated tests 425
Downloading diluent and detergent
information 427
Configuring mandatory reagents 429
Masks 431
Viewing system information 434
Saving and loading system parameters 436
Formatting external media 441
Saving reports to external media 443
Test profiles 445
Upload settings 450
Configuring barcode settings 454
Configuring print settings 456
Configuring the workflow 458
Configuring comment titles 460
Configuring display settings 462
Configuring the display order for results 464
Editing analyzer factors 466
Configuring the initial reaction cell 468
Configuring the number of wash cycles for the
wash rack 470
Troubleshooting
12 Alarms
About alarms 477
About system alarms 478
About data alarms 479
Configuring alarm settings 480
Configuring alarm sounds 482
Stopping the system 483
Resetting the mechanical parts of the
analyzer 484
13 Troubleshooting general problems
About troubleshooting 487
Checklists for handling materials 488
Troubleshooting problems with test results 490
Troubleshooting problems with calibration 494
Troubleshooting problems with QC 495
Handling unresolved problems 496
14 Troubleshooting data alarms
List of data alarms 501
About data alarm priority 503
Resolving data alarms for sample results 504
Resolving data alarms for calibrations 517
Resolving data alarms for QC results 521
Rerun list 526
15 Troubleshooting hardware
problems
Troubleshooting problems with the
analyzer 531
Table of contents 9
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
Software description
16 Global buttons and windows
Overview of the user interface 541
Rack Monitor window 550
Search Sample dialog box (extended) 554
Data Monitor window 556
Backup dialog box (Data Monitor window) 558
Shutdown dialog box 559
Logon dialog box 560
Sleep Conditions window 561
Shutdown Conditions dialog box 563
Change Password dialog box 564
Menu window 565
Workflow window 566
Maintenance window 572
Maintenance action windows 574
Maintenance check windows 583
Alarm window 589
Sound Setting window 592
Print window 593
History window 595
Backup dialog box (History window) 597
Print dialog box 598
Reports 599
Workplace tab (Print window) 600
Reagent tab (Print window) 610
Calibration tab (Print window) 614
QC tab (Print window) 625
Utility tab (Print window) 639
Automatic Data Monitor report 661
Automatic maintenance reports 667
Stop dialog box 673
S. Stop dialog box 674
Start Conditions window 675
Start Up Pipe Settings dialog box 679
Automatic Rerun dialog box 680
Host Settings dialog box 681
Rack Reception Settings dialog box 682
CLAS Settings dialog box 683
Default Profile dialog box 684
17 Workplace menu
Test Selection window 687
Sample Information dialog box 692
Barcode Read Error dialog box (barcode
mode) 694
Repeat Test Selection dialog box (non-barcode
mode) 696
Data Review window 697
Read Backup dialog box 702
Filter Setting window 703
Search Sample dialog box 705
Test Display Setting window 707
Send To Host dialog box 708
Backup Data dialog box 710
Reaction Monitor window 712
Scale Setting dialog box - Reaction
Monitor 715
Overlay window 716
Reagent Detail dialog box 718
Test Review window 719
Edit Result dialog box 723
18 Reagent menu
Reagent window 727
Reagent Registration dialog box 735
Bottle Set Information dialog box 736
19 Calibration menu
Calibration window 741
Start Up Setting window 744
Result window 746
Working Information window 750
Scale Setting dialog box - Working
Information 753
Reaction Monitor window 754
Scale Setting dialog box - Reaction
Monitor 756
Calibration Trace window 757
Scale Setting dialog box - Calibration Trace 760
Comment dialog box 761
Calibration Factor dialog box 762
Instrument Factor dialog box 764
Calibrators window 766
Rack Assignment window 768
Install window 770
Download Calibrator window 772
Registration dialog box 775
Rename Calibrator dialog box 777
Calibrator Detail window 778
10 Table of contents
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
20 QC menu
QC window 783
Standby Cassette QC window 785
Result window 787
Individual window 790
Individual QC Chart window 792
Comment dialog box 795
Select Chart window 796
Realtime QC window 797
Select Control window 800
Select Rules dialog box 802
Reject Data dialog box 803
Accumulate Data dialog box 804
Cumulative window 805
Cumulative QC Chart window 807
Controls window 810
Rack Assignment window 812
Install window 814
Add Control dialog box 816
Download Control window 817
Registration dialog box 819
Edit Control window 820
21 Utility menu
Utility window 825
Operator ID dialog box 829
Input ID dialog box 831
Access levels 832
Add Operator ID dialog box 838
Password Policy dialog box 839
Rack Assignment window 840
Tube Settings window 844
Barcode Settings dialog box 845
Communication Settings window 847
Review By Exception dialog box 848
Key Settings window 849
Test Key Setting dialog box 850
Profile Setting window 851
Print Settings dialog box 853
Comment Titles dialog box 855
Display Settings dialog box 856
Rack Reception dialog box 858
Analyzer Settings dialog box 859
Wash Racks dialog box 860
Maintenance Settings window 861
Edit Maintenance Type window 863
Pipe Settings window 866
Edit Pipe Setting window 868
Wake Up Settings window 870
Add Wake Up dialog box 872
Rack Delivery dialog box 874
Workflow Settings window 875
Alarm Settings dialog box 876
Result Upload Settings dialog box 878
QC Settings dialog box 880
Reagent Level window 882
Diluent and Cleaner window 884
Download Diluent and Cleaner window 888
Application window 890
Download Application window 906
Registration dialog box 908
System Configuration window 910
Test Assignment window 911
Appendix
22 Glossary
Index 923
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
11
Intended use
The cobas c 513 analyzer is a fully automated,
standalone clinical chemistry analyzer intended for the
in-vitro quantitative determination of analytes in body
fluids.
Symbols and abbreviations
Product names Except where the context clearly indicates otherwise, the
product names below are used.
Symbols used in the publication
GTIN icon used on the analyzer
Product name Term
cobas c 513 analyzer system system
cobas c 513 analyzer analyzer
cobas c 513 analyzer
software
software
cobas c pack reagent cassette
cobas
®
user assistance
User Assistance
y Product names
Symbol Explanation
o List item
u Related topics containing further information
q Tip. Extra information on correct use or useful
hints.
r Start of a task
I Extra information within a task
f Result of a user action within a task
c Frequency of a task
n Duration of a task
d Materials that are required for a task
j Prerequisites of a task
p Task. Used in cross-references to tasks.
w Figure. Used in figure titles and cross-references
to figures.
y Table. Used in table titles and cross-references to
tables.
y Symbols used in the publication
Symbol Explanation
Global Trade Item Number
y GTIN icon used on the analyzer
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
12
Content icon
Abbreviations The following abbreviations are used.
What is new in publication version 3.0
New maintenance actions There are new maintenance actions for cleaning racks
and rack trays.
u Cleaning the rack trays (387)
Cleaning the sample racks (390)
Individual clot detection settings The clot detection settings can now be individually
configured for the S1 probe and the S2 CTS probe by
your Roche Service representative.
u Configuring alarm settings (480)
Alarm Settings dialog box (876)
Symbol Explanation
Quantity in the package
y Content icon
Abbreviation Definition
ANSI American National Standards Institute
CLAS Clinical laboratory automation system
CSA Canadian Standards Association
CTS Closed tube sampling
EC European Community
EN European Standard
IEC International Electrical Commission
IVD In vitro diagnostic
LIS Laboratory information system
QC Quality control
RFID Radio frequency identification
RoHS Restriction of Hazardous Substances
SD Standard deviation
UL Underwriters Laboratories
WEEE Waste Electrical and Electronic Equipment
y Abbreviations
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
13
What is new in publication version 4.0
Connect to a CLAS The system can connect to a clinical laboratory
automation system (CLAS). A pre-analytical system can
perform time-consuming tasks before and after testing.
q Only Roche Service representatives are authorized
to install and configure the connection to a CLAS.
u Operating the system with a CLAS (416)
Program calculated tests Calculated tests are produced from applying a test
formula to the results of other tests performed on the
system.
u Programming calculated tests (425)
Data alarm priority When a data alarm occurs, the system attaches a flag to
the result. The system can only display one flag for a test
result in the software and on reports. If a test result has
multiple data alarms, the system prioritizes which flag to
display.
u About data alarm priority (503)
What is new in publication version 5.0
Updated maintenance actions There is an updated maintenance action for cleaning the
surfaces of theanalyzer.
u Cleaning the surfaces of the analyzer (273)
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
14
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
System description
1 Description of the cobas c 513 analyzer...................................................... 17
2 Specifications........................................................................................................... 49
1 Description of the cobas c 513 analyzer
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
17
Table of contents
Description of the
cobas c 513 analyzer 1
In this chapter
1
Overview of the cobas c 513 analyzer . . . . . . . . . . . . 19
About the control unit . . . . . . . . . . . . . . . . . . . . . . . . . . 21
About the analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
About the sampler unit . . . . . . . . . . . . . . . . . . . . . . 23
Components of the sampler unit . . . . . . . . . . . 24
Power switches on the sampler unit . . . . . . . . 26
Behind the front service doors of the sampler
unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
About the analytical unit . . . . . . . . . . . . . . . . . . . . . 28
Reagent area components of the analytical
unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Sample area components of the analytical
unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Reaction disk area components of the
analytical unit . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Photometric analysis workflow . . . . . . . . . . . . . 37
Behind the front service doors of the
analytical unit . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
At the front of the analytical unit . . . . . . . . . . . 40
Covers of the analytical unit . . . . . . . . . . . . . . . 40
At the rear of the analyzer. . . . . . . . . . . . . . . . . . . . 41
Sample types, containers, and racks . . . . . . . . . . . . . . 43
About sample types . . . . . . . . . . . . . . . . . . . . . . . . . 43
About sample containers. . . . . . . . . . . . . . . . . . . . . 43
About open sample tubes and closed sample
tubes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
About racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
1 Description of the cobas c 513 analyzer
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
18
Table of contents
1 Description of the cobas c 513 analyzer
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
Overview of the cobas c 513 analyzer 19
Overview of the cobas c 513 analyzer
The cobas c 513 analyzer is a fully automated,
standalone clinical chemistry analyzer. It is intended for
the in-vitro quantitative determination of analytes in body
fluids.
The main features of the system are below.
• Measurement of HbA1c in whole blood or hemolysate
• A throughput of 400 tests or more per hour
• Functionality for closed-tube sampling
The system components are below.
• Control unit
• Analyzer
- Sampler unit
- Analytical unit
A Monitor of the control unit B Sampler unit C Analytical unit
C
B
A
1 Description of the cobas c 513 analyzer
Roche Diagnostics
cobas c 513 analyzer · Software Version 02-04 · Operator's Manual · Version 5.0
20 Overview of the cobas c 513 analyzer
u Related topics
• About the control unit (21)
• About the sampler unit (23)
• About the analytical unit (28)
Unit Main Components
Control unit PC, located in the sampler unit
Touchscreen monitor
Mouse
Printer
Sampler unit Rack loading area and rack unloading area
STAT port
Barcode reader to read barcodes on racks and sample tubes
Rack rotor
Conveyor lines to transport racks
Power panel
Main circuit breakers
Analytical unit Reagent area:
o Reagent probes
o Reagent disk
o Rinse stations
o RFID reader units for the reagent cassettes
Sample area:
o Sample probes
o S1 probe
o S2 CTS probe
o Rinse stations
Reaction disk area:
o Reaction disk
o Incubation bath
o Ultrasonic mixers
o Reaction cell rinse unit
o Photometer
y Main components of the system
/